Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220771

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-07

    (113 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K203641, K201827

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stratasys TrueDent™ is a light-curable resin indicated for the fabrication of dental appliances including removable full and partial dentures, denture bases, denture teeth, bridges, crowns, inlays, and veneers in dental laboratories. The material is an alternative to traditional heat-curable and auto polymerizing resins. Stratasys TrueDent™ is intended exclusively for professional dental work.

    Device Description

    Stratasys TrueDent™ is a light-curable methacrylate-based resin, that enables fabrication of monolithic, high quality, functional, and aesthetic dental appliances, including full and partial removable dentures, denture bases, denture teeth, bridges, crowns, inlays, onlays and veneers in a CAD/CAM additive manufacturing process. This material is intended to be used by trained dental professionals for the fabrication of various dental appliances in a CAD/CAM additive manufacturing process that includes the following components: digital dental files based on a digital impression, a Stratasys Poly Jet 3D printer, and curing light equipment. It is an alternative to traditional heat cured and auto polymerization resins. The predesigned color of the dental appliance is achieved by digitally mixing colored resins. The resins are available in five (5) different base colors (clear, white, cyan, magenta, and yellow). Stratasys TrueDent can be used only with Stratasys J5 Series printers. Stratasys J5 Series printers include a 3D printing system that utilizes PolyJet technology, with drop-on-demand inkjet printing process, 3 print heads, and a UV LED (395nm) curing process. The dental appliance is then cured in the Stratasys TrueDent™ Cure curing chamber and sent back to the dentist for try-in and final adjustment.

    AI/ML Overview

    The provided text describes the regulatory clearance for the Stratasys TrueDent™ device and includes a comparison to predicate and reference devices. It details non-clinical performance data but does not include information about clinical studies with human subjects, AI assistance, or expert adjudication. Therefore, many of the requested items related to clinical studies and AI are not available in the given document.

    Here's the information that can be extracted or inferred from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by adherence to international standards for dental materials and biocompatibility. The reported device performance indicates that the Stratasys TrueDent™ meets these requirements.

    Acceptance Criteria (Standard)Reported Device Performance (Stratasys TrueDent™)
    ISO 10477 (Flexural Strength, Sorption, Solubility)Meets requirements; similar to predicate/reference devices
    ISO 4049 (Flexural Strength, Flexural Modulus, Sorption, Solubility)Meets requirements; similar to predicate/reference devices
    ISO 20795 (Flexural Strength, Flexural Modulus, Sorption, Solubility)Meets requirements; similar to predicate/reference devices
    ISO 7405 (Biocompatibility Categorization)Surface device with permanent contact; successfully evaluated
    ISO 10993 (Biocompatibility: Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Genotoxicity)Deemed biocompatible; meets requirements; similar to predicate/reference devices
    Shelf Life12 months (validated); 24 months (on-going validation)
    FDA's "Technical Considerations for Additive Manufactured Medical Devices" GuidanceEvaluation of relevant properties of printed resin on permitted machines; tests based on orientation during manufacturing were performed.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for the non-clinical performance and biocompatibility tests. The data provenance is "Non-clinical testing data submitted to demonstrate substantial equivalence." No country of origin for the data is specified, but the manufacturer is Stratasys Ltd. based in Israel. The tests are non-clinical (bench testing), so the terms "retrospective or prospective" do not directly apply in the same way as clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in a clinical context. The provided study is non-clinical (bench testing and biocompatibility assessments). The "ground truth" for these types of tests is the established scientific and regulatory standards (e.g., ISO standards, FDA guidance).

    4. Adjudication method for the test set

    Not applicable, as this was a non-clinical study focused on meeting specified standards. There was no mention of human-in-the-loop performance or expert adjudication of outputs.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was mentioned. The device is a dental material for fabrication of appliances, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This would not be applicable in the context of this device. The device itself (a resin) is the "standalone" component being evaluated for its physical, mechanical, and biocompatibility properties. The manufacturing process involves CAD/CAM, but the focus of the performance data is on the cured material, not an algorithm's diagnostic performance.

    7. The type of ground truth used

    The ground truth for the non-clinical performance data (mechanical and physical properties) and biocompatibility data is defined by established international standards (ISO 10477, ISO 4049, ISO 20795 for mechanical properties; ISO 7405, ISO 10993 for biocompatibility). These standards specify the test methodologies and acceptance criteria.

    8. The sample size for the training set

    This information is not applicable. The device is a physical material (resin), not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as the device is not an AI model requiring a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1