Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K980417

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-05-04

    (90 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K935668, K810624, K790062, K942391, K800823

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medrad Disposable MRI Kit with Check Valve is indicated for use in an MRI environment. The components of the kit are used to assist in filling and delivering intravascular contrast media and common flushing agents at various volumes and flow rates.

    Device Description

    The Medrad Disposable MRI Kit with Check valve is substantially equivalent to the current Medrad Qwik-Fit MRI Disposables Kit (K943189). The Medrad Disposable MRI Kit with Check valve has the same intended use as the predicate device - to assist in filling and delivering intravascular contrast media and common flushing agents at various volumes and flow rates in an MR application. The kit may consist of one or all of the following:

    • 2- 65 ml syringes (K935668)
    • 96" Low Pressure Connector Tube/T-connector assembly (K943189) .
    • . Fluid Delivery Device (Spike) - (K810624)
    • Normally closed check-valve B. Braun/Burron (K790062) ●
    • Female/Female Luer Adapter B.Braun/Burron (K942391) .
      The only difference between this kit and the predicate device is that this kit contains a normally closed check valve and a female/female luer Adapter. Device Verification and Validation Testing has been performed.
      The normally closed check-valve is a attached to the 96" LLPCT at the Tconnector (K943189) and is used to restrict flow in one of two directions. The device is normally closed requiring a minimum amount of force to open. The check-valve attaches to the 'A' syringe and remains closed during saline flush from the 'B' syringe to prevent contrast trickle during this process.
      The Female/Female Luer Adapter is a disposable device used to provide a fluid path between two vessels with male luer terminations. This device can be compared to the Medrad Ouick Fill Tube (K800823) which is also a filling device.
      All other components of the kit are identical to those in K943189.
      The devices are sterilized using Ethylene Oxide gas. Validation is performed according to AAMI guidelines.
    AI/ML Overview

    Here's an analysis of the provided text regarding the Medrad Disposable MRI Kit with Check Valve, focusing on acceptance criteria and the supporting study information:

    It's important to note that this document is a 510(k) summary, which is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. As such, the level of detail regarding comprehensive studies and acceptance criteria is different from what might be found in a full efficacy study report for a new drug or high-risk medical device. The focus here is on demonstrating that the new device does not raise new issues of safety or effectiveness compared to its predicate.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly present acceptance criteria in a quantitative, pass/fail table format, nor does it detail a standalone study proving the device meets specific performance metrics with numerical results for each criterion regarding its primary function (fluid delivery). Instead, it demonstrates substantial equivalence by comparing features and performing validation tests related to manufacturing and safety attributes.

    However, we can infer some "acceptance criteria" based on the validation tests performed and the comparison to the predicate devices. The "reported device performance" is primarily stated as compliance with established standards and equivalence to the predicate.

    Acceptance Criterion (Inferred from Validation/Comparison)Reported Device Performance
    Sterility (Microbial Challenge-Overkill Method)Achieves a Sterility Assurance Level (SAL) of $10^{-6}$ using biological indicators of Bacillus Subtilis var. niger ($1 \times 10^6$ spores per indicator).
    Residual Ethylene Oxide (EtO) Levels< 4 ppm after 8 days outgassing (via simulated-use aqueous extractions).
    Residual Ethylene Chlorohydrin Levels< 4 ppm after 8 days outgassing (via aqueous extraction).
    Residual Ethylene Glycol Levels< 18 ppm after 8 days outgassing (via aqueous extraction).
    PyrogenicityComplies with USP Rabbit Testing and/or USP Limulus Amebocyte Lysate (LAL) test methods.
    BiocompatibilityComplies with ANSI-AAMI(ISO) 10993-1 through 11 and US Pharmacopoeia XXII, <87 & 88> Biological Reactivity Test (Invitro and Invivo) for Cytotoxicity, Sensitization, Systemic Tox, Intracutaneous Tox, Pyrogen Mat'l Mediated, Hemolysis, Particulate analysis, Bacterio-/Fungistasis, Physicochemical analysis (aqueous/non-aqueous), and Infrared analysis.
    Packaging Compatibility with SterilizationThe polystyrene tray and Tyvek lid are heat-sealed and compatible with an EtO Sterilization Cycle.
    Shelf LifeConfirmed at least a two-year shelf life from the date of sterilization using an accelerated aging sequence (1 week at 130°F/70-80% RH, 1 day at -4°F, 1 week at 130°F/<15% RH, equivalent to 6 months normal aging). Ongoing testing aims for a five-year shelf life. Flow tests, pressure tests, and ARO tests (likely refers to "Air Removal" or similar functional test) are conducted to ensure product efficacy after accelerated aging. Specific results are not provided.
    Standard Luer Fitting (Female/Female Luer Adapter)"Yes - ANSI 70.1" (implies compliance with the standard).
    Standard Luer Fitting (LLPCT)"ANSI 70.1" (implies compliance with the standard).
    Functionality of Check-ValveRestricts flow in one of two directions, "normally closed requiring a minimum amount of force to open." Prevents contrast trickle during saline flush from the 'B' syringe.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sterility Validation:

      • Sample Size: 50 test products per half-cycle verification run. There were three half-cycle verification runs, totaling 150 test products.
      • Data Provenance: Not explicitly stated as retrospective or prospective, or country of origin, but it is implied to be internal testing conducted by Medrad, Inc. as part of their sterilization validation process for a new device submission. This would be prospective testing for the specific device.
    • Residual Levels:

      • Sample Size: Not specified for the pre-release testing.
      • Data Provenance: Implied as internal prospective testing.
    • Shelf Life Testing:

      • Sample Size: Not specified for the accelerated aging tests or the subsequent flow/pressure/ARO tests.
      • Data Provenance: Implied as internal prospective testing.
    • Biocompatibility/Pyrogenicity:

      • Sample Size: Not specified.
      • Data Provenance: Implied as internal prospective testing.

    For the functional aspects of the check-valve and luer adapter, the document states "Device Verification and Validation Testing has been performed" but does not provide details on sample sizes or specific test results, instead relying on a comparison of features to the predicate device and the fact that other components are identical to a previously cleared device (K943189).


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to the provided document. The Medrad Disposable MRI Kit with Check Valve is a hardware device (disposable kit) for fluid delivery, not a diagnostic algorithm or a device requiring expert interpretation of results. Therefore, there's no "ground truth" in the medical imaging or diagnostic sense established by human experts for the device's functional performance or safety. The "ground truth" for the device's properties (like sterility, biocompatibility, luer fitting compliance) is established through technical standards, laboratory tests, and reference protocols.


    4. Adjudication Method for the Test Set

    This is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., medical image reading) to resolve discrepancies. This device does not involve such human interpretation in its primary function or testing for a 510(k) submission.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This is not applicable. An MRMC comparative effectiveness study is used to evaluate the diagnostic performance of human readers, sometimes comparing performance with and without AI assistance for tasks like disease detection or classification. This device is a fluid delivery kit and does not involve AI or human diagnostic reading in its function.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. This device is a physical medical kit, not an algorithm, so a standalone algorithm performance study is irrelevant.


    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the established reference for the characteristics being tested. For this device, the ground truth types are:

    • Technical Standards and Reference Protocols: For sterility (ANSI/AAMI ST27-1988), residual levels (ANSI/AAMI ST29-1988 and ST30-1989), pyrogenicity (USP Rabbit Testing, USP LAL), biocompatibility (ANSI-AMMI/ISO 10993-1 through 11, USP XXII <87 & 88>), and luer fitting compliance (ANSI 70.1). These external, validated standards serve as the "ground truth" for compliance.
    • Predicate Device Characteristics: The existing Medrad Qwik-Fit MRI Disposables Kit (K943189) and its components serve as a ground truth for substantial equivalence, specifically for features that are identical or closely matched.
    • Functional Specifications: The intended behavior of the check-valve (restricting flow, preventing trickle) and the luer adapter (fluid transfer) define the functional "ground truth" against which device performance is validated internally.

    8. The Sample Size for the Training Set

    This is not applicable. A "training set" refers to data used to train a machine learning algorithm. This device is not an AI/ML product.


    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K970255

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-04-14

    (82 days)

    Product Code
    Regulation Number
    880.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K942391

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This device is intended to be used as a conduit for the delivery of I.V. fluids from a container to a patient's vascular system through a needle or cather inserted into a vein. Additionally, this device is intended to serve as an injection site for other influsion fluids. This device is intended to be used only by trained professionals in a clinical or hospital environment.

    Device Description

    The KippMed I.V. Manifold is substantially equivalent to the Braun Safesite Manifold, Minor differences in design specifications have been listed in Table One. Both devices are intended to provide connections to an I.V. Administration Set. Both devices provide a multi-port connector with a female leur-lock adapter, three normally closed backcheck valves and a connection adapter. The inlet connection adapter for both devices includes a free-spinsing hub/leur lock to connect to I.V. lines. Both devices are sterile and nonpyrogenic.

    AI/ML Overview

    This document (K970255) describes an I.V. Manifold, a low-risk medical device. The information requested in your prompt (acceptance criteria, study details, sample sizes, expert qualifications, etc.) is typically associated with studies for higher-risk devices, particularly those involving AI/ML components or complex diagnostic functions with human interpretation.

    For this I.V. Manifold, the "acceptance criteria" and "study" are focused on demonstrating substantial equivalence to a predicate device and confirming basic safety and performance requirements, not on diagnostic accuracy, reader performance, or AI efficacy.

    Here's a breakdown based on the provided text, addressing your points where applicable, and explaining why others are not relevant for this type of submission:

    1. A table of acceptance criteria and the reported device performance

    For this device, the "acceptance criteria" are essentially demonstrating substantial equivalence to the predicate device in terms of design, materials, intended use, and meeting basic safety standards (sterilization, biocompatibility). The "reported device performance" refers to the device meeting these equivalence points and safety validations.

    Acceptance Criteria CategoryKippMed I.V. Manifold Performance
    Intended UseConduit for I.V. fluids, injection site for infusion fluids. Substantially equivalent to predicate.
    Technological CharacteristicsMulti-port connector, female leur-lock, three normally closed backcheck valves, connection adapter with free-spinning hub/leur lock. Substantially equivalent to predicate (minor differences in design specifications detailed in Table 1, not provided in excerpt).
    Sterility Assurance Level (SAL)10⁻⁶
    Material BiocompatibilityAll tested components were biocompatible according to ISO 10993 and FDA G95-1.
    Non-pyrogenicDevice is non-pyrogenic.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and not provided in the document. The "tests" for this device are primarily related to:

    • Sterilization validation: This typically involves laboratory testing on a statistically significant sample of devices to demonstrate the Sterility Assurance Level (SAL). The specific sample size or provenance is not detailed in this summary but would be part of the full validation report.
    • Biocompatibility testing: This involves in vitro and potentially in vivo tests on the raw materials and finished components. Specific sample sizes for these material tests are not in the summary.
    • Bench testing for functional equivalence: This would confirm aspects like leak integrity, flow rates, and connection strength. The document mentions "minor differences in design specifications" and a comparison in "Table 1," implying such testing was done, but details are not provided. These are typically laboratory tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. There is no concept of "ground truth" or "expert readers" in the context of an I.V. manifold submission. This device performs a mechanical function, not a diagnostic one requiring human interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for the same reasons as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an I.V. manifold; it does not involve AI, human readers, or diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "truth" for this device is its physical integrity, sterility, biocompatibility, and functional equivalence to its predicate. These are established through validated testing methodologies, not expert consensus on diagnostic findings or clinical outcomes data in the way you might see for an AI diagnostic device.

    8. The sample size for the training set

    Not applicable. This device does not use machine learning or AI, and therefore has no "training set."

    9. How the ground truth for the training set was established

    Not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1