510(k) Data Aggregation

    K Number
    K964801

    Validate with FDA (Live)

    Date Cleared
    1997-06-02

    (185 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K951262, K961844, K954276

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optical Tracking System (OTS) is a graphical planning tool that allows for preoperative and operative planning of surgical procedures. The OTS is indicated for use in surgical procedures in which anatomical landmarks are not clearly visible or where a desired target is close to critical structures.

    Device Description

    The Optical Tracking System (OTS), addressed in this premarket notification, has the same intended use and similar technological characteristics as the commercially available Radionics Operating Arm System and the StealthStation Stereotactic System. The OTS is a modification of the Operating Arm System but substitutes a camera array for the operating arm to track the position of probes and various tools. This is the same camera array utilized by StealthStation. The OTS provides an interactive, image-guided means of localizing targets in surgical procedures. In addition to the camera array, the OTS consists of a computer workstation, application software, and probes. Upon calibration of the system in the surgical environment, the surgeon selects points on the patient using a probe which are interpreted by the computer and related to corresponding points on the image.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Optical Tracking System (OTS), based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Accuracy: The system is correctly encoded for stereotactic CT and MR localizer transformation equations, and all target coordinate input and displays are accurate.The camera array is accurate to a mean value of 0.3 mm.
    Stability: Accuracy is maintained over a period of extended operation.This accuracy is maintained over a period of extended operation.

    2. Sample Size Used for the Test Set and Data Provenance

    The text explicitly states: "The results from the OTS are compared to a phantom with targets of known position."

    • Sample Size for Test Set: The text does not specify a numerical sample size for the targets on the phantom. It only mentions "targets of known position."
    • Data Provenance: The data is generated through testing with a phantom (a physical model with known properties), not clinical patient data. Therefore, it is a simulated/laboratory setting, not retrospective or prospective clinical data from a specific country.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable.
    • Qualifications of Experts: Not applicable.

    The ground truth for this device's performance test is established by the known, pre-defined positions of targets on a phantom. Human experts are not involved in establishing this ground truth.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable.

    Since the ground truth is established by the known positions on a phantom, there is no need for human adjudication of the test results. The device's measurements are directly compared to these known values.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Comparative Effectiveness Study: No, an MRMC comparative effectiveness study was not done.

    This submission is for an image-guided surgical planning tool, not an AI diagnostic tool. The performance evaluation focuses on the system's accuracy in tracking and localization, not on changes in human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, a standalone performance evaluation was done.

    The reported accuracy of "0.3 mm" for the camera array appears to be a standalone measurement of the system's ability to track positions against a known phantom. While the system is intended for use with a surgeon ("human-in-the-loop"), the verification of its core accuracy is performed independently of a surgeon's interpretation or action, by comparing its output to predefined phantom targets. The statement "RSA Optical Tracking System (OTS) systems testing verifies that the stereotactic CT and MR localizer transformation equations are correctly encoded into the application software. Further, it verifies that all target coordinate input and displays are accurate. The results from the OTS are compared to a phantom with targets of known position. The camera array is accurate to a mean value of 0.3 mm. This accuracy is maintained over a period of extended operation" describes a standalone evaluation of the device's technical specifications.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: The ground truth used is based on known, pre-defined positions of targets on a phantom. This is a form of physical/engineering ground truth.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: The document does not provide information about a training set. This device is described as an "Optical Tracking System," not a machine learning or AI algorithm that would typically require a training set in the modern sense. Its functionality is based on established principles of optical tracking and geometry.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable, as no training set is mentioned or implied for this type of device in the provided text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K960714

    Validate with FDA (Live)

    Device Name
    ISG VIEWING WAND
    Date Cleared
    1996-05-29

    (98 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K942233,K951262,K944612

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Not Found

    Device Description

    The ISG Family of Viewing Wands is comprised of a medical imaging workstation that is integrated with a position-sensing user directed probe. This allows for the provision of information pertaining to the current position of the probe related to the surrounding anatomy by correlating pre-operative CT and/or MR imaging data with the intra-operative situation. Once a patient's CT or MR images have been transferred to the workstation, the patient is correlated to that image dataset. The system then displays the orientation of the probe and the position of the tip on the corresponding image, updating the display in real-time as the probe is moved.

    AI/ML Overview

    The provided 510(k) summary for the ISG Family of Viewing Wands does not contain the detailed information required to answer all parts of your request, specifically regarding acceptance criteria, a dedicated study proving performance against those criteria, or details about training and ground truth for such a study.

    This document describes a surgical navigation system that correlates pre-operative CT/MR images with intra-operative probe position. Its function is to display the orientation and position of a probe on corresponding images in real-time. It is not an algorithm designed to diagnose conditions or interpret medical images in an automated fashion, but rather a tool to aid a surgeon. Therefore, the concept of "device performance" in this context is likely related to accuracy of localization and real-time display, rather than diagnostic metrics like sensitivity, specificity, or AUC.

    Here's a breakdown of what can be extracted based on the input:


    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not explicitly state specific acceptance criteria or report performance metrics in a quantitative manner (e.g., accuracy, precision) for the Family of Viewing Wands. It states: "Standards are defined and tested under the specifications included in this submittal." However, these specifications are not provided in the excerpt.

    Given the device's function, potential implied performance characteristics would involve:

    • Localization Accuracy: How precisely the system can determine and display the probe tip's position relative to the imaging data.
    • Real-time Update Speed: How quickly the display updates as the probe moves.
    • Image Processing and Display Quality: How accurately the pre-operative images are rendered and how well the probe's position is superimposed.

    The remaining questions relate to a 'study' that would prove the device meets acceptance criteria. Since no such study with quantifiable performance metrics is described, many of these questions cannot be answered from the provided text.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not specified. The document does not describe a clinical or technical validation study with a test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not specified. As there's no described test set or study aiming to establish diagnostic performance, there's no mention of experts establishing a ground truth for interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not specified.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is a surgical navigation system, not an AI diagnostic algorithm for image interpretation. Therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant to this device's described function and is not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not specified in detail. The device is a system that works with a human operating a probe. Its function is to provide information to the human, not to act standalone in a diagnostic capacity. While its internal localization algorithms would function "standalone" in terms of computing coordinates, there's no standalone performance evaluation described as if it were a diagnostic AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not specified. For a surgical navigation system, "ground truth" would likely involve highly precise physical measurements of probe tip location compared to the system's displayed location, potentially using phantoms or cadaveric models. This is not detailed in the provided text.

    8. The sample size for the training set

    • Not applicable/Not specified. This device isn't described as using machine learning or AI that requires a "training set" in the conventional sense (e.g., for image classification). Its core functionality is based on known physical principles of position sensing and image display.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified. (See point 8).

    Summary of Device and its Purpose:

    The ISG Family of Viewing Wands is a surgical navigation system. Its primary function is to:

    • Process and display pre-operative radiographic images (CT and/or MR).
    • Provide real-time intra-operative tracking of a user-directed probe, correlating its position and orientation with the pre-operative imaging data.

    It is classified as a Class II diagnostic device accessory and is substantially equivalent to existing viewing wands and surgical navigation systems on the market. The submission focuses on describing its components (workstation, localizing device, communications, archive storage) and general functionality, rather than presenting a performance study with detailed acceptance criteria and results.