Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K030456

    Validate with FDA (Live)

    Device Name
    VASC-ALERT
    Manufacturer
    Date Cleared
    2003-07-11

    (150 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K960817, K980906, K002372

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vasc-Alert vascular data analysis software is intended for use by Healthcare professionals in a noncritical care setting for assessment of increased risk of access site stenosis in patients, as an aid for vascular access site management. The software is to be used with data generated from hemodialysis machines manufactured by Fresenius and Cobe/Gambro.

    Device Description

    Vasc-Alert is a software program for alerting dialysis center personnel of an increased risk of access site stenosis for individual hemodialysis patients. Vasc-Alert utilizes measurements routinely collected during a dialysis treatment by the hemodialysis machine, such as pressure and flow rate, and applies a previously published algorithm called the Vascular Access Pressure Ratio (VAPR) test to these measurements. The average VAPR test result for each treatment session is stored in a Vasc-Alert database. If a patient has a high reading in three consecutive dialysis sessions, a report is issued to the medical staff indicating that the patient should be examined more closely for the onset of stenosis. Hemodialysis center personnel can use the report as a tool to proactively monitor for incipient stenosis and prompt proactive intervention to avoid site closure.

    Vasc-Alert comprises five main components or modules:

    • A module for recording, transferring and parsing data collected by dialysis center machines.
    • A module for calculating the VAPR values from treatment data.
    • A module for identifying significant patterns in the calculated VAPR data that will prompt an alert (i.e., 3 high readings in a row for a patient).
    • A module for generating reports and sending these out to center personnel.
    • An internet-based data input module.
    AI/ML Overview

    The provided text does not contain specific acceptance criteria, reported device performance metrics, or details about a study designed to prove the device meets these criteria. The document is a 510(k) summary for the Vasc-Alert device, primarily focusing on its description, intended use, and substantial equivalence to predicate devices, rather than a detailed performance study report.

    However, based on the narrative, we can infer some aspects and highlight what information is missing:

    Inferred Information:

    The core function of Vasc-Alert is to identify an "increased risk of access site stenosis" when a patient has a "high reading in three consecutive dialysis sessions" based on the Vascular Access Pressure Ratio (VAPR) test. The "study" mentioned isn't a formalized clinical trial with defined endpoints and acceptance criteria, but rather a "side-by-side testing" for substantial equivalence.

    Here's an attempt to structure the information based on your request, noting significant gaps:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Substantial Equivalence Claim)Reported Device Performance (Inferred from Substantial Equivalence Claim)
    Identify patients at risk of stenosis.Vasc-Alert "identified more patients at risk than the other technologies."
    Patients identified as 'at risk' by Vasc-Alert develop complications within a reasonable timeframe (e.g., 3 months)."the majority of these patients [identified by Vasc-Alert as at risk] did in fact develop complications in their access sites within 3 months, proving that Vasc-Alert can be used as an indication of stenosis developing in the access site."
    Similar intended use to predicate devices (detection of stenosis buildup).Vasc-Alert has "the same intended use (detection of the buildup of stenosis in the access site)."
    Uses similar technology/measurements as predicate devices (pressure and flow)."uses essentially the same technology in their analysis (pressure and flow)."

    Missing Information:

    • Specific quantitative acceptance criteria: There are no defined thresholds for sensitivity, specificity, positive predictive value, negative predictive value, or accuracy for identifying "at risk" patients or predicting complications.
    • Detailed performance metrics: Beyond the qualitative statements, no numerical data (e.g., exact percentages of patients identified, or precise proportion of those who developed complications) is provided.

    2. Sample Size Used for the Test Set and Data Provenance

    Sample Size: Not specified. The document only refers to "side-by-side testing" without providing the number of patients or dialysis sessions included.

    Data Provenance: Not specified. It doesn't mention the country of origin, whether the data was retrospective or prospective, or from how many centers.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    Number of Experts: Not specified.

    Qualifications of Experts: Not specified. The phrase "medical staff" is used when referring to who examines the patient after an alert, but it's unclear if these or other experts were involved in establishing the ground truth for the "side-by-side testing."

    4. Adjudication Method for the Test Set

    Adjudication Method: Not specified. There is no information about how the "complications" were confirmed or how ground truth was established for the "side-by-side testing."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    MRMC Study: No, an MRMC comparative effectiveness study was not reported. The device is software that generates alerts, and the text describes its standalone performance and comparison to other technologies, not a human-AI comparison.

    Effect Size: Not applicable, as no MRMC study was performed.

    6. Standalone (Algorithm Only) Performance Study

    Standalone Study: Yes, a standalone (algorithm only) performance was implicitly reported as the "side-by-side testing" where Vasc-Alert identified patients at risk and predicted complications within 3 months. The software's ability to trigger an alert based on VAPR readings in three consecutive sessions constitutes its standalone performance.

    Missing Information: Quantitative results for this standalone performance (e.g., sensitivity, specificity, predictive values).

    7. Type of Ground Truth Used

    Type of Ground Truth: The ground truth used for performance inference was outcomes data, specifically whether "complications in their access sites" developed within 3 months after Vasc-Alert identified them as "at risk." This implies clinical diagnosis of stenosis or related complications.

    8. Sample Size for the Training Set

    Sample Size for Training Set: Not specified. The document describes the algorithm as "a previously published algorithm called the Vascular Access Pressure Ratio (VAPR) test," implying the algorithm itself was developed or validated prior to this submission, but does not provide details on Vasc-Alert's specific training set if any machine learning was involved (which is unlikely given the description of a fixed algorithm).

    9. How the Ground Truth for the Training Set Was Established

    How Ground Truth for Training Set Was Established: Not specified. Since the VAPR algorithm is stated to be "previously published," the details of its development and the establishment of ground truth for its initial formulation are outside the scope of this 510(k) summary. No specific training data or ground truth establishment for Vasc-Alert's software implementation is detailed.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1