Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K201268

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-21

    (162 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K992692, K983773, K161739

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Spherical and Aspheric

    Ethos Aquell (Hioxifilcon A) Spherical and Aspheric One-Day Soft (Hydrophilic) Contact Lenses (59% Water) are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +10.00 to -20.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less that does not interfere with visual acuity.

    Toric

    Ethos Aquell (Hioxifilcon A) Toric One-Day Soft (Hydrophilic) Contact Lenses (59% Water) are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +10.00 to -20.00 diopters and astigmatic corrections from -0.50 to -6.00 diopters.

    The lenses are intended for single-use disposable wear.

    Device Description

    The Ethos Aquell (Hioxifilcon A) One-Day Soft (Hydrophilic) Contact Lenses (59% Water) are clear, visibility-tinted and include UV blocker-containing. The lens is available in spherical, aspherical and toric designs. It is an non-ionic hydrogel lens Hioxifilcon derived from of 2-Hydroxyethylmethacrylate (2-HEMA) and 2,3-Dihydroxypropyl Methacrylate (GlycerolMethacrylate, GMA), cross-linked with ethylene glycol dimethacrylate (EGDMA) and made by photo-polymerization. It consists of 41% Hioxifilcon A and 59% water by weight when immersed in buffered solution. There are two types of buffer solution. one is borate solution (Saline 2); another one is borate solution with sodium hyaluronate (Saline 1). The lens polymer further contains a benzotriazole UV absorbing monomer and thus is able to block UV radiation. The lens is visibly tined with "Reactive Blue19" color additive, 21 CFR part 73.3121.

    AI/ML Overview

    This document is a 510(k) summary for a contact lens, not an AI/ML medical device. Therefore, the information typically required for an AI/ML device's acceptance criteria and study proving it meets those criteria (such as sample size for test/training sets, expert consensus, MRMC studies, etc.) is not present.

    The document focuses on demonstrating substantial equivalence to a previously cleared predicate device, which is the standard pathway for many medical devices, including contact lenses. This involves showing similar technological characteristics and performance through non-clinical testing.

    Here's an analysis of the provided text in relation to the requested information, highlighting why certain points cannot be addressed:

    1. A table of acceptance criteria and the reported device performance

    The document provides a comparative table of technological characteristics between the proposed device ("Ethos Aquell (Hioxifilcon A) One-Day Soft (Hydrophilic) Contact Lenses (59% Water)") and a predicate device (K040303). This isn't strictly "acceptance criteria" in the sense of performance metrics for an AI/ML algorithm, but rather a comparison for demonstrating substantial equivalence.

    CharacteristicProposed Device (Ethos Aquell)K040303 Predicate
    Production MethodCast-MoldedCast-Molded
    USAN NameHioxifilcon AHioxifilcon A
    Material ClassificationGroup 2 (high water non-ionic)Group 2 (high water non-ionic)
    Water Content (%)59%59%
    Refractive Index1.4001.404
    Oxygen Permeability (edge corrected) @ 35°C20.6 x 10-11 (cm²/sec)(mlO2/ml-mmHg)18 x 10-11 (cm²/sec)(mlO2/ml-mmHg)
    Transmittance: Visible light @ 380~780nm> 95%> 95%
    Transmittance: UVA @ 380-315nm< 50%N/A
    Transmittance: UVB @ 315-280nm< 5%N/A
    Lens designSpherical and Aspheric, ToricSpherical, Toric
    Indications for UseCorrection of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic/non-aphakic persons with non-diseased eyes. Daily wear.Correction of visual acuity (myopia, hyperopia, astigmatism) in aphakic/non-aphakic persons with non-diseased eyes. Daily wear.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states: "Clinical studies for Ethos Aquell (Hioxifilcon A) One-Day Soft (Hydrophilic) Contact Lenses (59% Water) of the present device are not required for the premarket notification," because the USAN name and process are the same as previously cleared predicate devices.

    Instead, non-clinical testing was performed. The document lists the types of non-clinical tests (e.g., Physicochemical Properties, Biocompatibility Testing, Shelf Life Test, Sterility Test), but does not provide sample sizes for these tests or data provenance (country of origin, retrospective/prospective). These types of details are usually found in the full submission, not the public 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as no clinical study or human expert review of "test sets" (in the AI/ML sense) was conducted or required for this 510(k) submission. The evaluation is based on material properties and non-clinical performance, and substantial equivalence to a predicate.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set requiring expert adjudication was part of this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is for an AI/ML device, not a contact lens.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is for an AI/ML device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For a contact lens, "ground truth" relates to material specifications and performance standards (e.g., ensuring water content is 59%, or that oxygen permeability meets a certain threshold). These are established physical and chemical properties, not clinical "ground truth" derived from patient data or expert consensus in the diagnostic sense. The document states that non-clinical performance tests were conducted "in accordance with the May 1994 FDA guidance title Premarket Notification 510(K) Guidance Document for Class II Contact Lenses." This guidance defines the expected methods and standards for these physical and chemical property assessments.

    8. The sample size for the training set

    Not applicable. This is for an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is for an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070308

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-03-12

    (39 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K983773, K964528

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexlens Multifocal Soft Contact Lens (hioxifilcon A) and (hioxifilcon B) for Daily Wear is indicated for daily wear use for the correction of refractive ametropia (myopia, hyperopia and/or presbyopia) in non-aphakic persons.

    Device Description

    The Flexlens Multifocal Soft Contact Lens (hioxifilcon A) and (hioxifilcon B) for Daily Wear is a lathe-cut soft lens multifocal with an aspheric front surface and a spherical base curve. In the dry (unhydrated) state the lens is machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. The progressive optics utilize the simultaneous vision concept which offers functional vision from distance to near throughout the viewing range. The pupil zone on the front surface varies with the size of the patient's pupil.

    The non-ionic lens material, hioxifileon A, is a copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2,3-Dihydrosypropyl Methacrylate (Glycerol Methacrylate, GMA). It consists of 41% hioxifilcon A and 59% water by weight. when immersed in normal saline solution buffered with sodium bicarbonate. Flexlens Multifocal Soft Contact Lenses for Daily Wear made out of hioxifilcon A are available in clear and with a blue visibility-handling tint, phthalocyanato (2) - (copper).

    The non-ionic lens material, hioxifileon B, is a copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2,3-Dihydrosypropyl Methacrylate (Glycerol Methacrylate, GMA). It consists of 52% hioxifilcon B and 48% water by weight when immersed in normal saline solution buffered with sodium bicarbonate. FlexIens Multifocal Soft Contact Lenses for Daily Wear made out of hioxifilcon B are available in clear and with a blue visibility-handling tint, phthalocyanato (2) - (copper).

    AI/ML Overview

    The provided document, K070308, is a 510(k) summary for the Flexlens Multifocal Soft Contact Lens (hioxifilcon A) and (hioxifilcon B) for Daily Wear. It clarifies that this is a premarket notification for a Class II medical device, and the FDA has determined it to be substantially equivalent to legally marketed predicate devices.

    Crucially, this document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and intended use, rather than presenting a de novo study with specific acceptance criteria and detailed performance data often seen in AI/algorithm-based device submissions.

    Therefore, most of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth establishment, which are typical for AI/ML device evaluations, are not explicitly present in this 510(k) summary. The document primarily refers to "Pre-Clinical Performance Data" in previous 510(k) submissions for the materials themselves (hioxifilcon A and B).

    Based on the provided text, here's what can be extracted:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a traditional contact lens submission, specific "acceptance criteria" and "device performance" metrics in the context of an AI/ML algorithm (e.g., sensitivity, specificity, AUC) are not described. Instead, the document focuses on the physical and optical properties of the lens materials.

    PropertyAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Flexlens Multifocal Soft Contact Lens)
    For hioxifilcon A:Should be comparable to predicate devices.
    Refractive Index (hydrated)N/A (implied by predicate)1.404
    Light Transmission (clear)N/A (implied by predicate)> 95% T
    Light Transmission (tinted)N/A (implied by predicate)> 95% T
    Water ContentN/A (implied by predicate)59%
    Specific Gravity (hydrated)N/A (implied by predicate)1.18
    Oxygen Permeability (Dk)N/A (implied by predicate)$18 \times 10^{-11}$ (cm2/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method)
    For hioxifilcon B:Should be comparable to predicate devices.
    Refractive Index (hydrated)N/A (implied by predicate)1.404
    Light Transmission (clear)N/A (implied by predicate)> 95% T
    Light Transmission (tinted)N/A (implied by predicate)> 95% T
    Water ContentN/A (implied by predicate)48 %
    Specific Gravity (hydrated)N/A (implied by predicate)1.136
    Oxygen Permeability (Dk)N/A (implied by predicate)$15 \times 10^{-11}$ (cm2/sec) (ml O2/ml x mm Hg @ 35°C), (revised Fatt method)

    2. Sample size used for the test set and the data provenance

    This information is not provided in the 510(k) summary. The document references "Pre-Clinical Performance Data" in other 510(k) submissions (K983773 and K964528) for the hioxifilcon A and B materials, but it does not detail the specific studies, sample sizes, or data provenance conducted for this particular multifocal lens design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the 510(k) summary. It's not relevant for a traditional contact lens submission focused on physical/optical properties and chemical composition.

    4. Adjudication method for the test set

    This information is not provided and is not relevant for the type of device and submission described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a MRMC comparative effectiveness study was not done. This is not an AI-assisted diagnostic device, but a contact lens.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This is not an AI/ML algorithm.

    7. The type of ground truth used

    The concept of "ground truth" as it applies to AI/ML diagnostics (e.g., pathology, outcomes data) is not applicable here. The "truth" for this device relates to its physical, chemical, and optical properties meeting established standards for contact lenses and being comparable to legally marketed predicate devices.

    8. The sample size for the training set

    This information is not provided and is not relevant as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    This information is not provided and is not relevant as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030593

    Validate with FDA (Live)

    Date Cleared
    2003-05-30

    (94 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K983773, K964528, K023349, K024045

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flexlens (hioxifilcon A), (hioxifilcon B), (acofilcon A), (acofilcon B) Soft Contact Lens for Daily Wear (lathe-cut) are indicated for daily wear use for the correction of refractive ametropia and specialized use such as atypical ametropia. Examples include, but are not limited to: adult and pediatric aphakia, and irregular astigmatism created by keratoconus, trauma, or post keratoplasty.

    The Flexlens (hioxifilcon A), (hioxifilcon B), (acofilcon A), (acofilcon B) Harrison Post Refractive Surgery Soft Contact Lens for Daily Wear (lathe-cut) are indicated for daily wear use for the correction of refractive ametropia and specialized use such as atypical ametropia following corneal refractive surgery.

    The lens may be disinfected with a chemical (not heat) disinfection system.

    Device Description

    The Flexlens and Harrison Post Refractive Surgery (hioxifilcon B), (acofilcon A), (acofilcon B) Soft Contact Lens for Daily Wear (lathe-cut) is fabricated from one of the above materials which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.

    In the hydrated state, the lens conforms to the curvature of the eve covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

    AI/ML Overview

    The provided text describes a 510(k) submission for new contact lens materials, not a study proving device performance against acceptance criteria in the typical sense of a clinical or AI performance study. The submission focuses on demonstrating substantial equivalence to predicate devices. Therefore, many of the requested categories are not applicable to this type of regulatory submission.

    Here's a breakdown of the relevant and non-applicable information based on the provided text:

    Acceptance Criteria and Device Performance

    The "acceptance criteria" in this context are the characteristics and performance values that demonstrate substantial equivalence to legally marketed predicate contact lenses. The device performance is assessed by comparing these characteristics of the new materials to those of the predicate materials.

    CharacteristicAcceptance Criteria (Predicate Device Range)Reported Device Performance (New Materials: hioxifilcon A, B; acofilcon A, B)
    INDICATIONDaily wear, Soft Contact LensDaily wear, Soft Contact Lens
    INTENDED USECorrection of refractive ametropia and specialized use for daily wear.Correction of refractive ametropia and specialized use (e.g., aphakia, irregular astigmatism due to keratoconus, trauma, post keratoplasty; and following corneal refractive surgery) for daily wear. (Matches predicate for both Flexlens and Harrison PRS)
    PRODUCTION METHODLathe-CutLathe-Cut
    HYDROPHILLIC MATERIAL/USAN(methafilcon A), (hefilcon A)(hioxifilcon A), (hioxifilcon B), (acofilcon A), (acofilcon B)
    Water Content(methafilcon A) - 55%(hefilcon A) - 45%(hioxifilcon A) - 59%(hioxifilcon B) - 48%(acofilcon A) - 58%(acofilcon B) - 49%
    Specific Gravity(methafilcon A) - 1.090(hefilcon A) - 0.979(hioxifilcon A) - 1.18 (hydrated)(hioxifilcon B) - 1.136 (hydrated)(acofilcon A) - 1.103 (hydrated)(acofilcon B) - 1.142 (hydrated)
    Oxygen Permeability*(methafilcon A) - 18.8(hefilcon A) - 16.0(hioxifilcon A) - 18.0(hioxifilcon B) - 15.0(acofilcon A) - 25.5(acofilcon B) - 15.8
    Light Transmittance(methafilcon A) - >95%(hefilcon A) - >95%(hioxifilcon A) - >95%(hioxifilcon B) - >95%(acofilcon A) - >93%(acofilcon B) - >96%
    Refractive Index(methafilcon A) - 1.40 (hydrated)(hefilcon A) - 1.43 (hydrated)(hioxifilcon A) - 1.40 (hydrated)(hioxifilcon B) - 1.40 (hydrated)(acofilcon A) - 1.40 (hydrated)(acofilcon B) - 1.42 (hydrated)

    *Note: Oxygen Permeability units are (cm²/sec)(ml O₂/ml x mm Hg @ 35°C), (revised Fatt method).


    Study Details Table

    This submission is a 510(k) Pre-Market Notification for substantial equivalence for new materials for an existing device, not a performance study (e.g., clinical trial) in the traditional sense that would generate new patient-specific data for acceptance criteria. The "study" here is primarily a comparison of material properties and intended uses to predicate devices.

    QuestionInformation from Text
    2. Sample size used for the test set and the data provenanceNot applicable. This is not a study with a "test set" of patient data. The "test" is a comparison of physical properties of the new materials to the established properties of predicate materials. The "data provenance" for the pre-clinical performance data (cytotoxicity, systemic injection, ocular eye irritation) for the new materials is by reference to previous 510(k) submissions from the material manufacturers (Benz Research and Development for hioxifilcon A and B; Contamac Ltd. for acofilcon A and B). These previous submissions would contain data from their respective studies.
    3. Number of experts used to establish the ground truth for the test set and their qualificationsNot applicable. There is no "test set" requiring ground truth established by experts. The properties presented (e.g., water content, oxygen permeability) are measured physical properties of the materials themselves, likely determined by standardized laboratory methods, not subjective expert assessment of clinical data.
    4. Adjudication method for the test setNot applicable. No "test set" requiring adjudication.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of human readers improve with AI vs without AI assistanceNot applicable. This is not an AI device, nor a comparative effectiveness study involving human readers.
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was doneNot applicable. This is not an AI algorithm.
    7. The type of ground truth usedNot applicable in the sense of clinical ground truth (e.g., pathology, outcomes data). The "ground truth" for the material properties are the experimentally measured physical and chemical characteristics of the materials, which are then compared to the established properties of the predicate devices for demonstrating substantial equivalence. For pre-clinical safety data, the "ground truth" refers to the results of standard biocompatibility tests (cytotoxicity, systemic injection, ocular irritation) referenced from prior 510(k)s.
    8. The sample size for the training setNot applicable. This is not a machine learning/AI device, so there is no training set.
    9. How the ground truth for the training set was establishedNot applicable. There is no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1