Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232857

    Validate with FDA (Live)

    Date Cleared
    2023-11-06

    (52 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201399, K202925, K212695

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.

    Upon intravenous administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for biliary duct visualization.

    Additionally, the KARL STORZ Endoscopic ICG System enables surgeon to perform minimally invasive cranial neurosurgery in adults and pediatrics and endonasal skull base surgery in adults and pediatrics > 6 years of age using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion using near infrared imaging.

    The KARL STORZ VITOM ICG System is intended for capturing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures in adults and pediatrics >Imonth of age. The VITOM ICG System is intended to provide a magnified view of the surgical field.

    Upon interstitial administration and use of ICG consistent with its approved label, the KARL STORZ Endoscopic ICG System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    The subject device KARL STORZ ICG System includes the following components:

    1. VITOM EAGLE (TH201): a 3D video exoscope with 4K resolution used during open procedures for the evaluation of tissue perfusion, related tissue-transfer circulation in tissue and free flaps used in plastic, micro and reconstructive surgical procedures. The subject device VITOM EAGLE System is being indicated for use in in adults and pediatrics >1month of age.
    2. Fiber Light Cable (495VTE): used to transmit visible and NIR light from the Power LED Rubina light source to the VITOM Eagle.
    3. IMAGE1 Pilot (TC014): used to control the optical functions of the VITOM EAGLE.
    4. Microscope Footswitch (TC019): alternatively used control the optical functions of the VITOM EAGLE
    5. The Power LED Rubina light source (TL400) along with the footswitch (UF101): previously cleared in K201399, K202925 and K212695.
    6. Imagel S Camera Control Unit (TC201US, TC304US): previously cleared in K201399, K202925 and K212695.
    AI/ML Overview

    The provided text describes the KARL STORZ ICG Imaging System and its acceptance criteria, along with a summary of the non-clinical performance data used to demonstrate substantial equivalence to a predicate device. However, it does not describe a study involving an AI algorithm. The device is an imaging system that uses Indocyanine Green (ICG) fluorescence for various surgical visualizations.

    Here's a breakdown of the requested information based only on the provided text, heavily noting limitations due to the absence of AI-specific study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a non-AI imaging system without specific AI performance metrics, the acceptance criteria are generally related to the technical performance of the imaging capabilities. The document states that the KARL STORZ ICG Imaging System (subject device) was compared to the predicate VITOM II ICG/NIR telescope of the KARL STORZ ICG Imaging System (K212695). The performance was demonstrated by testing for:

    Acceptance Criteria (Performance Metric)Reported Device Performance (Subject Device vs. Predicate)
    Spatial ResolutionSuccessfully demonstrated by comparison
    Signal to Noise Ratio and NoiseSuccessfully demonstrated by comparison
    Dynamic RangeSuccessfully demonstrated by comparison
    Geometric DistortionSuccessfully demonstrated by comparison
    Depth of FieldSuccessfully demonstrated by comparison
    Illumination Detection UniformitySuccessfully demonstrated by comparison
    LatencySuccessfully demonstrated by comparison
    Penetration DepthSuccessfully demonstrated by comparison
    Simultaneous Color ContrastSuccessfully demonstrated by comparison
    Minimum Detectable Concentration of ICGSuccessfully demonstrated by comparison
    3D Zoom and RotationSuccessfully demonstrated by comparison
    2D and 3D Mode TransitionSuccessfully demonstrated by comparison
    Image AlignmentSuccessfully demonstrated by comparison
    Photobiological SafetySuccessfully demonstrated by comparison
    Electrical Safety and EMC (IEC 60601-1, IEC 60601-1-2)Follows FDA recognized consensus standards and tested accordingly

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of an AI algorithm or patient data. The performance evaluation was based on non-clinical bench testing comparing the subject device's imaging capabilities to a predicate device. Therefore, there's no mention of sample size in terms of patient data or data provenance (country of origin, retrospective/prospective). The "sample" here would refer to the physical devices and controlled test scenarios.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As this is a non-AI imaging system being evaluated via bench testing, there are no "experts" establishing ground truth for a test set of images or patient data. The ground truth for the technical performance criteria would be established by validated measurement techniques and instrumentation during the bench tests.

    4. Adjudication method for the test set

    Not applicable, as there is no "test set" in the context of human expert review or an AI algorithm's output. The evaluation was based on objective technical measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications." Furthermore, this is not an AI-assisted device, so MRMC studies on AI assistance would not be relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is an imaging system, not a standalone AI algorithm. It produces images for human surgeons to interpret.

    7. The type of ground truth used

    For the non-clinical performance data, the ground truth was based on objective measurements from bench testing (e.g., measuring spatial resolution, signal-to-noise ratio, etc.) against established technical specifications or a predicate device's performance.

    8. The sample size for the training set

    Not applicable. The device does not involve an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no AI algorithm or training set.


    Summary of AI-related information (or lack thereof):

    The provided text describes a submission for a KARL STORZ ICG Imaging System, which is a medical device for real-time visible and near-infrared fluorescence imaging during surgery. The entire document focuses on demonstrating the substantial equivalence of this updated imaging system to a previously cleared predicate device through non-clinical bench testing. There is no mention of any artificial intelligence (AI) component, machine learning model, or any studies related to AI performance, human-in-the-loop improvements with AI, or standalone algorithm performance. Therefore, most of the questions regarding AI-specific criteria cannot be answered from the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1