510(k) Data Aggregation

    K Number
    K993643

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-12-28

    (61 days)

    Product Code
    Regulation Number
    870.2800
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K913523, K871720, K963035

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nexsystem ambulatory patient monitoring system is intended for use in the home or alternate care settings. It consists of a patient-worn sensor (Nexisensor), signal transfer unit (Nexender), communications module (Nextation), and a telemonitoring station computer based display and storage system (TMS) located at the health care professional's facility. The device stores and transmits ECG data, respiratory data, skin temperature, systolic and diastolic blood pressure (Non-invasive), and PEF and FEV1.

    The Nexsystem is not for use as a critical care monitoring system. The Nexsystem is not an apnea monitor.

    Device Description

    The Nexystem is a non-invasive ambulatory patient monitoring system for recording multiple physiological parameters from patients who may be located at home or in an alternate care setting. The Nexystem continuously gathers physiological data from a sensor band attached to the patient and transmits the data wirelessly via a Signal Transfer Unit (Nexender) to a Base Station Unit (Nextation) where the data are recorded and stored. Additionally, the Base Station Unit has interfaces for auxiliary sensors - spirometer and blood pressure monitor - for recording point in time lung function and blood pressure measurements. A Call Discriminator Unit enables incoming telephone calls to be correctly routed to either the Nextation or a telephone handset.

    The data are transmitted for display, monitoring and storage on a computer (Telemonitoring Station TMS running the Nexoft application software) at a distant location (Telemonitoring Centre). This data transfer is under the control of the Health Care Professional (HCP) at the TMS. Data may be transferred in real time to enable the HCP to check on the quality of the physiological data being recorded and/or the status of the patient. Normally data is transferred at a scheduled time after the end of a patient data recording session. Once transferred to the TMS the data can subsequently be displayed for analysis by the HCP. The Nexystem enables the HCP to print reports of raw or summary data.

    The Nexystem comprises a number of system components identified below.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Nexystem based on the provided K993643 Premarket Notification:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly define acceptance criteria in terms of numerical thresholds for all evaluated parameters. Instead, it relies heavily on compliance with recognized standards (ANSI/AAMI EC12 and EC38) and direct comparison to predicate devices, highlighting where the Nexystem meets or exceeds predicate performance.

    ParameterAcceptance Criteria (Implicit)Reported Device Performance (Nexystem)
    ECGCompliance with ANSI/AAMI EC12 for disposable electrodes. Compliance with ANSI/AAMI EC38 for ambulatory electrocardiographs (Type I). Consistent with predicate device performance (Sabratek, Propaq Encore).Compliant with ANSV/AAMI EC12. Compliant with ANSUV/AAMI EC38 (Type I, single lead, no ECG analysis). Records and stores extended monitor bandwidth (0.05 - 85 Hz). QRS detection. Heart rate range: 30 - 250 BPM. Heart rate accuracy: $\pm$ 3 BPM. Ag/AgCl custom sensing electrodes.
    RespirationConsistent with predicate device performance (Propaq Encore). Compliance with relevant standards for respiration monitoring (not explicitly stated for Nexystem, but predicates set a benchmark).Measures respiration rate by impedance. Range: 0 - 72 BPM. Not an apnoea monitor. Accuracy: $\pm$ 2 BPM. Ag/AgCl custom sensor.
    TemperatureConsistent with predicate device performance (Propaq Encore).Thermistor: Alpha Sensors medical 400 series. Range: 25.0 - 45.0 °C. Accuracy: $\pm 0.2 °C$.
    Blood PressureUses a legally marketed device (Omron IC) with established 510(k) clearance (K913523).Omron IC (HEM725CIC) - Non-invasive blood pressure.
    SpirometryUses a legally marketed device (Micromedical MicroPlus) with established 510(k) clearance (K963035).Micromedical MicroPlus.
    Wireless OperationProvides greater patient convenience than non-wireless predicates.Yes (Nexisensor to Nexender, Nexender to Nextation).
    SafetyCompliance with IEC 60601-1 for drip proofing. General electrical safety (Class I device, earthed).Tested to IEC 60601-1 (drip proofing). Nextation is Class I device and earthed.
    CommunicationsUses standard protocols (TCP/IP, PPP, Windows dial-up). Modem meets V34 standards.Nexender to Nextation radio link: Digital, frequency-shift keying. Power output: 50 mV/m at 3 m. Frequency: 433 MHz or 916 MHz. Nextation to Nexoft: V34 modem, TCP/IP and PPP protocol.
    FunctionalityProvides ability to view raw data, display a list of patient events, log events outside pre-set thresholds.Nexoft displays a list of all patient events; users can view raw data for each event. Events outside pre-set thresholds can be logged for review.

    2. Sample Size Used for the Test Set and Data Provenance

    The submission explicitly states: "14.7 Clinical performance data for equivalence: Not applicable". This indicates that no separate clinical test set of patient data was used to demonstrate the performance of the Nexystem in a clinical setting compared to a ground truth established from patient data. The primary method of demonstrating equivalence was through non-clinical testing against standards and direct comparison of specifications with predicate devices.

    Therefore, for a clinical test set:

    • Sample size: Not applicable.
    • Data Provenance: Not applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Since no clinical test set was utilized for demonstrating performance against a "ground truth" derived from patient data, this information is not applicable. The "ground truth" for the non-clinical tests was defined by the requirements of the standards (ANSI/AAMI EC12 and EC38) and the specifications of the predicate devices.

    4. Adjudication Method for the Test Set

    As no clinical test set with human-established ground truth was used, an adjudication method is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The submission explicitly states "Clinical performance data for equivalence: Not applicable". Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, in a way. The performance evaluation primarily involved standalone testing of the device's technical specifications and compliance with recognized standards. This is akin to standalone performance for the various physiological data acquisition and transmission functions. For example:

    • Compliance testing of the Nexisensor to ANSI/AAMI EC12 for disposable ECG electrodes.
    • Compliance testing of the Nexystem to ANSI/AAMI EC38 for ambulatory electrocardiographs.
    • Verification of heart rate range and accuracy, respiration range and accuracy, and temperature range and accuracy based on the device's internal functionality.
    • The use of legally marketed and separately cleared devices for blood pressure (Omron IC) and spirometry (Micromedical MicroPlus) implies their standalone performance was relied upon.

    However, it's important to note this is not a standalone performance of an AI algorithm, as this device does not appear to incorporate AI for diagnostic purposes. It's a standalone performance demonstration of monitoring hardware and software capabilities.

    7. The Type of Ground Truth Used

    The ground truth used in this submission is multifaceted and based on:

    • Industry Standards: The requirements and specifications outlined in ANSI/AAMI EC12 (Disposable ECG Electrodes) and ANSI/AAMI EC38 (Ambulatory Electrocardiographs).
    • Predicate Device Specifications: The technical parameters and performance characteristics of the legally marketed predicate devices (Sabratek PHMS, Propaq Encore, Mortara Ambulatory X-12 Telemetry Module, Reynolds Tracker/Pathfinder, Instromedix Poseidon Cardiac Monitoring System). The Nexystem demonstrated "substantial equivalence" by meeting or exceeding these established benchmarks.
    • Functional Testing: The inherent design specifications and measured performance of the Nexystem components (e.g., heart rate accuracy, temperature accuracy) as validated by engineers/testers against known input signals or reference devices.

    8. The Sample Size for the Training Set

    This information is not applicable. The Nexystem is an ambulatory patient monitoring system that primarily collects and transmits physiological data; it is not described as an AI-powered diagnostic or analysis device that would require a "training set" in the context of machine learning. The focus is on hardware and software functionality and compliance with performance standards for data acquisition and transmission.

    9. How the Ground Truth for the Training Set Was Established

    As no training set (in the context of machine learning/AI) was used, this information is not applicable.