Search Results
Found 5 results
510(k) Data Aggregation
(435 days)
Ultra Phonic® Free Conductivity Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic ultrasound imaging procedures. It is placed on the patient's skin or on the transducer prior to initiating an ultrasound examination.
Ultra/Phonic® Free Conductivity Gel is an aqueous, non-sterile ultrasound conducting medium for use during non-invasive ultrasonic medical procedures. It is placed between the ultrasound transducer and the patient, eliminating air at the contact surface while providing lubrication. This provides an acoustic pathway between the transducer and the skin. This uninterrupted pathway allows the transducer to send sound waves which enter the body, and to receive echoes from tissues and structures without unexpected interference. When the echoes are received by the scanner, it analyzes them to determine how strong they are, and how long they took to be received after transmission. That information is then processed and outputted to a display as an image which is used for diagnostic and therapeutic purposes by health care professionals.
It is packaged in 5L Cubitainers and 250 mL bottles.
Major characteristics include:
- Non-sensitizing, non-irritating
- Water soluble, non-staining, and easily cleanable
- Does not contain oil or fatty matter
- Does not damage the probe
- pH level is 6.25 ± 0.25 @ 25°C
- Resists thinning and decomposition when exposed to perspiration and body heat
- Sheer thinning gel is easy to apply, but doesn't run or drip
- High viscosity allows for thick application, adapting to contours and hair to minimize interference
This document describes the premarket notification (510(k)) for the Ultra/Phonic® Free Conductivity Gel, a diagnostic ultrasonic transducer accessory. The focus of the provided text is on demonstrating the substantial equivalence of the new device to existing predicate devices, rather than a detailed algorithm performance study for an AI-powered device. Therefore, many of the requested elements for AI/algorithm performance (e.g., test set sample size, ground truth details, MRMC studies) are not applicable or not present in this document.
Here's the information that can be extracted and a clear indication of what is not applicable based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the properties of the predicate and reference devices, as the goal is to demonstrate "substantial equivalence." The reported performance focuses on acoustic properties and biocompatibility.
| Property | Acceptance Criteria (Predicate/Reference) | Reported Performance (Ultra/Phonic® Free Conductivity Gel) | Met/Not Met |
|---|---|---|---|
| Acoustic Properties | Similar to Sonishield™ 100 & Ecogel 100 | Virtually identical to human skin; Similar to other coupling gels | Met |
| Sound velocity (m/sec) | Sonishield™ 100: 1497 ms-1 at 30°C | 1488 ms-1 at 22.5°C | Met (considered virtually identical despite temp diff) |
| Density (kg/m3) | Sonishield™ 100: 1023 kg/ m3 at 30°C | 1018 kg/ m3 at 22.5°C | Met (considered virtually identical despite temp diff) |
| Acoustic impedance (kg/m2 sec) | Sonishield™ 100: 1.53 MRayls at 30°C | 1.51 MRayls at 22.5°C | Met (considered virtually identical despite temp diff) |
| Attenuation coefficient (a/f) | Sonishield™ 100: 0.04 ± 0.0042 f | 0.0185 + 0.01335 $f^{1.0894}$ | Met (considered virtually identical) |
| Biocompatibility | Non-sensitizing, non-irritating | Non-sensitizing, non-irritating | Met |
| USP <51> Antimicrobial Effectiveness Test | Log reductions > 5 for bacterial, no increase yeast/molds in 14/28 days | Log reductions > 5 for all 14 and 28 day bacterial counts; yeasts and molds decreased | Met |
| In Vitro Cytotoxicity (ISO 10993-5:2009) | Pass | Not met (considered cytotoxic) | Not Met (but mitigated by RIPT) |
| Human-Repeated-Insult-Patch-Test (RIPT) | No potential for dermal irritation/allergic sensitization | No potential for dermal irritation or allergic contact sensitization | Met (mitigated cytotoxicity concern) |
| Physical/Chemical Properties | (Based on comparison table with predicates) | (See detailed characteristics in comparison table) | Met |
| Ingredients (Salt free, Dye free, Alcohol free, Formaldehyde free, Perfume Free) | Similar to Sonishield™ 100 (Ecogel had green coloring) | All free | Met |
| pH | Sonishield™ 100: 4.5 – 6.5; Ecogel 100: 6.5 ± 0.75 | 6.25 ± 0.25 @ 25°C | Met |
| Viscosity (CPS) | Sonishield™ 100: 80,000 – 120,000; Ecogel 100: 35,000 - 40,000 | 600,000 ± 300,000 | Met (difference acknowledged, but deemed equivalent in function) |
| Boiling Point | Sonishield™ 100: > 200°C; Ecogel 100: 100°C | > 200°C | Met |
2. Sample size used for the test set and the data provenance:
- Acoustic performance: The document states that the acoustic properties of the gel were evaluated and compared to human skin and other coupling gels. No specific sample size (number of gels tested, or number of measurements) is provided for this evaluation. The data provenance is implied to be laboratory testing.
- Antimicrobial Effectiveness Test: The test followed USP <51> (Category 2) standards, involving specific bacterial and fungal strains (Staphylococcus aureus ATCC 6538, Escherichia coli ATCC 8739, Pseudomonas aeruginosa ATCC 9027, Aspergillus brasiliensis ATCC 16404, Candida albicans ATCC 10231). The sample size typically refers to the number of replicates per microbial challenge, which are standard for this test method.
- Biocompatibility (Skin Irritation, Skin Sensitization, Cytotoxicity): These refer to ISO 10993-10 and ISO 10993-5:2009 standards. These are in vitro or animal tests that follow specific protocols for sample sizes of cells or animals. The document does not specify the exact number of animals or cell cultures used but confirms that tests were conducted according to the stated ISO standards.
- Human-Repeated-Insult-Patch-Test (RIPT): This is a human clinical study. The document does not specify the number of human subjects used but indicates that such testing was performed. Data provenance is likely the testing laboratory.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. This device is a medical accessory (ultrasound gel), not an AI-powered diagnostic algorithm requiring expert "ground truth" for classification. The "truth" is established by physical/chemical measurements and biological assays.
4. Adjudication method for the test set:
Not applicable (refer to point 3).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is not an AI-powered diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI-powered diagnostic device.
7. The type of ground truth used:
The "ground truth" for this device revolves around its physical, chemical, and biological properties:
- Acoustic Properties: Measured values (sound velocity, density, acoustic impedance, attenuation coefficient).
- Biocompatibility: Results of standardized in vitro and in vivo tests adhering to ISO 10993 guidelines, and the USP <51> antimicrobial test.
- Functional Properties: pH, viscosity, ingredients, etc., measured via standard laboratory methods.
8. The sample size for the training set:
Not applicable. This is not an AI-powered device that requires a training set.
9. How the ground truth for the training set was established:
Not applicable (refer to point 8).
Ask a specific question about this device
(435 days)
UltraPhonic® Scanning Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic and diagnostic ultrasound imaging procedures. It is placed on the patient's skin or on the transducer prior to initiating an ultrasound examination. The product is indicated for prescription use only.
Ultra/Phonic® Scanning Gel is an aqueous, non-sterile ultrasound conducting medium for use during non-invasive ultrasonic medical procedures. It is placed between the ultrasound transducer and the patient, eliminating air at the contact surface while providing lubrication. This provides an acoustic pathway between the transducer and the skin. This uninterrupted pathway allows the transducer to send sound waves which enter the body, and to receive echoes from tissues and structures without unexpected interference. When the echoes are received by the scanner, it analyzes them to determine how strong they are, and how long they took to be received after transmission. That information is then processed and outputted to a display as an image which is used for diagnostic and therapeutic purposes by health care professionals. It is packaged in 4L Cubitainers, 5L Cubitainers, and 250 mL bottles.
Here's a summary of the acceptance criteria and the study details for the Ultra/Phonic® Scanning Gel, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Ultra/Phonic® Scanning Gel Performance |
|---|---|
| Acoustic Properties: - Sound velocity - Density - Acoustic impedance - Attenuation coefficient as a function of frequency | - Sound velocity: 1590 ms-1 at 22.5°C - Density: 1035 kg/m3 at 22.5°C - Acoustic impedance: 1.65 MRayls at 22.5°C - Attenuation coefficient (a/f): 0.06 + 0.01116 f^1.3706 Conclusion: "Virtually identical to that of human skin" and "Similar to other coupling gels commonly used in the United States." Acoustic properties of the predicate gel (Sonishield™ 100) are also provided for comparison and are considered "virtually identical." |
| Antimicrobial Effectiveness (USP <51> Category 2) | - Log reductions > 5 for bacterial counts (Staphylococcus aureus, Escherichia coli, Pseudomonas aeruginosa) at 14 and 28 days. - Yeasts and molds (Aspergillus brasiliensis, Candida albicans) did not increase, and "considerably decreased." |
| Biocompatibility (ISO 10993-10): - Skin Irritation - Skin Sensitization | - Non-sensitizing - Non-irritating |
| In Vitro Cytotoxicity (ISO 10993-5:2009) | - Meets the requirements of the test and is "not considered cytotoxic." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state sample sizes for each specific test (acoustic, antimicrobial, biocompatibility, cytotoxicity). It refers to "evaluations," "tests conducted," and "studies." The provenance of the data is not specified in terms of country of origin, nor is it explicitly stated whether the studies were retrospective or prospective, though performance testing is generally prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
This information is not provided in the document. The studies performed are non-clinical (laboratory and animal) and do not involve human interpretation or subjective expert ground truth establishment in the traditional sense of diagnostic AI studies.
4. Adjudication Method for the Test Set
Not applicable, as this device does not involve subjective diagnostic interpretations requiring adjudication. The performance criteria are based on objective physical, chemical, and biological measurements.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (ultrasound scanning gel) and not an AI-powered diagnostic tool. Therefore, an MRMC study and evaluation of human reader improvement with AI assistance are irrelevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm or AI model. Its performance is evaluated through direct physical, chemical, and biological testing, not through standalone algorithm performance.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by recognized scientific and regulatory standards and validated laboratory methods.
- Acoustic Properties: Comparative measurements against human skin properties and other commercially available gels.
- Antimicrobial Effectiveness: Defined log reduction criteria established by USP <51>.
- Biocompatibility: Absence of irritation and sensitization as per ISO 10993-10.
- Cytotoxicity: Meeting negative control requirements and absence of cytotoxic effects as per ISO 10993-5:2009.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. This is a physical medical device, not an AI model.
Ask a specific question about this device
(434 days)
UltraPhonic® Conductivity Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic and diagnostic ultrasound imaging procedures. It is placed on the patient's skin or on the transducer prior to initiating an ultrasound examination.
Ultra/Phonic® Conductivity Gel is an aqueous, non-sterile ultrasound conducting medium for use during non-invasive ultrasonic medical procedures. It is placed between the ultrasound transducer and the patient, eliminating air at the contact surface while providing lubrication. This provides an acoustic pathway between the transducer and the skin. This uninterrupted pathway allows the transducer to send sound waves which enter the body, and to receive echoes from tissues and structures without unexpected interference. When the echoes are received by the scanner, it analyzes them to determine how strong they are, and how long they took to be received after transmission. That information is then processed and outputted to a display as an image which is used for diagnostic and therapeutic purposes by health care professionals. It is packaged in 4L Cubitainers, 5L Cubitainers, 250 mL bottles, and 250 mL tubes.
The request asks for specific details about the acceptance criteria and the study proving the device meets them. However, the provided document describes an ultrasonic conductivity gel, not an AI/ML device, and therefore does not contain information typically found in a study for AI/ML device performance.
Therefore, most of the requested fields cannot be filled from the provided text as they are not applicable to the type of device described.
Here's a breakdown of what can and cannot be extracted:
Device Type: Ultrasonic Conductivity Gel (not an AI/ML device)
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Acoustic properties "virtually identical to human skin" | Sound velocity: 1594.2 ms-1 at 22.5°C (compared to predicate's 1497 ms-1 at 30°C) |
| Acoustic properties "similar to other coupling gels" | Density: 1061 kg/m3 at 22.5°C (compared to predicate's 1023 kg/m3 at 30°C) |
| Acoustic impedance: 1.69 MRayls at 22.5°C (compared to predicate's 1.53 MRayls at 30°C) | |
| Attenuation coefficient a/f (dB/(cm-MHz)): 0.05 + 0.01148 f^1.213 (compared to predicate's 0.04 ± 0.0042 f) | |
| Antimicrobial effectiveness (USP <51> Category 2) | Passed with log reductions greater than 5 for bacterial counts (Staphylococcus aureus ATCC 6538, Escherichia coli ATCC 8739, Pseudomonas aeruginosa ATCC 9027) at 14 and 28 days. Yeasts and molds (Aspergillus brasiliensis ATCC 16404, Candida albicans ATCC 10231) did not increase. |
| Non-sensitizing (ISO 10993-10) | Found to be non-sensitizing. |
| Non-irritating (ISO 10993-10) | Found to be non-irritating. |
| Not cytotoxic (ISO 10993-5:2009) | Meets the requirement of the test and is not considered cytotoxic. |
| pH level (Product Characteristic) | 6.25 ± 0.25 @ 25°C (compared to predicate Sonishield™ 100: 4.5 - 6.5; Ecogel 100: 6.5 ± 0.75) |
| Viscosity (Product Characteristic) | 600,000 ± 300,000 CPS (compared to predicate Sonishield™ 100: 80,000 - 120,000 CPS; Ecogel 100: 35,000 - 40,000 CPS) |
| Density (Product Characteristic) | 1.061 g/mL (compared to predicate Sonishield™ 100: 1.009 g/mL; Ecogel 100: 0.99 g/mL) |
| Boiling point (Product Characteristic) | > 200°C (compared to predicate Sonishield™ 100: > 200°C; Ecogel 100: 100°C) |
| Other characteristics (non-sensitizing, non-irritating, etc.) | Non-sensitizing, non-irritating, water-soluble, non-staining, easily cleanable, no oil or fatty matter, no odor, does not damage the probe, resists thinning, shear-thinning gel, high viscosity for thick application. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not applicable. The performance data for this product is based on physicochemical and biological testing in a laboratory setting, not on a test set of patient data.
- Data Provenance: The acoustic testing used the Ultra/Phonic® Conductivity Gel and the predicate Sonishield™ 100 Antimicrobial Ultrasound Gel. Biocompatibility and cytotoxicity tests were conducted on the Ultra/Phonic® Conductivity Gel. The document does not specify the country of origin for the testing data. The studies are non-clinical performance evaluations.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. This is not an AI/ML device that requires expert ground truth labeling.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable in the context of AI/ML. The "ground truth" for this device's performance is based on established laboratory testing standards (e.g., USP <51> for antimicrobial effectiveness, ISO 10993 for biocompatibility and cytotoxicity) and physical property measurements.
8. The sample size for the training set
- Not applicable.
9. How the ground truth for the training set was established
- Not applicable.
Ask a specific question about this device
(434 days)
Other-Sonic™ Transmission Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic and diagnostic ultrasound imaging procedures. It is placed on the patient's skin or on the transducer prior to initiating an ultrasound examination. The product is indicated for prescription use only.
Other-Sonic™ Transmission Gel is an aqueous, non-sterile ultrasound conducting medium for use during non-invasive ultrasonic medical procedures. It is placed between the ultrasound transducer and the patient, eliminating air at the contact surface while providing lubrication. This provides an acoustic pathway between the transducer and the skin. This uninterrupted pathway allows the transducer to send sound waves which enter the body, and to receive echoes from tissues and structures without unexpected interference. When the echoes are received by the scanner, it analyzes them to determine how strong they are, and how long they took to be received after transmission. That information is then processed and outputted to a display as an image which is used for diagnostic and therapeutic purposes by health care professionals. It is packaged in 5L Cubitainers and 250 mL bottles.
The device under review is "Other-Sonic™ Transmission Gel," which is a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic and diagnostic ultrasound imaging procedures. The device is compared to a predicate device, "Sonishield™ 100 Antimicrobial Ultrasound Gel," and an additional reference device, "Ecogel 100 Ultrasound Gel."
Here's the breakdown of the acceptance criteria and the study details:
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" for each performance metric but rather presents a comparison of technological and non-clinical characteristics between the subject device and the predicate/reference devices to demonstrate substantial equivalence. The "performance" column below reflects the characteristics of the "Other-Sonic™ Transmission Gel" and the "Acceptance Criteria" column indicates that its performance should be substantially equivalent or similar to the predicate/reference devices for these characteristics.
| Acceptance Criteria (Compared to Predicate/Reference) | Reported Device Performance (Other-Sonic™ Transmission Gel) |
|---|---|
| Acoustic Performance: | |
| Sound velocity similar to predicate | 1601 ms-1 at 22.5°C (Predicate: 1497 ms-1 at 30°C) |
| Density similar to predicate | 1033 kg/ m3 at 22.5°C (Predicate: 1023 kg/ m3 at 30°C) |
| Acoustic impedance similar to predicate | 1.65 MRayls at 22.5°C (Predicate: 1.53 MRayls at 30°C) |
| Attenuation coefficient similar to predicate | 0.19 + 0.0000105 $f^{2.974}$ (Predicate: 0.04 ± 0.0042 $f$) |
| Antimicrobial Effectiveness: | |
| Pass USP <51> (Category 2) log reductions | Passed with log reductions > 5 for bacteria, yeasts and molds decreased |
| Biocompatibility: | |
| Non-sensitizing | Non-sensitizing |
| Non-irritating | Non-irritating |
| In Vitro Cytotoxicity: | |
| Not considered cytotoxic (per ISO 10993-5:2009) | Not considered cytotoxic |
| Physical/Chemical Properties: | |
| Salt free | Salt free |
| Alcohol free | Alcohol free |
| Formaldehyde free | Formaldehyde free |
| pH range (comparable to predicate) | 6.80 ± 0.25 |
| Density (g/mL) (comparable to predicate) | 1.033 |
| Viscosity (comparable to predicate) | 966,000 ± 483,000 CPS |
| Boiling Point (comparable to predicate) | > 200° C |
| Water soluble high MW polymers | Water soluble high MW polymers |
| No irritation | No irritation |
| Other: | |
| External Use | External |
| Pediatric and adult Target Population | Pediatric and adult |
| Multiple Uses | Multiple uses |
| Hospital Environment of Use | Hospital |
2. Sample Size Used for the Test Set and Data Provenance:
- Acoustic Testing: The document does not specify a separate "test set" sample size for acoustic testing. The acoustic properties are reported as values for "Other-Sonic™ Transmission Gel" and "Sonishield™ 100 Antimicrobial Ultrasound Gel." It implicitly compares the characteristics of the gel rather than a study on a specific number of instances.
- Antimicrobial Effectiveness Testing: The sample size is not explicitly stated as a number of distinct gel samples. However, the test was performed according to USP <51> (Category 2) using standard microbial challenges: Staphylococcus aureus ATCC 6538, Escherichia coli ATCC 8739, Pseudomonas aeruginosa ATCC 9027, Aspergillus brasiliensis ATCC 16404, and Candida albicans ATCC 10231. These represent specific strains used in lab testing.
- Biocompatibility (Skin Irritation and Sensitization): The document states "Animal tests" were conducted, referring to "Skin Irritation and Skin Sensitization." The number of animals or specific experimental setup (e.g., number of patches, individual subjects) is not provided.
- In Vitro Cytotoxicity: The test was conducted according to ISO 10993-5:2009. The sample size refers to the cell cultures used in the in vitro test, but specific numbers are not given.
- Data Provenance: The document does not specify the country of origin for the non-clinical test data. These tests are typically conducted in laboratory settings (i.e., in vitro or in vivo animal studies) rather than using retrospective or prospective human clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
This section is not applicable. The evaluations conducted for this device (acoustic properties, antimicrobial effectiveness, biocompatibility, cytotoxicity) are non-clinical laboratory tests. They do not involve expert interpretation of images or clinical data to establish a "ground truth" in the way a diagnostic algorithm might. The "ground truth" for these tests comes from established scientific methodologies and standards (e.g., USP <51>, ISO 10993).
4. Adjudication Method for the Test Set:
This section is not applicable as the studies were non-clinical laboratory tests that do not involve human adjudication of results in the context of diagnostic interpretation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:
No MRMC comparative effectiveness study was done. This device is an ultrasound transmission gel, not an AI-powered diagnostic tool requiring human reader assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
No standalone (algorithm-only) performance was done because this is a physical medical device (ultrasound transmission gel), not an algorithm or AI software.
7. The Type of Ground Truth Used:
The "ground truth" for the performance claims comes from:
- Acoustic Properties: Direct physical measurements of the gel's sound velocity, density, acoustic impedance, and attenuation coefficient.
- Antimicrobial Effectiveness: Quantitative microbial count reductions as measured by standard microbiological assays (USP <51>).
- Biocompatibility: Results from standardized animal (in vivo) irritation and sensitization tests (ISO 10993-10).
- In Vitro Cytotoxicity: Results from standardized cell culture tests (ISO 10993-5:2009) to assess cell viability.
8. The Sample Size for the Training Set:
This section is not applicable. This device is a physical product (ultrasound transmission gel) and does not involve AI or machine learning algorithms that require a training set.
9. How the Ground Truth for the Training Set Was Established:
This section is not applicable as there is no training set for this type of device.
Ask a specific question about this device
(319 days)
Konix® Ultrasound Gel is intended for general use as a nonsterile transmission media for acoustically coupling a transducer to a human body surface during external, diagnostic ultrasound imaging procedures. It is placed on the patient's skin prior to initiating an ultrasound examination. It is indicated for prescription use only.
Konix® Ultrasound Gel consists of deionized water, carbomer, triethanolamine, monopropylene glycol, 5-chloro-2-methyl-4-isothiazolin-3-one and 2-methyl-4-isothiazolin-3 one and is a type of conductive medium, i.e., scanning gel, used in ultrasound diagnostic techniques. A scanning gel acts as a couplant that provides an acoustic pathway between the transducer and the skin. In addition, the gel eliminates air (a disruptive influence) from the interface and adapts the contours of the probe to the skin.
The major characteristics of Konix® Ultrasound Gel include:
- . Hypoallergenic, non-irritating
- Water soluble, non-staining and easily cleanable .
- Does not contain oil and fatty matter .
- Free from formaldehyde and salt .
- No toxic effects .
- Produced as a completely harmless material .
- No smell ◆
- Vacuum treated production .
- Not-damageable to the probe .
- . Does not contain air bubbles
- pH level is 7 .
The Konix® Ultrasound Gel is a medical device designed for general use as a non-sterile transmission medium for acoustically coupling a transducer to a human body surface during external, diagnostic ultrasound imaging procedures. The device's acceptance criteria and studies are primarily focused on demonstrating its acoustic performance and equivalence to existing predicate devices.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for Konix® Ultrasound Gel were based on its acoustic properties being virtually identical to human skin and other commonly used coupling gels in the United States.
| Acceptance Criteria / Characteristic | Target Value (based on human skin/optimal average) | Reported Device Performance (Konix® Ultrasound Gel) |
|---|---|---|
| Sound velocity (m/sec) at 30°C | 1,518 (human skin), 1,520 (optimal average) | 1,516 |
| Density (kg/m³) at 30°C | Not specified as clear target, but comparable | 0.98 x 10-3 |
| Acoustic impedance (kg/m² sec) at 30°C | 1.6 (human skin) | 1.49 |
| Attenuation coefficient as a function of frequency, a/f (dB/(cm-MHz)) | Not explicitly stated, but "low" is desired | < 0.05 |
2. Sample Size and Data Provenance
The document does not specify a separate "test set" and corresponding sample size for the acoustic performance evaluation. The acoustic properties (Sound velocity, Density, Acoustic impedance, Attenuation coefficient) were measured for the Konix® Ultrasound Gel.
The data provenance for the comparison values (e.g., human skin sound velocity and acoustic impedance) is from published scientific literature:
- Ogura, I., Kidikoro, T., linuma, K., Tanaka, K. and Matsuda, A. "Measurement of Acoustic Impedance of Skin" Ultrasound in Medicine, Vol. 4, RC 78.7, U4 A 5a, page 535 (1978).
The study for Konix® Ultrasound Gel itself appears to be a prospective measurement of its physical properties. The country of origin of the device manufacturer is Turkey.
3. Number of Experts and Qualifications for Ground Truth
This type of device (ultrasound gel) does not typically involve expert review for establishing ground truth in the way medical imaging algorithms do. The "ground truth" for comparison for acoustic properties comes from established scientific literature regarding human skin properties. Therefore, there were no "experts" in the traditional sense involved in establishing ground truth for a test set of images or clinical outcomes.
4. Adjudication Method
Not applicable. The study involved direct measurements of the physical properties of the gel and comparison to published scientific data. There was no adjudication of expert opinions.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done. This type of study is suitable for AI algorithms that assist human readers in interpreting medical images, which is not applicable to an ultrasound gel.
6. Standalone Performance Study
Yes, a standalone study was performed. The non-clinical performance evaluation focused on directly measuring the acoustic properties of the Konix® Ultrasound Gel itself, rather than its performance in conjunction with human interpretation. The study states: "Konix® Ultrasound Gel was evaluated for its acoustic performance."
7. Type of Ground Truth Used
The ground truth used was established scientific literature and physical properties of human skin and other commonly used ultrasound coupling gels. For example, the reported longitudinal velocity of skin (1,518 m/sec) and acoustic impedance of skin (1.6 kg/m²s) were referenced from a scientific publication (Ogura, 1978).
8. Sample Size for the Training Set
Not applicable. This device is an ultrasound gel, not an AI algorithm that requires a training set. The "study" here involves physical property measurements and comparison, not machine learning model training.
9. How Ground Truth for the Training Set Was Established
Not applicable. As this is not an AI algorithm, there was no training set or ground truth established for such a set.
Ask a specific question about this device
Page 1 of 1