Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K033579

    Validate with FDA (Live)

    Date Cleared
    2005-07-11

    (606 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K771412

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed multiple clip appliers are intended for the ligation of vessels or other tubular structures.

    Device Description

    The reprocessed multiple clip applier is an automatic ligating clip applier. The device is preloaded with a minimum of 10 titanium ligating clips that individually advance after each clip application.

    AI/ML Overview

    Acceptance Criteria and Study Details for ClearMedical Reprocessed Multiple Clip Appliers

    The provided documents describe the 510(k) summary for the ClearMedical Reprocessed Multiple Clip Appliers, which are reprocessed versions of the Ethicon Ligaclip Multiple Clip Appliers. The submission focuses on demonstrating substantial equivalence to the predicate device through validation of reprocessing procedures, rather than a de novo clinical study with specific performance metrics for the reprocessed device's function in a surgical setting.

    Therefore, the "acceptance criteria" and "reported device performance" are primarily related to the successful validation of the reprocessing process and the maintenance of equivalence to the predicate device in terms of function and safety.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance (as per submission)
    Cleaning ValidationReprocessed devices must be effectively cleaned to remove biological and other contaminants to a pre-defined acceptable level."Validation of cleaning... demonstrate[s] reprocessed clip appliers perform as intended and are safe and effective." (No specific quantitative levels or tests are detailed in the summary, but this indicates successful completion of such validation.)
    Performance ValidationReprocessed devices must maintain the design, material, and functional performance characteristics equivalent to new predicate devices (e.g., clip advancement, clip formation, jaw closure, overall mechanical integrity)."The technological characteristics of design, material, and functional performance of reprocessed multiple clip appliers are unchanged and remain equivalent to the predicate devices." "Validation of... performance... demonstrate[s] reprocessed clip appliers perform as intended and are safe and effective." (Implies testing was conducted to confirm functional equivalence, though specific tests and thresholds are not provided in this summary.)
    Packaging ValidationReprocessed devices must be packaged to maintain sterility and physical integrity until point of use."Validation of... packaging... demonstrate[s] reprocessed clip appliers perform as intended and are safe and effective." (Implies testing was conducted to confirm shelf-life and integrity of packaging.)
    Sterilization ValidationReprocessed devices must be resterilized to a Sterility Assurance Level (SAL) appropriate for the intended use and equivalent to the predicate device."Validation of... sterilization... demonstrate[s] reprocessed clip appliers perform as intended and are safe and effective." (Implies successful validation of the sterilization process.)
    Biocompatibility TestingMaterials of construction, after reprocessing, must remain biocompatible and not leach harmful substances into the patient's body."Biocompatibility testing demonstrate[s] reprocessed clip appliers perform as intended and are safe and effective." (Implies successful completion of biocompatibility evaluations.)
    Substantial EquivalenceThe device must be demonstrated to be as safe and effective as a legally marketed predicate device, with equivalent intended use, technological characteristics, and performance."Based on information provided in this submission, ClearMedical Reprocessed Clip Appliers are substantially equivalent to the identified predicate devices and are safe and effective for their intended use." (This is the ultimate conclusion, supported by the validations mentioned above.)

    Study Details:

    The provided 510(k) summary is for a reprocessed medical device. The "study" described is primarily a series of validation tests for the reprocessing procedures and comparison to the predicate device. It is not a clinical study evaluating the surgical or diagnostic performance of the device in patients in the same way an AI/software device would be.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document does not specify the exact sample sizes used for each validation test (e.g., cleaning, performance, packaging, sterilization, biocompatibility).
    • Data Provenance: The data would have been generated internally by ClearMedical, Inc. during their reprocessing validation activities. This would be prospective testing conducted to support the 510(k) submission. No information on country of origin for this testing is provided, but it is implicit that it was conducted by or for ClearMedical, Inc. in the context of a US regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This type of information is generally not applicable to a reprocessing validation study for a manual surgical instrument. Ground truth would be defined by validated testing methods and specifications (e.g., cleanliness levels, sterility assurance levels, mechanical performance thresholds) rather than expert consensus on clinical findings or images. The "experts" would be individuals with expertise in:
    * Reprocessing validation (cleaning, sterilization)
    * Materials science and biocompatibility
    * Mechanical engineering and functional testing of surgical instruments

    The document does not detail the specific number or qualifications of these individuals.

    4. Adjudication Method for the Test Set:

    Not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies or studies involving human interpretation of data where there might be disagreement among experts. Here, the "truth" is determined by objective physical and chemical testing against pre-defined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    This is not applicable. The device is a reprocessed manual surgical instrument, not an AI or imaging diagnostic device. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This is not applicable. The device is a reprocessed manual surgical instrument, not an algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's validation is based on:

    • Engineering specifications and performance standards: For mechanical function (e.g., clip apposition, jaw closure, clip retention, tensile strength, fatigue resistance).
    • Microbiological standards: For sterility assurance and cleanliness levels.
    • Chemical standards: For residual levels of cleaning agents or material degradation products.
    • Biocompatibility standards: (e.g., ISO 10993 series) for material safety.

    Essentially, the ground truth is "the device successfully meets its pre-defined engineering, performance, safety, and regulatory specifications after reprocessing, demonstrating equivalence to the new predicate."

    8. The Sample Size for the Training Set:

    This is not applicable. The device is a reprocessed manual surgical instrument, not an AI or machine learning model that requires a training set. The validation involves testing samples of the reprocessed physical device.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1