510(k) Data Aggregation

    K Number
    K142812

    Validate with FDA (Live)

    Date Cleared
    2015-06-29

    (273 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K141546, K942779, K925085

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a spirometer which measures patient respiratory parameters including FVC, FEV1, FEV6, PEF, MVV and VC. The Vitalograph Pneumotrac is a desktop spirometer designed for lung function testing for use on adults and pediatrics, 5 years and older, in a variety of environments such as hospital wards, health centers and private homes under the supervision of a healthcare provider.

    Device Description

    The intended use of the Vitalograph Pneumotrac desktop spirometer, which connects to Vitalograph Model 7000 Spirotrac software, ref 510(k) K141546, shall provide a USB-powered desktop spirometer for creating, adding and recalling subjects and performing Spirometry testing on those subjects to aid in the measuring the effect of lung disease on pulmonary function. Spirometry is in the simple assessment of respiratory function through the measurement of dynamic lung volumes. Its primary functions are; 1. Standalone spirometric measures using single breath or multiple-breath testing techniques, display and record lung volumes and flow rates (including FVC, MVV, VC) and it's subdivisions, Display a single subject's demographic data entered by the user. 2. Produce reports In a clinical setting, the measurements obtained from a lung function test form part of the various findings of a physician in the detection, diagnosis and monitoring of chest diseases.

    AI/ML Overview

    The Vitalograph Model 6800 Pneumotrac is a diagnostic spirometer. The device was studied through validation bench testing against several international standards to demonstrate its performance.

    1. Table of Acceptance Criteria and Reported Device Performance:

    ParameterAcceptance CriteriaReported Device Performance
    Volume Range0-10 Liters (Predicate)0-10 Liters
    Max Flow Range0-16 Liters/second (Predicate)0-16 Liters/second
    Back Pressure<= 0.1 kPa/L/s (Predicate)< 0.1 kPa/L/s
    Accuracy FEV1+/- 3% (Predicate)+/- 3%
    Accuracy FEV6+/- 3% (Predicate)+/- 3%
    Accuracy PEF+/- 5% (Predicate)+/- 5%
    Operating Temp10-40°C (At least 17-37°C required per ATS 2005)10-40°C (At least 17-37°C required per ATS 2005)
    Storage Temp0-50°C (Predicate)0-50°C
    Humidity10-95% relative humidity (Predicate)10-95% relative humidity
    Performance StandardsATS ERS 2005, ISO 23747:2009 for PEF, EN ISO 26782:2009, ISO 10993-1 Fourth EditionATS ERS 2005, ISO 23747:2009 for PEF, EN ISO 26782:2009, ISO 10993-1 Fourth Edition
    ComplianceIEC / EN 60601 (EN 60601-1-1 and EN 60601-1-2)IEC / EN 60601 (EN 60601-1-1 and EN 60601-1-2)
    BiocompatibilityAll materials previously clearedAll materials previously cleared in K100687 & K925085
    Connectivity/CommunicationValidation with Vitalograph Model 7000 Spirotrac software to maintain integrity and successful download/storageValidation with Model 7000 Spirotrac (K141546) performed and deemed satisfactory

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample size of a "test set" for data provenance in terms of patient data. The study primarily involved validation bench testing against international performance standards. This implies controlled laboratory testing rather than a clinical study with patient data. Therefore, the data provenance would be from laboratory experiments rather than patient data from a specific country or collected retrospectively/prospectively.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the described study is a series of bench tests against engineering standards, not a study involving human interpretation of data where "ground truth" would be established by experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable as the study did not involve expert adjudication of results. The performance was assessed against predefined numerical and technical standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a spirometer, which measures physiological parameters. It is not an AI-assisted diagnostic tool that would be evaluated in a MRMC study involving human readers and AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a measurement instrument. The "standalone" performance was assessed through bench testing against international standards (ISO, ATS/ERS). This is the functional equivalent of standalone performance for such a device, ensuring its accuracy and precision in measuring respiratory parameters independent of human interpretation during the measurement process.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the bench tests, the "ground truth" was established by the international performance standards themselves (e.g., ISO 26782, ATS/ERS 2005, ISO 23747) which define acceptable ranges and accuracy for spirometry measurements. These standards are developed through scientific consensus and extensive validation.

    8. The sample size for the training set:

    This is not applicable as the device is not an AI/machine learning algorithm requiring a training set in the conventional sense. Its functionality is based on established physiological measurement principles.

    9. How the ground truth for the training set was established:

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100687

    Validate with FDA (Live)

    Date Cleared
    2010-10-16

    (220 days)

    Product Code
    Regulation Number
    868.1840
    Age Range
    All
    Why did this record match?
    Reference Devices :

    K912412, K854526, K873562, K925085

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a battery operated spirometer which measures three basic patient respiratory parameters {FVC, MVV and VC}. The model 2120 is a hand held spirometer designed for lung function testing in a variety of environments such as hospital wards, health centres and private homes. The model 2120 can be configured as a stand-alone spirometer or connected to a printer.

    Device Description

    The Vitalograph Model 2120 is intended to be used, as its predecessor, as a battery operated spirometer which measures three basic patient respiratory parameters {FVC, MVV and VC}. The model 2120 is a hand held spirometer designed for lung function testing in a variety of environments such as hospital wards, health centres and private homes. The model 2120 can be configured as a stand-alone spirometer or connected to a printer.

    The Model 2120 series has been available to the marketplace for over 15 years {ref K946075}. This device is a hand held electronic Spirometer, used in the simple assessment of respiratory function through the measurement of dynamic lung volumes i.e. Spirometry. This revised design is an upgrade to the existing product in terms of modernising its look and feel {i.e. changes to ergonomics and user interface}. The intended uses and operating principles in relation to its measurement of respiratory function remain unchanged.

    Its primary functions are:

    1. Spirometry measurements using single breath and multiple-breath testing techniques, the display and recording of measured lung volumes and flow rates (including FVC, VC, MVV) and it's subdivisions,
    2. Record subject data. Storage of data and test results on unit for later printing or export to Spirotrac {see K912412}.
    3. The flowhead will be a Fleisch Pneumotachograph type flowhead, the operating principle of which has been well established in the existing Model 2120 {K946075} and various other spirometers. {e.g. Vitalograph Compact Ref K854526. Vitalograph Alpha ref K873562, Vitalograph Escort K925085}

    All variants of the Model 2120 {e.g. including In2itive, e-Diary} will use very same operating principle, LCD, Buttons, optional Touchscreen, and Mouldings. Items that may vary within the range are the list of parameters that the different variants display.

    Navigation is allowed via the use of five buttons {Up, Down, Enter/Select, Cancel /Esc and power On/Off} or an optional touch screen.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Vitalograph Model 2120 Hand Held Spirometer, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria (Vitalograph Model 2120 Specifications)Reported Device Performance (Vitalograph Model 2120 Specifications)
    Volume Range0-10 Litres0-10 Litres
    Max Flow Range0-16 Litres/second0-16 Litres/second
    Back PressureLess than 0.1kPa/L/sLess than 0.1kPa/L/s
    Accuracy FEV1+/- 3%+/- 3%
    Accuracy FEV6+/- 3%+/- 3%
    Accuracy PEF+/- 5%+/- 5%
    Operating Temp10-40°C. (At least 17-37°C required per ATS 2005)10-40°C (At least 17-37°C required per ATS 2005)
    Storage Temp0 - 50 °C0 - 50 °C
    Humidity10 - 95% relative humidity10 - 95% relative humidity
    Performance StandardsATS ERS 2005, ISO 23747:2007 for PEF, EN ISO 26782:2009ATS ERS 2005, ISO 23747:2007 for PEF, EN ISO 26782:2009

    Study Proving Device Meets Acceptance Criteria:

    The document states that "Safety and effectiveness have been assured through the extensive testing in relation to IEC 60601 standards for electrical safety and EMC/EMI, as well as device specific performance testing to American Thoracic Society (ATS) and International Organisation for Standardisation (ISO) standards."

    The specific standards referenced for performance are:

    • ATS ERS 2005
    • ISO 23747:2007 for PEF (formerly EN13826:2003)
    • EN ISO 26782:2009

    The table above demonstrates that the device's specifications meet the listed accuracy and range requirements, indicating compliance with these standards. The conclusion explicitly states, "Based on the above, including the successful completion of all device testing, Vitalograph conclude that this device is substantially equivalent to the legally marketed predicate devices and is safe and effective for use."

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It refers to "extensive testing" but does not provide details on the testing methodology, number of subjects, or type of data used (patient data vs. simulated data). Given this is a spirometer, testing would typically involve controlled laboratory simulations with known volumes/flows and potentially human subjects, but these details are not provided.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the 510(k) summary. For device performance testing of a spirometer, ground truth is typically established by physical reference standards (e.g., calibrated syringes) rather than expert consensus on patient data.

    4. Adjudication Method for the Test Set

    This information is not applicable as the ground truth for a spirometer's performance is typically based on objective physical measurements and not on expert adjudication of subjective findings. The document does not describe any such adjudication process.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for imaging devices or diagnostic tools where human interpretation is a key component and the AI assists human readers. For a spirometer that directly measures physical parameters, an MRMC study is not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done

    Yes, implicitly. The device itself is a "standalone" spirometer designed to produce measurements directly. The performance testing "to American Thoracic Society (ATS) and International Organisation for Standardisation (ISO) standards" inherently assesses the algorithm's ability to accurately measure respiratory parameters without human interpretation in the loop. The device directly calculates and displays the FVC, MVV, and VC parameters.

    7. The Type of Ground Truth Used

    The ground truth used for validating the spirometer's performance would primarily be physical reference standards (e.g., highly accurate, calibrated syringes that deliver precise volumes and flows). The device's "Calibration" method is listed as "Using calibration syringe," which supports this. These standards are used to input known volumes and flows into the device, and its measured output is compared against these known values to determine accuracy.

    8. The Sample Size for the Training Set

    The concept of a "training set" in the context of machine learning is not applicable to this device. The Vitalograph Model 2120 is a rule-based or physics-based measurement device, using a Fleisch Pneumotachograph and established algorithms ("Same algorithm for spirometry parameters calculation"). It does not describe an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    As stated above, a "training set" in the context of AI/ML is not applicable to this device.