Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121599

    Validate with FDA (Live)

    Date Cleared
    2012-11-09

    (161 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K935958

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

    Device Description

    The intended use of the Crit-Line III (CLM III) Monitor is as a continuous real-time monitor for non-invasive hematocrit, oxygen saturation, and percent change in blood volume measurement. The CLM III is used primarily as a monitor for dialysis treatments.

    The CLM III uses the principle of light absorption and scattering through the blood under test to measure oxygen saturation (O2 SAT) and hematocrit (HCT). The HCT values are then used to calculate the related Blood Volume (BV) percentage change relative to the starting BV based on the beginning HCT.

    The CLM III consists of a microprocessor-controlled main circuit board located within the display housing. The display housing contains the sensor clip driver measurementcalculation circuitry and display screen circuitry. The sensor clip with optical system (LED array and a photodetector array) is tethered to the main housing and circuit board by a multi-conductor cable. The cable connects to a single-use, sterile cuvette (K935958) installed in-line with the arterial line of the dialysis tubing set.

    Calculations performed on the main circuit board within the display housing yield the hematocrit, oxygen saturation and the related value of blood volume measurement results and are displayed on the LCD display screen.

    Modifications to the previously cleared Crit-Line III Monitor include:

    1. Miniaturization/Re-packaging

    The CLiC Monitor incorporates all hardware/electronic components - miniaturized and contained within the sensor clip housing, for measuring indicated parameters (HCT O2 SAT). The primary modification of the device involves integrating a miniaturized microprocessor with the required electronic circuitry and software to drive the existing optical system (LED and photo-detector array) within the plastic sensor clip housing. The CLiC utilizes the same, previously cleared Crit-Line technology (K972470), sensor clip and optical system. The operating principles and fundamental measurement methods of the CLiC are the same as the predicate device.

    1. Distributed Process Technology

    A standard USB cable replaces the tethering multi-conductor cable used in the CLM III Monitor. USB-compatible hardware and software are incorporated in the CLiC housing. The CLiC can be hosted on a medical grade computer system (Windows OS) that can interface with, power a USB peripheral and operate supporting drivers. CLiC interlocks with a singleuse, sterile cuvette (K935958) which is located in-line with the arterial line of the dialysis tubing set for indicated measurements. The lightwave measurements are performed autonomously by the CLiC. Blood parameters are then calculated and transmitted to the host computer display microprocessor in text format by means of the USB interface. The accompanying display software installed on the host computer system facilitates the display of measured parameters: hematocrit, oxygen saturation and blood volume changes on the display screen.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study details for the CLiC Monitor (K121599) based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The 510(k) summary for the CLiC Monitor focuses on demonstrating substantial equivalence to its predicate device, the Crit-Line III Monitor (K972470), rather than establishing new, specific numerical acceptance criteria for its performance metrics (hematocrit, oxygen saturation, and percent change in blood volume).

    The key "acceptance criteria" presented are:

    • The CLiC Monitor's accuracy is equivalent to the predicate device specifications for both hematocrit and oxygen saturation measurements.
    • The percentage blood volume comparisons are also equivalent due to the equivalence in hematocrit.
    • The device functions as intended and met pre-determined acceptance criteria (though these specific criteria are not quantitatively laid out in the summary).
    • The modifications did not affect the essential performance of the device.
    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Hematocrit Measurement AccuracyEquivalent to predicate device specifications.Shown to be equivalent to predicate device specifications.
    Oxygen Saturation Measurement AccuracyEquivalent to predicate device specifications.Shown to be equivalent to predicate device specifications.
    Percent Change in Blood VolumeEquivalent to predicate device specifications (derived from HCT).Equivalent to predicate device specifications.
    Essential PerformanceModifications do not affect essential performance.Design verification and validation tests confirmed that modifications did not affect essential performance.
    Intended FunctionDevice functions as intended.Test results demonstrated the modified CLiC Monitor functions as intended and met pre-determined acceptance criteria.
    Safety and EffectivenessDevice is safe and effective for its intended use.Test results from performance testing (in vitro bench testing), safety testing, and usability testing indicate the modified CLiC device is substantially equivalent to the predicate and remains safe and effective for its intended use.
    ComplianceAdherence to relevant standards (UL 60601-1/IEC 60601-1, IEC 60601-1-2)Certified to comply with UL 60601-1/IEC 60601-1, and EMC standard 60601-1-2.
    Risk MitigationPotential hazards are identified and mitigated; risks acceptable.A risk analysis per ISO 14971 was completed, and all potential risks were deemed acceptable after mitigation.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document mentions "performance testing (in vitro bench testing), safety testing and usability testing," and "System verification and validation testing." However, no specific sample sizes for test sets (e.g., number of patients, number of blood samples, specific data points) are provided. The data provenance (country of origin, retrospective/prospective) is also not specified. The focus is on demonstrating equivalence to the predicate device through technical evaluation rather than large-scale clinical trials.

    3. Number of Experts and Qualifications for Ground Truth:

    The document does not mention using experts to establish ground truth for a test set in the context commonly associated with AI/algorithm performance studies. The evaluation is against the performance of a known predicate device and compliance with engineering standards.

    4. Adjudication Method:

    Given that there's no mention of experts establishing a ground truth for a test set, an adjudication method (like 2+1 or 3+1) is not applicable and therefore not mentioned in this 510(k) summary.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done or mentioned. This submission is for a device based on an existing technology (Crit-Line III) with modifications related to miniaturization, distributed process technology, and software, rather than a novel AI diagnostic algorithm requiring human-in-the-loop performance evaluation.

    6. Standalone Performance:

    The 510(k) summary describes the standalone performance of the CLiC Monitor relative to its predicate. The CLiC performs the lightwave measurements autonomously and transmits blood parameters for calculation and display. The core measurement functions (hematocrit, oxygen saturation, blood volume changes) are evaluated based on their equivalence to the predicate device's established performance specifications. The "algorithm" in this context refers to the inherent calculations performed by the device, and its stand-alone accuracy is assessed against the predicate's known accuracy.

    7. Type of Ground Truth Used:

    The ground truth used for performance validation is implicitly the established performance specifications and measurements of the legally marketed predicate device, the Crit-Line III Monitor. The CLiC Monitor is evaluated to ensure its measurements of hematocrit, oxygen saturation, and calculated blood volume changes are equivalent to those obtained from the predicate device under similar conditions. This involves direct comparison to a validated, previously cleared device, likely through in vitro bench testing (as mentioned).

    8. Sample Size for the Training Set:

    This device is not presented as an AI/machine learning device that requires a "training set" in the conventional sense. It uses existing, cleared technology and optical principles. Therefore, no specific sample size for a training set is applicable or mentioned.

    9. How the Ground Truth for the Training Set was Established:

    As this is not an AI/ML device with a training set, the question of how ground truth was established for a training set is not applicable. The device functions based on established light absorption and scattering principles and signal processing, akin to an updated version of an existing physical measurement device. Its performance is validated against the known performance characteristics of its predecessor.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1