Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K110046

    Validate with FDA (Live)

    Date Cleared
    2011-12-07

    (335 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Athena Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. The system is intended for posterior, pedicle fixation as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

    Device Description

    The Athena Pedicle Screw System is comprised of polyaxial pedicle screws in various diameters and lengths, spinal rods in various lengths, cross-connectors and set screws. The Athena Pedicle Screw System can be used for single or multiple level fixations. All implant components are manufactured from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Static and dynamic axial compression bending testing per ASTM F1717Performed; results demonstrate substantial equivalence to predicate devices.
    Static torsion testing per ASTM F1717Performed; results demonstrate substantial equivalence to predicate devices.
    Use of titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136Device manufactured from this material, consistent with predicates.
    Same intended use as predicate devicesConfirmed.
    Similar primary implant design as predicate devicesConfirmed.
    Equivalent material of manufacture as predicate devicesConfirmed.
    Overall substantial equivalence to predicate devicesConcluded based on non-clinical testing and shared characteristics.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text describes non-clinical mechanical testing of the device components. It does not refer to a "test set" in the context of patient data or clinical trials. Therefore, information regarding sample size for a test set and data provenance (e.g., country of origin, retrospective/prospective) is not applicable to this type of study as described. The testing involves physical samples of the pedicle screw system components.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    Given this is a non-clinical mechanical performance study rather than a clinical study evaluating diagnostic or treatment outcomes, there were no experts used to establish ground truth in the way one would for clinical data interpretation. The "ground truth" for this study is defined by the established standards of ASTM F1717 for mechanical testing. Compliance with these standards is objectively measured.

    4. Adjudication Method

    Since no experts were establishing ground truth for a clinical test set, an adjudication method is not applicable. The testing involves objective physical measurements against established engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This submission focuses on the mechanical substantial equivalence of a medical implant, not the effectiveness of a diagnostic or treatment method involving human readers.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone "algorithm only" performance study was not done. This is a physical medical device, not a software algorithm. The "standalone performance" here refers to the mechanical integrity of the device components in isolation (i.e., without being implanted in a patient, but using standardized test fixtures). The study conducted was precisely that – standalone mechanical testing of the device components.

    7. The Type of Ground Truth Used

    The "ground truth" used for this study is based on established engineering standards (ASTM F1717) for mechanical performance and material specifications (ASTM F136). The performance of the Athena Pedicle Screw System was compared directly against these standards and to the mechanical performance of predicate devices that previously met these standards.

    8. The Sample Size for the Training Set

    Since this is a non-clinical mechanical study rather than a machine learning or clinical study, the concept of a "training set" is not applicable. The device's design is developed through engineering principles, not through training on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    As the concept of a "training set" does not apply to this type of study, the question of how ground truth was established for a training set is not applicable. The design and manufacturing processes are validated against engineering specifications and industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K983904

    Validate with FDA (Live)

    Date Cleared
    1998-12-22

    (49 days)

    Product Code
    Regulation Number
    888.3050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K884263, K896106, K930298, K950865, K954696, K965224

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Non-Pedicle Screw Fixation

    When used as an anterior spinal system the Rogozinski Spinal Rod System is intended for the treatement of DDD (degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies), pseudoarthrosis, stenosis, deformities (scollosis, kyphosis), spondylolisthesis, fracture, previous failed fusion, or tumor resection. The Rogozinski Spinal Rod System is limited to non-cervical use.

    When used as a posterior, non-pedicle spinal system (consisting of hooks and sacral/iliac screws), the Rogozinski Spinal Rod System is intended for the treatement of DDD (degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies), pseudoarthrosis, deformities (scoliosis, kyphosis, lordosis), spondylolisthesis, fracture, previous failed fusion, or tumor resection. The Rogozinski Spinal Rod System is limited to non-cervical use.

    Pedicle Screw Fixation

    When used as a pedicle screw system, in the spine of skeletally mature patients, the Rogozinski Spinal Rod System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the LS-S1 igint: (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (d) who are having the device removed after the development of a solid fusion mass.

    In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Rogozinski Spinal Rod System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine. degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

    Device Description

    The Rogozinski Spinal Rod System consists of stainless steel rods attached to the spinal column through the use of interlaminal hooks and/or screws. Crossbars may be used to connect rods to rods to provide a more rigid construct, as well as screws to rods and hooks to rods.

    AI/ML Overview

    This is a medical device submission, specifically a Summary of Safety and Effectiveness for the Rogozinski Spinal System (K983904), dated December 22, 1998. This document does not contain information about acceptance criteria or specific study data.

    Medical device submissions like this primarily focus on demonstrating substantial equivalence to a predicate device based on design, materials, and intended use. They typically do not include detailed performance studies with acceptance criteria in the way an AI/software device would. The "study" here is more about asserting equivalency to previously cleared devices rather than a performance study measuring metrics like sensitivity, specificity, or accuracy against a ground truth.

    Therefore, most of the requested information cannot be extracted from the provided text.

    Here is what can be extracted based on the nature of this document:

    1. A table of acceptance criteria and the reported device performance

    • Not applicable / Not present. This document does not describe acceptance criteria in terms of performance metrics (e.g., accuracy, sensitivity, specificity) or report quantitative device performance data against such criteria. The device is a physical spinal fixation system, and its "performance" is generally established through material testing, mechanical testing, and clinical history of predicate devices, not through a statistical study with specific acceptance thresholds for diagnostics.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not present. This document does not detail a "test set" or clinical study data. It is a substantial equivalence submission for a physical implant.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not present. No ground truth establishment for a test set is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not present. No test set or adjudication process is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable / Not present. This is not an AI/software device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable / Not present. This is not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable / Not present. No ground truth for device performance is defined or used in the context of a "study" here. The "truth" for this device lies in its mechanical properties meeting standards and its clinical use being consistent with predicate devices that have established safety and effectiveness.

    8. The sample size for the training set

    • Not applicable / Not present. This is not a machine learning/AI device, so there is no training set mentioned.

    9. How the ground truth for the training set was established

    • Not applicable / Not present. See point 8.

    In summary: The provided text is a 510(k) summary for a spinal implant, focusing on its substantial equivalence to previously cleared devices rather than providing performance metrics from a study that would include acceptance criteria, test sets, or ground truth establishment in the context of diagnostic or AI devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1