Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151928

    Validate with FDA (Live)

    Date Cleared
    2015-11-06

    (115 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K013227, K023580, K933193

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PUR Implant System can be used in dental implant applications, for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. The PUR Implant is intended for delayed loading. It is also indicated for immediate loading with good primary stability and appropriate occlusal loading.

    The PUR Implant System is only intended for use with straight abutments. The PUR implant body is intended to be placed such no angle correction is necessary.

    Device Description

    The PUR is a self-tapping, double thread screw implant with a micro groove section, manufactured from Wrought Titanium 6AL-4V ELI. The implant surface is grit-blasted and acid etched. The implants are manufactured with four body diameters 3.5 mm, 4.3 mm, and 6.0 mm. The 3.5 and 4.3 mm implants have a Narrow Platform (NP) prosthetic head. The 5.0 and 6.0 mm implants have a Regular Platform (RP) prosthetic head. The PUR implants are substantially equivalent to the Zimmer Tapered Screw-Vent® 3.5 mmD Platform and the 4.5 mmD Platform implant systems. They provide for non- rotational single and multiple tooth restorations in both the maxilla and mandible. They are packaged sterile in a double blister, which contains: the implant in a titanium tube. A special driver is available for the implants which simplifies implant insertion by eliminating the need for an implant mount.

    This device uses straight abutments with no angles. Sterngold has SA Abutments) which are intended for multi-unit restorations. Sterngold also has UCLA abutments which are twopiece abutments composed of Wrought Titanium 6AL-4V ELI, conforming to ASTM F136-13 and Gold Alloy (64% gold, 22% palladium and 14.0% platinum). These UCLA abutments are intended for single and multiple unit use.

    The PUR implant system is not intended to include any angled abutments or allow for any correction of angled implant placement.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental implant system. This type of document is used to demonstrate that a new medical device is "substantially equivalent" to a legally marketed predicate device, meaning it is at least as safe and effective. The FDA's decision is based on comparing the new device's technological characteristics and intended use to those of the predicate.

    Crucially, a 510(k) summary for a dental implant system typically does not include "acceptance criteria" or a "study that proves the device meets the acceptance criteria" in the way one might find for an AI/ML-driven diagnostic device. The concept of "acceptance criteria" for performance metrics like sensitivity, specificity, or AUC is not directly applicable here. Instead, the "acceptance criteria" for a 510(k) are related to demonstrating substantial equivalence through various tests (mechanical, biocompatibility, sterilization validation) and design comparisons.

    Therefore, many of the requested points are not applicable to this type of regulatory submission. However, I will extract and describe the relevant information that is present.

    Here's an analysis based on the provided text, addressing the points where possible and indicating when they are not applicable for this kind of device and submission:


    1. A table of acceptance criteria and the reported device performance

    As explained above, this submission does not present acceptance criteria and performance in terms of diagnostic metrics (e.g., sensitivity, specificity, or F1 score). Instead, the "performance" demonstrated for dental implants revolves around mechanical properties, biocompatibility, and sterilization effectiveness to show substantial equivalence to predicate devices.

    The document states:

    • Performance Testing: "Non clinical Testing was performed following "Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments." Performance testing demonstrated that the device performs appropriately for the proposed indications for use."
    • Surface analysis: "SEM photographs confirm that all residue of blasting material had been removed from the treated implant surface, and the surface chemical analysis confirmed that no contaminants remained on the surface."
    • Sterilization, shelf life, and biocompatibility: "PUR Implants have the same sterilization process and radiation dose, same shelf life and biocompatibility as previous cleared Sterngold devices, therefore sterility, shelf life and biocompatibility testing performed on previous cleared Sterngold devices is applicable to the PUR Implants."

    Table of Comparison (extract of relevant parts from page 6, representing "performance" in the context of substantial equivalence):

    PropertySubject Device K151928Predicate Device K061410 (Primary Predicate)
    Intended UseThe PUR Implant System can be used in dental implant applications, for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. The PUR Implant is intended for delayed loading. It is also indicated for immediate loading with good primary stability and appropriate occlusal loading. The PUR Implant System is only intended for use with straight abutments. The PUR implant body is intended to be placed such no angle correction is necessary.For use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
    Implant DesignThreaded, Root-form implantThreaded, Root-form implant
    Implant ShapeCylinderCylinder
    Implant Body Diameter3.5mm, 4.3mm, 5.0mm, 6.0mm3.7mm, 4.1mm, 4.7mm, 6.0mm
    Implant Length8mm, 10mm, 12mm, 14mm10.0mm, 11.5mm, 13.0mm, 16.0mm
    Outer Thread0.136 - 0.140(Information not provided in table for predicate)

    The "acceptance criteria" here are that the subject device's characteristics and performance testing results are comparable to the predicates and meet the established guidance standards for dental implants, demonstrating that it is as safe and effective.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The testing described is "non-clinical testing," which typically refers to benchtop mechanical tests or material characterization rather than human subject trials. Therefore, concepts like "test set" in the context of patient data, data provenance, or retrospective/prospective studies are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The device is a physical dental implant, not a diagnostic AI/ML device requiring expert-labeled medical images for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable. As above, there is no "test set" in the context of expert adjudication for diagnostic accuracy.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is a physical medical device (dental implant), not an AI/ML software device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. This is a physical medical device (dental implant), not an AI/ML software device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a physical device like a dental implant, "ground truth" refers to its physical, mechanical, and biological properties meeting established standards and performance expectations. The document indicates:

    • "Non clinical Testing was performed following 'Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments.'" This guidance document essentially defines the "ground truth" or standard for performance.
    • "Surface analysis was conducted on the final device. SEM photographs confirm that all residue of blasting material had been removed from the treated implant surface, and the surface chemical analysis confirmed that no contaminants remained on the surface." This refers to chemical and microscopy analysis.
    • The reliance on "sterility, shelf life and biocompatibility testing performed on previous cleared Sterngold devices" based on the subject device having the "same sterilization process and radiation dose, same shelf life and biocompatibility." This means previous testing results and established biological safety are the "ground truth."

    8. The sample size for the training set

    This is not applicable. This is a physical medical device (dental implant), not an AI/ML software device that relies on a "training set."

    9. How the ground truth for the training set was established

    This is not applicable. As above, there is no "training set" for device performance in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1