510(k) Data Aggregation

    K Number
    K223923

    Validate with FDA (Live)

    Date Cleared
    2023-03-30

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K935279

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For telescopes with diameter ranging from 3mm-5mm
    The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults and pediatrics.
    For telescopes with diameter ranging from 5.5mm- 11mm
    The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults.

    Device Description

    The HOPKINS Telescopes are rigid telescopes that utilize the rod lens technology. At the distal end of the telescope's shaft is the lens and the other end of the shaft is attached to the eyepiece. Throughout the central lumen of the HOPKINS Telescopes, optical glass rods are used to transmit and magnify the image received from the lens. The HOPKINS Telescopes are available with 0°, 6°, 30° and 45° direction of view, diameters ranging from 3mm to 11mm and working lengths from 18cm- 50cm.

    AI/ML Overview

    The provided text describes the submission of a 510(k) summary for the HOPKINS Telescopes, comparing it to a predicate device. It details non-clinical performance data and references published literature for clinical performance, but it does not contain information about acceptance criteria or a specific study proving the device meets those criteria, especially in the context of an AI/ML device.

    The device described, "HOPKINS Telescopes," is a rigid endoscope that utilizes rod lens technology for visualization during surgery. There is no mention of AI or machine learning components. Therefore, many of the requested points, such as sample sizes for test/training sets, expert ground truth adjudication, MRMC studies, and standalone AI performance, are not applicable to the information provided in this document.

    However, based on the non-clinical performance data section, we can infer some "acceptance criteria" related to standards compliance and general performance.

    Here's an attempt to answer the questions based only on the provided text, recognizing that it does not pertain to AI/ML device performance:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML device, the "acceptance criteria" are related to compliance with recognized standards and successful bench testing. The document states:
    "the HOPKINS Telescopes has met all its design specification and is substantially equivalent to its predicate device."

    Criteria TypeAcceptance Criteria (Implied)Reported Device Performance
    Functional Standards (Endoscopy)Compliance with ISO 8600-1, ISO 8600-3, ISO 8600-5, ISO 8600-6Met these standards.
    BiocompatibilityCompliance with ISO 10993-5, -11, -10 (twice)Passed Cytotoxicity, Acute Systemic Toxicity, Intracutaneous Irritation, Maximization Sensitization tests.
    Electrical & Thermal SafetyCompliance with IEC 60601-2-18:2009 (3RD Edition)Met this standard.
    Reprocessing (Cleaning & Sterilization)Compliance with AAMI TIR12, TIR30, ST8, ST77, ST79, ST81, ISO 14937, ISO 17655-1Met these standards during reprocessing validation.
    General PerformanceSubstantial equivalence to predicate device in bench testingDemonstrated substantial equivalence; met all design specifications.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of data for an AI/ML model. The evaluation involves non-clinical bench testing and comparison to a predicate device. No specific sample sizes for these tests are mentioned, nor is there information about data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the device is a physical endoscope, not an AI/ML diagnostic tool requiring expert ground truth for image interpretation.

    4. Adjudication method for the test set

    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI-assisted device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical testing, the "ground truth" would be established by physical measurements, chemical analyses, and adherence to the specified performance parameters outlined in the referenced standards. For the pediatrics indication expansion, the ground truth is based on "published literature" supporting safety and effectiveness, which would implicitly rely on clinical data outcomes.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device with a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213194

    Validate with FDA (Live)

    Date Cleared
    2022-06-23

    (267 days)

    Product Code
    Regulation Number
    874.4720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K935279, K143640, K202272

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HD Mediastinoscope is intended to aid the surgeon in viewing the mediastinum and facilitate the introduction and removal of surgical instruments during ENT endoscopic surgical procedures.

    Device Description

    The HD Mediastinoscope is a reusable videoendoscope intended to be used in conjunction with the X-link module of the Image1 S Camera Control Unit. The HD Mediastinoscope is comprised of four main components: CMOS sensor at the distal end of the endoscope, an oval insertion portion (spatula), handle, and internal LED light source. The spatula can accommodate up to 2 x 5 mm instruments.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the HD Mediastinoscope:

    Due to the nature of the device (a physical surgical instrument rather than an AI/software device), the provided document does not contain information typically found in acceptance criteria and studies for AI/ML devices, such as sensitivity, specificity, or AUC.

    Based on the provided text, the "acceptance criteria" appear to be related to the device's technical specifications and safety/effectiveness in comparison to predicate devices, rather than specific performance metrics like those for an AI algorithm.

    Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" are implied through the comparison with existing predicate and reference devices and compliance with recognized standards. The "performance" is demonstrated through non-clinical testing, particularly a cadaver test study.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence to predicate device (K954910)Demonstrated through technical characteristics, intended use, indications for use.
    Compliance with electrical safety and EMC standardsANSI/AAMI ES:60601-1:2005+A1:2012, IEC 60601-1-2:2014, IEC 62471:2006
    Compliance with biocompatibility standardsISO 10993-10:2010, ISO 10993-11:2006/(R) 2010
    Compliance with reprocessing validation standardsAAMI TIR 12:2010, AAMI TIR 30:2011, AAMI/ANSI/ISO 11737-1:2006/(R)2011, ASTM E1837-96:2014
    Change in spatula design does not raise new questions of safety and effectivenessDemonstrated through a cadaver test study.

    Details of the Study (Non-AI/ML Device)

    Since this is a traditional medical device (mediastinoscope) and not an AI/ML device, many of the requested points related to AI/ML studies (sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size, ground truth for training set) are not applicable or not specified in this 510(k) summary.

    However, based on the information provided:

    1. Sample size used for the test set and the data provenance:

      • The primary non-clinical performance test mentioned is a cadaver test study.
      • The sample size (number of cadavers or tests performed) is not specified in the document.
      • The data provenance (country of origin, retrospective/prospective) is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document mentions "a cadaver test study was conducted to demonstrate that the change in spatula design...". It does not specify the use of "experts" to establish a "ground truth" in the way an AI/ML study would (e.g., radiologists labeling images). Instead, the study likely involved technical assessment and observation by qualified personnel (e.g., surgeons, engineers).
      • Not explicitly stated in terms of "ground truth experts" for an AI/ML context.
    3. Adjudication method for the test set:

      • Not applicable / Not specified for this type of device and study.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study was not conducted. This is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical surgical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the cadaver study, the "ground truth" would likely be the direct observation and technical assessment of the device's performance in a cadaveric model (e.g., ease of introduction, visualization, ability to facilitate instrument use, assessment of tissue damage, etc.) by the study investigators. This is not a "ground truth" in the sense of a diagnostic label for an AI.
    7. The sample size for the training set:

      • Not applicable. This device does not use a training set as it is not an AI/ML product.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202272

    Validate with FDA (Live)

    Date Cleared
    2021-05-11

    (273 days)

    Product Code
    Regulation Number
    874.4720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K935279, K143640

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HD Mediastinoscope is intended to aid the surgeon in viewing the mediastinum and facilitate the introduction and removal of surgical instruments during ENT endoscopic surgical procedures.

    Device Description

    The components subject of this submission are: HD Mediastinoscope (Part Number: 10973HD), and the IMAGE1 S CCU. The CCU consists of the IMAGE1 S Connect Module (Model Number: TC200US) and IMAGE1 S X-Link (Model Number: TC301US).

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the Karl Storz HD Mediastinoscope (K202272). The device is intended to aid surgeons in viewing the mediastinum and facilitating the introduction and removal of surgical instruments during ENT endoscopic procedures.

    The information provided focuses on the substantial equivalence determination based on non-clinical performance data, rather than a study proving the device meets AI-specific acceptance criteria. Therefore, most of the requested information regarding AI model performance, ground truth establishment, expert review, and clinical study details is not present in the provided text.

    Here is a breakdown of what can be extracted and what is not available based on the request:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantitative acceptance criteria in a table format for performance. Instead, it states compliance with recognized standards and successful completion of various bench tests.

    Test CategoryAcceptance Criteria (Implied by Compliance)Reported Device Performance
    Electrical Safety & EMCCompliance with ANSI/AAMI ES:60601-1:2005+A1:2012 and IEC 60601-1-2:2014In compliance with the stated standards.
    BiocompatibilityCompliance with ISO 10993-1, ISO 10993-5:2009/(R) 2014, ISO 10993-10:2010, ISO 10993-11:2006/(R) 2010, and FDA GuidanceEvaluation performed, tests conducted based on contact type and duration.
    Reprocessing ValidationCompliance with AAMI TIR 12:2010, AAMI TIR 30:2011, AAMI/ANSI/ISO 11737-1:2006/(R)2011, ASTM E1837-96:2014Validation activities performed for cleaning and sterilization; in compliance with standards.
    Optical PerformanceNot explicitly quantified, but successful completion of tests is implied for substantial equivalence.
    - Color Reproduction & Contrast(Implied successful performance for medical imaging)Tested.
    - Illumination Detection Uniformity(Implied successful performance)Tested.
    - Instantaneous Dynamic Range (IDR) & Detection Linearity(Implied successful performance)Tested.
    - Spatial Resolution & Depth of Field(Implied successful performance)Tested.
    - Distortion & Field of View(Implied successful performance)Tested.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications." This means there was no clinical test set in the traditional sense for assessing device performance in a human population. All testing was non-clinical bench testing. Therefore, sample size and data provenance (country, retrospective/prospective) related to a clinical test set are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical test set or ground truth in the context of diagnostic interpretation (e.g., for an AI model) was established. The ground truth for bench testing would be metrology standards and physical measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set involving human readers or interpretation was performed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mediastinoscope (a physical viewing instrument), not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical bench testing, the ground truth would be based on established engineering and optical metrology standards, and potentially manufacturer specifications for the device's components. For biocompatibility and reprocessing, the ground truth is compliance with the listed ISO and AAMI standards.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as this is not an AI/ML device.