Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K172850

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-11-17

    (58 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K792558, K944864, K053389, K112671

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap Sterilcontainer System is a reusable sterilization container system (consisting of solid & perforated bottoms, a perforated lid w/ filter retention plates, and reusable polytetrafluoroethylene (PTFE) filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in pre-vacuum steam and IUSS (Immediate Use Steam Sterilization modalities. The SterilContainer System for includes accessories such as silicone mats, baskets, trays, and racks.

    Device Description

    The SterilContainer with PrimeLine Pro Lid is used in conjunction with Aesculap's Sterilcontainer Systems (K792558/K944864/K053389/K112671). The SterilContainer with PrimeLine Pro Lid is designed to be compatible for use with pre-vacuum steam and IUSS sterilization. The lid is manufactured from anodized aluminum and utilizes a reusable polytetrafluoroethylene (PTFE) filter. The SterilContainer with PrimeLine Pro Lid is offered in a various colors and range of sizes as the predicates including full, ¾, and half sizes lids.

    AI/ML Overview

    The Aesculap SterilContainer with PrimeLine Pro Lid is a reusable sterilization container system designed to sterilize and maintain the sterility of medical devices. The device was found to be substantially equivalent to its predicate devices based on non-clinical performance testing.

    Here's a breakdown of the acceptance criteria and study information:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Sterilization EfficacyTesting demonstrated a 6 log reduction and a sterility assurance level (SAL) of $10^{-6}$ using the biological (BI) overkill method and half-cycle validation.
    Whole Package Microbial Aerosol ChallengeAfter exposure to a defined amount of aerosol microorganisms, the contents maintained sterility.
    Event Related Sterility MaintenanceThe test reports demonstrated that the SterilContainer with PrimeLine Pro Lid consistently maintained sterility of the contents after processing followed by 30, 180, and 365-day event-related storage under conditions which simulate hospital sterile package handling and storage conditions as long as the barrier is not compromised.
    Mechanical Cleaning (Protein Analysis)The study results validate that the mechanical cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a protein level of less than $6.4 \mu g/cm^2$ per device.
    Mechanical Cleaning (Hemoglobin Analysis)The study results validate that the mechanical cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a hemoglobin level of less than $2.2 \mu g/cm^2$ per device.
    Manual Cleaning (Protein Analysis)The study results validate that the manual cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a protein level of less than $6.4 \mu g/cm^2$ per device.
    Manual Cleaning (Hemoglobin Analysis)The study results validate that the manual cleaning instructions were effective in removing gross amounts of soil from the SterilContainer with PrimeLine Pro Lid to a protein level of less than $2.2 \mu g/cm^2$ per device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for each test. It mentions that "Test were performed in best/mid/worst case sized regarding vent to volume ratio and container size." The data provenance is not specified, but the testing was conducted by a "qualified testing laboratory" in accordance with FDA guidance and AAMI standards. This indicates a prospective and controlled experimental design.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is not an AI/imaging device requiring expert interpretation for ground truth. The performance testing involves objective measures of sterilization efficacy, microbial challenge, sterility maintenance, and cleaning effectiveness, which do not typically rely on expert consensus for ground truth establishment in the same way an AI diagnostic tool would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this device does not involve human interpretation or subjective assessment that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device, so MRMC studies and the concept of human readers improving with AI assistance are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device. The testing focuses on the physical and functional performance of the sterilization container system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance testing is based on established scientific and regulatory standards for sterilization, microbial challenge, sterility maintenance, and cleaning efficacy.

    • Sterilization Efficacy: Ground truth is a 6 log reduction and SAL of $10^{-6}$ as per biological indicator results and half-cycle validations.
    • Microbial Aerosol Challenge: Ground truth is the absence of microbial growth in the contents after exposure to a defined amount of aerosol microorganisms.
    • Event Related Sterility Maintenance: Ground truth is the absence of microbial growth in the contents after storage periods, indicating sterility was maintained.
    • Cleaning Effectiveness: Ground truth is a protein level of less than $6.4 \mu g/cm^2$ and a hemoglobin level of less than $2.2 \mu g/cm^2$ per device, as determined by laboratory analysis.

    These are objective, measurable outcomes based on established scientific principles and regulatory benchmarks.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for this device, a method for establishing its ground truth is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112671

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-04

    (233 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K073168, K053389, K944864, K792558

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap SterilContainer System is a reusable sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility for 360 days. The SterilContainer System includes accessories such as mats, baskets, trays, holders, organizers, filters, indicator cards and tamper proof locks.

    Device Description

    The Aesculap SterilContainer is designed as a container system that will allow for sterilization and storage of other medical devices. This system consists of full, three-quarter, half, quarter, and extra long and wide body container system. There are a variety of accessories also available for this container system. This container is designed to be compatible for use in pre-vacuum steam, pre-vacuum immediate use steam and EtO sterilization. The container bottom is manufactured in anodized aluminum and has stainless steel handles on each end. There are two types of lids for the container bottom. There are anodized aluminum lids that have removable retention plates to hold the filter and a stainless steel latch on each end to lock the lid only to the container bottom. There are also Prime Line lids available for use that are manufactured in the material commonly known as Radel R. The PrimeLine lids have a reusable filter system. The reusable filter is for Pre-vac steam and Immediate Use steam sterilization only.

    AI/ML Overview

    The provided text describes the Aesculap SterilContainer System for sterilization. Here's a breakdown of the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list "acceptance criteria" with pass/fail thresholds in a formal table separate from the performance results. Instead, it presents the validated sterilization cycle parameters and states that the device was "fully validated" according to FDA and AAMI standards. The reported device performance is that it meets these validated parameters and achieves a 360-day event-related shelf life.

    Acceptance Criteria CategorySpecific Criteria / Performance Target (Implied)Reported Device Performance
    Sterilization EfficacyMust achieve sterilization under the stated parameters for: - Pre-vacuum steam (270°F, 4 min. exposure, 30 min. dry) - Pre-vacuum immediate use steam (270°F, 3 min. (non-porous) / 4 min. (porous) exposure) - Ethylene Oxide (130°F, 60 min exposure, ≥ 50% RH, 725 mg/L gas pressure)Achieved: "fully validated for pre-vacuum steam, pre-vacuum immediate use steam, and EtO sterilization processes."
    Shelf LifeMaintain sterility for 360 days.Achieved: "360 days of event related shelf life testing has been conducted."
    CompatibilityCompatible with specified lids, filters, and accessories for each sterilization method.Achieved: Detailed tables provided for compatibility across various configurations and accessories.
    Max Load WeightAccommodate maximum load weights of 25 pounds or less (AAMI/AORN recommendation).Achieved: "Max total weight 25 pounds."
    Vent to Volume RatioMeet predetermined standards for the "worst case vent to volume ratio" to ensure sterilization.Achieved: "This container represents the worst case vent to volume ratio for the SterilContainer System," and validation was successful.
    Reuse TestingMaintain performance after repeat sterilization cycles.Achieved: "100 cycles minimum" for container, "2,200 cycles (lid)" for reusable filter.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the numerical sample size used for the test set in terms of number of containers or cycles. However, it does state:

    • "All required testing per 'Draft Guidance for the Preparation of Premarket Notifications 510(k)'s' for Aesculap SterilContainer was completed on this new three-quarter size 8" (187mm) container of the worst case vent to volume ratio..."
    • The testing was conducted by "a qualified testing laboratory."

    Data Provenance: The document does not specify the country of origin of the data or whether it was retrospective or prospective. Given that it's a 510(k) submission for a new device, the testing would inherently be prospective in nature, specifically designed to validate the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" for sterilization efficacy is established through standardized biological and physical indicator tests, not expert consensus on medical images or patient outcomes. The testing was performed in a laboratory setting per established standards (FDA and AAMI).

    4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 refer to a process where multiple human reviewers independently assess data (e.g., medical images) and disagreements are resolved by an expert panel. Sterilization and shelf-life testing relies on objective, measurable scientific and engineering principles (e.g., microbial kill, barrier integrity), not subjective human interpretation requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device for sterilization, not an AI-powered diagnostic or assistive tool for human readers/clinicians, so an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of a device on its own, without human intervention in its function during testing. In this context, the device (the SterilContainer) was tested in a standalone manner as its primary function is to achieve and maintain sterility independently when subjected to specified sterilization cycles. The performance data presented (sterilization cycle parameters met, 360-day shelf-life) reflects the container's inherent capabilities.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for sterilization efficacy is typically established through:

    • Biological Indicators (BIs): Contain a known number of highly resistant microorganisms (e.g., bacterial spores). A successful sterilization cycle renders the BI sterile, demonstrating microbial kill.
    • Chemical Indicators (CIs): React to one or more sterilization parameters with a visible change in color or state.
    • Physical Monitoring: Readings from sterilizer gauges, thermometers, pressure recorders, and timers confirm that the physical parameters (temperature, pressure, time) of the cycle were met.
    • Barrier Integrity Testing: For shelf-life, tests ensure the container maintains a sterile barrier over time, often involving microbial ingress challenges or visual inspection for damage.

    The document implicitly refers to these methods by stating: "These validations were conducted in accordance with FDA and AAMI standards by a qualified testing laboratory." These standards prescribe the use of BIs, CIs, and physical monitoring for sterilization efficacy, and specific tests for packaging integrity and shelf-life.

    8. The Sample Size for the Training Set

    Not applicable. Training sets are relevant for machine learning or AI models. This device is a physical container, and its performance is evaluated through conventional engineering and microbiological testing, not by training an algorithm.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073168

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-04-28

    (171 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K792558, K944864

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap Sterilcontainer System is a reusable sterilization container system (consisting of solid and perforated bottoms, a perforated lid w/ filter retention plates, and reusable polytetrafluoroethylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use in pre-vacuum steam and flash sterilization. The SterilContainer System includes accessories such as silicon mats, baskets, trays, and racks.

    Device Description

    The Aesculap SterilContainer with PrimeLine Lid is used in conjunction with Aesculap's Sterilcontainer System (K792558 & K944864). The PrimeLine Container Lid is designed to be compatible for use with pre-vacuum steam and pre-vacuum flash sterilization. The lid is manufactured from Radel R 5000/5008 and utilizes a reusable polytetrafluoroethylene (PTFE) filter.

    AI/ML Overview

    The provided document is a 510(k) summary for the Aesculap SterilContainer with PrimeLine Lids. It's a premarket notification for a medical device and describes its substantial equivalence to existing predicate devices. The document highlights the device's technological characteristics and performance data related to sterilization, but it does not contain the specific information required to answer all sections of your prompt.

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes performance data in terms of sterilization validation. However, it does not explicitly state "acceptance criteria" with numerical targets and then report performance against those targets in a formal table. Instead, it states that the device was "fully validated" against FDA guidance and AAMI standards for steam and flash sterilization.

    Acceptance Criteria (Inferred from indication for use and validation statement)Reported Device Performance (Summary)
    Compatibility with pre-vacuum steam sterilization"fully validated for the steam... sterilization in a prevacuum steam... cycle"
    Compatibility with pre-vacuum flash sterilization"fully validated for the... flash sterilization in a... prevacuum... flash cycle"
    Maintenance of sterility of enclosed device"intended to... maintain sterility of the enclosed device until used." (Implied by validation)
    Sterilization of enclosed device"intended to allow sterilization of the enclosed device" (Implied by validation)
    Specific cycle parameters for Flash (Non-porous, 270°F, 3 min)Validated for these parameters with specific lid/base sizes and lumen configurations (see Indications for Use table)
    Specific cycle parameters for Flash (Mixed, 270°F, 4 min)Validated for these parameters with specific lid/base sizes and lumen configurations (see Indications for Use table)
    Specific cycle parameters for Prevacuum (270°F, 4 min exposure, 15 min dry)Validated for these parameters with specific lid/base sizes and lumen configurations (see Indications for Use table)
    Maximum load for all container configurations"A combined maximum load validated for all container configurations is 35lbs."
    Number of lumens and their configuration (for various cycle types/sizes)Validated for 2 SS lumens, ≥3mm I.D. x ≤200mm L or ≤400mm L (see Indications for Use table)

    2. Sample size used for the test set and the data provenance:

    The document states, "The Aesculap Sterilcontainer system was fully validated for the steam and flash sterilization in a prevacuum steam and flash cycle." It also indicates that this validation was conducted "by a qualified testing laboratory." However, specific sample sizes (e.g., number of cycles, number of containers tested) for the validation study are not provided. The provenance of the data is that it was generated during product validation (likely in the US based on the FDA submission) by a qualified testing laboratory, and it would be prospective data collected specifically for the purpose of demonstrating device performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not available in the provided document. Sterilization validation typically involves scientific and engineering principles, often with industry standards (like AAMI) serving as the "ground truth" for successful sterilization. There is no mention of human experts establishing ground truth in the context of this device's performance data.

    4. Adjudication method for the test set:

    This information is not available in the provided document. Adjudication methods are typically relevant for studies involving human assessment or interpretation (e.g., medical imaging classification). For device performance validation related to sterilization, "adjudication" in the sense of expert consensus on an output is not typically a component. The success/failure of sterilization (e.g., through biological indicator testing) is typically objective.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable to this device. The Aesculap SterilContainer with PrimeLine Lids is a medical device for sterilization, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study related to AI assistance would not have been performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable to this device. This is a physical medical device. It does not involve an algorithm or AI that would have standalone performance.

    7. The type of ground truth used:

    The ground truth used for this type of device validation would be based on established scientific principles of sterilization effectiveness and compliance with recognized standards (e.g., FDA guidance, AAMI standards for sterilization). This would involve demonstrating the killing of microbial spores (e.g., using biological indicators) and maintaining sterility over time. It's not "expert consensus" in the sense of diagnostic interpretation, nor is it pathology or outcomes data in the clinical sense.

    8. The sample size for the training set:

    This is not applicable. This is a physical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    This is not applicable. As it's not an AI model, there is no training set or associated ground truth establishment process.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1