Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K151650

    Validate with FDA (Live)

    Date Cleared
    2016-04-04

    (291 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K052117, K071222, K081905, K991513

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SMARTeZ Pump (Long infusion time article) is intended for continuous infusions for general infusion use, including pain management.
    • Routes of administration: intravenous and subcutaneous.
    The SMARTeZ Pump (Short infusion time article) is intended for continuous infusions for general infusion use, including antibiotic delivery.
    · Route of administration: intravenous.
    The SMARTeZ Pump (Chemotherapy article) is intended for continuous infusion of chemotherapy medications.
    · Routes of administration: intravenous and intra-arterial.

    Device Description

    The SMARTeZ Pump, is a single-use disposable, non-electric infusion pump that consists of an elastomeric fluid reservoir as energy source with an integrated administration line (see Figure 5-1). The constriction of elastomeric fluid reservoir drives the medication through the tubing and eventually through a flow restrictor out into the patient connection. Drug products should be stored in their approved containers and closures.

    AI/ML Overview

    The SMARTeZ Pump, an elastomeric infusion pump, underwent a series of bench tests to demonstrate its performance and safety, aligning with ISO 28620:2010 standards. The study aimed to prove that the device consistently delivers medication within specified flow rate tolerances under various conditions, including nominal operation, changes in ambient temperature and solution viscosity, and after exposure to physical stresses like pressure, traction, and drops, as well as refrigeration. It also verified leak-proof integrity and the absence of retrograde flow.

    Here is a summary of the acceptance criteria and the reported device performance:

    1. Table of acceptance criteria and the reported device performance:

    Bench Test CompletedAcceptance CriteriaReported Device Performance
    Flow Rate Test (Nominal Condition)- Mean flow rate: ±15% compared to nominal. - At least 80% nominal volume delivered at instantaneous flow rate within ±50% of nominal.Passed. All samples met both criteria. Statistical analysis confirmed flow accuracy within +/- 15% at 99% confidence (lower- and upper-bounds of -0.400% to +0.151%) when filled with nominal volume and delivering normal saline at 31°C with the pump 40cm below the catheter site.
    Flow Rate Test (Change of Ambient Temperature)Mean flow rates should be no more than 15% slower when compared to those tested at 31°C.Passed. Mean flow rate was 14.07% slower, confirming the IFU claimed temperature effect.
    Flow Rate Test (Change of Solution Viscosity)Mean flow rates should be no more than 10% slower when compared to those tested with 0.9% NaCl as control solution.Passed. Mean flow rate was 9.81% slower, confirming the IFU claimed fluid viscosity effect.
    Flow Rate Test after Resistance to Pressure Test- Mean flow rate: ±15% compared to nominal. - At least 80% nominal volume delivered at instantaneous flow rate within ±50% of nominal.Passed. All samples met both criteria. Statistical analysis showed flow rate accuracy was not affected.
    Leak-Proof Test after Resistance to Pressure TestDevice shall remain watertight; solution shall not become colored.Passed. All samples met the criterion.
    Leak-Proof Test after Drop TestDevice shall remain watertight; solution shall not become colored.Passed. All samples met the criterion.
    Flow Rate Test after Resistance to Traction Test- Mean flow rate: ±15% compared to nominal. - At least 80% nominal volume delivered at instantaneous flow rate within ±50% of nominal.Passed. All samples met both criteria. Statistical analysis showed flow rate accuracy was not affected.
    Leak-Proof Test after Resistance to Traction TestDevice shall remain watertight; solution shall not become colored.Passed. All samples met the criterion.
    Flow Rate Test after Refrigeration- Mean flow rate: ±15% compared to nominal. - At least 80% nominal volume delivered at instantaneous flow rate within ±50% of nominal.Passed. All samples met both flow rate and leak integrity criteria. Statistical analysis showed flow rate accuracy was not affected.
    Leak-Proof Test after RefrigerationDevice shall remain watertight; solution shall not become colored.Passed. All samples met the criterion.
    Retrograde Flow of Infusate TestRetrograde flow should not be observed when back pressure applied ≤ 0.34bar (5 psi).Passed. Retrograde flow was observed at 0.8 bar (11.6 psi), which is double the acceptance criteria.
    Flow Rate Test under Non-Ambient Pressure (Influence of Routes of Administration)- Mean flow rate: ±15% compared to nominal. - At least 80% nominal volume delivered at instantaneous flow rate within ±50% of nominal, independent of IV, intra-arterial, and subcutaneous routes.Passed. Flow accuracy of +/- 15% was maintained, independent of the 3 routes of infusion administered.
    Performance/Functionality Testing for Chemotherapy Articles- Flow accuracies (nominal, after pressure/traction/refrigeration, non-ambient pressure): Mean flow rate ±15% of nominal, 80% nominal volume within ±50% instantaneous. - Watertight after pressure, traction, drop, and refrigeration. - No retrograde flow at back pressure ≤ 0.34bar (5 psi).Passed. All samples met all acceptance criteria.

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample size (number of devices) used for each individual bench test. However, it consistently refers to "All test samples" passing the criteria. It also mentions "The samples were tested."
    The data provenance is from bench testing, which implies a controlled laboratory environment. There is no information regarding the country of origin of the data or whether it was retrospective or prospective, as these types of studies are typically not applicable to physical bench tests of device functionality.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the study described is a series of bench tests for device performance against predefined engineering and regulatory standards (ISO 28620:2010), not a study involving expert assessment or clinical ground truth.

    4. Adjudication method for the test set:

    This is not applicable as the study involves objective physical measurements and functional checks against quantitative acceptance criteria, not subjective human evaluations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the device is a physical infusion pump, not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable as the device is a physical infusion pump, not an algorithm. The "standalone" performance here refers to the device's inherent mechanical and fluid dynamic performance.

    7. The type of ground truth used:

    The ground truth for the performance tests was established by objective measurements against quantitative engineering and regulatory standards (ISO 28620:2010), specifically defined numerical tolerances for flow rate, and observable physical properties like leak-proof integrity and retrograde flow.

    8. The sample size for the training set:

    This is not applicable as the device does not employ machine learning or AI models that require a training set.

    9. How the ground truth for the training set was established:

    This is not applicable as the device does not employ machine learning or AI models.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020660

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-03-20

    (19 days)

    Product Code
    Regulation Number
    880.5725
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K932740, K944692, K984502, K991513, K992072

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. The I-Flow Variable Rate Elastomeric Pump is intended for continuous and/or intermittent infusion of medications for general infusion use, including antibiotic delivery, chemotherapy and pain management. Routes of administration include the following: intravenous, intraarterial, subcutaneous, intramuscular and epidural.
    2. The I-Flow Variable Rate Elastomeric Pump is also intended for continuous and/or intermittent delivery of medication (such as local anesthetics or narcotics) to surgical wound sites and/or close proximity to nerves for preoperative, perioperative and postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural or percutaneous.
    Device Description

    The I-Flow Variable Rate Elastomeric Pump consists of an elastomeric pressure source with an integrated administration line. This special 510(k) proposes a new component that incorporates a variable flow rate mechanism.

    AI/ML Overview

    The provided document is a 510(k) summary for the I-Flow Variable Rate Elastomeric Pump. It establishes substantial equivalence to existing devices and approves the device for marketing. However, it does not contain any information about specific acceptance criteria, study data, or performance metrics that would be typically found in a clinical study report or a detailed validation summary.

    Therefore, I cannot provide the requested information. The document focuses on regulatory approval based on substantial equivalence, not on specific performance data against acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1