510(k) Data Aggregation

    K Number
    K050808

    Validate with FDA (Live)

    Date Cleared
    2005-07-28

    (120 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K955817, K031906, K003731, K042381

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioOptics Coronary Sinus Access (CSA™) System is intended to provide subclavian access to the right heart for accessing the coronary sinus and placing a quide sheath suitable for pacemaker lead implantation into the coronary sinus. The system includes visualization means to image anatomical structures to augment navigation.

    Device Description

    The Coronary Sinus Access (CSA™) System provides a catheter and sheath system for providing percutaneous access to the coronary sinus. In addition, the catheter provides visualization to augment the user's ability to locate the coronary sinus for placement of a guide sheath. The system includes a sterile, single use Coronary Sinus Access Kit, which includes a deflecting tip catheter with visualization capability (FLAIR™ CS Catheter), a Guide Sheath Set including a dilator and sheath slitters, and a Sterile Sleeve to drape over the camera pod. The system also includes an Imaging Acquisition System to collect and display video images present in front of the catheter tip.

    AI/ML Overview

    The CardioOptics Coronary Sinus Access (CSA™) System underwent performance testing, including mechanical, functional, and animal studies, to establish substantial equivalence to predicate devices. Biocompatibility testing was also conducted.

    Here is a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of quantitative acceptance criteria with corresponding device performance metrics. Instead, it states that various tests were performed to support "substantial equivalence" to predicate devices.

    Acceptance Criteria CategoryReported Device Performance
    Mechanical TestingPerformed to support substantial equivalence.
    Functional TestingPerformed to support substantial equivalence.
    Animal TestingPerformed to support substantial equivalence.
    BiocompatibilityAll specified biocompatibility requirements (consistent with ISO 10993) were met for blood and tissue contacting materials.
    SterilizationComponents will be sterilized using validated Ethylene Oxide (EtO) sterilization processes.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size for Test Set: The document does not specify the sample sizes used for the mechanical, functional, or animal testing.
    • Data Provenance: The provenance of the data (e.g., country of origin, retrospective or prospective) is not explicitly stated. Animal testing implies prospective data collection in a controlled environment.

    3. Number of Experts and Qualifications for Ground Truth

    This information is not provided in the document. The device's performance is described in terms of "substantial equivalence" to predicate devices and meeting general testing requirements, rather than an evaluation against human expert-defined ground truth in a clinical setting.

    4. Adjudication Method for Test Set

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The document focuses on demonstrating substantial equivalence through device testing.

    6. Standalone Performance Study

    A standalone performance study (i.e., algorithm only without human-in-the-loop performance) was not explicitly described or conducted, as the device is a system with visualization to augment human navigation rather than an autonomous algorithm. The visualization aspect of the FLAIR™ CS Catheter was shown to be substantially equivalent to angioscopes and associated video display systems.

    7. Type of Ground Truth Used

    The ground truth used for assessing the device's performance appears to be established through:

    • Engineering and pre-clinical testing standards: Mechanical and functional performance would be measured against predefined specifications or engineering benchmarks.
    • Biocompatibility standards: ISO 10993 serves as the ground truth framework for biocompatibility.
    • Comparison to predicate devices: The concept of "substantial equivalence" implies that the device's performance or characteristics are compared to those of already legally marketed predicate devices, which serve as a de facto "ground truth" for regulatory clearance.
    • Animal models: Animal testing provides a physiological environment to assess aspects like access and visualization, where observations by skilled operators in the animal model would serve as the ground truth.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI, as the device is not described as an AI/ML product. The testing performed relates to the physical and functional aspects of the medical device.

    9. How Ground Truth for the Training Set was Established

    As there is no mention of a training set for an AI/ML component, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K955817

    Validate with FDA (Live)

    Date Cleared
    1996-03-25

    (90 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K955817

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cordis Webster deflectable braided-tip electrode catheter is indicated for electrophysiological mapping of cardiac structures; i.e., endocardial stimulation and recording within the right side of the heart.

    Device Description

    The Cordis Webster deflectable braided-tip electrode catheter has been designed for electrophysiological mapping of cardiac structures. The catheter has a high-torque shaft with an internal stainless steel braid and a tip section containing platinum electrodes that can easily be seen under fluoroscopy. The tip section also employs an internal braid to provide additional lateral stability during mapping procedures.

    A piston in the handpiece is attached to an internal puller which changes the radius of curvature of the tip section. When the piston is pushed forward, the radius of curvature of the tip section is reduced; when the thumbknob is pulled back, the radius of curvature is increased until the tip section returns to its original state. The high-torque shaft allows the tip section to be maneuvered in order to facilitate accurate positioning within the heart.

    AI/ML Overview

    This 510(k) summary describes a medical device, a deflectable braided-tip electrode catheter, and its substantial equivalence to predicate devices. It focuses on the physical and technological characteristics and non-clinical performance, rather than a study involving acceptance criteria and device performance as typically understood for AI/software devices. Therefore, a direct response to your specific questions about acceptance criteria, sample sizes, ground truth, and human reader studies is not fully applicable.

    However, I can extract the relevant information from the provided text that describes the performance assessment conducted for this non-AI medical device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a non-AI physical device, the "acceptance criteria" are not explicitly quantitative thresholds like sensitivity or specificity. Instead, they are implied by the comparison to predicate devices and adherence to guidance documents.

    Acceptance Criterion (Implied)Reported Device Performance
    Basic electrical characteristics comparable to predicate devicesComparable to predicate devices.
    Basic mechanical characteristics comparable to predicate devicesComparable to predicate devices.
    Tip stiffness no greater than currently marketed devicesNo greater than other currently marketed devices.
    Side load capability comparable to/improved over predicate devicesGreater than that of the standard deflectable catheter.
    Adherence to "Electrode Recording Catheter Preliminary Guidance"Applicable tests were performed in accordance with this guidance.
    Sterilization and packaging identical to predicate devicesIdentical to Cordis Webster standard deflectable and fixed curve catheters.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes "non-clinical performance testing." This typically involves laboratory testing on physical samples of the device.

    • Sample Size: Not explicitly stated. The phrase "the braided-tip electrode catheter" and "predicate devices" implies testing of multiple units, but a specific number is not provided.
    • Data Provenance: Not applicable in the context of human data. The "data" comes from engineering and laboratory performance tests of the physical catheter.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • Experts: Not applicable. For a physical medical device, "ground truth" is established through standardized engineering tests, measurements, and comparisons to established device specifications or performance of predicate devices. There wouldn't be human experts establishing "ground truth" in the way it's done for diagnostic image analysis.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable. This concept is for resolving disagreements in human expert labeling or interpretation, which is not relevant here.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • MRMC Study: No. This type of study assesses human reader performance with and without AI assistance, which is not relevant for a physical electrophysiology catheter.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Standalone Performance: No. This question pertains to AI algorithms. The described device is a physical catheter.

    7. The Type of Ground Truth Used:

    • Type of Ground Truth: For this physical device, the "ground truth" is established by:
      • Engineering Specifications: Performance against pre-defined engineering standards for electrical and mechanical properties.
      • Comparative Performance Data: Direct comparison of the device's measured electrical, mechanical, and physical properties (e.g., tip stiffness, side load capability) against those of the identified predicate devices.
      • Regulatory Guidance: Adherence to "Electrode Recording Catheter Preliminary Guidance."

    8. The Sample Size for the Training Set:

    • Training Set Sample Size: Not applicable. This device does not involve machine learning or an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    • Training Set Ground Truth: Not applicable. As there is no training set for an AI algorithm, this question is not relevant.

    In summary: The provided text describes a non-clinical performance study of a physical medical device (catheter) to demonstrate substantial equivalence to predicate devices based on engineering characteristics and adherence to regulatory guidance, rather than a clinical study or an AI/software performance study. Therefore, many of your questions, which are framed for AI/software evaluations, are not directly applicable.