Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K962295

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1996-07-26

    (71 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K901613, K891695, K944195, K944195

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision QID Blood Glucose Test Strips are intended for in vitro diagnostic use (i.e., for external use only) for the quantitative measurement of glucose in fresh capillary whole blood. For home or professional use with the Precision QID Blood Glucose Sensor. Compatible with the MediSense 2 Card and Pen Blood Glucose Sensors, and the Companion 2 Card and Pen Blood Glucose Sensors.
    The product may also be used by healthcare professionals for quantitative measurement of glucose in venous or arterial whole blood, provided the sample is used within 15 minutes.

    Device Description

    The Precision QID Blood Glucose Test Strip utilizes amperometric biosensor technology to quantitatively measure glucose in whole blood and control solutions. Insertion of a test strip into the sensor, automatically turns the sensor on. A drop of whole blood or control solution is applied to the target area of the test strip and the assay is automatically initiated. A countdown begins and glucose oxidase catalyzes the oxidation of glucose to produce gluconic acid. During the reaction, electrons are transferred by an electrochemical mediator to the electrode surface, generating a current that is measured by the Precision QID Sensor. The size of the current is proportional to the amount of glucose present in the sample, thus giving an accurate reading of glucose concentration after 20 seconds.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Precision QID™ Blood Glucose Test Strip:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of specific correlation coefficient, slope, or y-intercept values. Instead, it presents the results of a comparison study and concludes that the performance is "acceptable and comparable" to the predicate device.

    However, based on the context of similar medical device submissions for blood glucose meters, common performance metrics are correlation coefficient, slope, and y-intercept when compared to a reference method. We can infer the reported device performance from the provided table.

    Performance MetricAcceptance Criteria (Inferred from "acceptable and comparable")Reported Device Performance (Arterial Whole Blood vs. Arterial Whole Blood Ref)Reported Device Performance (Arterial Whole Blood vs. Arterial Plasma Ref)Reported Device Performance (Capillary Whole Blood vs. Capillary Whole Blood Ref - Predicate)
    Correlation Coefficient (r)High correlation (e.g., > 0.95)0.9610.9670.984
    Slope (m)Close to 1 (e.g., 0.95 - 1.05)1.0050.9710.938
    Y-intercept, mg/dLClose to 0 (e.g., +/- 10 mg/dL)-6.20.710.6

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Sizes:

      • For the study comparing the Precision QID Test Strip using arterial whole blood against an arterial whole blood reference method: N = 320
      • For the study comparing the Precision QID Test Strip using arterial whole blood against an arterial plasma reference method: N = 306
      • For the study comparing the predicate Precision QID Blood Glucose Test Strip using capillary whole blood against a whole blood reference method: N = 311 (These results were reported in a previous 510(k) submission #K945887)
    • Data Provenance:

      • The current study (arterial whole blood) was performed "at a university medical center." No specific country of origin is mentioned beyond "university medical center."
      • The study appears to be prospective, as it describes clinical testing being "performed."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given text. While it states the study was done at a "university medical center" and compared to "hexokinase reference methods," it doesn't specify if experts were involved in establishing ground truth beyond running the reference method, nor does it detail their qualifications.

    4. Adjudication Method for the Test Set

    This information is not provided in the given text. Adjudication methods are typically relevant for subjective assessments or when multiple readers are interpreting results, which isn't the primary focus here (it's a quantitative measurement device).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This type of study (AI assistance for human readers) is not relevant for this device, which is a blood glucose test strip, a standalone diagnostic tool, not an AI-powered image interpretation or clinical decision support system.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The study directly assesses the performance of the Precision QID Blood Glucose Test Strip (algorithm + associated hardware to read the strip) against recognized laboratory reference methods (hexokinase). This is essentially the "algorithm only" performance for a quantitative measurement device, as the human-in-the-loop is simply applying the sample and reading the digital result.

    7. The Type of Ground Truth Used

    The ground truth used was reference laboratory methods:

    • Whole blood hexokinase reference method
    • Plasma hexokinase reference method

    Hexokinase methods are considered highly accurate and are a standard for glucose measurement in clinical laboratories.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided for this type of device. The Precision QID Blood Glucose Test Strip is based on a chemical reaction (biosensor technology) rather than a machine learning model that requires a training set. The "development" of the device would involve engineering and chemical optimization, not training on a dataset in the way an AI algorithm is trained.

    9. How the Ground Truth for the Training Set was Established

    As stated above, this question is not applicable as there is no mention or indication of a "training set" in the context of an AI/ML algorithm for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1