510(k) Data Aggregation

    K Number
    K043098

    Validate with FDA (Live)

    Date Cleared
    2004-12-15

    (36 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K031614, K973501, K022594

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    An electrically powered breast pump with adjustable vacuum and massage settings and 2 funnels for expressing milk from the breasts of a lactating woman.

    Device Description

    The Gerber Double Electric Breast Pump is a simple and effective system for expressing milk from a lactating women's breast(s). The device includes the following components: a pump base, flexible tubing, a funnel assembly, a carrying bag, and accessories (e.g. bottles, nipples, and disposable nursing pads). The device is powered by a variable speed power supply that drives a diaphragm pump creates the negative pressure required to extract breast milk. The breast pump is capable of providing a variable vacuum and massage like motion. A maximum vacuum of 240 mm Hg can be applied to one or both breasts.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the Gerber Double Electric Breast Pump, seeking substantial equivalence to existing predicate devices. It does not contain information about the acceptance criteria and a study proving device performance in the context of an AI/ML medical device.

    The document discusses "non-clinical tests" as "Bench testing of the device has demonstrated that the Gerber Double Electric Breast pump meets established requirements when used in the manner and environment specified in product labeling." It explicitly states, "No clinical tests have been conducted on this device."

    Therefore, I cannot provide the requested information as it pertains to a medical device that is not an AI/ML product and does not involve the types of studies described in your request. The device is a traditional medical device (breast pump), and its clearance is based on substantial equivalence to predicate devices rather than complex performance studies against specific acceptance criteria for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K022594

    Validate with FDA (Live)

    Date Cleared
    2002-11-01

    (88 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K973501, K950531, K950750

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    An electrically powered breast pump used to express milk from the breast of a lactating woman.

    Device Description

    The Playtex electric breast pump is a small, quiet, safe and effective system for expressing milk from a lactating mother's breast(s). This device is comprised of 4 major assemblies: a pump assembly, a breast cup assembly, a carry bag, and some commercially available items (i.e., bottles, bottle liners, etc). The device is designed with 5 pre-set suction levels and 5 pre-set speed settings, which are selectable by the user via a button pad. The device is powered by a 12V DC power supply, which is included with the package.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Playtex Breast Pump (K022594), addressing the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text for the Playtex Breast Pump (K022594) does not outline explicit acceptance criteria or report specific device performance metrics in the way one would expect for a diagnostic or AI device (e.g., sensitivity, specificity, accuracy). Instead, substantial equivalence for this medical device (a breast pump) is primarily demonstrated through comparison to predicate devices and adherence to material safety standards.

    However, based on the comparative table, we can infer some "performance" characteristics relative to the predicate devices:

    CharacteristicAcceptance Criteria (Inferred from Predicate Devices)Reported Playtex Breast Pump Performance
    Intended UseTo express milk from the breast of lactating women.To express milk from the breast of lactating women.
    Highest Vacuum SettingWithin range of predicate devices (e.g., 6.7 - 9.8 in Hg)9.0 in Hg
    Lowest Vacuum SettingWithin range of predicate devices (e.g., 0.0 - 3.7 in Hg)2.5 in Hg
    Range of Cycle SpeedsWithin range of predicate devices (e.g., 29 - 60 cycles/min for adjustable, 49 for fixed)45 - 75 cycles/min
    Power SourceDC Power Supply (some predicates also offer batteries)DC Power Supply
    Pump TypeReciprocating Piston or DiaphragmReciprocating Piston
    Single or Double PumpingBothBoth
    Adjustable Suction LevelsYesYes
    Adjustable Cycle SpeedYes (most, but not all predicates)Yes
    Overflow ProtectionYes (most, but not all predicates)Yes
    Breast Cup-to-Breast InterfaceRigid Plastic (some with partial silicone)Soft Silicone (difference highlighted as a feature)
    Active Breast MassageNo (for predicate devices)Yes (difference highlighted as a feature)
    Hospital Grade SystemYes (some predicates)Yes
    Usable on Tabletop or Inside BagTabletop (some predicates), Inside (one predicate)Both
    Material SafetyAll milk and human contact components manufactured from FDA food additive compliant materials.All milk and human contact components meet FDA food additive criteria (21 CFR Parts 176, 177, 178). Silicone breast cup insert tested for biocompatibility.
    FunctionalityProvides sufficient vacuum pressure to effectively express and collect milk."provides sufficient vacuum pressure to effectively express and collect milk from lactating women." (Conclusion)

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary explicitly states:

    • "No clinical tests have been conducted on this device."
    • "Testing of the device has demonstrated that the Playtex breast pump meets established requirements when used in the manner and environment specified in product labeling."

    Therefore, there was no "test set" in the context of clinical studies with human subjects or a dataset for algorithm evaluation. The testing mentioned refers to non-clinical, likely engineering or bench testing, to confirm specifications and material compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As no clinical tests or data-driven evaluations were performed for a "test set," no experts were used to establish ground truth in this context.

    4. Adjudication Method for the Test Set

    Not applicable. No "test set" or clinical evaluation requiring adjudication was described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a breast pump, not an AI-powered diagnostic or interpretive system that would involve human "readers" or an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical breast pump and does not involve an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's approval is primarily established through:

    • Predicate Device Comparison: The functional and performance characteristics (vacuum, cycle speeds, etc.) are deemed acceptable by virtue of being comparable to, or within the established range of, legally marketed predicate breast pumps.
    • Material Science and Biocompatibility Standards: Compliance with FDA food additive criteria and biocompatibility testing for human contact components serves as a "ground truth" for safety.
    • Engineering Specifications: The "testing of the device" mentioned in the non-clinical tests section implies verification against internal engineering specifications.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device, not an AI or machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As there is no training set for an AI/ML model, no ground truth was established in this manner.

    Summary of Approach for this Device:

    The 510(k) clearance for the Playtex Breast Pump relies on demonstrating substantial equivalence to existing legally marketed predicate devices. This approach primarily involves:

    • Comparing the new device's intended use, technological characteristics, and performance specifications to predicate devices.
    • Ensuring material safety and biocompatibility through standard testing and compliance with regulations.
    • Verifying that non-clinical tests (likely engineering/bench testing) confirm the device meets established requirements.

    The submission explicitly states that "No clinical tests have been conducted on this device," which is common for devices seeking 510(k) clearance where substantial equivalence can be demonstrated through non-clinical data and comparison to predicates.