    Intended Use

    The STERLINK Plus sterilizer with STERLOAD cassette is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal medical devices used in healthcare facilities. A pre-programmed sterilization lumen cycle operates at low temperature and is thus suitable for processing medical devices sensitive to heat and moisture.

    The STERLINK Plus can sterilize *:

    • · Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • · Medical devices with a single stainless steel lumen with:
    • o An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

    *The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 3 .97 lbs.

    Tyvek® Roll with CI for STERLINK Sterilizer, when used in CHAMBER mode, is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERLINK sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 1 month post sterilization.

    The materials compatibility for use in the Tyvek® Roll with CI for STERLINK Sterilizer, when used in CHAMBER mode, are as follows: Aluminum 6061, Stainless Steel 304, Stainless Steel 316L, Titanium, Acrylonitrile Butadiene Styrene (ABS), High Density Polyethylene (HPDE), Polypropylene (PP), Polytetrafluorethylene (PTFE) and Silicone (Hardness 50).

    The maximum load weight that can be placed in the Tyvek® roll is:

    • · 3.97 pounds (1.8kg) for CHAMBER mode of FPS-15s Plus
    • 3.97 pounds (1.8kg) for CHAMBER mode of STERLINK plus sterilizer
    • 1.54 pounds (0.7kg) for CHAMBER mode of STERLINK mini sterilizer

    The roll is printed with a chemical indicator bar which is a process indicator (ISO 11140-1 :2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINK sterilizer.

    The Tyvek® Roll with CI for STERLINK Sterilizer is offered in the follow 1 type: · Sterilization roll. Flat

    Terragene Bionova® SCBI (BT96) is a self-contained biological indicator inoculated with viable 10^6 Geobacillus stearothermophilus bacterial spores and is intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. BT96 has Super Rapid readout at 30 minutes at 60°C.

    Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio) incubate at 60°C and 37°C and read the Terragene Bionova® SCBI for fluorescent results at the times prescribed in the User Manuals.

    Terragene Chemdye® (CD42) is a chemical process indicator intended for monitoring the efficacy of vaporized hydrogen peroxide sterilization processes. The chemical indicator changes from red to yellow for CD42 to indicate that the conditions of the cycle have been met.

    Terragene Cintape® (CT40) is a chemical process indicator tape intended for monitoring the efficacy of Vaporized Hydrogen Peroxide sterlization processes. The indicating tape changes from purple to green when exposure to vaporized hydrogen peroxide.

    The self-contained biological indicator and chemical processing indicators are intended for use with.

    • · CHAMBER mode of FPS-15s Plus sterilizer
    • · CHAMBER mode of STERLink plus sterilizer
    • · CHAMBER mode of STERLINK mini sterilizer
    Device Description

    The STERLINK plus sterilizer with STERLOAD™ cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments. This sterilizer offers an effective, safe, fast, economical, easy to use, reliable, and flexible sterilization method.

    This system consists of a main device connected pump module and cassette which are called the STERLINK plus and STERLOAD™ cassette, respectively. The STERLOAD™ cassette contains 58-59.5% (weight concentration) of hydrogen peroxide (HzO2) which is utilized as the sterilant.

    Tyvek® Roll with CI for STERLINK™ Sterilizer is intended to be used to contain medical devices to be terminally sterilized in the STERLINK™ sterilization system. The medical devices are inserted into the roll, sealed, and then sterilized in the STERLINK™ sterilization system. After completion of the sterilization process, the roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices for up to 1 month post sterilization.

    The roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERLINKTM sterilizer.

    The Tyvek® Roll with CI for STERLINK™ Sterilizer is offered in one type as a Flat Sterilization roll. The sterilization roll is made from a Tyvek® sheet and a clear plastic film that are heat sealed together on opposite two sides. After being cut into a suitable length, the product to be sterilized is placed inside and the two open ends are heat sealed. The process indicator printed on the Tyvek® will exhibit a color change after the roll is exposed to hydrogen peroxide (H2O2).

    Terragene® Bionova® BT96 Fluorescence Super Rapid Readout Biological Indicators are singleuse Self-Contained Biological Indicators (SCBIs) that consist of a polypropylene tube, a spore carrier, and a glass ampoule with a culture medium, enclosed with a colored cap. Each tube contains a population of Geobacillus stearothermophilus ATCC 7953 spores inoculated on a spore carrier, a plastic cap with holes and a barrier permeable to Plasma or Vaporized Hydrogen Peroxide. Each BT96 has a Process Indicator on label that changes from purple to green when exposed to hydrogen peroxide. The Bionova® BT96 Biological Indicators have been designed for monitoring of Vaporized Hydrogen Peroxide sterilization processes when used in conjunction with Bionova® IC10/20FRLCD or Mini-Bio Auto-Readers Incubators.

    Chemdye® CD42 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of plastic strips printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes within loads, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items.

    Cintape® CT40 Process Indicators (Type 1 according to ISO 11140-1:2014 standard) are singleuse chemical indicators that consist of a roll of self-adhesive plastic tape printed with indicator ink. These indicators have been designed to monitor Plasma or Vaporized Hydrogen Peroxide sterilization processes, ensuring an adequate exposure to the sterilizing agent during the sterilization process and allowing to distinguish between processed and unprocessed items. The adhesive component of the tape allows the adhesion to different types of packaging and wraps, such as cloth, paper, and plastic.

    AI/ML Overview

    This document describes the acceptance criteria and supporting study for the STERLINK™ plus Sterilizer with STERLOAD™ Cassette and associated components (Tyvek® Roll with CI for STERLINK™ Sterilizer, Terragene Bionova® SCBI (BT96), Terragene Bionova® Reader Incubators (IC10/20FRLCD, Mini-Bio), Terragene Chemdye® (CD42), and Terragene Cintape® (CT40)).

    The document asserts substantial equivalence to a predicate device (STERLINK™ FPS-15s Plus Sterilizer with STERLOAD™ Cassette, K212200), and therefore, the acceptance criteria are generally met by demonstrating compliance with recognized standards and establishing equivalence in performance.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission demonstrating substantial equivalence to a predicate, specific numerical "acceptance criteria" for novel performance metrics are not explicitly stated in the provided text as they would be for a de novo device. Instead, the acceptance criteria are implicitly that the device performs as well as or better than the predicate device and meets established standards for sterilization, packaging, and indicators. The reported device performance is presented as "Pass" for compliance with these standards.

    Acceptance Criteria CategorySpecific Criteria / Standard MetReported Device Performance
    STERLINK™ plus Sterilizer with STERLOAD™ Cassette
    Risk managementISO 14971Pass
    Human factors and usability engineeringGuidance / IEC 62366-1Pass
    BiocompatibilityISO 10993-5Pass
    Software validationIEC 62304 / FDA GuidancePass
    Electrical safetyIEC 60601-1 / IEC 61010-1Pass
    Electromagnetic compatibility (EMC)IEC 60101-1-2Pass
    Resistance validation for biological indicator testISO 11138-1:2017Pass
    Lumen sterilizationISO 14937:2009Pass
    Surface sterilizationISO 11737-1:2018 / ISO 11737-2:2009Pass
    Mated surface sterilizationISO 14937:2009 / ISO 11737-1:2018 / ISO 11737-2:2009Pass
    Simulated use testASTM E1837-96(2014) / ISO 11737-1:2018Pass
    In-use testASTM E1837-96(2014)Pass
    Sporicidal activity testAOAC 966.04Pass
    Bacteriostasis testISO 11737-1:2018Pass
    Material compatibility testASTM D638, ASTM E8/E8M-16ae1, ASTM D790, ASTM E290-14, ASTM D256, ASTM E23-18, ASTM E1164, ASTM D3985, ASTM F1249Pass
    Delivery validationASTM D4169-14Pass
    Hydrogen peroxide gas detection testOSHA analytical method 1019Pass
    Tyvek® Roll with CI for STERLINK™ Sterilizer
    Sterilization efficacy test (for sterilant penetration)ISO 14937:2009Pass
    CI of sterilization roll validationANSI/AAMI/ISO 11140-1:2014Pass
    Shelf-life (Internal pressurization test)ASTM F1980-16:2016 / ASTM F1140/F1140M-13:2013Pass
    Shelf-life (Visual inspection test)ASTM F1980-16:2016 / ASTM F1886/F1886M-16:2016Pass
    Shelf-life (Dye penetration test)ASTM F1980-16:2016 / ASTM F1929-15:2015Pass
    Shelf-life (Tensile strength of Tyvek®)ASTM F1980-16:2016 / ASTM D5035-11:2019Pass
    Shelf-life (Tensile strength of plastic film)ASTM F1980-16:2016 / ASTM D882Pass
    Shelf-life (Seal strength)ASTM F1980-16:2016 / ASTM F88Pass
    Shelf-life (Tear resistance)ASTM F1980-16:2016 / ASTM D1922-20Pass
    Shelf-life (Microbial Barrier Test)ASTM F1980-16:2016 / DIN 58953-6:2016Pass
    CI of sterilization roll validationANSI/AAMI/ISO 11140-1:2014Pass
    Residual sterilant on Tyvek® validationManufacturer's internal test standardPass
    Sterilization Process Indicator (BT96, CD42, CT40)
    Resistance validation for biological indicator test (BT96)ISO 11138-1:2017Pass
    BI & Test pack validation testManufacturer's internal standardPass
    Chemical indicator validation (CD42, CT40)Manufacturer's internal standardPass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific test or the country of origin/provenance of the data. It refers to various ISO and ASTM standards, which typically specify test methodologies and sample size requirements. The studies are non-clinical, likely laboratory-based evaluations to demonstrate performance against these standards. The submission originates from Plasmapp Co., Ltd. in South Korea, suggesting studies were likely conducted either internally or by contract research organizations, potentially in South Korea, in accordance with international standards. The studies would be prospective in nature, as they are tests performed on the device to prove its compliance with standards and equivalence.

    • Sample Size: Not explicitly stated for individual tests.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The applicant is Plasmapp Co., Ltd., BVC-111, 125, Gwahak-ro, Yuseong-gu, Daejeon, 34141, Rep. of Korea (South Korea).
    • Retrospective or Prospective: Implied to be prospective as the tests were conducted to demonstrate device performance against defined standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This submission is for a sterilization system and associated components, which are evaluated based on defined physical, chemical, and biological performance standards (e.g., sporicidal activity, material compatibility, chemical indicator color change) rather than interpretation of complex medical images or patient data. Therefore, the concept of "experts establishing ground truth" in the diagnostic AI sense (e.g., radiologists for an imaging device) is not directly applicable here. The "ground truth" is established by the well-defined parameters and pass/fail criteria of the referenced ISO and ASTM standards for sterility, material science, and indicator performance.
    The experts involved would be laboratory technicians and engineers performing and verifying the tests, and potentially subject matter experts in sterilization science and engineering within Plasmapp and the regulatory body. Their qualifications are not specified but would be implicit in their roles and adherence to the testing standards.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is a technical performance study against objective standards, not a diagnostic study requiring adjudication of expert interpretations.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a sterilization system and its accessories, not an AI-based diagnostic device intended for use by human readers.


    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical sterilization system, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance evaluation of sterilization devices and accessories typically refers to:

    • Biological inactivation: Complete kill of test microorganisms (e.g., Geobacillus stearothermophilus spores) for sterilizers and biological indicators, confirmed through culture methods.
    • Physical and chemical parameters: Confirmation that specified physical conditions (temperature, pressure, time) were met and chemical indicators changed appropriately, as defined by industry standards and internal specifications.
    • Material integrity: No degradation or adverse effects on device materials after sterilization, as per material compatibility tests.
    • Packaging integrity: Maintenance of sterility barrier, confirmed by physical and microbial barrier tests.

    These 'ground truths' are established through validated laboratory tests against international and national standards (ISO, ASTM, AOAC).


    8. The sample size for the training set

    Not applicable. This is a hardware device (sterilizer, packaging, indicators) and not an AI/machine learning model that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI/machine learning model, there is no training set and therefore no ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223015

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-09

    (222 days)

    Product Code
    Regulation Number
    880.6860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K212200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The STERLINK™ FPS-15s Plus sterilizer is a vaporized hydrogen peroxide sterilizer intended for use in the terminal sterilization of cleaned, rinsed, and dried reusable metal medical devices used in healthcare facilities. A preprogrammed sterilization lumen cycle operates at low pressure and is thus suitable for processing medical devices sensitive to heat and moisture.

    The STERLINK™ FPS-15s Plus can sterilize*:

    • Instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors
    • Medical devices with a single stainless steel lumen with:
      • STERPACK™ plus: An inside diameter of 1.6 mm or larger and a length of 200 mm or shorter
      • STERLOAD™: An inside diameter of 2.4 mm or larger and a length of 280 mm or shorter

    *The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load with a total weight of 1 1b and 3.97 lbs with STERPACK™ plus and STERLOAD™, respectively,

    Device Description

    The STERLINK™ FPS-15s Plus sterilizer with STERPACK™ plus cassette is a low temperature sterilizer which uses the STERLINK™ process to inactivate microorganisms on a broad range of medical devices and instruments.

    This system consists of a main device and cassettes which are called the STERLINK™ FPS-15s Plus and STERPACK™ plus or STERLOAD™, respectively. The STERPACK™ plus and STERLOAD™ cassette contain 58-59.5% (weight concentration) of hydrogen peroxide (H2O2) which is utilized as the sterilant.

    AI/ML Overview

    Based on the provided text, the device in question is the STERLINK™ FPS-15s Plus sterilizer with STERPACK™ plus cassette, a vaporized hydrogen peroxide sterilizer. The acceptance criteria and the study proving the device meets these criteria are primarily focused on the non-clinical performance of the sterilizer, specifically its ability to sterilize medical devices. The information provided is typical of a 510(k) summary for a medical device.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally "Pass" for compliance with various standards and guidance documents. The reported device performance is that it "Passes" these tests.

    Test CategoryStandard/Guidance DocumentAcceptance CriteriaReported Device Performance
    Human factors and usability engineeringGuidance for Industry and Food and Drug Administration Staff, "Appling Human Factors and Usability Engineering to Medical Devices", issued on February 3, 2016; IEC 62366-1:2015PassPass
    Shelf-life test (Sterilant Preservation Test)Manufacturer's internal standardPass1 year
    BiocompatibilityISO 10993-5:2009PassPass
    Software validationIEC 62304:2006 + AMD1:2015; Guidance for Industry and Food and Drug Administration Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device", issued on May 11, 2005PassPass
    Electrical safetyEN 61010-1:2010; EN 61010-2-040:2015; IEC 60601-1:2005 (3rd Ed.) + CORR.1:2006 + CORR.2:2007 + A1:2012; EN 60601-1-2:2015PassPass
    RF disturbanceEN 55011:2009/A1:2010PassPass
    Electromagnetic compatibility (EMC)EN 61326-1:2013; EN 61000-3-2:2014; EN 61000-3-3:2013; EN 60601-1-2:2015PassPass
    Resistance validation for biological indicator testISO 11138-1:2017PassPass
    Lumen sterilizationISO 14937:2009PassPass
    Surface sterilizationISO 14937:2009; ISO 11737-1:2018; ISO 11737-2:2009PassPass
    Mated surface sterilizationISO 14937:2009; ISO 11737-1:2018; ISO 11737-2:2009PassPass
    Simulated use testASTM E1837-96:2014; ISO 11737-1:2018PassPass
    In-use testASTM E1837-96:2014PassPass
    Sporicidal activity testAOAC 966.04 (AOAC 6.3.05:2013)PassPass
    Bacteriostasis testISO 11737-1:2018PassPass
    Material compatibility testASTM D638; ASTM E8/E8M-16ae1; ASTM D790; ASTM E290-14; ASTM D256; ASTM E23-18; ASTM E1164; ASTM E313; ASTM D3985; ASTM F1249PassPass
    Delivery validation (Sterilizer)ASTM D4169-14PassPass
    Delivery validation (Sterilant)Manufacturer's internal standardPassPass
    Hydrogen peroxide gas detection testOSHA analytical method 1019PassPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific "sample size" in terms of number of medical devices or test runs for each individual performance test (e.g., lumen sterilization). Instead, it states for lumen sizes: "The validation testing for all lumen sizes was conducted using a maximum of five (5) lumens per load." It also mentions "a validation load with a total weight of 1 lb and 3.97 lbs with STERPACK™ plus and STERLOAD™, respectively."

    • Sample Size for performance tests: For lumen sterilization, it mentions a maximum of five lumens per load. Other tests don't specify a detailed sample size breakdown within this document. The "Pass" results indicate that the testing met the minimum requirements for the specific standards.
    • Data Provenance: The manufacturer is "Plasmapp Co., Ltd." in "Rep. of Korea (South Korea)." The document implies that the testing was conducted to meet the requirements of the U.S. FDA, but the location where the actual tests were performed is not explicitly stated. The nature of these tests (e.g., sterilization efficacy) strongly suggests they were performed as prospective validation studies to demonstrate the device's performance.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    This type of information (number and qualifications of experts establishing ground truth for a test set) is generally relevant for AI/ML-driven diagnostic devices where human expert consensus forms the ground truth for image interpretation or diagnosis. For a sterilizer, the "ground truth" for its performance is typically established through objective, laboratory-based biological and physical testing using standardized methods and biological indicators (e.g., Geobacillus stearothermophilus). Therefore, the concept of "experts establishing ground truth for a test set" in the context of human interpretation of data is not applicable here. The "experts" would be the scientists and technicians conducting the validation tests according to the specified standards.

    4. Adjudication Method for the Test Set

    As the "ground truth" for a sterilizer's performance is objective (e.g., kill/no kill of microorganisms, meeting sensor specifications), adjudication methods like 2+1 or 3+1 (common in image interpretation studies) are not applicable. The device's performance is measured against established scientific and engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC study was not done. MRMC studies are typically performed for diagnostic devices (especially those involving image interpretation) to evaluate the impact of a device (e.g., AI assistance) on human reader performance. This document is for a medical device sterilizer, which does not involve human interpretation of cases or AI assistance in that context.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the performance tests described are standalone in the sense that they evaluate the sterilizer's ability to sterilize medical devices independently of human subjective assessment. The "algorithm" here is the sterilization cycle itself, and its effectiveness is measured directly through biological and chemical indicators. The results are based on the device's inherent design and operational parameters.

    7. The Type of Ground Truth Used

    The ground truth used for proving the sterilizer's efficacy is based on:

    • Biological Activity: The ability to inactivate microorganisms (e.g., Geobacillus stearothermophilus spores) as measured by standard biological indicator tests (e.g., ISO 11138-1:2017, ISO 11737-1:2018, AOAC 966.04).
    • Physical Parameters: Confirmation that the sterilizer operates within specified parameters (temperature, pressure, time) and that the sterilant is delivered effectively.
    • Material Compatibility: Ensuring that the sterilization process does not degrade the materials of the medical devices being sterilized.
    • Engineering Standards: Compliance with electrical safety, EMC, software validation, and other relevant engineering standards (e.g., IEC, EN, ASTM).

    8. The Sample Size for the Training Set

    This document does not refer to a "training set." The concept of training sets (and validation/test sets derived from them) is specific to machine learning or AI models that learn from data. This document is about a hardware device (sterilizer) whose performance is validated through defined physical and biological tests, not through training a model on a dataset.

    9. How the Ground Truth for the Training Set was Established

    As there is no "training set" in the context of an AI/ML model for this device, this question is not applicable. The sterilizer's performance is established through controlled, laboratory-based validation studies against regulatory and industry standards for sterilization efficacy.