    Reference Devices :

    K970718

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nichols Advantage® Chemiluminescence Soluble Transferrin Receptor Immunoassay is intended for use on the Nichols Advantage® Specialty System for the quantitative determination of Soluble Transferrin Receptor in human serum, EDTA and Heparin plasma as an adjunct in the diagnosis of Iron Deficiency Anaemia and for the differential diagnosis of Iron Deficiency Anaemia and Anaemia of Chronic Disease.

    Device Description

    The Nichols Advantage® Soluble Transferrin Receptor Assay is a two-site chemiluminescence assay for use with the Nichols Advantage® Specialty System.

    AI/ML Overview

    The provided text describes the 510(k) notification for the Nichols Advantage® Chemiluminescence Soluble Transferrin Receptor Immunoassay and its comparison to a predicate device. However, it does not include specific acceptance criteria, performance data, or details of a study designed to prove the device meets acceptance criteria. The document focuses on regulatory approval based on substantial equivalence.

    Therefore, I cannot fulfill most of your request as the information is not present in the provided text.

    Here's what can be extracted based on the information provided:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the text. The document doesn't state specific statistical acceptance criteria for accuracy, precision, or other performance metrics for the Nichols Advantage® assay. It only compares its characteristics (e.g., sensitivity, sample size, incubation) to the predicate device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the text. There is no mention of a test set, sample size, or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the text. There is no mention of experts or ground truth establishment for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not provided in the text. This device is an in-vitro diagnostic immunoassay, not an AI-assisted diagnostic tool for human readers, so an MRMC study in this context would not be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a standalone immunoassay system. Its performance is determined by the assay itself, not an algorithm that assists a human. The text describes the device's characteristics and its equivalence to a predicate device, which implies standalone performance was evaluated for regulatory purposes. However, specific standalone performance data (e.g., accuracy, precision) against defined acceptance criteria is not provided. The "Sensitivity" listed in the table is a device characteristic, not necessarily a performance outcome against a clinical ground truth.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided in the text. For an immunoassay, the "ground truth" would typically refer to a gold standard method for measuring soluble transferrin receptor or clinical diagnoses against which the assay's results are validated. This is not detailed in the provided summary.

    8. The sample size for the training set

    This information is not provided in the text. Training sets are typically relevant for machine learning algorithms or for initial assay development/optimization. This document does not describe such studies.

    9. How the ground truth for the training set was established

    This information is not provided in the text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991157

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-08-02

    (117 days)

    Product Code
    Regulation Number
    866.5880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K970718

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The N Latex sTfR Reagent is an in vitro diagnostic reagent for the quantitative determination of sTfR in human serum or heparinized plasma using the Behring Nephelometer Systems (BNA, BN100, & BNII), and aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

    Device Description

    N Latex sTfR Reagent is intended to be used together with the Behring Nephelometer Systems for the quantitative determination of soluble transferrin receptor (sTfR) in human serum or heparinized human plasma.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Dade Behring Inc. N Latex sTfR Reagent, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary for the N Latex sTfR Reagent primarily focuses on establishing substantial equivalence to a predicate device rather than defining new, explicit acceptance criteria in the traditional sense. The performance characteristics described are used to demonstrate that the device performs comparably to the predicate. Therefore, the "acceptance criteria" can be inferred from what the manufacturer chose to report as sufficient for equivalence.

    Performance CharacteristicAcceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (N Latex sTfR Reagent)
    CorrelationStrong positive correlation (e.g., r > 0.90) with the predicate device.r = 0.96 with a commercially available sTfR assay.
    Slope (Correlation)Close to 1.0 (indicating similar overall scale).0.81
    Y-intercept (Correlation)Close to 0 (indicating similar bias at low concentrations).-0.19
    Inter-assay PrecisionLow variability (e.g., < 5% CV).0.8% to 1.2%
    Intra-assay PrecisionLow variability (e.g., < 5% CV).1.4% to 2.1%

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: 62 samples
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the study involved comparing the N Latex sTfR Reagent to a "commercially available sTfR assay," implying human serum/plasma samples. The type of samples (ranging from 0.41 to 3.5 mg/l) suggests clinical or laboratory-collected samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given text. For this type of in vitro diagnostic device, "ground truth" for the test set would typically be established by the results from the predicate device (Quantikine IVD human sTfR assay) or a reference method. The text does not mention any expert review of these results for the purpose of establishing ground truth for the comparison.

    4. Adjudication Method for the Test Set

    This information is not specified in the provided text. Adjudication methods (like 2+1, 3+1) are usually relevant when expert reviewers are determining a diagnosis or outcome from images or complex clinical data. For a quantitative assay comparison, the "ground truth" is typically the measurement from the reference method or predicate, and adjudication in the traditional sense is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size of human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic reagent, not an AI-powered diagnostic imaging tool. Therefore, a study involving human readers and AI assistance for interpreting results is not applicable.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done. The study directly assessed the performance of the N Latex sTfR Reagent itself (the "algorithm only") by comparing its measurements to those of a predicate device and evaluating its precision. This "standalone" assessment of the device's analytical performance is the core of the submission. The "human-in-the-loop" component for this type of device would involve laboratory technicians performing the test and clinicians interpreting the results, but the study itself focuses on the device's inherent measurement capabilities.

    7. The Type of Ground Truth Used

    The ground truth for the correlation study was established by the measurements obtained from the predicate device (or a "commercially available sTfR assay"). For precision studies, the ground truth refers to the expected value of the control materials and human serum pools used, which would have established ranges or target values based on prior validation.

    8. The Sample Size for the Training Set

    This information is not provided in the text. This type of device (a reagent for an immunoassay) typically does not involve a "training set" in the machine learning sense. The performance characteristics (correlation, precision) are determined through analytical validation studies, not by training an algorithm on a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set in the machine learning sense for this device, this question is not applicable. The device's formulation and operational parameters would be developed through R&D, and its performance validated through studies like those described, rather than "trained."