Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K993885

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-02-10

    (86 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K791137, K954633

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UltrAblator™ Electrode is intended to be used in arthroscopic applications of resection, ablation, tissue modification, excision of soft tissue, hemostasis of blood vessels and in coagulating soft tissues. Arthroscopic surgery includes the following:
    Knee

    1. Meniscectomy
    2. Lateral Release
    3. Chondroplasty
    4. Synovectomy
    5. ACL Debridement
    6. Plica Removal
    7. Meniscal Cystectomy
      Ankle
    8. Fracture Debridement
    9. Excision of Scar Tissue
    10. Synovectomy
    11. Chondroplasty
      Wrist
    12. Synovectomy
    13. Cartilage Debridement
    14. Fracture Debridement
      Shoulder
    15. Labral Tear Resection
    16. Synovectomy
    17. Excision of Scar Tissue
    18. Acromioplasty
    19. Bursectomy
    20. Subacromial Decompression
    21. Chondroplasty
      Elbow
    22. Synovectomy
    23. Tendon Debridement
    24. Chondroplasty
    Device Description

    The UltrAblator™ Electrode is a monopolar electrode which will be packaged individually or in a kit containing an UltrAblator™ Electrode, ConMed electrosurgical pencil, and a ConMed electrosurgical dispersive pad. The electrode is connected to an electrosurgical generator via the ConMed electrosurgical pencil. The device is inserted into the joint. Upon activation of the generator, the tip of the device electrosurgically ablates and coagulates tissue.
    The UltrAblator™ Electrode is a modification of Linvatec's Ablator Electrode which received FDA clearance under 510(k)# K983652 on March 23, 1999. The electrode will be made available in a range of sizes from 2.4" to 6.3" inches in working length, and in three angles, 0°, 30°, & 90°. The insulating material will be changed from polyolefin shrink to Vitek powder coat.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria for a device's performance, nor does it describe a study proving the device meets specific performance criteria.

    The document is a 510(k) premarket notification for a medical device (UltrAblator™ Electrode) seeking clearance from the FDA based on substantial equivalence to a predicate device. This type of submission typically focuses on ensuring safety and efficacy through comparison to an already marketed device, rather than through a detailed study proving performance against explicit acceptance criteria.

    Here's what the document does provide and why the requested information is absent:

    1. Acceptance Criteria and Reported Device Performance: Not present. The document describes the device, its intended use, materials, manufacturing processes (sterilization, packaging), and identifies a predicate device (Linvatec's Ablator Electrode, K983652). It also mentions "Engineering testing will be performed per ANSI/AAMI American National Standard for Electrosurgical Devices HF-18/1993" for tests like Dielectric Withstand, Voltage Withstand, Run Test, and Resection Effectiveness Test, but it does not provide the acceptance criteria for these tests nor the results that demonstrate the new device meets them. It merely states that the protocols are included in an exhibit (which is not provided in the input text).

    2. Sample size used for the test set and the data provenance: Not present. Since no performance study is detailed, there's no mention of a test set, its sample size, or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not present. This information would be relevant for clinical performance studies, which are explicitly stated as "not required" for this submission.

    4. Adjudication method: Not present, for the same reasons as point 3.

    5. Multi Reader Multi Case (MRMC) comparative effectiveness study: Not present. This type of study is more common for diagnostic imaging AI systems to assess human reader improvement with AI assistance, which is not applicable to this electrosurgical electrode.

    6. Standalone (algorithm only without human-in-the-loop performance) study: Not present. This is a physical electrosurgical device, not an algorithm.

    7. Type of ground truth used: Not applicable, as there's no clinical performance study involving ground truth.

    8. Sample size for the training set: Not applicable, as there's no AI/algorithm being trained.

    9. How the ground truth for the training set was established: Not applicable, as there's no AI/algorithm being trained.

    In summary, the provided document is a 510(k) notification focused on demonstrating substantial equivalence, not a detailed performance study with explicit acceptance criteria and results for the device's functional performance. It references engineering tests and biocompatibility tests but does not include the results or specific acceptance criteria within the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1