Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K103128

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-02-23

    (121 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K050325, K063729, K992869, K062549

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cadwell Disposable Stimulator Probe is used to perform localized stimulation of neural tissue and to locate, identify and monitor cranial motor nerves, peripheral nerves and spinal nerve roots during surgery. The Cadwell Disposable Stimulator Probe is a single patient use device.

    Device Description

    Cadwell Disposable Stimulator Probes come sterilized in three electrical configurations and multiple lengths to accommodate stimulating through cannulas during minimally invasive surgeries: Disposable Concentric Probes -Localized stimulation due to coaxial nature of active and reference electrodes. Tip diameter approx 1mm. Particularly useful when working with a microscope. Disposable Concentric Probes may be bent by user to facilitate viewing access under a microscope. Disposable Bipolar Probes -Both active and return electrode are built into the probe. Both tips of the Disposable Bipolar Probe must be in contact with tissue for stimulation current to flow. The Bipolar Probe will stimulate a small amount of tissue, more than Concentric Probes, but less than Monopolar probes. Disposable Bipolar Probes may be bent to facilitate viewing access under a microscope. Disposable Monopolar Probe -The Monopolar Probe has a single active electrode and must be used in conjunction with a remote return electrode connected to the patient. The Monopolar probe is insulated to the tip with only a stimulating "ball" or "tip" exposed. Disposable Monopolar Probes may be bent to facilitate viewing access under a microscope.

    AI/ML Overview

    The information primarily discusses the substantial equivalence of Cadwell Disposable Stimulator Probes to predicate devices. It provides detailed comparisons of physical characteristics.

    Here's an analysis of the provided text in relation to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" in terms of performance metrics (e.g., sensitivity, specificity, accuracy) for a device that provides an output like a diagnostic algorithm. Instead, it defines "substantial equivalence" as the acceptance criterion, which is met by demonstrating that the Cadwell probes have similar physical and material characteristics to existing predicate devices.

    FeatureAcceptance Criteria (Predicate Devices)Reported Device Performance (Cadwell Probe)
    Shaft length80-340mm (varies by probe type)80-340mm (varies by probe type)
    Handle Length110mm110mm
    Lead Length2000mm2000mm
    Tip Diameter/ExposureVaries by probe type (e.g., Ø2.3mm, Ø0.8x0.8mm exp.)Varies by probe type (e.g., Ø2.3mm, Ø0.8x0.8mm exp.)
    Shaft MaterialSST 316SST 316
    Shaft InsulationPTFEPTFE
    Handle MaterialMedical Grade ABSMedical Grade ABS
    Lead Wire MaterialTin Plated CopperTin Plated Copper
    Lead Wire InsulationMedical Grade PVCMedical Grade PVC
    Sterility Assurance Level (SAL)Not explicitly stated for predicates, but implied safeAt least 10^-6 (validated following EN ISO11135:2007)

    2. Sample Size Used for the Test Set and Data Provenance:

    The study is not a clinical trial or algorithm performance study with a "test set" in the traditional sense of AI/diagnostic device evaluation. It is a comparison of physical device specifications. Therefore, information about "sample size for the test set" and "data provenance" (country of origin, retrospective/prospective) is not applicable in this context. The "test" involved bench testing of the Cadwell probes themselves against the specifications of the predicate devices.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Ground truth in this context means the established specifications of the predicate devices. These specifications are likely derived from engineering designs and regulatory approvals, not from expert consensus on medical images or patient outcomes. Therefore, information about "number of experts" and their "qualifications" for establishing ground truth is not applicable.

    4. Adjudication Method:

    Since the study is a comparison of physical specifications rather than an interpretation task, an "adjudication method" (like 2+1 or 3+1) is not applicable. The comparison is against objective measurements and declared specifications.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This study is focused on the physical characteristics and materials of a stimulator probe, not on the interpretative performance of human readers or AI algorithms. As such, there is no "effect size of how much human readers improve with AI vs without AI assistance" to report.

    6. Standalone (Algorithm Only) Performance Study:

    No, a standalone performance study was not done. This device is a physical medical instrument (a stimulator probe), not an AI algorithm or software device.

    7. Type of Ground Truth Used:

    The "ground truth" for this study is the established specifications and characteristics of legally marketed predicate devices as declared in their 510(k) notifications or product documentation, combined with bench testing measurements of the Cadwell device's physical attributes and functionality (like sterility). It is not expert consensus, pathology, or outcomes data, as this is a device for stimulation, not diagnosis.

    8. Sample Size for the Training Set:

    This concept is not applicable as the document describes the substantial equivalence of a physical medical device, not an AI model or algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This concept is not applicable for the same reason as point 8.

    In summary, the provided document describes a 510(k) submission for substantial equivalence based on a comparison of physical device characteristics and materials, supported by bench testing and sterilization validation, rather than a performance study of a diagnostic algorithm or AI system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062996

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-12

    (102 days)

    Product Code
    Regulation Number
    882.1350
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K050325, K992869

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To locate, identify and monitor cranial motor nerves, peripheral nerves and spinal nerve roots during surgery.

    Device Description

    Stimulator probes are used as the medium to deliver electrical stimulation to tissue during intraoperative neurological monitoring. The probes are available in various monopolar and bipolar configurations according to the required application. The probes are supplied sterile and are for single use only. The probes are connected to an electrical stimulator using a flexible lead wire(s) and a "touch-proof" safety connector(s) on the distal end. Monopolar electrodes require a separate stimulator return electrode. Stimulator probes are used by the surgeon to locate and identify motor nerves and spinal nerve roots and to assess nerve function. Bipolar probes may also be used to record nerve action potentials directly from the nerve. The probes are designed with a plastic handle and stainless steel active electrode shaft insulated to the tip. The probe shaft may be bent to allow viewing access under a microscope.

    AI/ML Overview

    This submission pertains to Axon Systems' Disposable Stimulator Probes, which are devices used to deliver electrical stimulation to tissue during intraoperative neurological monitoring. The document provided is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and performance study results. As such, the requested information (acceptance criteria, specific performance metrics, sample sizes, expert qualifications, etc.) is not present in the provided text.

    The 510(k) summary only states the following:

    1. A table of acceptance criteria and the reported device performance: This information is not provided. The summary focuses on technological characteristics and substantial equivalence, not specific performance metrics against pre-defined acceptance criteria.

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): No information on test set sample sizes or data provenance is available.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): This information is not provided.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable and not mentioned, as there is no test set or ground truth establishment process described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not applicable. The device described is a medical instrument (stimulator probe), not an AI diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable, as there isn't a diagnostic algorithm undergoing validation against a ground truth.

    8. The sample size for the training set: Not applicable and not mentioned.

    9. How the ground truth for the training set was established: Not applicable and not mentioned.

    Summary of the 510(k) Conclusion:

    The document concludes that "Axon Systems' Disposable Stimulator Probes are substantially equivalent to the predicate devices. No new questions of safety or effectiveness are raised or evident." This equivalence is based on the description of technological characteristics, such as:

    • Identical materials: "Materials used are the same as in the predicate devices."
    • Similar design: Insulated probe shaft of various lengths mounted to a plastic handle, electrically connected to a "touch-proof" safety connector.
    • Intended Use: "To locate, identify and monitor cranial motor nerves, peripheral nerves and spinal nerve roots during surgery," which is presumably consistent with the predicate devices.

    The predicate devices mentioned are:

    • Technomed Europe - Disposable Probes (K050325)
    • Xomed Surgical Products - Ball Tip Monopolar stimulating probe (K992869)

    In essence, the "study" proving the device meets acceptance criteria in a 510(k) submission like this is primarily a demonstration of substantial equivalence to already cleared devices, rather than a de novo clinical trial with defined acceptance criteria and performance outcomes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031003

    Validate with FDA (Live)

    Date Cleared
    2003-04-11

    (11 days)

    Product Code
    Regulation Number
    874.1820
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K014165, K992869

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Stimulus-Dissection Instruments are indicated for tissue dissection and stimulation of cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.

    Device Description

    The designs of the Stimulus - Dissection Instruments are similar to existing stainless steel manual surgical instruments. The instruments consist of scissors, retractors, awls, taps, needles, probes, and hooks with biocompatible electrical insulation applied to selected portions, and proximal connectors provided to attach the instruments to a monopolar stimulator. The distal surfaces of the instruments are non-insulated stainless steel to provide for mechanical, manual dissection / resection, and tissue stimulation. The Stimulus - Dissection Instruments are a protected pin design and meet the requirements of IEC 60601-1:1988 /A1:1991 /A2:1995 Clause 56.3(c) per 21 CFR 898.12. Accessories include monopolar cables available in lengths up to 3 M.

    AI/ML Overview

    The provided 510(k) summary (K031003) for the Medtronic Xomed "Stimulus - Dissection Instruments" does not contain a detailed study with specific acceptance criteria and reported device performance in the way a clinical trial or algorithm performance study would.

    Instead, this 510(k) is a substantial equivalence submission, meaning the manufacturer is demonstrating that their new device is as safe and effective as a legally marketed predicate device, without requiring new clinical performance data in the same way. The "acceptance criteria" here relate to meeting the characteristics of the predicate device and relevant standards.

    Here's a breakdown based on the provided text, addressing your points as much as possible within the context of a 510(k) summary for this type of device:


    1. Table of Acceptance Criteria and Reported Device Performance

    For this specific 510(k), the "acceptance criteria" are not performance metrics in a clinical sense (e.g., sensitivity, specificity for a diagnostic device). Instead, they are the design, material, functional, and safety characteristics that allowed the device to be deemed substantially equivalent to its predicate. The "reported device performance" is implicitly that it meets these characteristics, making it comparable to the predicate.

    Acceptance Criteria (Characteristic of Predicate Device)Reported Device Performance (Proposed Device)
    Indications For Use: Tissue dissection and stimulation of cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.Meets: Tissue dissection and stimulation of cranial and peripheral motor nerves for location and identification during surgery, including spinal nerve roots.
    Construction: Stainless steelMeets: Yes
    Electrical insulation: On all surfaces not intended to provide electrical contact with the patientMeets: Electrical insulation on all surfaces not intended to provide electrical contact with the patient
    Distal patient contact surface: Stainless steelMeets: Yes
    Proximal stimulator connector: YesMeets: Yes
    IEC 60601-1 Protected Pin design: YesMeets: Yes (Meets IEC 60601-1:1988 /A1:1991 /A2:1995 Clause 56.3(c) per 21 CFR 898.12)
    Biocompatible: YesMeets: Yes
    Use: Single Use, SterileMeets: Yes

    Implicit Acceptance Criteria (not explicitly in the comparison table but mentioned in device description):

    • Design: Similar to existing stainless steel manual surgical instruments.
    • Function: Provide for mechanical, manual dissection/resection, and tissue stimulation.
    • Safety: The instruments are a protected pin design and meet the requirements of IEC 60601-1:1988 /A1:1991 /A2:1995 Clause 56.3(c) per 21 CFR 898.12.

    The "study" that proves the device meets the acceptance criteria is simply the detailed comparison provided by Medtronic Xomed in their 510(k) submission, confirming that their proposed device shares the critical characteristics and intended use of the predicate device (K014165) and adheres to relevant safety standards.


    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable in the context of this 510(k) summary. This submission relies on demonstrating substantial equivalence to a predicate device, not on a new clinical performance study with a "test set" of patients or data. The "data provenance" would refer to the technical specifications and design documentation provided by Medtronic Xomed for their new device, comparing it to the specifications of the predicate device.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. There was no "test set" requiring expert ground truth establishment for this 510(k) submission, as it's not a performance study of a diagnostic or AI device. The "ground truth" for showing substantial equivalence resided in the technical specifications and regulatory clearances of the predicate devices.


    4. Adjudication Method for the Test Set

    Not applicable, as there was no test set requiring expert adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. This is not an MRMC study and does not involve human readers interpreting data. It's for a surgical instrument.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    No. This device is a manual surgical instrument, not an algorithm or AI system.


    7. The Type of Ground Truth Used

    The "ground truth" for this 510(k) submission is the pre-existing regulatory classification, safety, and effectiveness profile of the predicate devices (Medtronic Xomed Stimulus Dissection Instruments cleared via K014165 and Medtronic Xomed Monopolar Stimulator Probes originally described in K992869). The manufacturer's submission's "truth" is that their new device aligns with these established characteristics and design standards.


    8. The Sample Size for the Training Set

    Not applicable. This is not an AI or algorithm-based device that requires a training set.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1