510(k) Data Aggregation

    K Number
    K201395

    Validate with FDA (Live)

    Date Cleared
    2020-10-28

    (154 days)

    Product Code
    Regulation Number
    870.1650
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The proposed device is intended for the injection of contrast media or saline. This syringe use with US legally marketed angiographic injectors.

    Device Description

    The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1. The connection tubes, which is used to connect the syringe and the catheter. The tubes are available in two configurations, which are Type Y and Type T tube. The different between these tubes is tube shape, which is available in Y shape for Y tube and T shape for T tube. J shape tube, which is used to draw contrast media/ saline into the syringe barrel. Spikes, which are used to draw contrast media/saline into the syringe barrel.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: "Sterile High-pressure Angiographic Syringes for Single-use". This type of submission focuses on demonstrating "Substantial Equivalence" to a legally marketed predicate device, rather than proving novel effectiveness or safety through extensive clinical trials.

    As such, the document does not describe a study to prove the device meets acceptance criteria in the way you've outlined for an AI-powered diagnostic device. Instead, it details non-clinical tests to show the proposed device performs comparably to the predicate and meets relevant engineering standards.

    Therefore, many of your requested points regarding Acceptance Criteria and Study Proof (especially related to AI, MRMC, expert consensus, and ground truth establishment) are not applicable to this 510(k) submission for a non-AI medical device.

    However, I can extract the relevant information from the document regarding the "acceptance criteria" (in the sense of standards compliance and comparative performance) and the "study" (non-clinical tests) that proves these criteria are met.

    Here's a breakdown based on the provided document:


    Device: Sterile High-pressure Angiographic Syringes for Single-use

    Device Type: Non-AI Medical Device (Angiographic Syringes)

    Study Purpose: To demonstrate Substantial Equivalence (SE) to a predicate device (Shenzhen BaoAn Medical Supplies Co., Ltd. Sterile High-pressure Angiographic Syringes for Single-use, K151960) through non-clinical testing and comparison.


    1. A table of acceptance criteria and the reported device performance

    For a non-AI device like this syringe, "acceptance criteria" primarily refer to adherence to recognized international standards and demonstration of comparable performance to a predicate device. "Reported device performance" is the outcome of non-clinical tests demonstrating compliance.

    Acceptance Criteria (Standard / Property)Basis / Test MethodReported Device Performance (Result & SE Justification)
    Performance
    Compliance with Syringe StandardsISO 7886-1 (Sterile hypodermic syringes for single use), ISO 594-1 (Conical fittings with a 6% (Luer) taper), ISO 594-2 (Lock Fitting)Complied with ISO 7886-1, ISO 594-1, and ISO 594-2. This demonstrates performance equivalent to the predicate.
    Seal StrengthASTM F88/F88M-15 (Standard test method for seal strength of flexible barrier materials)Test results met the standard.
    Seal Leaks (Porous Medical Package)ASTM F1929-15 (Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration)Test results met the standard.
    Connection Tube PerformanceISO 594-1 and ISO 594-2Performance test results of connection tube meet the criteria. Even with more product specifications (various lengths), the performance is maintained.
    J-shape Tube LeakageNon-standardized test, implied to ensure functionality.Performance test results showed no leaks.
    Biocompatibility
    Biocompatibility AssessmentISO 10993-1: 2018 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process)Overall biocompatibility demonstrated; differences in materials (e.g., Piston, Filter membrane) from predicate are deemed not to affect SE due to successful biocompatibility testing (Analysis 2).
    CytotoxicityISO 10993-5:2009 (Tests for In Vitro cytotoxicity)No Cytotoxicity. (Same as predicate)
    IrritationISO 10993-10:2010 (Tests for irritation and skin sensitization)No Irritation. (Same as predicate)
    SensitizationISO 10993-10:2010 (Tests for irritation and skin sensitization)No Sensitization. (Same as predicate)
    Systemic ToxicityISO 10993-11:2017 (Tests for systemic toxicity)No Acute Toxicity. (Same as predicate)
    HemolysisASTM F 756-17 (Standard practice for assessment of hemolytic properties of material)No Hemolysis. (Same as predicate)
    PyrogenicityUSP 41-NF36:2018 <151> (Pyrogen Test (USP Rabbit Test))No Pyrogen. (Same as predicate)
    Sterility & Particulate Matter
    Bacterial Endotoxins LimitUSP 41-NF36 2018 <85> (Bacterial Endotoxins Limit)Test results met the standard.
    Sterility Assurance (Post-sterilization)ISO 11737-2:2009 (Sterilization of medical devices- Microbiological methods- Part 2: Test of sterility performed in the definition, validation and maintenance of a sterilization process)Test results met the standard. (EO sterilized, same as predicate).
    Ethylene Oxide Sterilization ResidualsISO 10993-7:2008 (Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals)Test results met the standard. (EO sterilized, same as predicate).
    Particulate MatterUSP <788> (Particulate matters)Complied with USP 788. (Same as predicate)
    Material & Physical Compatibility
    Syringe Volume & Configuration Range Compatibility with Predicate's PerformanceComparison of various single and dual shot volumes (50ml to 200ml) and accessory types (connection tubes, J-shape tube, spikes) with the predicate device.While the proposed device has a wider range of sizes/configurations than the predicate, non-clinical tests confirmed that all these configurations meet the performance criteria of relevant ISO standards. This difference is deemed not to affect SE. (Analysis 1)
    Patient-Contact Material CompatibilityComparison of materials used for various components (Barrel, Piston, Connection Tube, Luer Connectors, Spike, Filter Membrane, J-shape tube) with the predicate. Explicitly, changes in Piston (Medical rubber vs. Polyisoprene rubber) and Filter Membrane (Medical rubber vs. Polypropylene) from the predicate.Differences in Piston and Filter Membrane materials exist. However, the contact classification (blood path, indirect limited contact) remains the same as the predicate. Extensive biocompatibility testing (Cytotoxicity, Irritation, Sensitization, Acute Toxicity, Hemolysis, Pyrogen) on the
    proposed device's materials showed no adverse effects. Therefore, these material differences are not considered to affect substantial equivalence. (Analysis 2)
    Labeling21 CFR Part 801 (Medical Device Labeling)Complied with 21 CFR Part 801. (Same as predicate)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not explicitly state the number of units/samples used for each non-clinical test (e.g., how many syringes were tested for seal strength, or how many batches for endotoxin). It only states that "Non clinical tests were conducted to verify that the proposed device met all design specifications." The specific number of test replicates/samples would typically be detailed in the full test reports, which are not part of this summary document.
    • Data Provenance: The tests are non-clinical (laboratory/bench testing) performed to verify compliance with standards. The company "Youwo (Guangzhou) Medical Device Co., Ltd." is based in Guangzhou, P.R. China, implying the testing was conducted in China, likely at their own facilities or certified labs in China. The data is entirely lab-generated, not from human subjects. The nature of the testing is prospective, in the sense that the tests were performed specifically for this submission on new device samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a non-AI, non-diagnostic medical device. There is no concept of "ground truth" established by human experts in the context of diagnostic interpretation. The "ground truth" for the performance of these syringes is defined by the objective pass/fail criteria of the recognized engineering and biocompatibility standards (e.g., is the seal strong enough? Is there pyrogen presence?). Compliance is measured by laboratory equipment and established protocols, not expert human review of an image or signal.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As per point 3, there's no human diagnostic interpretation to adjudicate. Test results are objective measurements against pre-defined thresholds from standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is not an AI diagnostic tool, nor is it intended to assist human readers in interpreting medical cases. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or AI device. Standalone performance refers to the device's ability to meet its specifications independently, which is what the non-clinical tests implicitly demonstrate (e.g., a syringe's seal strength is measured independently of human operation).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Standards-Defined Ground Truth: The "ground truth" here is the adherence to internationally recognized performance and safety standards (e.g., ISO, ASTM, USP). The device is "true" if it passes the specified tests according to the criteria defined within those standards. For example, a "true" biocompatible device produces no cytotoxicity according to ISO 10993-5. This is determined by laboratory measurements and chemical/biological assays, not clinical outcomes or expert diagnostician consensus.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device. There is no training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/machine learning device. No ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K980417

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1998-05-04

    (90 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K935668, K810624, K790062, K942391, K800823

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medrad Disposable MRI Kit with Check Valve is indicated for use in an MRI environment. The components of the kit are used to assist in filling and delivering intravascular contrast media and common flushing agents at various volumes and flow rates.

    Device Description

    The Medrad Disposable MRI Kit with Check valve is substantially equivalent to the current Medrad Qwik-Fit MRI Disposables Kit (K943189). The Medrad Disposable MRI Kit with Check valve has the same intended use as the predicate device - to assist in filling and delivering intravascular contrast media and common flushing agents at various volumes and flow rates in an MR application. The kit may consist of one or all of the following:

    • 2- 65 ml syringes (K935668)
    • 96" Low Pressure Connector Tube/T-connector assembly (K943189) .
    • . Fluid Delivery Device (Spike) - (K810624)
    • Normally closed check-valve B. Braun/Burron (K790062) ●
    • Female/Female Luer Adapter B.Braun/Burron (K942391) .
      The only difference between this kit and the predicate device is that this kit contains a normally closed check valve and a female/female luer Adapter. Device Verification and Validation Testing has been performed.
      The normally closed check-valve is a attached to the 96" LLPCT at the Tconnector (K943189) and is used to restrict flow in one of two directions. The device is normally closed requiring a minimum amount of force to open. The check-valve attaches to the 'A' syringe and remains closed during saline flush from the 'B' syringe to prevent contrast trickle during this process.
      The Female/Female Luer Adapter is a disposable device used to provide a fluid path between two vessels with male luer terminations. This device can be compared to the Medrad Ouick Fill Tube (K800823) which is also a filling device.
      All other components of the kit are identical to those in K943189.
      The devices are sterilized using Ethylene Oxide gas. Validation is performed according to AAMI guidelines.
    AI/ML Overview

    Here's an analysis of the provided text regarding the Medrad Disposable MRI Kit with Check Valve, focusing on acceptance criteria and the supporting study information:

    It's important to note that this document is a 510(k) summary, which is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. As such, the level of detail regarding comprehensive studies and acceptance criteria is different from what might be found in a full efficacy study report for a new drug or high-risk medical device. The focus here is on demonstrating that the new device does not raise new issues of safety or effectiveness compared to its predicate.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly present acceptance criteria in a quantitative, pass/fail table format, nor does it detail a standalone study proving the device meets specific performance metrics with numerical results for each criterion regarding its primary function (fluid delivery). Instead, it demonstrates substantial equivalence by comparing features and performing validation tests related to manufacturing and safety attributes.

    However, we can infer some "acceptance criteria" based on the validation tests performed and the comparison to the predicate devices. The "reported device performance" is primarily stated as compliance with established standards and equivalence to the predicate.

    Acceptance Criterion (Inferred from Validation/Comparison)Reported Device Performance
    Sterility (Microbial Challenge-Overkill Method)Achieves a Sterility Assurance Level (SAL) of $10^{-6}$ using biological indicators of Bacillus Subtilis var. niger ($1 \times 10^6$ spores per indicator).
    Residual Ethylene Oxide (EtO) Levels< 4 ppm after 8 days outgassing (via simulated-use aqueous extractions).
    Residual Ethylene Chlorohydrin Levels< 4 ppm after 8 days outgassing (via aqueous extraction).
    Residual Ethylene Glycol Levels< 18 ppm after 8 days outgassing (via aqueous extraction).
    PyrogenicityComplies with USP Rabbit Testing and/or USP Limulus Amebocyte Lysate (LAL) test methods.
    BiocompatibilityComplies with ANSI-AAMI(ISO) 10993-1 through 11 and US Pharmacopoeia XXII, <87 & 88> Biological Reactivity Test (Invitro and Invivo) for Cytotoxicity, Sensitization, Systemic Tox, Intracutaneous Tox, Pyrogen Mat'l Mediated, Hemolysis, Particulate analysis, Bacterio-/Fungistasis, Physicochemical analysis (aqueous/non-aqueous), and Infrared analysis.
    Packaging Compatibility with SterilizationThe polystyrene tray and Tyvek lid are heat-sealed and compatible with an EtO Sterilization Cycle.
    Shelf LifeConfirmed at least a two-year shelf life from the date of sterilization using an accelerated aging sequence (1 week at 130°F/70-80% RH, 1 day at -4°F, 1 week at 130°F/<15% RH, equivalent to 6 months normal aging). Ongoing testing aims for a five-year shelf life. Flow tests, pressure tests, and ARO tests (likely refers to "Air Removal" or similar functional test) are conducted to ensure product efficacy after accelerated aging. Specific results are not provided.
    Standard Luer Fitting (Female/Female Luer Adapter)"Yes - ANSI 70.1" (implies compliance with the standard).
    Standard Luer Fitting (LLPCT)"ANSI 70.1" (implies compliance with the standard).
    Functionality of Check-ValveRestricts flow in one of two directions, "normally closed requiring a minimum amount of force to open." Prevents contrast trickle during saline flush from the 'B' syringe.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sterility Validation:

      • Sample Size: 50 test products per half-cycle verification run. There were three half-cycle verification runs, totaling 150 test products.
      • Data Provenance: Not explicitly stated as retrospective or prospective, or country of origin, but it is implied to be internal testing conducted by Medrad, Inc. as part of their sterilization validation process for a new device submission. This would be prospective testing for the specific device.
    • Residual Levels:

      • Sample Size: Not specified for the pre-release testing.
      • Data Provenance: Implied as internal prospective testing.
    • Shelf Life Testing:

      • Sample Size: Not specified for the accelerated aging tests or the subsequent flow/pressure/ARO tests.
      • Data Provenance: Implied as internal prospective testing.
    • Biocompatibility/Pyrogenicity:

      • Sample Size: Not specified.
      • Data Provenance: Implied as internal prospective testing.

    For the functional aspects of the check-valve and luer adapter, the document states "Device Verification and Validation Testing has been performed" but does not provide details on sample sizes or specific test results, instead relying on a comparison of features to the predicate device and the fact that other components are identical to a previously cleared device (K943189).


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to the provided document. The Medrad Disposable MRI Kit with Check Valve is a hardware device (disposable kit) for fluid delivery, not a diagnostic algorithm or a device requiring expert interpretation of results. Therefore, there's no "ground truth" in the medical imaging or diagnostic sense established by human experts for the device's functional performance or safety. The "ground truth" for the device's properties (like sterility, biocompatibility, luer fitting compliance) is established through technical standards, laboratory tests, and reference protocols.


    4. Adjudication Method for the Test Set

    This is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., medical image reading) to resolve discrepancies. This device does not involve such human interpretation in its primary function or testing for a 510(k) submission.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This is not applicable. An MRMC comparative effectiveness study is used to evaluate the diagnostic performance of human readers, sometimes comparing performance with and without AI assistance for tasks like disease detection or classification. This device is a fluid delivery kit and does not involve AI or human diagnostic reading in its function.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. This device is a physical medical kit, not an algorithm, so a standalone algorithm performance study is irrelevant.


    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the established reference for the characteristics being tested. For this device, the ground truth types are:

    • Technical Standards and Reference Protocols: For sterility (ANSI/AAMI ST27-1988), residual levels (ANSI/AAMI ST29-1988 and ST30-1989), pyrogenicity (USP Rabbit Testing, USP LAL), biocompatibility (ANSI-AMMI/ISO 10993-1 through 11, USP XXII <87 & 88>), and luer fitting compliance (ANSI 70.1). These external, validated standards serve as the "ground truth" for compliance.
    • Predicate Device Characteristics: The existing Medrad Qwik-Fit MRI Disposables Kit (K943189) and its components serve as a ground truth for substantial equivalence, specifically for features that are identical or closely matched.
    • Functional Specifications: The intended behavior of the check-valve (restricting flow, preventing trickle) and the luer adapter (fluid transfer) define the functional "ground truth" against which device performance is validated internally.

    8. The Sample Size for the Training Set

    This is not applicable. A "training set" refers to data used to train a machine learning algorithm. This device is not an AI/ML product.


    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.