    K982696

    Validate with FDA (Live)

    Device Name
    INTERPLANT
    Date Cleared
    1999-04-01

    (241 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K924240, K875241, K963302, K972672

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Burdette Medical Systems, Inc. Interplant™ system is used in planning and performing a prostate brachytherapy implant procedure to provide ultrasound probe alignment and radioactive seed implantation. The Interplant™ device will use standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds for the treatment of prostate cancer. The system will allow volume measurement and dose calculation, relevant prostate anatomy and dose visualization, and post-treatment seed localization.

    Device Description

    The Interplant™ system for brachytherapy was developed to address the need for a comprehensive system and procedure for prostate seed implants. The Interplant in system compresses the requirement for separate multiple procedures into one online setting, This product integrates all of the delivery system components (needles, template guide, probe stepper) with the dose planning system for 125-Iodine (125 I) and 103-Paladium (103 Pd) implants. The system is designed for real-time dose planning in the Operating Room (OR) and for performing post-implant seed localization assessment. Real time ultrasound image plane acquisition with automatic probe position determination makes it possible to complete the image acquisition needed for dose planning within about a minute. The interplant™ system provides the needed components for seed implant procedures except the seeds themselves and uses existing ultrasound imaging equipment. This will enable the physician to focus on the procedure, providing immediate feedback for improved quality assurance.

    The Real-Time Brachytherapy Radiation Seed Implant Program and System includes all technology, computer software, and hardware, and methods associated with the implant of radioactive "seeds" or pellets in the prostate. System elements include the user interface, real-time image acquisition, image contouring, dose calculation and display software, dose volume histograms, isodose surface contours, post-implant seed localization, and the patient scheduling spreadsheet software.

    The System consists of a two-dimensional and three-dimensional image visualization and brachytherapy dose system employing 2D ultrasound imaging for use in radioactive seed implants of the prostate. The software for the brachytherapy seed implant and dose calculation system was developed on a Pentium-based processor with supporting graphics and digitizing hardware.

    AI/ML Overview

    This 510(k) summary does not contain information on acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) submission for Burdette Medical Systems, Inc.'s Interplant™ system, which is a brachytherapy dose planning and implant system. It primarily focuses on demonstrating substantial equivalence to already marketed predicate devices.

    Here's what can be extracted from the provided text, and what is missing:

    Missing Information:

    The provided text does not contain any of the following information:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for the test set.
    • Data provenance (country of origin, retrospective/prospective).
    • Number of experts used to establish ground truth for the test set and their qualifications.
    • Adjudication method for the test set.
    • Whether a multi-reader, multi-case (MRMC) comparative effectiveness study was done.
    • Effect size of human reader improvement with AI vs. without AI assistance.
    • Whether a standalone (algorithm only) performance study was done.
    • Type of ground truth used (expert consensus, pathology, outcomes data).
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    Information Extracted from the Document:

    While the core request about acceptance criteria and a study proving their fulfillment cannot be answered from this document, here's what the document does provide:

    • Device Name: Interplant™
    • Intended Use: To provide precision ultrasound probe alignment and radioactive seed implantation in the treatment of prostate cancer. The system allows volume and dose calculation, image visualization, and post-treatment seed localization. It uses standard 125-Iodine (125 I) or 103-Paladium (103 Pd) isotope permanent implant radiation seeds.
    • Technological Characteristics:
      • Integrates delivery system components (needles, template guide, probe stepper) with a dose planning system for 125-I and 103-Pd implants.
      • Designed for real-time dose planning in the Operating Room (OR) and post-implant seed localization assessment.
      • Real-time ultrasound image plane acquisition with automatic probe position determination allows image acquisition for dose planning within about a minute.
      • Includes a user interface, real-time image acquisition, image contouring, dose calculation and display software, dose volume histograms, isodose surface contours, post-implant seed localization, and patient scheduling spreadsheet software.
      • Employs 2D and 3D image visualization and brachytherapy dose system using 2D ultrasound imaging for prostate seed implants.
      • Software developed on a Pentium-based processor with supporting graphics and digitizing hardware.
    • Regulatory Classification: Class II, 21 CFR 892.5730/Procode: 90 MUJ (Medical charged-particle radiation therapy system).
    • Predicate Devices:
      1. Multimedia Medical Systems, Inc. B3DTUA TherapacPLUS dose planning
      2. SSGI Prowess 500 and Prowess 2000 dose planning K924240
      3. CIVCO Medical Instruments Ultra-Step stepper K875241
      4. B&K Medical Systems HP-100 stepper/stand
      5. Tayman Medical Accuseed stand/stabilizer K963302
      6. COTAN stand/stabilizer
      7. Northwest stand/stabilizer
      8. Maroon Bels, Inc. Brachystand and Brachystepper K972672
    • Submission Type: 510(k) Premarket Notification, seeking substantial equivalence.

    Conclusion:

    This document is a regulatory submission for premarket notification (510(k)), not a study report. The primary goal of a 510(k) is to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than to present detailed clinical study results with acceptance criteria. Therefore, the specific information requested about acceptance criteria and a study proving performance against them is not typically found in this type of summary. Such information would be expected in a more comprehensive clinical or validation study report, which is usually referenced or summarized within a 510(k) but not fully detailed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K982821

    Validate with FDA (Live)

    Date Cleared
    1998-11-05

    (86 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K930506, K924240

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MMS . TherpacPLUS B3DTUI is a software application for pre-operative planning and post-operative evaluation of permanent implant brachytherapy procedures for the treatment of prostate cancer. It assists physicians and medical physicists in pre-olanning the patient's implant procedure and evaluating the actual post-operative results of the procedure.

    Device Description

    Hardware Platform and Operating System: The application runs on standard Intel PCs under Microsoft Windows® 32-bit operating systems.

    The application is available on an OR-safe hardware platform. This hardware platform has electrical safety, emission control and cleaning provisions.

    Peripherals and Accessories: The application interfaces with an electromagnetic digitizer, video capture cards, printers, film scanners, as well as DICOM and other image importer and data drives and external sources.

    Software Features: The application supports image capture and importation from clinical imaging modalities sources, structure contouring and template registration, seed placement and dosimetry display for procedure planning and post-plan evaluation. Planning functions are supported by auto seed placement and dose optimization routines. The application supports post-procedure evaluation through automatic seed identification. Both planning and evaluation functions allow plans to be printed on reports, support visualization in 2D and 3D views. The applications has two user support modules - seed source specification and database manager.

    AI/ML Overview

    The provided text describes a medical device, MMS-TherpacPLUS B3DTUI, a software application for prostate brachytherapy planning and evaluation. However, the document is a 510(k) summary for regulatory clearance and does not contain any information about acceptance criteria or a study proving that the device meets such criteria. It focuses on device description, intended use, and substantial equivalence to predicate devices, which is the basis for its FDA clearance.

    Therefore, I cannot provide the requested information from the given text. The details about acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or comparative effectiveness studies are not present within these documents.