510(k) Data Aggregation

    Why did this record match?
    Reference Devices :

    K221140, K191467

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging system intended for use by sufficiently trained healthcare professionals for ultrasound imaging, measurement, and analysis of the human body, which is intended to be used in a hospital or medical clinic.

    The system is intended for use in the following clinical applications: Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), OB/Gyn and Urology.

    Modes of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Directional Power Doppler, Tissue Harmonic Imaging, Tissue Doppler Imaging, 3D/4D Imaging mode, Strain Elastography, Contrast imaging and Combined modes: B/M, B/PWD, B/THI, M/Color M, B/Color Doppler, B/Color Doppler/PWD, B/Power Doppler/PWD.

    Device Description

    This X11 Exp/X11 Elite/X11 Pro/X11 Plus/X11 Super/X11 Senior/X11/E11/E11 Pro/E11 Plus/E11 Elite/X11T/X11U/X11R/X11i/X11s/E11T/E11i/E11s/XR1/XR2/XR3/ER1/ER2/ER3/X10 Exp/X10 Elite/X10 Pro/X10 Plus/X10 Super/X10 Senior/X10/X10T/X10U/X10R/X10i/X10s/E10/E10 Pro/E10 Plus/E10T/E10i/E10s/SU11A EXP/SU11A PRO/SU11A AD/SU11A CU/SU11B EXP/SU11B PRO/SU11B AD/SU11B CU/SU11C EXP/SU11C PRO/SU11C AD/SU11C CU Digital Color Doppler Ultrasound System (hereafter as "X11 Exp Series Digital Color Doppler Ultrasound System") is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

    The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

    This system is a Track 3 device that employs a wide array of probes that include linear array, convex array phased array and etc.

    This system consists of a console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.

    This system is a portable, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, contrast, 3D/4D.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K241949

    Validate with FDA (Live)

    Date Cleared
    2025-01-17

    (198 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K222596, K221140

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by appropriately trained healthcare professionals in hospital for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Cerebral Vasculoskeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph (Cardiac), Laparoscopic, OB/Gyn and Urology.

    Device Description

    This SonoScape P60 Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

    The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

    This system is a Track 3 device that employs a wide array of probes that include linear array, convex array phased array and etc.

    This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.

    This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler, Power Doppler, Direct Power Doppler, or the combination of these modes, Compound, 3D/4D, Elastography and Contrast.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the S-FetusAI feature of the Sonoscape Digital Color Doppler Ultrasound System (P60 Series), based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Device Performance
    S-Fetus-RecognitionAccuracy: Not less than 85%.90.01%
    S-Fetus-MeasureAccuracy: Not less than 80%.85%

    Clinical Performance:

    Evaluation MetricAcceptance CriteriaReported Device Performance
    Identification Precision of the 14 Standard Sections (Primary Evaluation Indicator)Non-inferiority design: The lower limit of the 95% CI for the difference of Identification Precision must be greater than -10% (S-FetusAI compared to manual recognition).The recognition precision of the standard sections of S-FetusAI was not inferior to that of the manual scan.
    Measurement Error of Growth Parameters (Secondary Evaluation Indicator)- Relative errors of measurement for 11 Growth Parameters.- Precision of Sp (spinal cord cone end positioning).S-FetusAI has a good consistency with the manual measurement.The Sp (spinal cone end positioning) precision was basically equal between the trial group and control group.

    2. Sample Size Used for the Test Set and Data Provenance

    • S-Fetus-Recognition Test Set:
      • More than 200 images for each of the 14 standard sections.
      • Total: 3717 images.
    • S-Fetus-Measure Test Set:
      • More than 250 images for each of the 9 standard sections.
      • Total: 2929 images.
    • Data Provenance: A mix of retrospective and prospective data collection in clinical practice. The test sets were collected at three hospitals in China.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Initial Ground Truth Drawing: Five OB/GYN experts with more than 3 years' experience.
    • Cross-checking: Two experts with more than 20 years' experience.
    • Re-checking: Four experts with more than 30 years' experience.
    • Clinical Performance Study (Adjudication of results): Two Readers from an independent evaluation group, consisting of qualified professionals (senior experts who had been involved in prenatal ultrasound examination).

    4. Adjudication Method for the Test Set

    The ground truth was established through a multi-level expert consensus process:

    1. Initial Drawing: 5 OB/GYN experts.
    2. Cross-checking: 2 senior experts.
    3. Re-checking: 4 very senior experts.
      This indicates a robust, multi-stage consensus-based adjudication, effectively a form of "expert consensus with multiple layers of review."

    For the clinical performance testing, a separate "independent evaluation group" with 2 senior experts evaluated the section recognition results and precision of Sp from both the trial (AI-assisted) and contrast (manual) groups.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    • A clinical performance testing was conducted, which involved comparing the S-FetusAI function (Trial group) with manual recognition and measurement by investigators (Contrast group). This can be considered a comparative effectiveness study.
    • Research Group Composition (Contrast Group): Investigators (physicians with more than 3 years of specialized training in obstetric ultrasound examination; total 7 investigators) performing manual recognition and measurement.
    • Effect Size: The study focused on demonstrating non-inferiority rather than a specific improvement percentage for human readers with AI assistance. The conclusion states that "the recognition precision of the standard sections of S-FetusAI was not inferior to that of the manual scan." This implies that the AI-assisted system performs at least as well as, or comparably to, manual recognition. It does not explicitly quantify how much human readers improve when using AI, but rather that the AI's standalone performance (for recognition) and consistency/precision (for measurement) are comparable to manual methods.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance testing was done for the S-FetusAI feature for both S-Fetus-Recognition and S-Fetus-Measure. The results (90.01% accuracy for recognition and 85% accuracy for measurement) are reported as entirely algorithmic performance against established ground truth.


    7. The Type of Ground Truth Used

    The ground truth used was expert consensus. It was established by multiple OB/GYN experts with varying levels of experience, going through a process of initial drawing, cross-checking, and re-checking. Any images not meeting inclusion/exclusion criteria were excluded from the ground truth set.


    8. The Sample Size for the Training Set

    The document explicitly states: "The test set used to test the standalone performance of S-FetusAI are independent of the training data set and tuning data set". However, the sample size for the training set is not provided in the given text.


    9. How the Ground Truth for the Training Set Was Established

    While the document implies that training and tuning datasets were used, the specific method for establishing their ground truth is not mentioned in the provided text. It only details the ground truth establishment for the test set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222596

    Validate with FDA (Live)

    Date Cleared
    2023-02-10

    (168 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K171000, K221140, K171233, K202785, K210053

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The S90 Exp Series Digital Color Doppler Ultrasound System (S90 Exp, S80 Exp, S80 Exp, S80 Plus, S80 Elite, S80 Senior, S80 Super, S70i, S100 Exp, P90i, P80, P80 Exp, P80 Plus, P80 Elite, P80 Senior, P80 Super, P70i, I80-Endo, I80-Surg, 175-Endo, 170-Endo) is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician or sonographer with sufficient clinical ultrasound training for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic. The system is applicable for people who need clinical ultrasound examination.

    The system is intended for use in the following clinical applications: Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esophageal (Cardiac), Laparoscopic, OB/Gyn and Urology.

    Modes of operation include: B, M, PW Doppler, CW Doppler, Color M Doppler, Power Doppler, Power Doppler, Directional Power Doppler, Tissue Harmonic Imaging, Tissue Doppler Imaging, 3D/4D Imaging mode, Strain Elastography, Shear Wave Elastography, Contrast and Combined modes: B/M, B/PWD, B/THI, M/Color M, B/Color Doppler, B/Color Doppler/PWD, B/Power Doppler/PWD.

    Device Description

    This SonoScape S90 Exp Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

    The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

    This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array.

    This system consists of a mobile console with touch screen and keyboard control panel, power supply module, color LCD monitor and optional probes.

    This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler, Power Doppler and Directional Power Doppler Imaging, or the combination of these modes, Contrast Imaging, Strain Elastography, Shear Wave Elastography (S-SWE, P-SWE), SonoFusion, 3D/4D.

    AI/ML Overview

    The provided text describes the Sonoscape Medical Corp.'s S90 Exp Series Digital Color Doppler Ultrasound System and its substantial equivalence to predicate devices, but does not contain information about specific acceptance criteria related to a performance study for the device's clinical or algorithmic performance, nor does it detail a study that proves the device meets such criteria.

    The document primarily focuses on:

    • Indications for Use: What the device is intended for (various clinical applications and modes of operation).
    • Comparison to Predicate Devices: Explaining how the S90 Exp Series is similar to previously cleared devices in terms of intended use, regulations, safety standards, acoustic output, probes, biopsy brackets, and technical characteristics.
    • Non-Clinical Tests: Listing compliance with electrical safety, EMC, acoustic, and software verification standards.

    Therefore, I cannot provide the requested information. The text explicitly states:

    "No clinical testing was required."

    This indicates that, for the purpose of this 510(k) submission, a clinical study demonstrating the device's performance against specific clinical acceptance criteria was not performed or submitted. The substantial equivalence argument is based on the device's similarity to predicate devices and compliance with non-clinical technical standards.