510(k) Data Aggregation

    K Number
    K251012

    Validate with FDA (Live)

    Date Cleared
    2025-08-14

    (134 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K221775, K203271, K203214

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TRUDERMAL Pro (Model: ZLD-390) is an over-the-counter(OTC) device that delivering targeted phototherapy using three wavelengths(blue:415nm, red:630nm, Infrared(IR):850nm):

    Red light: Reduces full-face wrinkles.
    Blue light: Treats mild to moderate inflammatory acne.
    IR light: Provides topical heating for the purpose of elevating tissue temperature, temporarily relieving minor muscle and joint pain, arthritis and muscle spasms; relieving stiffness; promoting relaxation of muscle tissues and temporarily increasing local blood circulation.

    Mixed lights:
    Red+IR light: Reduces full-face wrinkles.
    Blue+Red light:Treats mild to moderate inflammatory acne.

    Device Description

    The TRUDERMAL Pro is an over-the-counter LED light therapy medical device. The device emits specific wavelengths of low level, narrow band light for the treatment of certain dermatological and musculoskeletal indications. The wavelengths used in the TRUDERMAL Pro system are Blue 415nm, Red 630nm and Near Infrared 850nm.

    The device consists of a TRUDERMAL Pro (with LED flexible silicone panel, Controller and AC/DC adapter), power cable, and 2 base seats, a pair of eye protectors. The TRUDERMAL Pro enables treatment of the face and the body via a LED flexible silicone panel.The system is operated by a controller and treatments can be performed by selecting different wavelengths. The light is generated by Light Emitting Diodes(LED's) that are contained within the LED flexible silicone panel. The LED panel can be placed into the Base, holding in the appropriate position to administer a facial treatment. For use on the body,the LED panel is removed from the Base unit and placed over the appropriate body part.

    The device is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The device does not require checks on the light output as the LEDs do not dim with age to any practical extent.

    AI/ML Overview

    I apologize, but I cannot fulfill your request to create a table of acceptance criteria and reported device performance, nor can I describe a study that proves the device meets specific acceptance criteria based solely on the provided FDA 510(k) Clearance Letter for TRUDERMAL Pro (ZLD-390).

    Here's why:

    • No Performance Data or Acceptance Criteria in the Document: The provided 510(k) clearance letter and its associated summary (510(k) Summary) do not contain any information about specific performance acceptance criteria or the results of clinical performance studies that demonstrate the device meets such criteria.
      • The document states, "Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively." This explicitly indicates that no clinical performance data (which would typically include performance against acceptance criteria) was provided or evaluated for this 510(k) submission.
      • The non-clinical tests mentioned (electrical safety, EMC, biocompatibility, software V&V) are regulatory compliance tests, not performance studies assessing the therapeutic efficacy against specific clinical endpoints or acceptance criteria.

    Therefore, I cannot extract or infer the following information from the provided text:

    1. Table of acceptance criteria and reported device performance: This data is not present. The document explicitly states clinical testing was not needed.
    2. Sample size used for the test set and data provenance: Since no clinical performance study was conducted or reported, there is no test set, sample size, or data provenance information.
    3. Number of experts used to establish ground truth & qualifications: Not applicable, as no clinical performance study data relying on expert ground truth is provided.
    4. Adjudication method for the test set: Not applicable.
    5. MRMC comparative effectiveness study: Not conducted or reported. The device is a phototherapy device, not an AI-powered diagnostic tool, so an MRMC study is generally not relevant.
    6. Standalone (algorithm only) performance study: Not applicable. This is a physical phototherapy device, not an algorithm.
    7. Type of ground truth used: Not applicable.
    8. Sample size for the training set: Not applicable, as this is a phototherapy device, not an AI model that would have a "training set" in the traditional sense.
    9. How ground truth for the training set was established: Not applicable.

    What the document does provide regarding "performance" is:

    • Comparison to Predicate Devices: The submission argues for substantial equivalence by comparing the TRUDERMAL Pro's technical characteristics (wavelengths, irradiance, treatment time) to legally marketed predicate devices. The claim is that "The subject device's wavelengths, treatment time, irradiance and radiation dose are similar and close to those of the predicate devices," and this, combined with non-clinical tests, "supports that the subject device is safe and as effective as the predicate devices for its intended use."
    • Non-Clinical Test Summary: This section lists the electrical safety, electromagnetic compatibility, biocompatibility, and software verification/validation tests performed, demonstrating compliance with relevant industry standards (IEC, ISO). These tests ensure the device is safe and functions as designed electrically and mechanically, but they do not prove clinical efficacy against specific performance metrics related to wrinkle reduction, acne treatment, or pain relief.

    In summary, the provided FDA document is a clearance letter and 510(k) summary that emphasizes substantial equivalence to existing predicate devices based on design, intended use, and non-clinical safety testing, rather than presenting a performance study with defined acceptance criteria and clinical results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230351

    Validate with FDA (Live)

    Device Name
    LED Facial Mask
    Date Cleared
    2023-05-05

    (85 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K203271, K190443

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LED Facial Mask is an over the counter device that is intended to use LED light for the treatment of wrinkles and mild to moderate acne.

    Device Description

    The LED Facial Mask is an over-the-counter, facemask-shaped design light emitting diode (LED) device that emits light energy in the red, blue, and amber spectrum for the treatment of wrinkles or mild to moderate inflammatory acne on the face. The device consists of a mask body unit, controller, and AC adapter. The mask body unit contains light emitting diodes (LEDs) that emit visible red light (650nm+/-5nm), amber light (605nm+/-5nm), or blue light (465nm+/-5nm) to help reduce the appearance of wrinkles or mild to moderate acne. A controller is connected to the mask body unit to control the device, such as turn on/off the device, switch LED color output. An AC adapter is used to power the device. To use the device, user should place the mask body unit over the face and use the controller to operate. The device will automatically turn off after each treatment.

    The LED Facial Mask includes MZ-01, NEWKEY-01, and SP-FM-01 models. Their intended use, performance, structure design, and operation are identical, with the difference being the product appearance, but these differences do not affect or change the intended use of the device.

    AI/ML Overview

    The provided text is a 510(k) summary for an LED Facial Mask, a medical device for over-the-counter use in treating wrinkles and mild to moderate acne. However, it does not contain information about studies that prove the device meets acceptance criteria related to its clinical efficacy.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through technical specifications, safety data (biocompatibility, electrical safety, eye safety), and software verification and validation. There's no mention of clinical trials or performance studies evaluating the device's effectiveness in treating wrinkles or acne in accordance with specific acceptance criteria beyond general safety and design similarity to other cleared devices.

    Therefore, I cannot provide the requested information regarding acceptance criteria and the study proving the device meets them for its intended clinical effects because such information is not present in the provided text. The document primarily details technical and safety compliance, not clinical performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230597

    Validate with FDA (Live)

    Date Cleared
    2023-04-28

    (56 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K203271, K202390

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of Ila - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.

    Device Description

    The Aduro Comb is (Model: SZ-22A) a comb-shaped device that emits led light with the intention to promote hair growth. The device provides distributed led to the scalp at 650nm (±10nm) while the comb teeth simultaneously part the user's hair to ensure the user's scalp. The device is designed as a handheld product, and it consists of the main unit, the charging dock and a power cable, as well as it is powered by the built-in rechargeable lithium battery. The device has only one key for switching on and off the device and it will automatically shut down after a 10-minute treatment is completed.

    AI/ML Overview

    The Aduro Comb (Model: SZ-22A) is indicated to treat Androgenetic Alopecia and promote hair growth. The following information outlines its acceptance criteria and the studies performed.

    1. Table of Acceptance Criteria and Reported Device Performance

    CriteriaAcceptanceReported Performance
    Electrical SafetyCompliance with AAMI/ANSI ES60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2Complies with all listed standards
    Photobiological SafetyCompliance with IEC 62471Complies with IEC 62471
    Battery SafetyCompliance with IEC 62133-2Complies with IEC 62133-2
    Biocompatibility (for patient-contacting materials)Compliance with ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-10 (Irritation)Complies with all listed ISO 10993 standards, based on equivalence to previously cleared devices.
    Software Verification and ValidationCompliance with IEC 62304 and FDA Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"Documentation provided and considered sufficient for a "moderate" level concern software.

    2. Sample size used for the test set and the data provenance

    The document explicitly states, "Clinical testing was not needed for this 510(k)." Therefore, there was no clinical test set in the traditional sense, and thus no sample size or data provenance from a clinical study for the device's efficacy. The performance evaluation relied solely on non-clinical testing and comparison to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set was used to assess the device's efficacy or establish ground truth from experts. The evaluation focused on non-clinical safety and performance.

    4. Adjudication method for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an infrared lamp for hair growth, not an AI-assisted diagnostic or imaging device that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a physical infrared light therapy device, not an algorithm.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" was compliance with established international and national standards for electrical safety, photobiological safety, battery safety, and biocompatibility.

    8. The sample size for the training set

    Not applicable, as this device does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable, as there was no training set.