Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K213951, Shoreline ACS and RT Interbody System, SeaSpine, K233414, Meridian Interbody System, SeaSpine, K233694
    , WaveForm A Interbody System, SeaSpine, K233694

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reef TO/TA System: When used as an intervertebral body fusion device, the Reef TO/TA System with NanoMetalene® surface technology is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fission and supplemental fixation.

    Regatta Lateral System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Regatta Lateral System with NanoMetalene® surface technology is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD, defined as back pain of discogenic origin, with degeneration of the disc confirmed by history and radiographic studies). It is intended for use at ether one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of DDD with up to Grade 1 spondylolisthesis at the involved level(s). The interior of the interbody spacer component may be packed with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The Regatta Lateral System is intended for use with supplemental fixation. TruProfile Interbody Implants: The Regatta Lateral System assembled with the TruProfile Lateral Plate, when used with Screws, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The interior of the spacer component may be packed with autogenous bone graff and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The Regatta Lateral System assembled with the 1 - hole TruProfile Lateral Plate, when used with Screws, is intended for use with supplemental fixation.

    Explorer TO System: When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in sketally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation.

    WaveForm C Interbody System: The WaveForm C Interbody System are interbody fusion devices intended for use in skeletally mature patient with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. When used as a standalone system, the WaveForm C Interbody System, which includes the 2,3,4-hole TruProfile plates and 2-hole No-profile interfixated spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. When the WaveForm C Interbody (excluding 2-hole No-profile interfixated spacer) is used with supplemental fixation, such as anterior cervical plates, the WaveForm C Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

    WaveForm TO Interbody System: When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cortical, cancellous, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation.

    WaveForm TA Interbody System: When used as an intervertebral body fusion device, the system is intended for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had six months of non-operative treatment. The device is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cortical, cancellous, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation.

    FORZA XP Expandable Spacer: FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s). FORZA XP Expandable Spacer System is intended for use with autograft comprised of cancellous, cortical, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation. Patients must have undergone a regimen of at least six months of non-operative treated with FORZA XP Expandable Spacer System.

    Shoreline ACS and RT Interbody System: Shoreline ACS Interbody System: The Shoreline ACS Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature had at least six (6) weeks of non-operative treatment. These devicesare to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion. When used as a standalone system, the Shoreline ACS Interbody System, which includes the TruProfile plates and Noprofile spacer, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline ACS interbody spacers with > 20° lordosis are intended to be used with supplemental fixation. When the Shoreline ACS Interbody (excluding the No-profile spacer) is used with supplemental fixation, such as anterior cervical plates, the Shoreline ACS Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1). Shoreline RT Interbody System: The Shoreline RT Interbody System with NanoMetalene® surface technology are interbody fusion devices intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) for multiple contiguous levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. These devices are to be filled with autograft bone and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. When used as a standalone system, the Shoreline RT Interbody System, which includes the TruProfile plates, is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the (C2-T1) and must be used with bone screw fixation and locking covers. The Shoreline RT interbody spacers with ≥ 20° lordosis are intended to be used with supplemental fixation. When the Shoreline RT Interbody System is used with supplemental fixation, such as anterior cervical plates, the Shoreline RT Interbody System is intended to be used as an adjunct to spinal fusion procedures at multiple contiguous levels of the cervical spine (C2-T1).

    Meridian Interbody System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The Meridian System with NanoMetale@ surface technology when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interded to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The Meridian Interbody is intended for use with supplemental fixation. No-Profile Interbody Implants with Screws: The Meridian System 2-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion. The Meridian System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. The Meridian System 4-Hole No-profile Interbody, when used with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The Meridian System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw traiectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. No-Profile Interbody Implants with Inline Fixation Anchors: The Meridian System No-profile Interbody, when used with Inline Fixation Anchors and a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The Meridian System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. TruProfile Interbody Implants: The Meridian System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

    WaveForm A Interbody System: Interbody Device (IBD) Implants (i.e., interbody implants used alone): The WaveForm A System Interbody, when used with or without a Spin Plate, is indicated for use as an adjunct to fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm A System Interbody is intended for use with supplemental fixation. No-profile Interbody Implants with Screws: The WaveForm A System 2-Hole and 3-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed cancellous, cortical, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm A System 2-Hole and 3-Hole No-profile Interbody must be used with the maximum number of screws allowed. Hyperloriosis sizes (greater than 20°) are intended for use with supplemental fixation. The WaveForm A System 4-Hole No-profile Interbody, when used with Screws and with or without a No-profile Locking Cover, is a standalone interbody implant indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-overative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The WaveForm A System 4-Hole No-profile Interbody may be used as a standalone device only when at least two Screws are inserted with one inferior and one superior screw traiectory. If the surgeon chooses to use the 4-Hole No-profile Interbody with fewer than two Screws inserted with one inferior and one superior screw trajectory, then additional supplemental fixation system must be used. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation. No-profile Interbody Implants with Inline Fixation Anchors: The WaveForm A System No-profile Interbody, when used with Inline Fixation Anchors and with a No-profile Locking Cover, is indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The intended to be used with autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The WaveForm A System No-profile Interbody with Inline Fixation Anchors is intended for use with supplemental fixation. TruProfile Interbody Implants: The WaveForm A System TruProfile Interbody assembled with the Anterior Plate, when used with Screws, an Anterior Plate Locking Cover, and with or without a Spin Plate, is a standalone indicated for use as an adjunct to fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the device. The interbody is intended to be used with autogenous bone graft composed of cancellous, cortical. and/ or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. Hyperlordotic sizes (greater than 20°) are intended for use with supplemental fixation.

    Device Description

    Reef TO/TA System: The Reef TO/TA Interbody System featuring NanoMetalene® surface technology consists of straight, rectangular-shaped (Reef TO) and curved, banana-shaped (Reef TA), posterior lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface laver of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical. and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    Regatta Lateral System: The Regatta Lateral System featuring NanoMetalene surface technology consists of lateral lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (ASTM F2026) with from tantalum (per ASTM F560) radiographic markers and a one-micron thick surface layer of commercially pure titanium (per ASTM F67). The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include central graft windows for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation. The spacers are individually packaged and gamma sterilized. The Regatta Lateral interbody spacers can be used alone with supplemental fixation or in combination with fixation implants (i.e., plates, screws, and locking covers) to create the TruProfile Interbody configuration. The fixation implants are manufactured from titanium alloy (Ti-6A1-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    Explorer TO System: The Explorer TO System is a lumbar intervertebral body fusion device used to act as a disc spacer and hold bone graft to promote fusion in the lumbar spine. The system consists of non-sterile implants manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136). The system is comprised of height expandable and lordotic expandable interbody spacers, and both configurations are available in a range of lengths, widths, heights, and lordotic angles. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as travs and caddies that mav be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    WaveForm C Interbody System: The WaveForm C Interbody System consists of anterior cervical intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft material, are individually packaged and gamma sterilized. The WaveForm C Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with the Shoreline ACS Interbody System non-sterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile and No-profile configurations. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136). The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    WaveForm TO Interbody System: The WaveForm TO Interbody System consists of posterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The straight, rectangular-shaped interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    WaveForm TA Interbody System: The WaveForm TA Interbody System consists of posterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The curved, banana-shaped interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. The spacers are individually packaged and gamma sterilized. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    FORZA XP Expandable Spacer System: The FORZA XP Expandable Spacer System is comprised of an assortment of single-use spacers made from titanium alloy and PEEK polymer with height expansion capability. The implants feature a bulleted nose for ease of insertion and anti-migration ripples on both the inferior and superior surfaces to provide increased stability and help prevent anterior/posterior movement of the device. FORZA XP spacers are provided in both sterile and non-sterile configurations. The implants are offered in non-lordotic, lordotic, and hyperlordotic configurations to help restore the natural curvature of the spine. The implants can be used in single placement or pairs with typical approaches being transforaminal lumbar interbody fusion (TLIF) and posterior lumbar interbody fusion (PLIF). The FORZA XP Expandable Spacer System is not intended to be used as a stand-alone device and must be used with a supplemental fixation system.

    Shoreline ACS Interbody System: The Shoreline ACS Interbody System featuring NanoMetalene® surface technology consists of anterior cervical intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a onemicron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft material or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation, are individually packaged and gamma sterilized. The Shoreline ACS Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with non-sterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile configurations. The fixation implants are manufactured from titanium allov (Ti-6Al-4V ELI per ASTM F136). The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as travs and caddies that mav be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    Shoreline RT Interbody System: The Shoreline RT Interbody System featuring NanoMetalene® surface technology consists of anterior cervical intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and/or titanium alloy (Ti-6A1-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a onemicron thick surface layer of commercially pure titanium (per ASTM F67) that provides a microscopic roughened surface with nano-scale features. The interbody spacers are offered in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy. The spacers, which include a central graft window for receiving autogenous bone graft and/or allogenic bone graft material or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation, are individually packaged and gamma sterilized. The Shoreline RT Interbody System can be used with supplemental fixation, such as an anterior plate, or as a standalone construct in combination with the Shoreline ACS Interbody System nonsterile fixation implants (i.e., plates, bone screws, and locking covers) to create the TruProfile configuration. The fixation implants are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136). The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    Meridian Interbody System: The Meridian Interbody System featuring NanoMetalene® surface technology consists of anterior lumbar intervertebral body fusion devices manufactured from polyetheretherketone (PEEK) (per ASTM F2026) with tantalum (per ASTM F560) and titanium alloy (Ti-6Al-4V ELI per ASTM F136) markers for radiographic visualization and NanoMetalene, which is a one-micron thick surface layer of commercially pure titanium (per ASTM F67). NanoMetalene surface technology provides a microscopic roughened surface with nano-scale features. The interbody spacers are available in a variety of footprints, heights, and lordotic options to accommodate variations in pathology and patient anatomy and include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation. The spacers are individually packaged and gamma sterilized. The Meridian Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6A1-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as travs and caddies that mav be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    WaveForm A Interbody System: The WaveForm A Interbody System consists of anterior lumbar intervertebral body fusion devices additively manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F3001). The interbody spacers are offered in a variety of footprints. heights, and lordotic options to accommodate variations in pathology and patient anatomy as well as with and without a sodium hydroxide (NaOH) surface treatment that provides a microscopic, roughened surface with nanoscale features. The interbody spacers, which include a central graft window for receiving autogenous and/or allogenic bone graft composed of cancellous, cortical, and/or corticocancellous bone or a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion and supplemental fixation, are individually packaged and gamma sterilized. The WaveForm A Interbody System spacers can be used alone with supplemental fixation or in combination with fixation implants to create the TruProfile and No-profile Interbody configurations. Fixation implants include Anterior Plates, Screws, Inline Fixation Anchors, Anterior Plate Locking Covers, and No-profile Locking Covers, all of which are manufactured from titanium alloy (Ti-6A1-4V ELI per ASTM F136) and provided non-sterile. The system includes the associated non-sterile instruments that facilitate the placement, adjustment, and removal, if necessary, of the system implants as well as trays and caddies that may be used for storage, protection, and organization prior to and during the steam sterilization process for the non-sterile components.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and study information:

    Based on the provided text, the submission is not about a medical device with an artificial intelligence/machine learning (AI/ML) component or software. The text describes various intervertebral body fusion devices, their indications for use, and their technological characteristics. The "performance testing summary" explicitly states:

    "The purpose of this submission is to expand the indications for use of the Orthofix Medical Inc. Interbody System devices to include use in the use of a bone void filler as cleared by FDA for use in intervertebral body fusion to facilitate fusion. Because the subject device is identical to the previously cleared primary predicate devices (K231076. K221751. K203714. K231735), it was determined that existing testing data is adequate to support the expanded indications proposed in the current submission."

    This indicates that the submission relies on comparison to predicate devices and existing testing data for mechanical and material properties, rather than new studies demonstrating software performance, AI efficacy, or human-in-the-loop improvements. Therefore, most of the requested information regarding acceptance criteria for AI models, sample sizes for test/training sets, expert ground truth, adjudication methods, and MRMC studies are not applicable.

    However, I can extract the relevant information about the "acceptance criteria" and the "study" that proves the device meets them in the context of this 510(k) submission, even without an AI component.

    Acceptance Criteria and Device Performance (in the context of a 510(k) "Substantial Equivalence" claim):

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (as stated in the document)
    Mechanical properties (strength, stiffness, durability, fatigue resistance) are equivalent to predicate devices."The subject devices and predicate devices are identical in all respects except for the indications for use (as described above)." "existing testing data [from predicate devices] is adequate to support the expanded indications proposed in the current submission."
    Biocompatibility is equivalent to predicate devices.Devices are made from PEEK (ASTM F2026), Tantalum (ASTM F560), Titanium (ASTM F67), or Titanium Alloy (Ti-6Al-4V ELI per ASTM F136/F3001), which are well-established biocompatible materials for implants. NanoMetalene surface technology is also referenced.
    Sterilization methods are validated and equivalent to predicate devices.Devices are individually packaged and gamma sterilized (e.g., Reef TO/TA System, Regatta Lateral System, WaveForm C Interbody System, WaveForm TO Interbody System, WaveForm TA Interbody System, Shoreline ACS Interbody System, Shoreline RT Interbody System, Meridian Interbody System, WaveForm A Interbody System). Non-sterile components (instruments) are steam sterilized.
    Functional characteristics (e.g., ability to hold graft, facilitate fusion, accommodate patient anatomy) are equivalent to predicate devices.Devices are designed with central graft windows for autogenous/allogeneic bone graft/bone void filler. They come in various footprints, heights, and lordotic options.
    Indications for Use are supported by established clinical practice and predicate device performance.The primary "study" is a demonstration of substantial equivalence by comparing the subject devices to multiple predicate devices that already have similar indications for use, thereby covering the expanded use with bone void filler.

    Study Information (Based on the provided Text):

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • Not Applicable in the traditional sense for testing an AI model. This submission relies on a claim of "substantial equivalence" to predicate devices. There is no new test set of patient data being evaluated for the device's performance because the device itself (hardware) hasn't changed, only the allowable bone graft material. The "testing data" referenced is existing data from the predicate devices. The document does not specify the provenance of this existing data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Not applicable. There is no "test set" in the context of an AI model requiring expert-established ground truth. The acceptance is based on engineering principles, material science, and comparison to legally marketed predicate devices.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. See point 2.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is not an AI/ML device, so an MRMC study is not relevant to this submission.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not applicable for an AI ground truth. The "ground truth" for this type of submission is the established safety and effectiveness of the predicate devices for their stated indications, combined with engineering and material testing data demonstrating their mechanical properties and biocompatibility.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established:

      • Not applicable. This is not an AI/ML device.

    Summary regarding AI/ML:

    The provided document describes a 510(k) submission for intervertebral body fusion devices. The core of the submission is to expand the indications for use of these existing devices to include the use of bone void fillers. The approval is based on demonstrating substantial equivalence to previously cleared predicate devices, asserting that the subject devices are "identical in all respects except for the indications for use" and that "existing testing data is adequate." Therefore, this document does not contain information related to AI/ML acceptance criteria, performance studies, or ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1