    Intended Use

    The SI Handheld Gamma Finder is indicated to detect and quantify the gamma photons in the body intra-operatively or extra-operatively, when gamma emitting radiopharmaceuticals are administered into the human body.

    Device Description

    The gamma-detector, energy discriminator, amplifier, controls and energy supply have been assembled into a single hand-held unit SI-Handheld Gamma Finder® (HGF), which is shielded by a biocompatible polymer. The probe's shape is conical; its shaft tapers off towards a tip that holds the collimator, the crystal and the PINdiode while a head of larger diameter houses the display, menu buttons, and other components. The device is held in one hand by the middle of the shaft. Outer diameters and weight are designed so as to allow single-handed use over periods of time longer than those of routine sentinel biopsies. The polymer casing of SI HGF is water resistant and thus allows extended cleansing with commercial surface disinfectants. The probe tip may even be immersed in disinfectants for several hours. This fulfills standard requirements of hospital hygiene. Disinfection of the probe is comparable to other, larger probes that are being used like Gamma Finder in conjunction with disposable sterile plastic covers. The HGF can be switched on or off by any key. Results are numerically shown in the liquid crystal display in counts per second (CPS), and updated every second. The HGF has only four keys. The currently measured value can be retained in the display for 5 seconds by pressing Key A. The HGF is packaged and transported in a small suitcase. The HGF can only be operated by a Sonnenschein Lithium Battery (SL2770/T), with a maximum charging current of 15 mA. It is not user replaceable. The display of the HGF will display an battery symbol, warning the user the battery life is approximately 20 hours from exhaustion. If the battery is exhausted, the display will show "LobAt". The only accessory is a Sterile Sleeve (No. SI-FOL-101) that accompanies each device, along with the User Manual.

    AI/ML Overview

    Here's an analysis of the provided text regarding the "SI-Handheld Gamma Finder (HGF)" acceptance criteria and study, structured to answer your specific questions:

    Note: The provided document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed clinical study results for novel acceptance criteria. As such, some of your requested information (e.g., specific acceptance criteria for performance metrics, detailed sample sizes for training sets, MRMC study details) is not explicitly present in the document. Where information is not available, it will be noted.


    Acceptance Criteria and Device Performance

    The document does not explicitly state numerical acceptance criteria for a novel device, as it is a 510(k) submission seeking substantial equivalence to a predicate device (Neoprobe Model 1500). The "study" mentioned refers to verification testing to ensure the HGF performs comparably and meets safety standards, rather than proving performance against specific acceptance metrics in a clinical trial.

    Acceptance Criteria CategoryReported Device Performance (HGF)
    Energetic Range Detection140 keV to 360 keV (e.g., Tc-99M and I-131)
    Operational ControlSingle-handed use, four-key operation, display of counts per second (CPS), acoustic signal ON/OFF, LED ON/OFF, menu mode (dynamics, volume).
    Battery Life WarningBattery symbol displayed; "LobAt" when exhausted. Warns approximately 20 hours from exhaustion.
    DisinfectionWater-resistant polymer casing; allows extended cleansing with commercial surface disinfectants. Probe tip can be immersed for several hours.
    Dimensions/WeightDesigned for single-handed use over prolonged periods (compared to routine sentinel biopsies). Specific dimensions/weight not given, but implied to be comparable to predicate for this use case.
    AccessoriesCompatible with Sterile Sleeve (No. SI-FOL-101).
    Safety StandardsTested to appropriate electrical standards, biocompatibility, and sterilization standards.

    Study Details

    The document refers to "verification testing conducted by a university as to both system and technical evaluations" rather than a single, comprehensive "study" with a defined clinical endpoint.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated. The verification testing seems to be internal technical and system evaluations, not a clinical trial with patient-specific "data provenance."
      • Data Provenance: The verification testing was conducted by "a university." The country of origin is not explicitly stated for the testing, but the company is based in Berlin, Germany. The nature of the testing (technical/system evaluation) suggests it involved devices and possibly radioactive sources, not patient data in the sense of retrospective or prospective clinical data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable as the verification testing described is technical/system evaluation, not a clinical study requiring expert ground truth for diagnostic accuracy.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. The reported testing is technical verification, not a clinical study requiring adjudication of expert interpretations.
    4. If a multi-reader multicase (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is a gamma detection probe, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the device itself is a standalone measurement instrument. Its performance is evaluated on its ability to detect and quantify gamma radiation, independently of human interpretation of a complex algorithm's output. The "verification testing" would assess this standalone performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the technical verification testing, the "ground truth" would likely be established through calibrated radioactive sources of known energy and activity, and adherence to relevant electrical, biocompatibility, and sterilization standards. It is not expert consensus, pathology, or outcomes data related to patient diagnosis.
    7. The sample size for the training set:

      • Not applicable. This device is a hardware instrument with firmware, not a machine learning model that requires a "training set" in the conventional AI sense.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reasons as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K961172

    Validate with FDA (Live)

    Date Cleared
    1996-04-23

    (29 days)

    Product Code
    Regulation Number
    862.1370
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K971167

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DSL IGF-I IRMA assay is intended for the quantitative determination of IGF-I in human serum or EDTA plasma. The measurement of serum/plasma IGF-I is used as a diagnostic aid in the evaluation of growth status disorders.

    Device Description

    The DSL Active IGF-I IRMA kit was developed for the quantitative measurement of Insulin-like Growth Factor-I in human serum and plasma. This IRMA format is a capture assay. Mouse monoclonal antibody to IGF-I is immobilized to the inner surface of the coated test tubes. IGF-I in the standards or samples is "sandwiched" between this monoclonal and the goat-anti-IGF-I polyclonal antibody labelled with lodine-125.

    AI/ML Overview

    The provided document describes a medical device, the "DSL 2800 IGF-I IRMA Kit," for the quantitative measurement of Insulin-like Growth Factor-I (IGF-I). It details the device's intended use and compares it to a substantially equivalent device, the "DSL 5600 IGF-I IRMA."

    However, the provided text does not contain information related to acceptance criteria, detailed study designs, expert involvement, or any aspects of AI/algorithm performance. The document is a summary of safety and effectiveness, focusing on demonstrating substantial equivalence between two kits by comparing their measurement results on patient samples.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or answer most of the requested questions because the necessary information is not present in the input text.

    Here's what I can extract and what is missing based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: Not explicitly stated in the document.
    • Reported Device Performance:
      • Comparison with DSL 5600 IGF-I IRMA: Y = 1.19(X) - 22.07 (Linear regression equation)
      • Correlation Coefficient (r): 0.98

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: 146 patient samples.
    • Data Provenance:
      • Country of Origin: Not specified.
      • Retrospective or Prospective: Not specified, but samples were "collected and assayed simultaneously by both methods," suggesting a prospective collection or at least simultaneous testing of existing samples. The selection based on "expected IGF-I levels" implies a targeted, rather than purely random, selection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. This is a measurement assay comparison, not an expert-driven diagnostic interpretation. "Ground truth" in this context is established by the reference assay (DSL 5600 IGF-I IRMA) on the same samples.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. There's no adjudication process described as it's a quantitative measurement comparison, not an interpretation task.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This document describes a comparison between two lab assay kits, not an AI system or human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a diagnostic assay kit, not an algorithm or AI system. Its performance is inherent to the kit's chemical and immunological reactions.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the purpose of demonstrating substantial equivalence, the "ground truth" was defined by the results obtained from the DSL 5600 IGF-I IRMA kit on the same patient samples. This kit is presented as the substantially equivalent device to which the new device (DSL 2800 IGF-I IRMA Kit) is being compared.

    8. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI device that requires a training set. The assay's performance is based on its chemical and immunological design.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set for an AI model is described.