510(k) Data Aggregation

    K Number
    K053124

    Validate with FDA (Live)

    Date Cleared
    2006-01-30

    (84 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K943177,K980741,K000795,K013762

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BOSTON ES® (enflufocon A), BOSTON EO® (enflufocon B) and BOSTON XO® (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, and presbyopia) in aphakic and not-aphakic persons with nondiseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only.

    Device Description

    The BOSTON ES (enflufocon A) and BOSTON EO (enflufocon B) are rigid gas permeable Contact Lens materials composed of aliphatic fluoroitaconate siloxanyl methacrylate copolymer with an ultraviolet absorher.

    BOSTON XO (hexafocon A) is a rigid gas permeable contact lens material, composed of siloxanyl fluoromethacrylate copolymer containing an ultraviolet absorber.

    The color additives conform to 21 CFR Part 74 and/or 21 CFR Part 73. The lenses may also be supplied clear (no tint).

    AI/ML Overview

    The provided text describes information for the "BOSTON ES® (enflufocon A), BOSTON EO® (enflufocon B) AND BOSTON XO® (hexafocon A) RIGID GAS PERMEABLE CONTACT LENSES". However, it is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study proving device performance against those criteria in the way a clinical trial would for a novel medical device.

    Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, training set size) are not applicable or not provided in this type of document.

    Here's a breakdown of the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a pass/fail sense against pre-defined thresholds. Instead, it reports the physical/optical properties of the lenses, which are implicitly accepted given their substantial equivalence claim. The reported properties are presented below:

    PropertyBOSTON ESBOSTON EOBOSTON XO
    Specific Gravity1.221.231.27
    Refractive Index1.4431.4291.415
    Visible Light Transmittance> 85%> 85%> 92%
    Water Content<1%<1%<1%
    Wetting Angle52°49°49°
    Oxygen Permeability (Dk)***36*18**82*58**140*100**
    • gas to gas method
    • polarographic method (ISO/Fatt)
    • ***(x 10⁻¹¹ (cm³ O₂ • cm)/ (cm² • sec • mmHg) @ 35° C}

    2. Sample size used for the test set and the data provenance

    Not applicable. This is a 510(k) submission based on substantial equivalence to existing predicate devices, not a clinical trial with a "test set" in the context of AI/software performance. The safety and efficacy were previously demonstrated in earlier 510(k) clearances.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a 510(k) submission.

    4. Adjudication method for the test set

    Not applicable. This is a 510(k) submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a contact lens device, not an AI-based diagnostic tool requiring MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a contact lens device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document supports the safety and effectiveness of the device by demonstrating its physical/optical properties and clinical indications are substantially equivalent to previously cleared predicate devices. The "ground truth" here is regulatory acceptance based on comparisons to established, safe, and effective devices, as well as adherence to recognized material property standards.

    8. The sample size for the training set

    Not applicable. This is a 510(k) submission for a physical device, not an AI algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This is a 510(k) submission for a physical device, not an AI algorithm.


    Summary of the "Study" (Demonstration of Substantial Equivalence):

    The document itself is the "study" in the context of a 510(k). It states:

    • "The safety and efficacy of BOSTON ES (enflufocon A), BOSTON EO (enflufocon B), and BOSTON XO (hexafocon A) Rigid Gas Permeable Contact Lenses was demonstrated in 510(k) Premarket Notifications as follows: K943177 cleared on August 25, 1994; K980741 cleared on May 11, 1998; and K000795 cleared on May 25, 2000, respectively. The most recent Premarket Notification for all three materials (Boston ES, EO, and XO) was 510(k) K013762, cleared April 3, 2002, for a new indication for keratoconus."
    • "BOSTON ES (enflufocon A), BOSTON EO (enflufocon B), and BOSTON XO (hexafocon A) Rigid Gas Permeable Contact Lenses are substantially equivalent to Lens Dynamics Inc., Dyna Intra-Limbal Lens (enflufocon A or hexafocon A) Rigid Gas Permeable Contact Lens, cleared in 510(k) Premarket Notification K020006, and Rose K Post Graft (hexafocon A or enflufocon B) Rigid Gas Permeable Contact Lens, cleared in 510(k) Premarket Notification K013646, including an indication for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplastv."

    This means the "study" is a historical chain of prior 510(k) clearances and the demonstration that the current devices' physical properties and indications are substantially equivalent to those already cleared. There is no new clinical trial data presented in this specific 510(k) summary (K053124) to validate performance against novel acceptance criteria. Instead, the "acceptance criteria" are effectively met by demonstrating that the device is the same or very similar to already approved predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013762

    Validate with FDA (Live)

    Date Cleared
    2002-04-03

    (141 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K943177, K980741, K000795

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BOSTON ES® (enflufocon A), BOSTON EO® (enflufocon B) and BOSTON® XO (hexafocon A) Rigid Gas Permeable Contact Lenses are indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia and keratoconus) in aphakic and not-aphakic persons with non-diseased eyes. The lenses may be disinfected using a chemical disinfection system only.

    Device Description

    The BOSTON ES® (enflufocon A) and BOSTON EO® (enflufocon B) are Rigid Gas Permeable Contact Lens materials composed of aliphatic fluoroitaconate siloxanyl methacrylate copolymer with an ultraviolet absorber.

    BOSTON® XO (hexafocon A) is a rigid gas permeable contact lens material, composed of siloxanyl fluoromethacrylate copolymer containing an ultraviolet absorber.

    The color additives conform to 21 CFR Part 74 and/or 21 CFR Part 73. The lens may also be supplied clear (no tint).

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the BOSTON ES®, BOSTON EO®, and BOSTON® XO Rigid Gas Permeable Contact Lenses.

    It's important to note that this document is a 510(k) summary, which often points to prior submissions and substantial equivalence rather than detailing a new comprehensive clinical study for the specific device being cleared.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for these devices are primarily based on demonstrating substantial equivalence to a predicate device (BOSTON® II itafocon A). This means that the new devices are considered safe and effective because they are as safe and effective as a legally marketed device.

    The "performance" for these new devices is presented not as a comparison against specific numerical targets, but as their fundamental physical and optical properties, which are then implicitly considered acceptable because they fall within the parameters of what is deemed safe and effective for rigid gas permeable contact lenses, similar to the predicate.

    Acceptance Criteria Category (Implicit)Specific Criteria (from predicate/general RGP lens standards)Reported Device Performance (BOSTON ES, EO, XO)
    Substantial Equivalence to Predicate DeviceSafety and efficacy equivalent to BOSTON® II (itafocon A) lens.Achieved through similarity in materials, indications for use, and general design.
    Material CompositionAppropriate for RGP contact lenses, including UV absorber.BOSTON ES/EO: aliphatic fluoroitaconate siloxanyl methacrylate copolymer with UV absorber. BOSTON XO: siloxanyl fluoromethacrylate copolymer with UV absorber. Color additives conform to 21 CFR Part 74/73.
    Physical/Optical PropertiesWithin acceptable ranges for RGP lenses (specific gravity, refractive index, light transmittance, water content, oxygen permeability).BOSTON ES: Specific Gravity: 1.22 Refractive Index: 1.443 Light Transmittance: C.I.E. Y value - at least %
    Indications for UseCorrection of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia and keratoconus) in aphakic and non-aphakic persons with non-diseased eyes; chemical disinfection only.Matched to predicate device's indications, as stated in the submission.

    Notes:

    • Specific numerical acceptance criteria for each physical/optical property are not explicitly stated in this 510(k) summary, but are implied to be acceptable based on the predicate and general RGP lens standards.
    • The "at least %" for Light Transmittance in BOSTON ES and EO is an incomplete entry in the original document.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "The safety and efficacy of BOSTON ES® (enflufocon A), BOSTON EO® (enflufocon B) and BOSTON® XO (hexafocon A) Rigid Gas PermeableContact Lenses was demonstrated in 510(k) Premarket Notifications: K943177 cleared on August 25, 1994; K980741 cleared on May 11, 1998; and K000795 cleared on May 25, 2000, respectively."

    This means that the current 510(k) (K013762) is leveraging previous clearances. The specific details of the test sets (sample sizes, prospective/retrospective, country of origin) for those earlier 510(k)s are not provided in this summary document. This 510(k) relies on the fact that those previous submissions adequately demonstrated safety and efficacy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the summary. For 510(k) submissions, particularly for devices with a clear predicate, detailed expert adjudication for ground truth (as might be seen in AI/diagnostic device studies) is generally not required unless there's a novel aspect requiring new clinical data. The primary "ground truth" here is the established safety and efficacy of the predicate device.

    4. Adjudication Method for the Test Set

    This information is not provided in the summary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done or is mentioned. This type of study is specifically relevant for AI-powered diagnostic devices where human interpretation interacts with AI output. These contact lenses are a medical device, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a contact lens, not an algorithm. Its performance is inherent in its physical and material properties, and its interaction is directly with the human eye, not through an algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The primary "ground truth" for these contact lenses, as presented in this 510(k) summary, is:

    • Substantial Equivalence to a Predicate Device: The established safety and efficacy of the BOSTON II (itafocon A) Rigid Gas Permeable Contact Lens, which was approved under a Premarket Application (PMA P820065). This implies that the predicate device underwent comprehensive clinical trials and demonstrated safety and efficacy through clinical outcomes data, which then serves as the benchmark.
    • Physical and Optical Property Characterization: The physical properties of the new lenses (e.g., oxygen permeability, refractive index) are measured and implicitly compared against accepted standards for RGP lenses, which are themselves derived from years of clinical experience and outcomes data.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical device, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, this is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020006

    Validate with FDA (Live)

    Date Cleared
    2002-03-29

    (86 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K943177, K000795

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dyna Intra-Limbal Lens (enflufocon or hexafocon A) rigid gas permeable contact lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons. The lens may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration or following penetrating keratoplasty. The lens may be disinfected using a chemical disinfection system.

    Device Description

    The Dyna Intra-Limbal Lens (enflufocon A) or (hexafocon A) rigid gas permeable spherical contact lens is intended for the correction of highly irregular comeas occurring naturally or from disease such as Pellucid Marginal Degeneration and keratoconus or from post surgery complications such as tilted grafts. The design is larger in diameter, with a 11.2 mm standard diameter that is sized to fit within the limbus. A spherical peripheral system is employed to insure maximum exchange of tears and adequate cleansing of the cornea from flushing action behind the lens. The lens material attributes and lens parameters are found in the package insert.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on acceptance criteria and supporting study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided FDA 510(k) summary does not explicitly state numerical acceptance criteria for the Dyna Intra-Limbal Lens. Instead, it relies on the concept of "substantial equivalence" to predicate devices (Boston ES and Boston XO RGP lenses) and the demonstration of efficacy through a small clinical study for its new indication.

    Therefore, the "acceptance criteria" can be inferred as:

    • Safety profile equivalent to predicate devices.
    • Manufacturing adherence to specified standards (ANSI Z80.20-1998) and cGMP.
    • Clinical data supporting the efficacy of the new indication.

    Since specific quantitative acceptance metrics are not provided, I cannot populate a table with "reported device performance" against them in the traditional sense. The document states:

    Implicit Acceptance CriterionReported Device Performance
    Safety ProfileEquivalent to predicate devices (Boston ES/XO RGP lenses) based on pre-clinical toxicology and manufacturing/chemistry data.
    Manufacturing StandardsManufactured according to ANSI Z80.20-1998 and cGMP quality assurance program.
    Efficacy for New IndicationSupported by clinical data from 20 randomly selected patients with various corneal conditions.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 20 patients.
    • Data Provenance: The document does not specify the country of origin. It is a "clinical data presented" which implies a prospective study, though this is not explicitly stated. The nature of the device (contact lens) and the typical regulatory pathway for such devices often involve prospective clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number or qualifications of experts used to establish ground truth for the clinical data. It simply mentions "clinical data," implying a clinical trial where patient outcomes and observations by eye care professionals would serve as the basis for efficacy.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, an MRMC comparative effectiveness study was not done or reported. The study described is a direct clinical evaluation of the device's performance in patients, not a comparison of human readers with and without AI assistance. The device itself is a contact lens, not an AI diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable as the device is a contact lens, not an AI algorithm. No standalone algorithm performance study was mentioned.

    7. The Type of Ground Truth Used

    The ground truth used for evaluating the efficacy of the Dyna Intra-Limbal Lens in the clinical study was likely patient outcomes and clinical observations by eye care professionals. This would include metrics like visual acuity improvements, comfort, fit, and successful management of irregular corneal conditions, as assessed by the prescribing and follow-up clinicians. The document refers to "efficacy of the Dyna Intra-Limbal Lens" supported by "clinical data."

    8. The Sample Size for the Training Set

    This question is not applicable. The Dyna Intra-Limbal Lens is a medical device (contact lens), not an AI algorithm that requires a training set. The "training" in this context would refer to the historical data and experience used to design the device, which is not quantified in terms of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reason as #8. No ground truth for a "training set" of an AI algorithm was established. The "ground truth" for the device's design would stem from established ophthalmic knowledge, contact lens science, and the performance of predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K000795

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-05-26

    (74 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K943177

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BOSTON XO (hexafocon A) Contact Lens Material is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia, astigmatism and nresbyopia) in aphakic and not-aphakic persons with nondiseased eyes. The lens may be disinfected using a chemical disinfecting system only.

    Device Description

    The BOSTON XO Contact Lens Material, hexafocon A, is composed of aliphatic fluoroitaconate siloxanyl methacrylate copolymer. The color additives conform to 21 CFR part 74.3206. The hexafocon A material has an oxygen permeability, DK of 100, a specific gravity of 1.26, and the lens visible light transmittance of at least 70%. The hexafocon A name has been adopted by the United States Adopted Names Council (USAN).

    AI/ML Overview

    The provided text describes the BOSTON XO (hexafocon A) Contact Lens Material, and its submission for 510(k) clearance, asserting substantial equivalence to the predicate device, BOSTON ES (enflufocon A). The document focuses on preclinical and clinical studies conducted to support this claim.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" for the clinical study in the format of a target value to be met. Instead, it concludes equivalence to the predicate device. The preclinical tests had requirements that were met.

    Test CategoryAcceptance Criteria/RequirementReported Device Performance
    Preclinical Testing
    In-Vitro CytotoxicityMeets requirements of USP XXII for Agar Diffusion Test"The test article meets the requirements of the Agar Diffusion Test."
    Acute Ocular IrritationNo ocular irritation produced"produced no ocular irritation."
    Systemic InjectionMeets requirements of Systemic Injection Test; considered non-toxic"The lens material meets the requirements of the Systemic Injection Test and is considered non-toxic."
    Shelf LifeNot required for clearance (due to <1% water content and dry shipping)"shelf-life studies are not required for clearance of this material."
    Solution CompatibilityNo significant changes to lens parameters after 30 cycles of cleaning and conditioning"There were no significant changes to lens parameters after 30 complete cycles."
    Clinical Testing
    Overall Safety & EfficacyEquivalent to predicate device (Boston ES)"The sponsor concludes that BOSTON XO (hexafocon A) contact lens material is equivalent in safety and efficacy to the predicate device, Boston ES (enflufocon A)."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study):
      • Enrolled: 128 eyes (64 patients)
      • Completed Study: 102 eyes (51 patients)
    • Data Provenance: The document does not explicitly state the country of origin. It does state that "3 Investigators" were involved, which often implies a multi-site study, but specific locations are not provided. The study was prospective, as patients were "entered into the study" and followed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts or their qualifications used to establish "ground truth" for the clinical study in the way one might assess an AI diagnostic device. Instead, the "safety and efficacy measures" were assessed by the clinical investigators. The document mentions "3 Investigators" who entered patients into the study. Their specific qualifications are not detailed beyond "Investigators."

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method for the clinical study's safety and efficacy measures. The results are presented as conclusions from the study, without mentioning independent adjudication or consensus review of specific findings.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not conducted. This document describes the evaluation of a contact lens material, not an AI-assisted diagnostic system. Therefore, assessing how much human readers improve with AI vs. without AI assistance is not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Performance Study

    No, a standalone performance study in the context of an "algorithm only" or AI system was not conducted. This document describes the evaluation of a contact lens material, which does not involve an algorithm or AI.

    7. Type of Ground Truth Used

    For the preclinical tests, the "ground truth" was based on established standard testing methodologies (e.g., USP XXII, Systemic Injection Test requirements) and observable physical changes.

    For the clinical study, the "ground truth" for safety and efficacy was established through clinical observations and measurements by the investigators. These included:

    • Adverse Events (self-reported and observed)
    • Positive Slit Lamp Findings
    • Symptoms/Complaints (patient-reported)
    • Keratometry Changes (objective measurement)
    • Refractive Changes (objective measurement)
    • Lens Visual Acuity (objective measurement)
    • Lens VA Line Changes (objective measurement)
    • Lens Deposits (objective observation/grading)
    • Lens Wettability (objective observation/grading)

    This constitutes a form of expert assessment and objective measurement in a clinical setting.

    8. Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI. This is a medical device (contact lens material) and its evaluation methods do not involve AI training. Therefore, this information is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set for an AI system, this question is not applicable.