510(k) Data Aggregation

    K Number
    K040433

    Validate with FDA (Live)

    Device Name
    SINON
    Date Cleared
    2004-05-19

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K991285, K910422, K9909002, K971814

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. In the Q-Switch Mode, for the cutting, vaporization, or ablation of soft tissue. This includes the removal of tattoos and treatment of benign pigmented lesions.
    2. In the Free-Running Mode, for the removal of unwanted hair in patients with Fitzpatrick skin types of I and II.
    Device Description

    The SINON is a 694 nm Ruby laser system which can be operated in two different modes. The quality-switch or Q-switch or QS Operating Mode is characterized by extremely short pulse widths (ns = 103s) and high peak power (MW = 106 W). The QS Operating Mode in the SINON is used for the removal of tattoos and treatment of pigmented lesions. The second mode is the Free-Running Operating Mode which is used for the removal of hair.

    AI/ML Overview

    The provided document, K040433 for the WaveLight Laser Technologie, AG SINON laser system, does not contain information about acceptance criteria or a study proving the device meets them.

    The document explicitly states "Performance Data None presented." in section 6. This means there are no details available within this submission regarding:

    • Specific acceptance criteria for device performance.
    • Any studies conducted to demonstrate the device meets such criteria.
    • Sample sizes for test sets or their provenance.
    • Information about experts, ground truth, or adjudication methods for any hypothetical test sets.
    • Any multi-reader multi-case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance.
    • Training set details or how their ground truth was established.

    This 510(k) submission primarily focuses on establishing substantial equivalence to predicate devices based on device description and intended use, rather than presenting detailed performance study data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032297

    Validate with FDA (Live)

    Date Cleared
    2003-08-26

    (32 days)

    Product Code
    Regulation Number
    886.4370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K990900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OASIS N-PE Microkeratome Blades are designed as replacement blades for the Nidek MK-2000 Microkeratome for lamellar resection of the comea.

    Device Description

    The OASIS Disposable N-PE Microkeratome Blades are replacement stainless steel blades for the Nidek MK-2000 Microkeratome blade. The Disposable N-PE Microkeratome Blades are made of 400 series low carbon stainless steel, packaged and sterilized using the same methods. The OASIS Disposable N-PE Microkeratome Blades are single-use, disposable blades.

    AI/ML Overview

    The acceptance criteria for the OASIS Disposable N-PE Microkeratome Blades and the study proving its performance are detailed below. It's important to note that this device is a Class I medical device (keratome blade) and the submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than complex clinical studies typically associated with higher-risk devices or software.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Functional Equivalence:
    Intended UseSame as predicate device (Nidek MK-2000 Microkeratome blade)
    Target PopulationSame as predicate device
    PerformanceComparable to Nidek MK-2000 Microkeratome blade
    Fit into Nidek MK-2000Tested and shown to be acceptable
    Material/Design Equivalence:
    Blade MaterialLow carbon stainless steel (same as predicate)
    BiocompatibilityFor Stainless Steel Blades (same as predicate)
    Sterilization MethodEthylene Oxide (same as predicate), SAL of 10-6
    Performance Measurements:
    Flap DiameterMeasured, shown to be Equivalent to predicate device
    Flap ThicknessMeasured, shown to be Equivalent to predicate device
    DimensionsMeasured, shown to be Substantially equivalent to predicate device
    SharpnessTests show performance as well as the predicate device
    Safety:
    Mechanical SafetyAssured (same as predicate)
    Adverse Safety IndicationsNo adverse safety indications when used per manufacturer's instructions

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a precise numerical sample size for the "test sets" of blades used in the dimensional equivalency, sharpness, and fit tests. It implies that these tests were performed on a sufficient number of OASIS blades to draw conclusions about their equivalence to the predicate.
    • Data Provenance: The data provenance for the non-clinical testing is "non-clinical testing on porcine eyes." This indicates that the testing was performed on animal models. There is no mention of country of origin for this data specifically, nor is it specified if it was retrospective or prospective, though non-clinical tests are typically prospective for device validation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information (number and qualifications of experts for ground truth) is typically not applicable for a 510(k) submission of a Class I device focused on substantial equivalence to a predicate. The evaluation relies on direct material and performance comparisons, not expert interpretation of results that would require consensus for a "ground truth" in the way it's used in AI/diagnostic studies.

    4. Adjudication Method for the Test Set

    Not applicable. The tests performed are objective measurements (dimensions, sharpness, fit) or direct comparisons, not requiring expert adjudication for ambiguous results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices or AI algorithms where human interpretation is involved. This submission is for a surgical blade, where performance is assessed through physical and functional equivalence.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical surgical tool, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the established characteristics and performance of the predicate device (Nidek MK-2000 blade). The OASIS blades are directly compared against these known attributes through:

    • Objective Measurements: Such as dimensional equivalency.
    • Functional Demonstrations: Such as sharpness testing and fit into the microkeratome.
    • Performance Equivalence in a Model: Confirmed by producing "corneal lamellar sections equivalent to the predicate devices" on porcine eyes.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.