    Intended Use

    Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in patient transport and hospital environments.

    Device Description

    The picoSAT II SpO2 pulse oximetry module for use in host patient monitors.

    AI/ML Overview

    The Philips Medical Systems picoSAT II SpO2 pulse oximetry module (K030973) is intended for monitoring, recording, and alarming of multiple physiological parameters (SpO2 and pulse rate) in adults, pediatrics, and neonates within hospital and patient transport environments.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    ItemAcceptance Criteria (Specification)Reported Device Performance (Provided as Specification)
    SpO2 Accuracy (functional) over range of 70% to 100% for neonates through adults (reusable probes)Not explicitly stated as "acceptance criteria", but implied by device specification.M1191A and M1192A: ± 2.5%
    M1193A and M1195A: ± 3.0%
    M1194A: ± 3.0% (adult only)
    SpO2 Accuracy (functional) over range of 70% to 100% for adults and neonates (disposable probes)Not explicitly stated as "acceptance criteria", but implied by device specification.M190xA and Nellcor®: ±3.0%
    Perfusion Indicator>0.3 indicates that >95% of the time signals are good enough for valid measurements. At 0, no measurement is made.An indicator of SpO₂ signal quality>0.3 indicates that >95% of the timesignals are good enough for validmeasurements.At 0, no measurement is made. (This is a description, and implies the device meets this rule.)
    NBP cuff inflation detection suppression of SpO2 INOPs & parameter outputSpO₂ and pulse rate parameter output and the SpO₂ Non-pulsatile and Pleth Non-pulsatile INOPS are suppressed for adjustable period of 30 s to 60 s when picoSAT II SpO2 pulse oximetry module detects that an NBP measurement is in progress. SpO2 EXTD. UPDATE INOP generated after 30 s of suppression.SpO₂ and pulse rate parameter output and theSpO₂ Non-pulsatile and Pleth Non-pulsatileINOPS are suppressed for adjustable periodof 30 s to 60 s when picoSAT II SpO2 pulseoximetry module detects that an NBPmeasurement is in progress. SpO2 EXTD.UPDATE INOP generated after 30 s ofsuppression. (This is a description, and implies the device meets this rule.)
    Technical alarm conditions (INOPS)Generates various specified technical alarm conditions (SpO₂ EQUIP MALF, SpO₂ TEST SIGNAL, etc.).Generates the following technical alarmconditions:SpO₂ EQUIP MALFSpO₂ TEST SIGNALSpO₂ SENSOR MALFNO SpO₂ SENSORSpO₂ INTERFERENCESpO₂ LEARNINGSpO₂ NOISY SIGN.SpO₂ NON-PULSAT.SpO₂ ERRATICSpO₂ EXTD. UPDATESpO₂ LOW PERF (This means it is designed to achieve this.)
    SpO2 AlgorithmPhilips FAST SpO2 algorithm motion and low perfusion tolerant (Implied by specification)Philips FAST SpO2 algorithmmotion and low perfusion tolerant
    SpO2 displayed range0% to 100% (Implied by specification)0% to 100%
    SpO2 parameter resolution1% (Implied by specification)1%
    Pulse Rate parameter range & resolution30 bpm to 300 bpm ±2% or 1 bpm whichever is greater (Implied by specification)30 bpm to 300 bpm ±2% or 1 bpm whichever isgreater
    FAST SpO2 parameter averaging5 s to 20 s (Implied by specification)5 s to 20 s
    SpO2 parameter data update period1 s (Implied by specification)1 s
    Pleth wave height requirement32 pixels, minimum (Implied by specification)32 pixels, minimum
    Input power1.8 V to 11.5 Vdc, 300 mW max (Implied by specification)1.8 V to 11.5 Vdc, 300 mW max
    Serial data interface3/5 V logic levels, switchable, 8-bit word with stop, start & parity bit (Implied by specification)3/5 V logic levels, switchable'8-bit word with stop, start & parity bit

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Clinical validation studies were also conducted," but does not provide any specific sample sizes for the test set or the data provenance (e.g., country of origin, retrospective or prospective nature of the data).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    This information is not provided in the document.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not mentioned in the document. The document focuses on the technical specifications and clinical validation of the device itself, not a comparison of human reader performance with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, a standalone performance evaluation of the algorithm is implied. The document states "Clinical validation studies were also conducted" in the context of establishing the performance characteristics of the new device (the picoSAT II module, which includes the Philips FAST SpO2 algorithm). The specifications for SpO2 accuracy (e.g., ± 2.5%, ± 3.0%) are direct measurements of the device's ability to accurately determine SpO2 levels.

    7. The Type of Ground Truth Used

    The type of ground truth used for the clinical validation studies is not explicitly stated, but for SpO2 accuracy measurements, it is conventionally established through co-oximetry readings of arterial blood samples, which is considered the gold standard for blood oxygen saturation. While not explicitly mentioned, this is the expected method for establishing ground truth in pulse oximetry validation studies.

    8. The Sample Size for the Training Set

    This information is not provided in the document. The document refers to "Philips FAST SpO2 algorithm," implying a pre-existing or internal development and training process, but details are not disclosed in this summary.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the document. Similar to the test set, it would likely involve co-oximetry readings of arterial blood samples if applicable to the algorithm's development.