    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K971938

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Partial replacement of articulating surfaces of the knee when only one side of the joint is affected due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.

    The devices covered in this 510(k) are intended to be used with Repicci™ Femoral Components.

    The device is a single use implant intended for implantation with bone cement.

    Device Description

    The Repicci II® Unicondylar Knee System consists of a femoral and tibial component. The Repicci™ tibial bearing is used with the Repicci™ femoral components (K971938).

    The all polyethylene tibial components are anatomical in geometry with right and left medial/lateral allocations. Located on the inferior surface of the implant is a face with fixation holes and a keel.

    AI/ML Overview

    The provided document does not describe acceptance criteria for a device, nor does it detail a study that proves a device meets such criteria.

    Instead, this document is a Summary of Safety and Effectiveness and a 510(k) clearance letter for a medical device called the "Repicci™ Locked Keel Tibial Bearing." This type of submission to the FDA is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving that a device meets specific performance acceptance criteria through detailed studies.

    Here's a breakdown of why this information is not present in the provided text, and what is present:

    Key Takeaways from the Document:

    • Device: Repicci™ Locked Keel Tibial Bearing (part of a unicompartmental knee system).
    • Purpose of Submission: Seek 510(k) clearance from the FDA by demonstrating substantial equivalence to a predicate device (Worland Unicondylar All Poly Tibial Bearing -- K011795).
    • Basis for Substantial Equivalence:
      • Non-Clinical Testing: An engineering justification determined that the Repicci™ Locked Keel tibial component presented no new risks and was substantially equivalent. This suggests design and material comparisons, but no specific performance metrics or studies are detailed. The change from the predicate device is a modification to the keel and removal of a waffle pattern.
      • Clinical Testing: "None provided as a basis for substantial equivalence." This explicitly states that no clinical trials or studies were conducted to prove the device meets specific performance criteria.
    • FDA Decision: The FDA granted 510(k) clearance, concluding that the device is substantially equivalent to legally marketed predicate devices for the stated indications for use.

    Addressing your specific points based on the absence of the requested information in the text:

    1. A table of acceptance criteria and the reported device performance

      • Not Available. The document does not specify any quantitative acceptance criteria or report device performance against them. The basis for clearance was "engineering justification" and "substantial equivalence" to a predicate, not performance against defined metrics.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Not Applicable/Not Available. No specific test set or data from a study is mentioned, as clinical testing was not provided for substantial equivalence. The "engineering justification" would involve design comparisons and possibly bench testing, but details on sample size, provenance, or type are not in this summary.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not Applicable/Not Available. There was no clinical study described that would require experts to establish ground truth for a test set.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not Applicable/Not Available. No test set or related adjudication method is described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not Applicable. This device is a knee implant, not an AI-powered diagnostic tool. MRMC studies involving human readers and AI are irrelevant to this product type.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not Applicable. This is a physical orthopedic implant, not a software algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Not Applicable/Not Available. No clinical study result requiring ground truth is presented.
    8. The sample size for the training set

      • Not Applicable/Not Available. No algorithm or machine learning component is involved in this device, therefore no training set exists.
    9. How the ground truth for the training set was established

      • Not Applicable/Not Available. As above, no training set or ground truth for one is relevant here.

    In summary, the provided document is a regulatory submission demonstrating "substantial equivalence" based on engineering justification and comparison to a predicate device, rather than a study proving performance against defined acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020264

    Validate with FDA (Live)

    Date Cleared
    2002-02-08

    (14 days)

    Product Code
    Regulation Number
    888.3530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K971938

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indications for use for the Repicci II™ Unicondylar Knee System are for painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved; the correction of varus, valgus, or posttraumatic deformity; and the correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure. This device is for use with bone cement.

    Device Description

    The Repicci II™ Unicondylar Porous Knee is the same as the predicate, the Repicci III™ Unicondylar Knee (K971938), except for the additions of an intermediate sized femoral component -- 51mm and Ti-6Al-4V plasma spray porous coating to the interior of the components.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (Repicci II™ Unicondylar Knee System) and does not contain information about acceptance criteria or a study proving the device meets specific performance criteria.

    Instead, it's a submission for substantial equivalence to a legally marketed predicate device. This means the manufacturer is asserting their device is as safe and effective as a device already on the market, rather than presenting novel performance data against predefined criteria.

    Here's why the requested information cannot be extracted from this document:

    1. No Acceptance Criteria or Reported Device Performance: The document doesn't define any specific performance metrics (e.g., accuracy, sensitivity, specificity, mechanical thresholds) that the device needs to meet, nor does it report any such performance data for the Repicci II™ Unicondylar Knee System.
    2. No Clinical Testing Provided as a Basis for Substantial Equivalence: The document explicitly states: "Clinical Testing: No clinical testing was provided as a basis for substantial equivalence."
    3. Non-Clinical Testing was a Finite Element Analysis (FEA): The non-clinical testing mentioned was an FEA combined with engineering justification to demonstrate no new risks compared to the predicate device. This is a computational analysis, not a study involving a test set, ground truth, or human readers.

    Therefore, I cannot populate the table or answer the specific questions about sample size, data provenance, experts, adjudication, MRMC studies, standalone performance, or training set details because this type of information is not present in the provided text.

    The core of this submission is establishing substantial equivalence to a predicate device (Repicci II™ Unicondylar Knee, K971938) based on similarities in materials, design, sizing, and indications for use, with the primary difference being an added intermediate sized femoral component and a porous coating, which were justified through non-clinical FEA.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011795

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-06-13

    (5 days)

    Product Code
    Regulation Number
    888.3530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K971938

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Partial replacement of articulating surfaces of the knee when only one side of the joint is affected due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.

    The devices covered in this 510(k) are intended to be used with Repicci™ Femoral Components.

    The device is a single use implant intended for implantation w/ bone cement.

    Device Description

    The Repicci IITM Unicondylar Knee System consists of a femoral and tibial component. The Worland tibial bearing is used with the Repicci II™ femoral components (K971938).

    The all polyethylene tibial components are anatomical in geometry with right and left medial/lateral allocations. Located on the inferior surface of the implant is a patterned face with holes and a fin.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the "Worland Unicondylar Tibial Bearing." This type of submission is for demonstrating substantial equivalence to a predicate device, not for a novel device requiring extensive clinical trials to prove efficacy against acceptance criteria. Therefore, most of the requested information (acceptance criteria, specific study details, ground truth, expert involvement, MRMC studies, etc.) is not applicable or available in this document.

    Here's a breakdown based on the information provided:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of performance metrics. The underlying acceptance criteria for a 510(k) submission is "substantial equivalence" to a predicate device. This means the device is as safe and effective as a legally marketed device.
    • Reported Device Performance:
      • Non-Clinical Testing: "Fatigue testing and stress analysis determined that the Worland Unicondylar Tibular components presented no new risks and were; therefore, substantially equivalent to the predicate device."
      • Clinical Testing: "None provided as a basis for substantial equivalence."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Not applicable. No clinical test set or data provenance is mentioned as substantial equivalence was based on non-clinical testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not applicable. No clinical test set or ground truth was established. The "ground truth" for substantial equivalence was the predicate device's established safety and effectiveness.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set requiring adjudication was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-based device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm or AI-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the purpose of this 510(k) submission, the "ground truth" or benchmark for safety and effectiveness was the predicate device (Repicci II™ Unicondylar All Poly Tibial Bearing - K980665). The new device was deemed substantially equivalent based on similarities in materials, design, sizing, indications, and non-clinical testing results.

    8. The sample size for the training set

    • Not applicable. This is not a machine learning device, and no training set was used.

    9. How the ground truth for the training set was established

    • Not applicable. No training set was used.

    In summary, this document describes a device (a unicondylar tibial bearing) seeking market clearance through the 510(k) pathway, which primarily relies on demonstrating substantial equivalence to an existing predicate device rather than novel clinical performance studies against specific acceptance criteria. The equivalence was established through non-clinical fatigue testing and stress analysis, showing no new risks compared to the predicate.