    K Number
    K250112

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-10

    (84 days)

    Product Code
    Regulation Number
    888.3020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K203399, K220234

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FitboneTM Trochanteric is indicated for limb lengthening of the femur. FitboneTM Trochanteric is indicated for adult and pediatric (greater than 12 through 21 years of age) patients.

    Device Description

    FitboneTM Trochanteric is a fully implantable intramedullary lengthening nail and accessories. The subject Fitbone Trochanteric consists of the implantable intramedullary lengthening nail and accessories (Locking screws, Trial nails K-wire and Convenience kits). The subject device is implanted into the medullary canal of the femur and connected to the primary predicate intracutaneous Receiver (K203399) by a bipolar feed line. The external FITBONE Control Set is the same as previously cleared for the reference device Fitbone TAA (K203399) and consists of a control electronics station and transmitter. The power required for the distraction process is controlled by hermetically enclosed motor which draws the telescope apart. The electro-magnetic field sent from the Transmitter to the Receiver is converted in the Receiver into DC-Voltage to supply the motor of the subject Fitbone Trochanteric Nails with voltage, when actioned. The subject Fitbone Trochanteric Nails are available in two different diameter models (D09mm, D11mm), different lengths and lengthening capabilities. The subject nail is anchored to the bone by locking screws. The locking screws to be used with the subject nails are the same as cleared for the primary predicate Fitbone Trochanteric (K233867). Trial nails accessories are available for each variant of the Fitbone Trochanteric nails and are used to simulate the shape of the implant. The Fitbone Trochanteric nail and K-wire are provided in sterile conditions only. The trial nails are provided in non-sterile version only. The locking screws are available in both sterile and non-sterile versions. The subject Fitbone Trochanteric Nails and their accessories are made from implant grade stainless steel 1.4441 (AISI 316LVM) and Silicone Nusilmed.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the FitboneTM Trochanteric, an intramedullary fixation rod used for limb lengthening. This type of device is classified as a Class II medical device. The document focuses on demonstrating substantial equivalence to a predicate device through engineering and mechanical testing, rather than clinical studies involving human efficacy data. Therefore, many of the requested elements pertaining to clinical study design, such as human reader performance, ground truth establishment for a training set, and multi-reader multi-case studies, are not applicable here.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are established through comparison to a predicate device and bench testing against recognized standards. The "performance" described primarily refers to mechanical properties.

    Acceptance Criteria / CharacteristicReported Device Performance (Subject Device)
    Intended Use & Indications for Use"Fitbone™ Trochanteric is indicated for limb lengthening of the femur. Fitbone Trochanteric is indicated for adult and pediatric (greater than 12 through 21 years of age) patients." (Identical to predicate)
    Anatomical SitesFemur (Identical to predicate)
    Intended EnvironmentClinic or Home environment (Identical to predicate)
    Nail MaterialImplant Grade Stainless Steel (1.4441, AISI 316LVM) and Silicone Nusilmed (Identical to predicate)
    Nail Size Range217-357mm in length; 9 and 11mm diameters.
    Maximum Distraction PossibleFrom 40mm (with nail length 217mm) to 80mm (with longer nails) (Equivalent to predicate)
    Tail Nail Geometry2 holes (vs. 3 holes for predicate, assessed via bench testing)
    Method of Distraction/Energy SourceInternal motor electro-magnetically induced by an external transmitter with signal received through a receiver placed just under skin (Identical to predicate)
    Sterilization MethodGas Plasma (Identical to predicate)
    Static Cantilever Bending TestNot explicitly quantified, but stated to "prove substantial equivalency with predicate devices" and "demonstrated not to raise different questions of safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a "sample size" in terms of patient data for a test set, as this was not a clinical study assessing patient outcomes. The testing described is bench testing (mechanical testing). For mechanical testing, samples of the physical device would be tested. The number of such samples is not explicitly stated but would typically involve multiple units for each configuration tested.
    • Data Provenance: The data provenance is from bench testing conducted on the subject device, its primary predicate (K233867), and reference devices (K203399, K220234). This testing would have been done in a laboratory setting, likely at the manufacturer's facility or a third-party testing lab. The origin of the device is Italy (Orthofix S.r.l.). The data is prospective in the sense that the tests were conducted specifically for this submission to evaluate the design change.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Ground truth, in the context of this 510(k), is established through engineering specifications, material standards (e.g., ASTM F138-13), and validated mechanical testing methodologies, rather than human expert interpretation of clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is used for diagnostic devices where human readers interpret medical images or data, and an AI would assist in that interpretation. The FitboneTM Trochanteric is an implantable intramedullary lengthening nail, not a diagnostic AI software.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. The FitboneTM Trochanteric is a physical medical device, not a software algorithm or AI. Its function is mechanical distraction, controlled by an external unit, but it does not have a "standalone algorithm" performance to report in this context.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance evaluation is based on engineering specifications, material properties, and established mechanical testing standards. The primary method for establishing substantial equivalence involves comparing these aspects to a legally marketed predicate device and demonstrating that any differences do not raise new questions of safety or effectiveness.

    8. Sample Size for the Training Set

    Not applicable. This device does not involve a "training set" in the context of AI or machine learning. The device's design and materials are based on established engineering principles and prior device history, not on learning from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233867

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-18

    (195 days)

    Product Code
    Regulation Number
    888.3020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K220234, K955848, K190388, K200246, K974186

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FITBONE® Trochanteric is indicated for limb lengthening of the femur. FITBONE® Trochanteric is indicated for adult and pediatric (greater than 12 through 21 years of age) patients.

    Device Description

    The FITBONE® Trochanteric is a fully implantable intramedullary lengthening nail and accessories. The subject FITBONE® Trochanteric consists of the implantable intramedullary lengthening nail and accessories (Locking screws, Trial nails, K-wire and Convenience kits). The Subject device is implanted into the medullary canal of the femur and connected to the primary predicate intracutaneous Receiver (K163368) by a bipolar feed line. The external FITBONE Control Set is the same as previously cleared for the primary predicate Fitbone TAA device (K203399) and consists of a control electronics station and transmitter. There are no changes to the previously cleared Control Sets and Receiver as a result of this submission. The power required for the distraction process is controlled by hermetically enclosed motor which draws the telescope apart. The electro-magnetic field sent from the Transmitter to the Receiver is converted in the Receiver into DC-Voltage to supply the motor of the subject Fitbone Trochanteric Nails with voltage, when actioned. The subject Fitbone Trochanteric Nails are available in two different diameter models (D09mm, D11mm), different lengths and lengthening capabilities. The subject nail is anchored to the bone by subject locking screws. The locking screws are available in two variants (standard locking screws and revision locking screws), in two diameters, D4.5mm and D4mm, and in multiple lengths. The energy needed for the distraction process is transmitted from the outside by placing the external transmitter over the implanted receiver, which is placed in the subcutaneous tissue during surgery. There is no transcutaneous contact between the implanted intramedullary nail and the outer surface of the patient's body. The subject trial nails accessories are available for each variant of the Fitbone Trochanteric nails and are used to simulate the shape of the implant. The subject Fitbone Trochanteric nail and K-wire are provided in sterile conditions only. The trial nails are provided in non-sterile version only. The bone screws are available in both sterile and non-sterile versions. The subject Fitbone Trochanteric Nails and their accessories are made from, as follows: Nail: implant grade stainless steel 1.4441, according to ASTM F138-13 "Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)", and Silicone Nusilmed (NuSil MED-4870, NuSil MED-1511, Nusil MED 4750, NUSIL MED1-161, NUSIL MED2-4502). Trial nail: implant grade stainless steel 1.4441, according to ASTM F138-13 "Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)" and ASTM F899-20 Standard Specification for Wrought Stainless Steels for Surgical Instruments. Locking screws: implant grade stainless steel 1.4441, according to ASTM F138-13 "Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)" K-wire: implant grade stainless steel 1.4441, according to ASTM F138-13 "Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)" The Subject, as the primary predicate, will be implanted only by Healthcare Professionals (HCP), with full awareness of the appropriate orthopedic procedures

    AI/ML Overview

    The provided text describes the regulatory clearance for the Orthofix Fitbone Trochanteric, an intramedullary lengthening nail. However, this document does not contain information about acceptance criteria, device performance, ground truth establishment, sample sizes for training or testing sets, expert qualifications, adjudication methods, or MRMC studies for an AI/ML powered medical device.

    The document is a 510(k) summary for a traditional medical device (an intramedullary lengthening nail), not an AI/ML-powered device. Therefore, the specific types of studies and criteria outlined in your request are not relevant to this submission.

    The "Performance Analysis" section describes mechanical testing performed on the implantable nails and screws to demonstrate their safety and effectiveness, based on established ASTM and ISO standards for medical devices. This is a standard non-clinical performance evaluation for mechanical orthopedic implants.

    In summary, none of the requested information regarding acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, or study types (MRMC, standalone) for an AI/ML device is present in the provided text because the device is not an AI/ML product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230765

    Validate with FDA (Live)

    Date Cleared
    2023-08-29

    (162 days)

    Product Code
    Regulation Number
    888.3020
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K200430, K220234

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precice Ankle Salvage System is indicated for tibio-talo-calcaneal fusions in adults. When used for tibio-talo-calcaneal fusion, the Precice Ankle Salvage System may be used for open and closed fracture fixation, pseudarthrosis, mal-unions, non-unions, or bone transport of long bones adjacent to the fusion site. The device may be used for subsequent limb lengthening once tibio-talo-calcaneal fusion has been achieved.

    Device Description

    The Precice Ankle Salvage System is a tibio-talo-calcaneal fusion system that consists of the Precice Ankle Salvage nail, locking screws, end caps, and associated general instruments. The Precice Ankle Salvage nail is compatible with an accessory External Remote Controller (ERC). The Precice Ankle Salvage nail, endcap, and locking screws are supplied sterile by gamma radiation. The Precice Ankle Salvage nail contains an enclosed rare earth magnet, telescoping distraction rod, and planetary gearing which allows the length of the nail to be adjusted non-invasively by the ERC. Retraction of the Precice Ankle Salvage nail can be utilized to maintain compression across the tibio-talo-calcaneal ioints post-operatively. The Precice Ankle Salvage nail can also be used to subsequentially correct a limb length discrepancy that may result after the tibio-talo-calcaneal fusion procedure. The subject device is manufactured from medical grade Titanium alloy (Ti-6Al-4V) per ASTM F136. The Precice Ankle Salvage nail is available in various nail styles, diameters, and lengths to accommodate a variety of patient anatomies and surgeon preference. The locking screws are available in a variety of diameters, lengths, and thread styles. The ERC is available in several compatible models.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Precice Ankle Salvage System. It primarily focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed clinical study results of a device's performance against specific acceptance criteria in a human-in-the-loop or standalone AI context.

    Therefore, most of the requested information regarding acceptance criteria, human reader improvement with AI assistance, sample size for test/training sets, expert qualifications, ground truth establishment, and adjudication methods cannot be extracted from the provided text because this document is a regulatory submission for a physical medical device (an intramedullary fixation rod system), not an AI/software as a medical device (SaMD).

    The document describes non-clinical performance verification testing for the physical device's mechanical properties, not clinical performance for diagnostic or prognostic purposes typically associated with AI/SaMD.

    Here's a breakdown of what can be extracted and what cannot:

    Information that can be extracted:

    • Device Name: Precice Ankle Salvage System
    • Regulatory Class: Class II
    • Product Code: HSB (Intramedullary Fixation Rod), HWC (Smooth or threaded metallic bone fixation fastener)
    • Intended Use/Indications for Use: For tibio-talo-calcaneal fusions in adults, for fracture fixation, pseudarthrosis, mal-unions, non-unions, or bone transport of long bones adjacent to the fusion site. May be used for subsequent limb lengthening once fusion is achieved.
    • Predicate Devices: Precice Ankle Salvage System (K200430) and Precice Intramedullary Limb Lengthening System (K220234).
    • Material: Titanium alloy (Ti-6Al-4V) per ASTM F136.

    Information that CANNOT be extracted from this document, as it pertains to AI/SaMD studies:

    • Acceptance Criteria Table and Reported Device Performance (as pertains to AI/SaMD): Not applicable. The document lists mechanical tests for the physical device, not AI-driven performance metrics.
    • Sample sizes used for the test set and data provenance: Not applicable for an AI study. The document refers to non-clinical mechanical testing, not a clinical study with patients/data.
    • Number of experts used to establish the ground truth for the test set and qualifications of those experts: Not applicable. Ground truth for an AI diagnostic device is not relevant here.
    • Adjudication method for the test set: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, not for this type of device.
    • Effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    • The sample size for the training set: Not applicable.
    • How the ground truth for the training set was established: Not applicable.

    Acceptance Criteria and Study for the Physical Device (as present in the document):

    The document defines acceptance criteria for the mechanical performance of the physical implant and demonstrates its substantial equivalence to predicate devices through non-clinical performance verification testing.

    1. A table of acceptance criteria and the reported device performance

    Testing DescriptionApplicable StandardAcceptance Criteria (Implicit)Reported Device Performance
    Dynamic Compression Bending StrengthASTM F1264 - Standard Specification and Test Methods for Intramedullary Fixation DevicesDevice meets the requirements of ASTM F1264 for dynamic mechanical loads typical of intramedullary fixation devices.Results demonstrate substantial equivalence to the primary predicate.
    Static Compression Bending StrengthASTM F1264 - Standard Specification and Test Methods for Intramedullary Fixation DevicesDevice meets the requirements of ASTM F1264 for static mechanical loads typical of intramedullary fixation devices.Results demonstrate substantial equivalence to the primary predicate.
    TorsionASTM F1264 - Standard Specification and Test Methods for Intramedullary Fixation DevicesDevice meets the requirements of ASTM F1264 for torsional strength typical of intramedullary fixation devices.Results demonstrate substantial equivalence to the primary predicate.
    Tensile StrengthASTM F1264 - Standard Specification and Test Methods for Intramedullary Fixation DevicesDevice meets the requirements of ASTM F1264 for tensile strength typical of intramedullary fixation devices.Results demonstrate substantial equivalence to the primary predicate.
    Axial Pullout StrengthASTM F543 - Standard Specification and Test Methods for Metallic Medical Bone ScrewsScrews meet the requirements of ASTM F543 for axial pullout strength.Results demonstrate substantial equivalence to the primary predicate.
    Insertion TorqueASTM F543 - Standard Specification and Test Methods for Metallic Medical Bone ScrewsScrews meet the requirements of ASTM F543 for insertion torque.Results demonstrate substantial equivalence to the primary predicate.
    Ultimate TorqueASTM F543 - Standard Specification and Test Methods for Metallic Medical Bone ScrewsScrews meet the requirements of ASTM F543 for ultimate torque.Results demonstrate substantial equivalence to the primary predicate.
    Distraction and Retraction ForceN/A (Internal Test)Mechanical forces required for distraction and retraction are within acceptable limits for the intended function of the device's lengthening mechanism.Results demonstrate substantial equivalence to the primary predicate.
    Wear Debris TestingN/A (Internal Test)Minimal and acceptable wear debris generation.Results demonstrate substantial equivalence to the primary predicate.
    Corrosion AssessmentN/A (Internal Test)Device materials demonstrate acceptable corrosion resistance.Results demonstrate substantial equivalence to the primary predicate.

    Study that proves the device meets the acceptance criteria:

    The study proving the device meets its acceptance criteria is the "Nonclinical performance verification testing" conducted by NuVasive Specialized Orthopedics, Inc.

    • Type of Study: Non-clinical (benchtop) mechanical and material testing.
    • Purpose: To demonstrate that the subject Precice Ankle Salvage System, with its new titanium material and minor design changes, is substantially equivalent in terms of mechanical performance to its predicate devices.
    • Methodology: The device components were subjected to various mechanical tests (dynamic/static compression bending, torsion, tensile, axial pullout, insertion/ultimate torque) according to recognized ASTM standards (F1264, F543) and internal methods (Distraction and Retraction Force, Wear Debris Testing, Corrosion Assessment).
    • Results/Conclusion: "The results demonstrate that the subject Precice Ankle Salvage System is substantially equivalent to the primary predicate." This implies that the device's performance metrics in these tests were comparable to or better than the predicate device, thereby meeting the necessary performance benchmarks for regulatory clearance.
    Ask a Question