Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241492

    Validate with FDA (Live)

    Date Cleared
    2024-08-23

    (87 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K182865, K223662

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Zirconia Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Zirconia Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Zirconia Guided Surgery Kit Cases requre the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterlization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Dynamic Air Removal (pre-vacuum) – Exposure at 132°C for 4 minutes, 20-minute dry time. Gravity displacement – Exposure at 132°C for 15 minute dry time. Ziroonia Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The Zirconia Guided Surgery Kit Case and the associated instruments is 342.2 g. The weight of the empty Kit is 236.8 grams. Zirconia Guided Surgery Kit Case are recommended not to be stacked during sterilization.

    Nuvo CF Guided Surgery Kit Cases are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Nuvo CF Guided Surgery Kit Cases are intended to allow sterilization of the enclosed medical devices. Nuvo CF Guided Surgery Kit Cases require the use of FDA-cleared wrap to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles: Fractionated vacuum): exposure at 132°C for 4 minutes, drying for 20 minutes. Gravity displacement: exposure at 132°C for 15 minutes, drying for 30 minutes. Nuvo CF Guided Surgery Kit Cases are intended for sterilization of non-porous loads. The combined weight of the Nuvo CF Guided Surgery Kit Cases and the associated instruments is approximately 693 g. The weight of the empty case is 487 grams. Nuvo CF Guided Surgery Kit Cases are not recommended to be stacked during sterilization.

    Device Description

    The Zirconia and Nuvo CF Guided Surgery Kit Cases subject to this submission are reusable rigid containers, composed by a case bottom, a removable inner tray and a lid, which are made of autoclavable polymer. They are indicated to hold various dental instruments (not subject of this submission) in order to organize, sterilize and protect them. Sterilization via moist heat (steam) is performed by the healthcare provider.

    AI/ML Overview

    This document is a 510(k) Summary for the Zirconia and Nuvo CF Guided Surgery Kit Cases. It describes the non-clinical performance data used to evaluate the device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Standard or Test MethodPurpose of the TestingAcceptance CriteriaReported Device Performance
    ANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2Steam sterilization validation, including sterilant penetration and drying timeAll Biological Indicators must be incubated for at least 7 days at 55-60°C. All positive controls for SAL testing must show characteristic growth of the indicator organism.Passed
    AAMI TIR 30:2011Manual cleaning validation• Test Soil: Blood Soil (BLSO)• Cleaning Method: Manual• Residuals Tested: Hemoglobin and Protein• Visual Inspection: No visible soil• Hemoglobin Test: < 2.2 µg/cm²• Protein Test: < 6.4 µg/cm²Passed
    Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff (Life cycle testing)Life cycle (simulate usage) testingThe tested samples must withstand 100 cycles of use (cleaning, sterilization and functional tests) without compromising their functionalities.Passed
    ANSI/AAMI/ISO 10993-5Biocompatibility of subject devices via cytotoxicity testingLess than 30% cell proliferation inhibitionPassed

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes for each test set (e.g., number of kit cases tested for sterilization, cleaning, or life cycle). It mentions "representative samples" for biocompatibility.

    The data provenance is not explicitly stated in terms of country of origin but is from the manufacturer (JJGC Indústria e Comércio de Materiais Dentários S.A. dba Neodent, Brazil, and Straumann USA, LLC). The studies appear to be prospective, specifically designed to validate the new devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This document describes non-clinical performance data involving laboratory testing against established standards. Therefore, it does not involve "experts" in the sense of clinicians establishing ground truth for diagnostic accuracy. The ground truth is defined by the technical specifications and performance requirements established by the listed ISO and AAMI standards.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is laboratory testing against objective criteria defined by standards, not subjective evaluation by human experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is not a study assessing the effectiveness of an AI or diagnostic device with human reader interaction. It relates to the sterilization and reprocessing of medical device cases.

    6. Standalone Performance Study:

    Yes, the studies described are standalone performance studies of the device itself (the Zirconia and Nuvo CF Guided Surgery Kit Cases). The tests evaluate characteristics like sterilization effectiveness, cleaning efficacy, durability over multiple cycles, and biocompatibility, without human-in-the-loop performance being a factor.

    7. Type of Ground Truth Used:

    The ground truth used is based on established international and national medical device standards (ISO 17665-1, ISO 17665-2, AAMI TIR 30:2011, ISO 10993-5) and regulatory guidance documents (FDA Guidance document for ISO 10993-1). These standards define objective criteria for successful performance (e.g., specific microbial reduction, acceptable residual levels, physical integrity over cycles, cytotoxicity limits).

    8. Sample Size for the Training Set:

    Not applicable. These are non-clinical validation studies for physical devices, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set involved. The "ground truth" for the device's performance is established by its compliance with the specified performance standards in the validation testing.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1