Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K020945

    Validate with FDA (Live)

    Date Cleared
    2002-05-31

    (67 days)

    Product Code
    Regulation Number
    862.1545
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K896294

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AxSYM® Intact PTH is a Microparticle Enzyme Immunoassay (MEIA) for the in vitro quantitative determination of intact human parathyroid hormone (I PTH) in human serum or plasma on the AxSYM® system. The AxSYM® I PTH assay is intended for use as an aid in the differential diagnosis of hypercalcemia and hypocalcemia.

    Device Description

    AxSYM® I PTH is based on the Microparticle Enzyme Immunoassay (MEIA) technology. The AxSYM® I PTH assay uses paired monoclonal and polyclonal antibodies, each reactive with epitopes in the N-terminal portion of intact human PTH (1-84). The polyclonal antibody is biotin labeled. I PTH in the sample serves to bridge the two antibodies.

    AI/ML Overview

    The provided text does not contain a study that proves the acceptance criteria of the device. Instead, it details the substantial equivalence of the "AxSYM Intact PTH" device to a predicate device, the "DSL-8000 ACTIVE™ Intact PTH IRMA," through method correlation and comparison of various performance characteristics.

    Therefore, the following points address the questions based on the information available regarding the substantial equivalence study, rather than a study specifically proving acceptance criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria in this context are not explicitly stated numerical targets for device performance, but rather the demonstration of "substantial equivalence" to a predicate device. This is primarily shown through method correlation and comparable performance characteristics.

    FeatureAcceptance Criteria (Implied by Predicate Performance / Substantial Equivalence Goal)Reported AxSYM® I PTH Performance
    Method CorrelationVery high correlation (r value close to 1) between AxSYM® I PTH and predicate device.r = 0.97 (p < 0.0001)
    Regression SlopeSlope close to 1.0 (95% CI covering 1.0)0.99 (95% CI: 0.95, 1.05)
    Regression InterceptIntercept not statistically significantIntercept not statistically significant (p < 0.35)
    Precision (Within Run)Comparable to predicate device.Level I: 5.9 % CV; Level II: 6.2 % CV; Level III: 6.5 % CV
    Precision (Inter)Comparable to predicate device.Level I: 9.4 % CV; Level II: 8.1 % CV; Level III: 8.8 % CV
    LinearityComparable to predicate device.93.8 ± 8.9 %
    RecoveryComparable to predicate device.95.3 ± 8.9 %
    Analytical SensitivityEqual to or better than predicate device.≤ 2.0 pg/mL I PTH
    SpecificityNo detection of human PTH fragments 53-84, 44-68 and 39-84.No detection of human PTH fragments 53-84, 44-68 and 39-84
    InterferenceAcceptable interference levels across various substances.Bilirubin (20 mg/dL): < 10 %; Triglycerides (1500 mg/dL): < 10 %; Total Protein (3 mg/dL): < 10 %; Hemoglobin (1000 mg/dL): < 10 %; Red Blood Cells (0.2 % v/v): < 10 %
    Dynamic Range0 - 2000 pg/mL0 - 2000 pg/mL
    Expected Values8 - 57.8 pg/mL or comparable to predicate.8 - 57.8 pg/mL

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 120 samples were used for the method correlation study.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). However, given that it's a comparison for regulatory submission, it is likely data collected specifically for this purpose. The samples were "serum and plasma samples."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not relevant or provided for a chemical assay. The "ground truth" for a quantitative diagnostic assay like this is typically established by measurements from a reference method or a predicate device, as seen here. There are no "experts" in the sense of human adjudicators for establishing ground truth values of PTH in samples.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this involves a comparison to a predicate device, not subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is a quantitative diagnostic immunoassay, not an AI-driven imaging or diagnostic system that involves human readers or AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance characteristics and method correlation described are for the "AxSYM® Intact PTH" device operating in a standalone capacity (i.e., algorithm/device only). The device provides quantitative results directly from the sample. It is stated that "The I PTH assay is not used as a stand-alone test. Clinical interpretation requires simultaneous measurement of serum calcium levels." This refers to clinical utility, not device performance. The device itself performs its measurement independently.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluation, in this context of a substantial equivalence claim, is the performance and measurements obtained from the predicate device, DSL-8000 ACTIVE™ Intact PTH IRMA. The study compared the AxSYM I PTH results to those of the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This is a chemical immunoassay, not an AI/machine learning device that typically requires a "training set" in the computational sense. The device's calibration and internal parameters are set through manufacturing processes and control procedures, not through training on a dataset of clinical cases.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device technology.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1