510(k) Data Aggregation

    K Number
    K971939

    Validate with FDA (Live)

    Date Cleared
    1997-08-06

    (71 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K971939

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smith & Nephew Suture Collet is intended for use in the management of soft vessel ligation and / or fixation of soft tissue structures during open and endoscopic procedures.
    The Smith & Nephew Suture Collet is indicated for use in the management of soft vessel ligation and / or fixation of soft tissue structures during open and endoscopic procedures, including thoracoscopic surgery, laparoscopic procedures, and general surgery. The device is not indicated for use in contraceptive tubal ligation.

    Device Description

    The Smith & Nephew Suture Collet System comprises three main components:
    • the suture collet implant (ring and pin),
    • the disposable cartridge assembly with threader, and
    • the reusable delivery instrument.

    AI/ML Overview

    The provided text describes a 510(k) summary for the "Smith & Nephew Suture Collet" device. This is a submission demonstrating substantial equivalence to a predicate device, not a study proving the device meets acceptance criteria in the sense of performance metrics against a medical condition.

    Therefore, many of the requested sections about acceptance criteria, study design, expert ground truth, etc., are not applicable to this type of regulatory submission as they are focused on performance equivalence to a legally marketed device rather than clinical efficacy or diagnostic accuracy.

    Here's an attempt to answer the questions based on the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criterion for a 510(k) submission is typically substantial equivalence to a legally marketed predicate device. This is primarily demonstrated by comparing design, materials, intended use, indications, and performance characteristics.

    Acceptance Criterion (Implicit for 510(k))Reported Device Performance (Summary of Equivalence)
    Substantial Equivalence to PredicateThe Smith & Nephew Suture Collet is substantially equivalent to the Suture Lock, and silk 2-0 suture. Comparative strength testing demonstrates the equivalence of the Suture Lock to the predicate devices.
    Similar designYes, similarities in design identified.
    Similar materialsSuture Collet implant: Implant grade polyacetal. Suture Lock implant: Implant grade polyacetal. Delivery Instrument: Aluminum and Stainless Steel (both devices).
    Similar intended useSuture Collet: Management of Soft Tissue. Suture Lock: Management of Soft Tissue.
    Similar indications for useSuture Collet: Open and Endoscopic/Laparoscopic/Thoracoscopic Surgical Procedures. Suture Lock: Open and Arthroscopic Surgical Procedures. (While "Thoracoscopic" is added for the Suture Collet, the general scope is similar).
    Similar sterilization methodImplant: Ethylene Oxide (both devices). Delivery Instrument: Steam autoclave or ethylene oxide (both devices).
    Performance (e.g., strength)Comparative strength testing demonstrates the equivalence of the Suture Collet to the predicate devices (Suture Lock and silk 2-0 suture). Specific numeric values for strength are not provided in this summary.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated. The submission mentions "Comparative strength testing," but does not provide details on the number of samples or tests performed.
    • Data Provenance: Not explicitly stated. The testing would have likely been conducted by the manufacturer (Smith & Nephew, Inc.) as part of their product development and regulatory submission process. It is not described as involving external clinical data or retrospective/prospective studies on patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth in the clinical sense (e.g., disease diagnosis) is not relevant to this type of device submission, which focuses on mechanical equivalence. The "ground truth" here is the performance of the predicate device and USP standard suture.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods (like 2+1, 3+1 for clinical consensus) are not relevant to demonstrating substantial equivalence for a medical device through comparative strength testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for tissue fixation/ligation, not an AI-assisted diagnostic or imaging device for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to the established characteristics and performance of the legally marketed predicate devices (Suture Lock and 2-0 silk suture), against which the new device's properties (materials, design, intended use, indications, and strength) were compared. This is a regulatory "ground truth" for demonstrating equivalence rather than a clinical ground truth for a medical condition.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device, not an AI model.