    Intended Use

    This product is indicated for temporary use (maximum of 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.

    This product is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). During the implantation period, patients have to use traditional mobility assist devices (e.g. crutches, walkers, canes) for daily activities.
    With pre-clinical validations of antibiotic elution, tibial spacer fatigue test, and knee spacer wear test, only Palacos MV+G bone cement may be used for preparing the spacers.

    Device Description

    Cellbrick Knee Spacer is a temporary knee spacer product, it's suitable for the patients with mature skeleton and required to perform two-stage knee joint prosthesis procedure for infection control. It consists of three components, which are (1) Femoral Spacer, (2) Tibial Spacer, and (3) Canal Rod. The Femoral Spacers and the Tibial Spacers are designed with multiaperture features, which can act as carrier of antibiotic-loaded bone cement to release antibiotic at infection site. The Canal Rod is optional.

    • Femoral Spacer
      The Femoral Spacer is made of UHMWPE (ASTM F648/ISO5834-2). The geometrical appearance of Femoral Spacer is designed and modified based on the 510(k) cleared cruciate-retaining femoral component (K140073). Their articular surface geometry and curvature is identical. In addition, the Femoral Spacer is designed with multiaperture feature. Such scaffold-like features can use as a carrier of antibiotic-loaded cement to release antibiotic at the infection site for infection control.
      The Femoral Spacer can be used for both of left and right knee because of the symmetric condyle design. Five femoral spacer sizes have been prepared for different anatomical demands with anteroposterior (AP) and mediolateral (ML) dimensions. The Femoral Spacer is designed “same/ one up" sizing options with the tibial spacers.
    • Tibial Spacer
      The Tibial Spacer is made of UHMWPE (ASTM F648/ISO5834-2). The bearing surface's curvature and geometrical design of the tibial spacer is the same as the 510(k) cleared ultra-congruent tibial insert (K132752, K150829, and K222700). The Tibial Spacer is designed with multiaperture feature as well. This scaffold-like feature is used as a carrier of antibiotic-loaded cement to release antibiotic at the infection site for infection control.
      Five Tibial Spacer sizes have been prepared for different anatomical demands with anteroposterior (AP) and mediolateral (ML) dimensions.
    • Canal Rod
      The canal rod is made of Ti-6Al-4V alloy (ASTM F136/ISO 5832-3). As a carrier, the rod provides the area for cement to attach so that to increase the surface area of cement. The canal rod is an optional device, it could be applied to patient with deeper infection problem, applying Canal Rod can make surgeon placing antibiotic-loaded cement into the deeper infection site. The Canal Rod is provided with one dimension: Ø 4(diameter) ×80mm (length).
    AI/ML Overview

    The provided text is a 510(k) summary for the "Cellbrick Knee Spacer". It describes the device, its intended use, and compares it to predicate devices. However, the document does not contain any information about acceptance criteria or a study proving that an AI/ML device meets acceptance criteria.

    The requested information (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set details) is all highly relevant to the evaluation of AI/ML-driven medical devices. Since the Cellbrick Knee Spacer is described as a physical implant (femoral spacer, tibial spacer, canal rod made of UHMWPE and Ti-6Al-4V alloy) and not a software or AI/ML product, the submission details focus on pre-clinical validations for physical properties like antibiotic elution, wear tests, fatigue tests, range of motion, constraint performance, and endotoxin testing.

    Therefore, I cannot provide the requested information as it is not present in the provided document. The document describes a physical medical device clearance, not an AI/ML device validation.