510(k) Data Aggregation

    K Number
    K260265

    Validate with FDA (Live)

    Date Cleared
    2026-02-23

    (26 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K252838

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, and that displays, depending on optional local coils that have been configured with the system, the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

    The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

    Device Description

    MAGNETOM Flow.Ace & MAGNETOM Flow.Plus with software Syngo MR XB10 include new and modified hardware and software compared to the predicate devices, MAGNETOM Flow.Ace & MAGNETOM Flow.Plus with software syngo MR XA70A.

    New compared to predicate:

    • Spine support respiratory (Cushion) as a part of BM Spine Coil Set 1.5T (including new surface material)
    • Gradient Configuration Upgrade

    Modified same as predicate, but with new claim introduced:

    • PETRA (new claim for the existing sequence)

    In addition, the following hardware and software are transferred from the reference device MAGNETOM Flow.Neo with software Syngo MR XB10 (K252838), to the subject devices without any modifications:

    Hardware (New compared to predicate, same as reference):

    • BioMatrix Dockable Table with / without eDrive
    • Comfort Sound: Cushion

    Hardware (Modified compared to predicate, same as reference):

    • Comfort Sound: BM Head/Neck Coil
    • Relocatable Option

    Software (New compared to predicate, same as reference):

    • Open Workflow

    Software (Modified compared to predicate, same as reference):

    • BioMatrix Motion Sensor (SAMER)
    • CS_VIBE
    • SPAIR FatSat Improvements: SPAIR "Abdomen & Pelvis" mode and SPAIR Breast mode
    • Deep Resolve Boost for FL3D_VIBE and SPACE
    • Deep Resolve Sharp for FL3D_VIBE and SPACE
    • Preview functionality for Deep Resolve Boost
    • myExam Implant Suite
    • GRE_PC
    • Open Recon 2.0
    • Deep Resolve Boost for TSE
    • "MTC Mode" for SPACE

    Other Modifications and / or Minor Changes (New compared to predicate, same as reference):

    • Eco Power Mode Pro

    Other Modifications and / or Minor Changes (Modified compared to predicate, same as reference):

    • Off-Center Planning Support
    • Flip Angle Optimization (Lock TR and FA)
    • ID Gain (re-naming)
    • Marketing bundle "myExam Companion"
    AI/ML Overview

    Acceptance Criteria and Study Details for Siemens MAGNETOM Flow.Ace and Flow.Plus

    Based on the provided 510(k) clearance letter, the acceptance criteria and study details are as follows. It's important to note that this document primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed de novo clinical trial for device efficacy. Therefore, specific metrics like sensitivity, specificity, or AUC for diagnostic performance are not explicitly stated as acceptance criteria in the typical sense for a new diagnostic claim.

    The acceptance criteria are generally focused on demonstrating that the new and modified features of the MAGNETOM Flow.Ace and Flow.Plus systems maintain equivalent safety and performance to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Software Verification and Validation: New and modified software features conform to design specifications and perform as intended.Testing demonstrated that the new and modified software features performed as intended, supporting substantial equivalence.
    Functionality of New/Modified Hardware: New hardware ("Spine support respiratory (Cushion)"), and modified hardware ("BioMatrix Dockable Table with/without eDrive", "Comfort Sound: BM Head/Neck Coil", "Relocatable Option") perform as intended and safely.Testing demonstrated that the new and modified hardware features performed as intended and safely, supporting substantial equivalence.
    Biocompatibility: Surface of applied parts (Spine support respiratory cushion and Comfort Sound Cushion) in contact with patients is biocompatible.Biocompatibility testing (per ISO 10993-1) was conducted, demonstrating compliance.
    Electrical Safety and Electromagnetic Compatibility (EMC): The complete system complies with relevant safety and EMC standards.Electrical safety and EMC testing (per IEC 60601-1 and related collateral standards) was conducted, demonstrating compliance.
    Acoustic Noise: The device meets acoustic noise limits.Acoustic noise measurement procedures (per NEMA MS 4-2010), were followed.
    Compliance with General Standards: All modifications comply with recognized industry standards (e.g., IEC 60601-1 series, ISO 14971, IEC 62304, NEMA, DICOM).The device conforms to the listed FDA recognized and international IEC, ISO, and NEMA standards.
    Clinical Equivalence (through sample images and comparative literature): New features (e.g., Deep Resolve Boost, CS_VIBE, PETRA, BioMatrix Motion Sensor) provide information that assists in diagnosis, maintaining the existing indications for use.Sample clinical images were provided as claim evidence. Clinical publications were referenced to support the use and performance of specific features (SAMER, CS_VIBE, PETRA). The conclusion was that the features bear an equivalent safety and performance profile.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific sample size for a test set in the context of clinical images or patient data.

    • Data Provenance: "Sample clinical images" were provided as "claim evidence." The origin (e.g., country) and whether this data was retrospective or prospective are not specified in the provided text. The clinical publications referenced are peer-reviewed articles, which would typically involve patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not provide information regarding the number of experts, their qualifications, or how ground truth was established for any "test set" of images. The phrase "interpreted by a trained physician" is used in the Indications for Use, which is a general statement about MR diagnostic devices.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for a test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not explicitly done or reported in this document. The submission relies on "sample clinical images" as "claim evidence" and references existing clinical publications for certain features to demonstrate substantial equivalence, rather than a direct comparative study showing improvement with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Study was Done

    The document discusses "software verification and validation" and "nonclinical tests" which would imply standalone performance testing of the algorithms and features. However, it does not provide specific metrics or results for standalone algorithm performance (e.g., sensitivity, specificity, or AUC if applicable to specific features). The focus is on the performance of the integrated system.

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for any specific evaluation. The "Indications for Use" mention that images "when interpreted by a trained physician yield information that may assist in diagnosis." For the referenced clinical publications, the ground truth would depend on the methodology of those studies (e.g., pathology, clinical follow-up, expert consensus in a research setting), but this information is external to the 510(k) summary itself.

    8. The Sample Size for the Training Set

    The document does not provide any information regarding the sample size for a training set. This is a 510(k) submission for an updated MR system, not a de novo AI/ML device where training data details are typically prominent. While features like "Deep Resolve Boost" and "Deep Resolve Sharp" likely leverage AI/ML, the details of their development and training are not disclosed in this summary.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide any information on how ground truth was established for a training set.