Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K014282

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-06-26

    (181 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K881657, K945359, K941381, K941536

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with ventilators, anesthesia machines, and open flow systems where filtration of inspired and / or expired gases is desired and to add and retain moisture in the circuit as required.

    For patients with a tidal volume 150-500 ml (pediatrics) and >250 ml (adults).

    Single patient use. Duration of use up to 24 hours.

    Device Description

    The Munktell Bacstop Filter and Filter / HME are available in multiple sizes and incorporate standard 15 / 22 mm connectors with a gas sampling luer port. The depth filter uses electrostatic media for filtration and a foam media for the HME media.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Munktell Bacstop Filter and Filter/HME:

    Acceptance Criteria and Device Performance

    AttributeAcceptance Criteria (from Predicate/Standard)Reported Device Performance (Munktell Bacstop)
    Resistance to Flow (Adult)< 2.1 cm H2O @ 60 lpm< 2.1 cm H2O @ 60 lpm
    Resistance to Flow (Pediatric)< 0.82 cm H2O @ 15 lpm< 0.82 cm H2O @ 15 lpm
    Bacterial Filtration (BFE)(Not explicitly stated as a separate criterion, but predicate devices' performance would set the bar)99.999% (Nelson Lab. testing)
    Viral Filtration (VFE)(Not explicitly stated as a separate criterion, but predicate devices' performance would set the bar)99.99+% (Nelson Lab. testing)
    Humidification Output (Adult)(Not explicitly stated as a separate criterion, but predicate devices' performance would set the bar)31 mg H2O /L at TV of 1000 ml
    Humidification Output (Pediatric)(Not explicitly stated as a separate criterion, but predicate devices' performance would set the bar)32 mg H2O /L at TV of 250 ml
    Tidal Volume Range (Pediatric)(Implied by predicate devices and intended use)150 - 500 ml
    Tidal Volume Range (Adult)(Implied by predicate devices and intended use)> 250 ml
    Dead Space(Implied by predicate devices' performance for similar devices)23 to 85 ml
    Intended UseFor use with ventilators, anesthesia machines, and open flow systems where filtration of inspired/expired gases is desired and to add and retain moisture. For pediatric (150-500ml TV) and adult (>250ml TV) patients. Up to 24 hours use.Same
    Single Patient UseYesYes
    Environment of UseHome, Hospital, Sub-acute Institutions, Emergency ServicesSame
    Gas Sampling PortYes (as per predicate design, implicitly)Yes
    Standard 15/22 mm connectorsYes (as per predicate design, implicitly)Yes
    ISO 5356-1 Conical 15/22Yes (Standard conformance)Yes
    ISO 594-2 Luer FittingsYes (Standard conformance)Yes
    ISO 9360- HME moisture outputYes (Standard conformance)Yes

    Study Information

    Based on the provided Non-Confidential Summary of Safety and Effectiveness (510(k) K014282), the "study" is primarily a bench-testing and comparison to predicate devices to demonstrate substantial equivalence, rather than a clinical trial with human subjects.

    Here's the breakdown of the requested information based on the document:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Sample size: Not explicitly stated for specific tests. The document refers to "Nelson Lab." for bacterial and viral filtration, implying laboratory testing. For other parameters like resistance and humidification, it would involve testing multiple units to ensure consistency, but specific numbers are not provided.
      • Data provenance: "Nelson Lab." is mentioned for BFE and VFE, which is a known independent testing laboratory. The overall data likely originates from testing performed by or for Munktell Filter AB (Sweden). The studies appear to be prospective bench tests designed to evaluate specific performance attributes.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • This question is not applicable to this type of device submission. Medical device filters do not typically use expert ground truth for performance metrics in the same way an imaging AI algorithm would. Performance is measured against physical standards and laboratory methods (e.g., measuring pressure drop, bacterial retention, water vapor content). The "ground truth" is defined by the physical or microbiological measurement standards themselves.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • None. Adjudication methods are typically used in clinical studies or expert review of image interpretation, which is not relevant for the bench testing of a breathing circuit filter.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • No. An MRMC study is relevant for an AI-powered diagnostic imaging device involving human readers. This submission is for a medical filter, which does not involve human readers or AI assistance in its function.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      • Yes, in spirit. The performance metrics (BFE, VFE, resistance, humidification) are "standalone" in that they measure the device's intrinsic physical and biological filtering capabilities without human intervention during its operation. There is no "algorithm" in the sense of software interpreting data, but the filter itself performs its function independently.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • The "ground truth" for the performance claims in this submission is the measured output from standardized laboratory testing. For example:
        • Bacterial Filtration (BFE) and Viral Filtration (VFE): Ground truth is established by standardized microbiological testing methods (e.g., aerosol challenge tests) that measure the percentage of bacteria/viruses retained by the filter.
        • Resistance to flow: Ground truth is established by physical measurement of pressure drop across the filter at specified flow rates using calibrated equipment.
        • Humidification Output: Ground truth is established by standardized tests (e.g., ISO 9360) that measure the absolute humidity delivered by the HME.
    7. The sample size for the training set

      • Not applicable. This device is hardware (a filter), not an AI algorithm that requires a "training set."
    8. How the ground truth for the training set was established

      • Not applicable. As above, there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013122

    Validate with FDA (Live)

    Date Cleared
    2002-04-19

    (212 days)

    Product Code
    Regulation Number
    868.5260
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K941536, K941381, K002201

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with ventilators, anesthesia machines, and open flow systems where filtration of inspired and / or expired gases is desired and to add maintain and retain moisture for the exhaled breathe of the patient. Model with tidal volumes ranging from 20-70 cc for neonate, 70-250 cc for pediatric and >150 cc for adults. Use up to 24 hours.

    Device Description

    The EMS Filter and combined Filter / HME are available in multiple sizes and shapes, rectangular and round, and incorporate standard 15 / 22 mm connectors with a gas sampling luer port. The depth filter uses electrostatic media for filtration and a foam media for the HME media.

    AI/ML Overview

    Here's an analysis of the provided text regarding the EMS Electra Filter and Filter/HME device, focusing on acceptance criteria and the study proving it meets those criteria:

    The provided document is a 510(k) Premarket Notification, which is a submission to the FDA to demonstrate that the device to be marketed is substantially equivalent to a legally marketed predicate device. This type of submission relies on comparing the new device's characteristics and performance to those of existing, legally marketed devices, rather than comprehensive clinical trials for novel devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly define "acceptance criteria" as a separate, formal list against which the device was tested as one might see in a clinical trial. Instead, it presents general technical characteristics of the proposed device and demonstrates their equivalence to predicate devices. The "acceptance criteria" are implicitly defined by the performance of the predicate devices and existing standards for such devices.

    AttributeAcceptance Criteria (Implied by Predicate/Standard)Reported Device Performance (EMS Electra)
    Dead Space (ml)- (Should be comparable to predicate devices)10 to 65 ml
    Resistance to flow (Adult)≤3 cm H2O @ 60 lpm≤3 cm H2O @ 60 lpm
    Resistance to flow (Pediatric)≤1.0 cm H2O @ 20 lpm1.0 cm H2O @ 20 lpm
    Resistance to flow (Neonate)≤0.5 cm H2O @ 5 lpm0.5 cm H2O @ 5 lpm
    Bacterial Filtration Efficiency (BFE)99.999+% (Based on Nelson Lab. results for predicate)99.999+%
    Viral Filtration Efficiency (VFE)99.99+% (Based on Nelson Lab. results for predicate)99.99+%
    Weight (gm)- (Should be comparable to predicate devices)10 to 35 gm
    Humidification Output (Adult, 1000cc TV)32 mg H2O/L32 mg H2O/L
    Humidification Output (Pediatric, 250cc TV)32 mg H2O/L32 mg H2O/L
    Humidification Output (Neonate, 50cc TV)30 mg H2O/L30 mg H2O/L
    Tidal Volume Range (Neonate)20-70 cc20-70 cc
    Tidal Volume Range (Pediatric)70-250 cc70-250 cc
    Tidal Volume Range (Adult)>150 cc>150 cc
    Standards ComplianceISO 5356-1, ISO 594-2, ISO 9360Yes (Compliance asserted)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document details performance data derived from laboratory testing, specifically mentioning "Nelson Lab." for Bacterial and Viral filtration efficiency. This indicates the data is likely from professional testing laboratories rather than human clinical trials.

    • Sample Size: Not explicitly stated for each test. For filtration efficiency, usually multiple samples of the filter media are tested. For resistance and humidification, often a set number of devices are tested according to a standard protocol.
    • Data Provenance: The document states "Nelson Lab." for BFE and VFE, implying a specialized testing facility. The country of origin is not specified, but Nelson Labs is a well-known US-based testing laboratory. The data is prospective in the sense that the new device was fabricated and then tested to determine its performance characteristics. It is not retrospective in the sense of analyzing past patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    This information is not applicable as the evaluations are based on objective physical and biological performance tests, not subjective expert assessment of images or patient outcomes. The "ground truth" for these tests are the established scientific methodologies and standards by which filter performance, resistance, and humidification are measured.

    4. Adjudication Method for the Test Set:

    This is not applicable as the evaluation relies on direct measurement of physical properties and performance characteristics, not on subjective assessments that would require adjudication (e.g., medical image interpretation).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    This is not applicable. The device is a filter/heat and moisture exchanger (HME), not an imaging or diagnostic AI system. Therefore, an MRMC study related to human reader improvement with AI assistance is irrelevant to this device.

    6. Standalone (Algorithm Only) Performance:

    This is not applicable. The device is a physical medical device (a filter), not an algorithm or AI system.

    7. Type of Ground Truth Used:

    The ground truth used is objective laboratory measurement data and established performance standards. For example:

    • For resistance, the ground truth is the measured pressure drop across the filter at specified flow rates.
    • For filtration efficiency, the ground truth is the measured percentage of bacteria/viruses removed according to standardized laboratory testing protocols (e.g., using bacterial aerosols like Staphylococcus aureus).
    • For humidification output, the ground truth is the gravimetric measurement of water vapor produced by the HME under specific temperature, humidity, and tidal volume conditions as per ISO 9360.

    8. Sample Size for the Training Set:

    This is not applicable. Since this device is a physical product and not an AI/machine learning model, there is no "training set." Its design and performance capabilities would be based on engineering principles, materials science, and testing iterations, not data training.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable for the same reason as point 8. The "ground truth" for the device's design would stem from engineering specifications, industry standards (like ISO for connectors and HME moisture output), and the performance characteristics observed in predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1