Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172839

    Validate with FDA (Live)

    Device Name
    ENDOSOLV
    Manufacturer
    Date Cleared
    2018-02-13

    (147 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K780003

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Root canal sealers solvent when canal retreatment is required.

    Device Description

    ENDOSOLV (new device, K172839/S001) is a solvent used for eugenate-type or phenolic resin-based endodontic cements in the context of endodontic retreatment. This solvent is able to soften conventional zinc oxide eugenol cements and phenolic resin type sealers in the context of endodontic retreatment. It is indicated for use as a root canal sealer solvent when canal retreatment is required. ENDOSOLV (new device, K172839/S001) composition consists of ethyl acetate, amyl acetate and thymol. Ethyl acetate and amyl acetate are the two main components of the formulation that are effective in dissolving zinc oxide-eugenol-based and resin based filling material.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for ENDOSOLV (K172839) and compares it to a predicate device, ENDOSOLV E (K980633). The study focuses on demonstrating substantial equivalence, primarily through performance testing of the new device's ability to dissolve root canal sealers.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The text doesn't explicitly state "acceptance criteria" in a quantitative manner (e.g., "must dissolve within X seconds"). Instead, the performance studies aim to show that the new device, ENDOSOLV, is "as effective as" or "comparable to" the predicate device, ENDOSOLV E, and a positive control (ENDOSOLV R) in softening root canal sealers.

    Therefore, the reported device performance is framed as a comparison to the predicate and positive control.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (ENDOSOLV K172839)
    Dissolution of Eugenate-based Root Canal SealersAs effective as ENDOSOLV E (predicate device) in softening eugenate-based sealers.For Endomethasone SP: 47 seconds (compared to 70 seconds for ENDOSOLV E). For PCS Pulp Canal Sealer: 41 seconds (compared to 68 seconds for ENDOSOLV E). For Sealite regular: 141 seconds (compared to 68 seconds for ENDOSOLV E). Conclusion stated: "ENDOSOLV (new device, K172839/S001) is as effective as ENDOSOLV E (predicate device, K980633) to dissolve eugenate based root canal sealer."
    Dissolution of Resin-based Root Canal SealersComparable to ENDOSOLV R (positive control) in softening resin-based sealers.For resin-based root canal filling material: 50 seconds (compared to 83 seconds for ENDOSOLV E and 68 seconds for ENDOSOLV R). Also, against Acroseal (resin-based root canal sealer): 7 minutes 10 seconds (compared to 19 minutes 57 seconds for ENDOSOLV E). Conclusion stated: "This study showed ENDOSOLV (new device, K172839/S001) results comparable to those obtained with ENDOSOLV R."
    BiocompatibilityNot cytotoxic, not sensitizer, negligible irritation.Cytotoxicity: Not cytotoxic (ISO 10993-5) Hypersensitivity: Not sensitizer (ISO 10993-10) Irritation: Negligible irritation (ISO 10993-10)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Performance Testing: Not explicitly stated. The tables provide softening times for different types of cements, suggesting multiple tests were performed for each, but the exact number of samples or replicates for each test is not provided.
    • Data Provenance: The performance testing was performed by Septodont, the manufacturer. No information on the country of origin of the data or whether it was retrospective or prospective is provided beyond the location of the study site (Septodont itself did the performance testing, biocompatibility was outsourced to Envigo and Phycher).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The performance testing involves direct measurements of softening time for root canal sealers, not interpretation by human experts. Biocompatibility testing follows ISO standards and lab protocols.

    4. Adjudication Method for the Test Set

    Not applicable, as the tests involve objective measurements (softening time) and standardized biocompatibility assays, not expert interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This study is for a medical device (solvent) and involves laboratory performance testing (softening time) and biocompatibility, not an imaging or diagnostic device requiring human reader analysis.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    Not applicable, as the device is a chemical solvent and does not involve an algorithm. Performance is assessed directly through its chemical and physical properties.

    7. The Type of Ground Truth Used

    The ground truth for the performance testing is the objectively measured "softening time" of various root canal sealers when exposed to the solvent. For biocompatibility, the ground truth is established by the results of standardized ISO 10993 tests (cytotoxicity, hypersensitivity, irritation).

    8. The Sample Size for the Training Set

    Not applicable. This device is a chemical solvent, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1