    K984447

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-01-08

    (25 days)

    Product Code
    Regulation Number
    890.3850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K952641, K941989

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Model Action AT II Manual Wheelchair is to provide mobility to persons limited to a sitting position.
    To provide mobility to persons limited to a sitting postion.

    Device Description

    The Invacare Corporation Model Action AT II is a manually operated, self propelled, manual, mechanical wheelchair. It's intended function and use is to provide mobility to persons who may be restricted to a sitting position. This product may also be used as an attendant propelled patient transport device in a health care environment such as a hospital, nursing home or extended care facility.
    The product consists basically of the wheelchair frame, larger rear wheels with handrims for propelling the chair, and smaller front pivoting casters for steering and turning.
    The wheelchair frame is constructed from both 1/4" and 1" outside diameter (OD) round, mechanical, steel tubing. The side frames are of welded construction and are secured to the remaining frame members using screws and bolts. This device is a rigid, "non folding" type wheel chair that incorporates a solid seating surface. This type of seat makes the chair easily adaptable to the various types of wheelchair cushions and seating systems currently available in the market.
    The Action AT II also includes a "Tilt in Space" feature which allows the seat and back of the wheelchair to be tilted. This feature is used for those patients who require a tilt feature for stability, comfort or head control. It also serves as an attendant aid in those situations where a patient need to be tilted to be fed, or attended to in some fashion. The attendant version of the chair also is available with a stroller type push handle for ease of attendant pushing.
    The "Tilt in Space" feature is manually operated and is activated by two release levers located at the rear of the wheelchair. Adjusting the tilt is achieved by squeezing the two release lever handles inward simultaneously. When the levers are squeezed, flexible cables attached to each release lever diseneage gear rack segments from independent actuator rods which are notched to match the teeth of the gear segments. Once the desired tilt angle has been obtained, the handles are released and the chair will remain at the angle chosen. The chair has a tilt angle range of 55 degrees.

    AI/ML Overview

    The provided text is a 510(k) summary for the Invacare Model Action AT II Manual Wheelchair. It focuses on demonstrating substantial equivalence to predicate devices and adherence to performance standards rather than clinical study results involving human readers and AI. Therefore, much of the requested information regarding AI performance, human expert involvement, and specific clinical study designs is not available in this document.

    However, I can extract the information related to acceptance criteria and the study that proves the device meets those criteria based on the provided text.

    Here's the breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Standards Met)Reported Device Performance
    Rehabilitation Engineering Society of North America (RESNA) Standard ANSI/RESNA WC/Vol.1-1998 "Requirements and Test Methods for Wheelchairs (Including Scooters)"The Invacare Action AT II manual wheelchair meets the applicable performance requirements specified in this standard.
    DIS ISO 7176 (1992) Part 8: Requirements and Test Methods for Static, Impact and Fatigue StrengthsThe Invacare Action AT II manual wheelchair meets the applicable performance requirements specified in this standard.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document refers to the device meeting performance requirements and test methods of established standards (RESNA and ISO). It does not describe a clinical test set in the traditional sense of patient data. Instead, it indicates that the device was subjected to engineering and materials testing as outlined by these standards. Therefore:

    • Sample Size: Not applicable in the context of patient data. The "sample" would refer to the physical wheelchair unit(s) subjected to the engineering tests. The specific number of units tested is not stated.
    • Data Provenance: Not applicable in the context of patient data. The "tests" are engineering performance tests, not clinical data from patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    Not applicable. The ground truth for performance is established by the specifications and test methodologies defined in the RESNA and ISO standards, which are engineering and material science standards, not medical expert consensus.

    4. Adjudication Method for the Test Set:

    Not applicable. This concept applies to human expert review of clinical data, not engineering performance testing against established standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a manual wheelchair, not an AI-powered diagnostic or assistive device where human reader performance would be a factor.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This is a manual wheelchair, and there is no algorithm or AI component involved.

    7. The Type of Ground Truth Used:

    The "ground truth" for the device's performance is established by the engineering and performance requirements specified in the RESNA and ISO standards. These standards define the acceptable limits for static, impact, and fatigue strengths, as well as general wheelchair functionality.

    8. The Sample Size for the Training Set:

    Not applicable. This document describes a medical device (manual wheelchair) which is evaluated against engineering performance standards. There is no AI or machine learning component, and thus, no training set for an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for an algorithm.