Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K240070

    Validate with FDA (Live)

    Device Name
    Profound Matrix
    Date Cleared
    2024-07-01

    (174 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K211217

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Profound Matrix System is intended for dermatological procedures, as follows:

    The Matrix Pro applicator is indicated for general dermatological procedures for electrocoagulation and hemostasis. The Matrix Pro applicator is indicated for the percutaneous treatment of facial wrinkles in patients with Fitzpatrick Skin Types I-IV.

    The Sublative RF applicator is indicated for dermatological procedures requiring ablation and resurfacing of the skin, and for the treatment of facial wrinkles. At higher energy levels greater than 62mJ/pin, the Sublative RF applicator is limited to Fitzpatrick Skin Types I-IV.

    The Sublime applicator is indicated for non-invasive wrinkle treatment. At higher energy levels, greater than 100 J/cm³, the Sublime applicator is limited to Fitzpatrick Skin Types I-IV.

    Device Description

    Profound Matrix is a system that combines three existing technologies and applications for dermatological procedures. Each respective technology is used in one of the three applicators designed for the Profound Matrix System. These three applicators are named: Matrix Pro, Sublative RF and Sublime.

    The system console consists of a CPU, power supply, RF pulse generator, I/O board, splitter/distribution board and footswitch board. When the system is powered up, the user can select between user mode and service/maintenance mode. After selecting user mode and entering the required password, the user is requested to choose between the Matrix Pro, the Sublative RF or the Sublime applicators. Only one applicator can be used to treat a patient at any given time.

    The user interfaces with the Profound Matrix System and applicators' settings through the 15″ touch screen. Each of the applicators have a finger trigger that can be used to activate the applicator or alternatively the end user can use the footswitch to activate the applicator.

    Matrix Pro: The Matrix Pro applicator is designed to deliver bipolar, non-ablative radiofrequency energy through a 7x7 matrix/grid of sterile microneedles to the skin in a fractional manner. The microneedles can be set up to pulse up to a maximum of three depths during each insertion of the matrix/grid of microneedles with a maximum depth of up to 3.5mm. The Matrix Pro applicator delivers the right amount of energy at the precise depth, every time. The technology of the Matrix Pro applicator combines adjustable depth, energy, and real time impedance monitoring for customizable and consistent results. The Profound Matrix system monitors tissue impedance across the 3 modalities, in real-time, while simultaneously adjusting RF energy pulse duration and power output. The result is an accurate dose of RF energy delivery with every treatment application for consistent treatment outcomes, independent of patient-to-patient, body location, skin hydration, and procedural-related impedance variations.

    Sublative RF: The Sublative RF applicator delivers bipolar radiofrequency (RF) to the skin surface in a fractional manner via an array of multi-electrode pin tips, which results in heating of both the demarcated spots to temperatures that ablate and resurface at contact points of the multi-electrode pins while also delivering heat in a wide diffuse manner to the dermis.

    Sublime: The Sublime applicator uses a combination of an infrared (IR) light source and bipolar radiofrequency (RF) to bulk heat the dermis and affect dermal collagen in order to treat wrinkles. The bulk heating of the dermal layers refers to the gradual and gentle accumulation of heat with each additional pass performed.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the Profound Matrix System with the Matrix Pro Applicator.

    Context: The Candela Corporation sought clearance for a new indication for their Profound Matrix System (specifically the Matrix Pro applicator) – the percutaneous treatment of facial wrinkles in patients with Fitzpatrick Skin Types I-IV. This new indication builds upon existing clearances for general dermatological procedures and electrocoagulation/hemostasis for the Matrix Pro, and similar wrinkle treatment indications for other applicators within the Profound Matrix system (Sublative RF and Sublime) and a predicate device (Lutronic Infini).


    1. Table of Acceptance Criteria and Reported Device Performance

    The core acceptance criterion for the new wrinkle treatment indication was the successful identification of post-treatment photographs by a majority of blinded evaluators, indicating an improvement in wrinkle appearance.

    Acceptance CriterionReported Device Performance
    At least 70% of subjects were to have their post-treatment photographs correctly identified by at least 2 of 3 blinded reviewers.75% of subjects (24 of 32) had their post-treatment photographs correctly identified by at least 2 of 3 blinded reviewers.

    Additional Efficacy & Safety Metrics (Supporting Data, not explicit acceptance criteria but demonstrating effectiveness and favorable risk-benefit profile):

    Supporting MetricReported Device Performance
    Proportion of subjects identified as responders (at least 1 point improvement on FWES by $\ge$ 2 of 3 blinded reviewers post-hoc)62.5% (20 of 32) of subjects were responders.
    Statistical significance of improvement in combined blinded evaluator FWES ratings (mean change from baseline to 3 Months Follow-Up)Statistically significant (p<0.005).
    Subject-perceived improvement in overall facial appearance at 1-Month Follow-Up (1MFU)91% reported 'improved' facial appearance; 41% reported 'much improved' or 'very much improved'.
    Subject-perceived improvement in overall facial appearance at 3-Month Follow-Up (3MFU)94% reported 'improved' facial appearance; 53% reported 'much improved' or 'very much improved'.
    Mean pain score during treatment4.7 ($\pm$ 1.67) on an 11-point Numerical Rating Scale (0=no pain, 10=worst pain).
    Adverse EventsFour possibly related AEs, all mild, resolved, and did not lead to withdrawal. No serious or unanticipated AEs.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): Data from 32 healthy subjects.
    • Data Provenance: Retrospective analysis of a source study (FUFT2002) which was a multi-site prospective clinical trial sponsored by Candela Medical. The document does not specify the country of origin of the data, but it's likely from the US given the FDA submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three (3) blinded evaluators were used.
    • Qualifications: The document states they were "blinded evaluators" or "blinded reviewers." It does not explicitly mention their specific professional qualifications (e.g., dermatologists, plastic surgeons, radiologists) or years of experience. However, given the nature of wrinkle assessment and clinical trials for aesthetic devices, it's highly probable they were medical professionals with expertise in dermatology or aesthetics.

    4. Adjudication Method for the Test Set

    • Adjudication Method: A majority consensus approach was used. Treatment was considered a success if at least 2 of 3 blinded reviewers correctly identified the post-treatment photograph (for the primary efficacy endpoint).
    • For the post-hoc assessment of individual subject responders, a participant was a responder if their "after photos were rated at least 1 point improved on the validated Fitzpatrick Wrinkle and Elastosis Scale (FWES) by at least two of three blinded reviewers."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was not done.
      • This study was a clinical trial evaluating the device's effectiveness in treating wrinkles, not an AI-assisted diagnostic device performance study. The "blinded evaluators" were assessing the clinical outcome on the subjects, not providing diagnoses based on AI output.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is a physical electrosurgical system that delivers RF energy for treatment, not an AI algorithm for diagnostic purposes. Therefore, there's no "standalone algorithm performance" in this context. The study evaluates the direct clinical outcome of using the device on patients.

    7. The Type of Ground Truth Used

    • Expert Consensus (Blinded Evaluators): The primary efficacy endpoint and supporting assessments relied on the consensus ratings of independent, blinded expert evaluators who compared photographs of subjects before and after treatment.
    • Validated Scale: The Fitzpatrick Wrinkle and Elastosis Scale (FWES) was used by the blinded evaluators, indicating a structured and standardized method for assessing wrinkle improvement.
    • Patient Reported Outcomes (PROs): The Global Aesthetic Improvement Scale (GAIS) was used to capture subjects' perceived improvement, which serves as an additional form of outcome data.

    8. The Sample Size for the Training Set

    • The document does not mention a training set in the context of an AI/algorithm. This device is an energy-based medical device, not an AI-powered diagnostic tool. The "training" for such a device would refer to the development and engineering process, not a data-driven machine learning training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there was no AI/algorithm training set. The "ground truth" in this context pertains to the clinical evaluation of the device's effect on subjects, as assessed by experts and subject self-reporting for the efficacy study.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1