Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K964715

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-04-25

    (151 days)

    Product Code
    Regulation Number
    862.3830
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K771805

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Paramax® Salicylate Reagent is intended for use as quantitative determination of salicylate in plasma and serum.

    Device Description

    The Paramax® Salicylate Reagent is based on colormetric methodology.

    AI/ML Overview
    {
      "acceptance_criteria_and_performance": {
        "table": [
          {
            "criterion": "Correlation with predicate device",
            "acceptance_level": "High correlation coefficient",
            "reported_performance": "Correlation coefficient = 0.999"
          },
          {
            "criterion": "Regression equation",
            "acceptance_level": "Regression equation indicating substantial equivalence",
            "reported_performance": "Paramax® value = (1.16 x aca value) - 1.8 mg/dL"
          }
        ],
        "study_summary": "The study demonstrated substantial equivalence between the Paramax® Salicylate Reagent and the DuPont aca Salicylate Reagent based on the high correlation coefficient (0.999) and a regression equation of Paramax® value = (1.16 x aca value) - 1.8 mg/dL."
      },
      "sample_size_test_set": 49,
      "data_provenance": "Not specified (likely clinical samples, but country/retrospective/prospective not mentioned in snippet).",
      "number_of_experts_ground_truth": "Not applicable; the ground truth for this comparison study was established by the predicate device (DuPont aca Salicylate Reagent).",
      "qualifications_of_experts": "Not applicable",
      "adjudication_method": "Not applicable",
      "mrmc_comparative_effectiveness_study": "No, this was a device-to-device correlation study, not an MRMC study involving human readers.",
      "standalone_performance_study": "Yes, the study evaluated the standalone performance of the Paramax® Salicylate Reagent by comparing its results to a predicate device's results on the same samples.",
      "type_of_ground_truth": "Predicate device's results (DuPont aca Salicylate Reagent). The predicate device's measurements are considered the reference for establishing substantial equivalence.",
      "sample_size_training_set": "Not applicable; this is a device validation study, not a machine learning model training.",
      "how_ground_truth_for_training_set_established": "Not applicable"
    }
    
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1