Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152675

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-10-29

    (41 days)

    Product Code
    Regulation Number
    876.5630
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K882498, K894783

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MiniCap Extended Life PD Transfer Sets: This set is used during Peritoneal Dialysis therapy to transfer peritoneal dialysis solution to the patient from the source solution container.

    Locking Titanium Adapter for Peritoneal Dialysis Catheter: The Locking Titanium Adapter for Peritoneal Dialysis Catheter is intended to secure the peritoneal catheter tubing to the Baxter transfer set used during Peritoneal Dialysis therapy.

    Locking Cap for Peritoneal Dialysis Catheter Adapter: This device is indicated for use in the treatment of patients receiving peritoneal dialysis therapy, to cap the Locking Titanium Adapter for Peritoneal Dialysis Catheter between Baxter Transfer Set installations.

    Device Description

    The MiniCap Extended Life PD Transfer Sets are single use, sterile, non-pyrogenic devices for use with Baxter peritoneal dialysis systems. A Transfer Set is connected to a Titanium Adapter that is at the end of an implanted peritoneal catheter. The Transfer Sets stay connected to the patient and allows for the exchange of peritoneal dialysis solution into and out of the peritoneal cavity as prescribed.

    The Locking Titanium Adapter for Peritoneal Dialysis Catheter (Titanium Adapter) and Locking Cap for Peritoneal Dialysis Catheter (Locking Cap) are single use, sterile, non-pyrogenic devices for use with Baxter peritoneal dialysis systems. The Titanium Adapter is a device that is secured to the end of a peritoneal dialysis catheter and is used to connect the peritoneal dialysis catheter to the Transfer Sets. The Locking Cap is used to cap the end of the Titanium Adapter between Transfer Set installations.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for two sets of medical devices from Baxter Healthcare Corporation:

    1. MiniCap Extended Life PD Transfer Sets
    2. Locking Titanium Adapter for Peritoneal Dialysis Catheter and Locking Cap for Peritoneal Dialysis Catheter Adapter

    The submission declares that these devices are substantially equivalent to legally marketed predicate devices. The majority of the information provided pertains to non-clinical performance testing and biocompatibility.


    1. Acceptance Criteria and Reported Device Performance

    The documents state that "All results meet their acceptance criteria, and support that the proposed device is appropriately designed for its intended use." However, specific numerical acceptance criteria and the detailed reported performance values (e.g., exact leak rates, torque values, flow rates) are not provided in the given text. Instead, the document lists the functional tests performed.

    Table of Acceptance Criteria and Reported Performance (Based on provided text - specific values are not available):

    Test PerformedAcceptance CriteriaReported Device Performance
    MiniCap Extended Life PD Transfer Sets:
    Leak Test Twist Clamp ClosedNot specified, implied to be "no leak"Met acceptance criteria (implied)
    Twist Clamp Torque to Open/CloseNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Leak Test Transfer Set to Patient ConnectorNot specified, implied to be "no leak"Met acceptance criteria (implied)
    Patient Connector Torque On to Transfer SetNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Patient Connector Torque Off from Transfer SetNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    MiniCap Test Attachment Torque On/Off to Transfer SetNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Leak Test MiniCap to Transfer SetNot specified, implied to be "no leak"Met acceptance criteria (implied)
    Flow Test CAPD Therapy and APD TherapyNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Locking Titanium Adapter & Locking Cap for Peritoneal Dialysis Catheter:
    Titanium Adapter/Patient Catheter Tubing Seal TestNot specified, implied to be "no leak"Met acceptance criteria (implied)
    Tensile Pull Test Titanium Adapter to Patient Catheter TubingNot specified, implied to be above a minimum thresholdMet acceptance criteria (implied)
    Locking Cap Removal Torque TestNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Locking Cap Torque On TestNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Tensile Strength Adapter Catheter to Luer End of Titanium AdapterNot specified, implied to be above a minimum thresholdMet acceptance criteria (implied)
    Adapter Catheter to Titanium Adapter - Torque On TestNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Adapter Catheter to Titanium Adapter - Torque Off TestNot specified, implied to be within functional rangeMet acceptance criteria (implied)
    Leak Test Adapter Catheter to Titanium AdapterNot specified, implied to be "no leak"Met acceptance criteria (implied)
    Simulation TestingNot specified, general functional performanceMet acceptance criteria (implied)
    Biocompatibility (for all devices):
    CytotoxicityNon-cytotoxicMet acceptance criteria (implied)
    Systemic Toxicity (Acute and Sub-chronic)No significant systemic toxic effectsMet acceptance criteria (implied)
    Irritation (Intracutaneous Reactivity)Non-irritatingMet acceptance criteria (implied)
    SensitizationNon-sensitizingMet acceptance criteria (implied)
    HemocompatibilityBiocompatible with blood componentsMet acceptance criteria (implied)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the functional and biocompatibility test sets. It generically mentions "nonclinical tests." The data provenance is not specified regarding country of origin or whether it was retrospective or prospective, but as these are non-clinical (laboratory/bench) tests, typical clinical study provenance details would not apply. The tests were presumably conducted by Baxter Healthcare Corporation.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable as the provided documentation describes non-clinical performance and biocompatibility testing of physical devices, not assessment of diagnostic accuracy or a similar task that would require expert-established ground truth.


    4. Adjudication Method for the Test Set

    This section is not applicable for the same reasons as point 3. Adjudication methods are typically relevant for clinical studies involving human interpretation or subjective assessments.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. The devices are physical medical instruments (transfer sets, adapters, caps) used in peritoneal dialysis, not AI-powered diagnostic or assistive technologies. Therefore, an MRMC study related to AI assistance for human readers is irrelevant to this submission.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable as the devices are not algorithms or AI systems.


    7. The Type of Ground Truth Used

    For the functional tests, the "ground truth" is established by predefined engineering specifications and performance standards for the device's operation (e.g., no leaks, specific torque values, adequate flow rates). For biocompatibility, the ground truth is established by the biological responses observed in standardized tests performed according to ISO 10993 standards. These are objective measures rather than expert consensus, pathology, or outcomes data in a clinical sense.


    8. The Sample Size for the Training Set

    This section is not applicable as the devices are physical medical instruments, not machine learning algorithms that require training data.


    9. How the Ground Truth for the Training Set was Established

    This section is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1