    Intended Use

    The RFA Medical InCircle™ Bi-Polar RF Ablation System is intended to be used in conjunction with radiofrequency (RF) generators for the thermal coagulation of soft tissues.

    Device Description

    The RFA Medical InCircle™ Bi-Polar RF Ablation System is a sterile, disposable, bi-polar, Radio Frequency (RF), hand-held electrosurgical device that utilizes opposing sets of Electrodes to coagulate/ablate a region of soft tissue using bi-polar RF energy. The device is designed for in percutaneous, laparoscopic, and intraoperative surgical use procedures. The device is designed for use with standard FDA cleared generators, such as the Boston RF Scientific Corporation (RadioTherapeutics) RF3000 (K000241) and the RITA Medical Systems 1500X (K983214).

    AI/ML Overview

    The provided text describes the "InCircle™ Bi-Polar RF Ablation System" and its 510(k) submission. However, it does not contain information about specific acceptance criteria for a study, a detailed study description, performance data, sample sizes, expert qualifications, or ground truth establishment relevant to the questions asked.

    The document is a 510(k) summary for a medical device seeking clearance based on substantial equivalence. It focuses on:

    • Device identification and contact information.
    • Device description and intended use.
    • Comparison to predicate devices.
    • Statements about bench testing/functional testing to ensure substantial equivalence and safety/efficacy.

    Therefore, I cannot answer the specific questions about acceptance criteria, study details, performance data, sample sizes, expert qualifications, or ground truth establishment based solely on the provided text.

    The document states: "Bench testing/functional testing was performed on the RFA Medical InCircle™ Bi-Polar RF Ablation System to ensure that the product is substantially equivalent to the predicate device and to ensure that the new device does not raise new questions of safety and efficacy." This implies testing was done, but the details of those tests, their acceptance criteria, and specific results are not included in this summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021329

    Validate with FDA (Live)

    Device Name
    RITA SYSTEM
    Date Cleared
    2002-10-09

    (166 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Predicate For
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RITA® System (RF generator and electrosurgical devices) supplies energy for use in electrosurgery and is indicated for use in percutaneous, laparoscopic, or intraoperative coagulation and ablation of soft tissue, including:

    • the partial or complete ablation of non-resectable liver lesions and
    • the palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard pain therapy.
    Device Description

    The RITA System (RF Generator and electrosurgical devices) is designed to create coagulative necrotic lesions in soft tissue, and allow localized delivery of fluid to the lesion. The RF Generator is specifically designed for use with RITA Electrosurgical Devices (Accessories). The RF Generator provides multiple temperature sensors, impedance, and power monitoring to assist the user in delivering the desired energy to the target tissue. The electrosurgical devices consist of monopolar electrosurgical devices that include disposable electrosurgical probes with deployable needle arrays that deliver RF power with temperature and impedance monitoring to ablate predictable volumes of tissue.

    AI/ML Overview

    The provided text describes the RITA® System, an electrosurgical device, and its intended use and performance data. However, it does not contain the detailed information necessary to fully answer all aspects of your request regarding acceptance criteria and a specific study proving device performance against those criteria.

    Here's what can be extracted and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria (e.g., target specificity, sensitivity, or similar metrics for a diagnostic device). Instead, it describes a clinical study and its observed outcomes as a measure of effectiveness.

    Acceptance CriteriaReported Device Performance
    (Not explicitly defined in terms of specific performance metrics for this device, which is a therapeutic device for ablation and pain palliation)Effectiveness for Pain Palliation (Worst Pain): 75% of patients experienced at least a two-point reduction in "worst pain" from baseline to week four.
    Effectiveness for Pain Palliation (Average Pain): 80% of patients experienced at least a two-point reduction in "average pain" from baseline to week four.
    Safety/Morbidity: Low attendant morbidity (no reports of death related to use of the devices).
    Adverse Events: Three adverse events potentially related to RF ablation.
    (Bench study findings on ablation characteristics/parameters)Bench studies determined the effect of the bone environment on ablation characteristics and parameters. (Specific quantitative results not provided).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the total number of patients in the clinical study. It refers to percentages of "patients" (75% and 80%) but not the base number.
    • Data Provenance: The study was conducted on "patients with lesions that had metastasized to one or more locations in the skeleton." It is a prospective clinical study assessing safety and effectiveness. The country of origin for the data is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. For a therapeutic device like this, "ground truth" would typically relate to clinical outcomes rather than expert-derived labels on images.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given the nature of pain palliation as an endpoint, "adjudication" in the sense of expert consensus on image interpretation is less relevant. The effectiveness was measured using a "validated instrument for evaluating cancer pain," suggesting a standardized assessment.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    This information is not provided. The study described is a clinical study evaluating the device's effectiveness in alleviating pain, not a comparative effectiveness study involving human readers with/without AI assistance. The device itself is an RF ablation system, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

    This question is not applicable to the RITA® System as described. It is a physical medical device (RF generator and electrosurgical devices) that requires direct human operation, not a standalone algorithm. The "performance" refers to the device's clinical efficacy and safety when used by a clinician.

    7. The Type of Ground Truth Used

    The "ground truth" for the effectiveness of the RITA® System's pain palliation indication was based on:

    • Outcomes Data: Specifically, patient-reported pain levels assessed using an "instrument validated for evaluating cancer pain."
    • Safety Data: Observed adverse events and mortality related to the device's use.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The RITA® System is a physical electrosurgical device, not a machine learning algorithm that requires a "training set" in the computational sense. The "bench studies" mentioned would involve laboratory testing, not a data training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reasons as #8. There is no "training set ground truth" in the context of this device.

    In summary, the provided text describes a clinical study demonstrating the RITA® System's effectiveness in pain palliation and its safety profile. However, it does not detail specific, quantitative acceptance criteria in a tabular format, nor does it address aspects related to AI/algorithm performance (like training sets, expert adjudication, or MRMC studies) as the device is a therapeutic electrosurgical system.