510(k) Data Aggregation

    K Number
    K142658

    Validate with FDA (Live)

    Date Cleared
    2014-11-13

    (56 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K062352, K971069, K962233, K082320, K070039

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biomet Headless Compression Screws and Twist-Off Screws are indicated for fixation of a joint (arthrodesis) or bone reconstruction (osteotomy) of the mid-foot bones, metatarsal and phalanges of the foot or the phalanges, metacarpals and carpals of the hand. In the foot, these include procedures to correct Hallux Valgus (bunions), Hallux Varus and Hallux Rigidus, Hammer toe, Claw toe and Mallet toe.

    Device Description

    The Biomet Headless Compression and Twist-Off Screws consist of bone screws of various lengths and diameters. The Biomet Headless Compression and Twist-Off Screws are also accompanied by dimensionally optimized corresponding instruments which are used to aid in the alignment and stabilization of fractures to the skeletal system. The Biomet Headless Compression Screws are a cannulated headless screw, which is inserted below the bone surface. The Biomet Headless Compression Screw is designed to minimize soft tissue irritation and provides compression due to a dual thread design. The Biomet Twist-Off Screw is a solid one piece screw that has a direct connection to a drill or large diameter pin driver, this allows the screw to break-off cleanly upon contact. The Biomet Twist-Off Screw also has compression capabilities with a thread-free segment that achieves compression at the osteotomy site.

    AI/ML Overview

    This document does not describe a study that proves the device meets specific acceptance criteria in the way you've outlined for an AI/algorithm-based device.

    The provided text is a 510(k) premarket notification summary for Biomet Headless Compression and Twist-Off Screws, which are physical medical devices (bone screws), not an AI algorithm or software. Therefore, many of the requested categories (like MRMC studies, ground truth for training/test sets, expert adjudication, AI improvement effect size) are not applicable to this type of device submission.

    However, I can extract the relevant information from the document regarding the device's performance assessment:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (from ASTM F543)Reported Device Performance
    Torsional properties (e.g., torque to failure)Met requirements, found substantially equivalent to predicate devices
    Axial pullout strengthMet requirements, found substantially equivalent to predicate devices
    Driving torqueMet requirements, found substantially equivalent to predicate devices

    Note: The document states "Results indicate that the subject screws are substantially equivalent to legally marketed devices offering a reasonable assurance of safety and effectiveness." It does not provide specific numerical values for the performance criteria, only that they met the standards of ASTM F543 and were comparable to predicate devices.

    The following questions are NOT APPLICABLE to this document as it pertains to a physical medical device (bone screws) and not an AI or algorithm-based device:

    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective) - N/A (mechanical testing on screws themselves)
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts - N/A
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set - N/A
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance - N/A
    • 6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done - N/A
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc) - N/A (Ground truth for mechanical testing is typically the physical measurement itself against established standards).
    • 8. The sample size for the training set - N/A
    • 9. How the ground truth for the training set was established - N/A
    Ask a Question

    Ask a specific question about this device

    K Number
    K971069

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-07-09

    (107 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K962233

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fixation of fracture, fusion of a joint or bone reconstruction (osteotomy) of the phalanges, metacarpals and carpals of the hand.

    Device Description

    The Twist off Screw is made of titanium alloy and consists of a screw integrated with separates from the screw when the head meets cortical bone. It is 2.0mm in diameter and 11 to 14mm in length and has a selftapping tip.

    AI/ML Overview

    This document describes the safety and effectiveness of the Medinov-AMP Twist-off™ Screw (K971069).

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Maximum torque at rupture for separation of holder and screwAverages 0.17 Nm
    Maximum torque at rupture of the Twist-off™ ScrewAverages 0.37 Nm

    Note: The document does not explicitly state "acceptance criteria" through specific threshold values or ranges. Instead, it reports the average performance values obtained from testing. The implication is that these reported values were deemed acceptable for market clearance.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not explicitly stated in the provided text, only "Testing of the Twist-off Screw was performed for the K962233." This indicates that tests were conducted, but the number of screw units tested is not specified.
    • Data Provenance: Not specified. The manufacturer (Medinov-AMP) is located in Roanne, France, but the location where the testing was performed is not mentioned. The data is retrospective as the testing was completed prior to the 510(k) submission.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. The ground truth for this device's performance relies on physical testing of mechanical properties (torque at rupture), not on expert opinion or interpretation of images/data.

    4. Adjudication Method for Test Set

    Not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in studies involving subjective interpretation (e.g., image reading by multiple experts). The testing performed here involved objective mechanical measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic tools where human readers interpret images or data, and the device aims to assist or replace human performance. The Twist-off™ Screw is a physical implant, not a diagnostic or AI-driven device.

    6. Standalone Performance Study

    Yes, a standalone study was performed in the sense that the mechanical properties of the device (torque at rupture) were tested independently. The reported values (0.17 Nm and 0.37 Nm) are the standalone performance of the screw under specific test conditions. This is a direct measurement of the device's physical capabilities, not an algorithm's performance.

    7. Type of Ground Truth Used

    The ground truth used was physical mechanical measurement. Specifically, the torque required to induce rupture, both for the separation of the holder and for the screw itself, was directly measured during testing.

    8. Sample Size for the Training Set

    Not applicable. This device is a mechanical implant, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for a mechanical implant device.