510(k) Data Aggregation

    K Number
    K210711

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-22

    (287 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K200805, K191363, K191605, K173402, K182776, K202100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous adult patients who qualify for dental implants.

    When Yomi Plan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. Yomi Plan provides pre-operative planning for dental implantation procedures. The output of Yomi Plan is to be used with the Neocis Guidance System (NGS).

    Device Description

    The purpose of subject device is for modification of the Neocis Guidance System (K202264) to allow for Wi-Fi to be continuously active in the Yomi Plan v2.0.1 while it is powered on. All other software and hardware features/functions remain identical to the predicate. In terms of FDA regulations, the Neocis Guidance System (NGS) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120).

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    Major components include: Monitor, Planning Station Laptop PC, Lift Column, Base Cart, Robotic Guide Arm, and Patient Tracker.

    The system allows the user to plan the surgery virtually in Yomi Plan (K191363-cleared for use alone on third party PCs for preplanning). The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in our planning software. The plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The NGS robotic arm holds and guides a standard FDAcleared third party powered bone cutting instrument (K191605).

    The patient tracking portion of the NGS is comprised of linkages from the patient to the NGS, which include the Chairside Patient Splint (CPS) (K173402) or Edentulous Patient Splint (EPS) (K200805), the End Effector (EE) and the Patient Tracker (PT).

    AI/ML Overview

    The request is to describe the acceptance criteria and the study that proves the device meets the acceptance criteria for the Neocis Guidance System (NGS) with Yomi Plan v2.0.1.

    Based on the provided text, the submission does not contain a study proving that the device meets specific performance acceptance criteria for its clinical function. Instead, it modifies an existing, previously cleared device (NGS with Yomi Plan v2.0) to allow continuous Wi-Fi activity and describes the testing conducted to ensure this modification maintains safety and effectiveness, primarily through software verification, wireless coexistence testing, and EMC testing.

    Therefore, the following information is extracted and presented in relation to the modification and associated testing, rather than a clinical performance study with acceptance criteria for the device's primary function of dental implantation guidance.


    Acceptance Criteria and Study Proving Device Meets Criteria

    The Neocis Guidance System (NGS) with Yomi Plan v2.0.1 is a modification of a previously cleared device (Neocis Guidance System (NGS) with Yomi Plan v2.0, K202264). The primary change in this submission is to allow Wi-Fi to be continuously active in the Yomi Plan v2.0.1 while it is powered on. Therefore, the "acceptance criteria" and "study" described below relate to the safety and effectiveness of this change and the overall system's compliance with relevant standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Category / TestAcceptance CriteriaReported Device Performance
    Software VerificationCompliance with relevant software and risk management standards.Fully executed according to: - ANSI AAMI ISO 14971: 2019 (Risk Management) - ANSI AAMI IEC 62304:2006/A1:2016 (Software Life Cycle Processes) - FDA Guidance for Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) - FDA Guidance for Cybersecurity in Medical Devices (Oct 2, 2014; Dec 28, 2016) - Cybersecurity for Networked Medical Devices (Jan 14, 2005) - AAMI TIR57: 2016 (Medical Device Security)
    Wireless CoexistenceSafe and effective operation with continuous Wi-Fi in the presence of other radio-frequency devices.Tested according to: - AAMI TIR69: 2017 (Risk management of radio-frequency wireless coexistence for medical devices and systems) - IEEE ANSI C63.27-2017 (American National Standard for Evaluation of Wireless Coexistence)
    Electromagnetic Compatibility (EMC)Compliance with EMC standards for medical electrical equipment.Tested according to: - IEC 60601-1-2 Edition 4.1 2020-09 (Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests)
    Clinical Performance (Primary Function)Not explicitly detailed in this 510(k) summary for the subject device. This submission is for a modification to a previously cleared device, assuming its fundamental clinical performance is already established."Our performance testing demonstrates substantially equivalent performance of the subject device as compared to the predicate." (This statement refers to equivalence, not necessarily a new clinical performance study for the modified device).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify sample sizes for test sets related to clinical performance. The testing described relates to technical compliance (software, wireless, EMC).

    • Data Provenance: Not applicable in the context of clinical data for this specific 510(k) submission, as it doesn't describe a clinical study for performance. The testing pertains to engineering and regulatory standards for the device modification.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not applicable. This submission focuses on engineering testing (software, wireless, EMC) rather than a clinical study requiring expert-established ground truth for device performance validation in a patient setting.

    4. Adjudication Method for the Test Set

    • Not applicable. This submission focuses on engineering testing, not a study requiring adjudication of clinical outcomes or interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done as described in the provided text. The device is a computerized navigational system for dental implantation, not an AI-assisted diagnostic or interpretation tool for human readers. This submission focuses on a software modification (continuous Wi-Fi) and its impact on technical safety and performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not explicitly detailed. The described testing pertains to the integrated system's technical compliance and software behavior (Yomi Plan v2.0.1) in the context of continuous Wi-Fi. While individual software components would undergo standalone verification, the submission does not describe a standalone clinical performance study. The device is described as providing "navigational guidance of the surgical instruments" and "haptic feedback to the surgeon by constraining the motion of the bone cutting instrument to the plan," implying human-in-the-loop operation.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Not applicable. For the technical testing (software verification, wireless coexistence, EMC), the "ground truth" is adherence to established engineering standards, cybersecurity protocols, and functional specifications, rather than clinical ground truth (e.g., pathology, expert consensus on images).

    8. The Sample Size for the Training Set

    • Not applicable. The submission does not describe a machine learning or AI model that requires a training set. The software is a planning and guidance system not explicitly described as employing AI in a way that requires a training set for model development.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As no training set is described for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K211129

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-28

    (103 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K202348, K202100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments.

    Device Description

    In terms of FDA regulations, the Neocis Guidance System (NGS) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120). The NGS is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. The system allows the user to plan the surgery virtually in Yomi Plan. The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in our planning software. The plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The NGS robotic arm holds and guides a standard FDA-cleared third party powered bone cutting instrument. The patient tracking portion of the NGS is comprised of linkages from the patient to the NGS, which include the Chairside Patient Splint (CPS) or Edentulous Patient Splint (EPS), the End Effector (EE) and the Patient Tracker (PT). A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the CPS prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the CPS on the patient, which relays information to the NGS in order to track patient movement. The subject of this submission is adding Intraoral Fiducial Array (IOFA) as a new device accessory based on a previously cleared design. The IOFA is intended for use in clinical sites with reduced scan volumes. The IOFA is designed to reside with all the fiducial beads within the patient's mouth. The IOFA can only be used with the C-CPS.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Neocis Guidance System (NGS) with Intraoral Fiducial Array (IOFA), based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Technological Characteristics)Reported Device Performance (NGS with IOFA Subject Device)
    System Lateral AccuracyRMS < 1 mm
    System Depth AccuracyRMS < 1 mm
    System Angular AccuracyRMS < 6.0°
    CT Scan Quality Requirements0.3 mm Voxel, 0.3 mm Slice Thickness, Matrix 512 x 512, Full 13 cm 21 sec, Multi 2 DICOM format.
    F/T Sensor Force Measurement Range+/- 30 N
    F/T Sensor Torque Measurement Range+/- 2 Nm
    F/T Sensor Single Axis Force Overload Limit200 N
    F/T Sensor Single Axis Torque Overload Limit20 Nm
    Upper limit specification for Guidance Arm Translation Speed1.25 m/s
    Storage RequirementsStore powered at Room Temperature (68°F to 76°F or 20°C to 24.4°C) and standard ambient humidity (5% to 95%) in a dust free, clean environment.
    BiocompatibilityYes (ISO 10993-1, -5, 10, -12)
    SterilizationSteam (ISO 17665-1)
    Software Level of ConcernModerate

    Notes on Acceptance Criteria: The document primarily uses "Same as the subject device" or similar phrasing for the predicate/reference devices, indicating that these performance metrics are consistent across the devices and establish the acceptance criteria for the new IOFA.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not explicitly state the sample size for the test set for all verification activities. However, it mentions an "IOFA End User Validation" which is described as a "nonclinical surrogate that simulates the process of applying, qualitatively evaluating rigidity, and removing a Intraoral Fiducial Array directly to a patient." The specifics of this "validation" and its sample size are not detailed.

    The data provenance (country of origin, retrospective/prospective) is not mentioned in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    For the "IOFA End User Validation," it states the validation was "as performed by a surgeon (end user)." The specific number of surgeons or their qualifications are not specified. For other performance tests like accuracy, the ground truth is likely established through a combination of manufacturing specifications, calibrated measurement tools, and engineering standards, rather than expert consensus on a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    The document does not describe any adjudication method for establishing ground truth from multiple experts. Given that most performance metrics are objective measurements (e.g., RMS accuracy), formal adjudication methods typically used for subjective clinical assessments are unlikely to be applied.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not mentioned or performed. The device is a "computerized navigational system" providing guidance, not an AI for image interpretation or diagnosis that would typically involve a multi-reader study. The text explicitly states, "Animal or clinical testing was not conducted for the subject device."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device (NGS with IOFA) is described as a "computerized navigational system intended to provide assistance" and "provides navigational guidance of the surgical instruments." It also emphasizes that the "surgeon may modify the plan intraoperatively...and is always in control of the surgical instrument." This indicates a human-in-the-loop system. Therefore, a standalone (algorithm only) performance assessment as would be done for an AI diagnostic algorithm is not applicable in the same way. The performance metrics focus on the accuracy of the system's guidance (lateral, depth, angular accuracy), which are inherently tied to the interaction with the human user and surgical instruments.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for the verification tests (Registration Testing, Kinematic Mount Repeatability, Scan Artifact Testing, Fit Clearance Simulation, Total System Accuracy, IOFA Deflection Test) appears to be based on:

    • Engineering measurements and calibrated standards: For system accuracy (lateral, depth, angular), sensor ranges, and physical properties.
    • Established industry standards: For electrical safety, electromagnetic disturbances, ingress protection, biocompatibility (ISO standards), and sterilization (ISO standards).
    • Design specifications: For CT scan quality requirements, guidance arm speed, and storage requirements.

    For the "IOFA End User Validation," the ground truth is implied to be a qualitative assessment of rigidity and usability by a surgeon, simulating real-world application.

    8. The sample size for the training set:

    The document does not mention a training set size. The device is a "computerized navigational system" and its software (Yomi Plan v2.0) was cleared previously. This type of device relies on engineering design, calibration, and deterministic algorithms for guidance, rather than a machine learning model that requires a "training set" in the conventional AI sense.

    9. How the ground truth for the training set was established:

    As no training set is described for an AI/machine learning model, the process for establishing ground truth for a training set is not applicable or mentioned. The system's functionality is verified against established engineering and safety standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203401

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-15

    (26 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K161399, K173402, K202100, K182776, K190059, K014263

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The NGS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    Device Description

    The Neocis Guidance System (NGS) (K161399) is a dental stereotaxic instrument (Product Code PLV) and a powered surgical device for bone cutting (21 CFR 872.4120). The Neocis Guidance System (NGS) is a computerized navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides precise and accurate navigational guidance of surgical instruments, with regard to planning in dental implantation procedures. The system allows the user to plan the surgery virtually in software using a cone beam computed tomography (CBCT) scan of the patient, and the plan is used by a guidance system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The holds and guides a standard FDA-cleared powered bone cutting instrument.

    The implant process occurs in two phases. First, the dental surgeon plans the surgical procedure with the planning software. A virtual implant is placed at the desired location in the CT scan, allowing the dental surgeon to avoid interfering with critical anatomical structures during implant surgery. Second, when the implant plan is optimally positioned, the NGS provides accurate guidance of the dental surgical instruments according to the pre-operative plan. The NGS provides haptic feedback to the surgeon by constraining the motion of the bone cutting instrument to the plan. This allows the surgeon to feel resistance to attempts at motions that may deviate from the plan.

    The patient tracking portion of the NGS is comprised of linkages from the NGS, which for partially edentulous patients include the Chairside Patient Splint (CPS) (K173402) or the Clamped Chairside Patient Splint (CCPS) (K202100), the End Effector (EE) and the Patient Tracker (PT). The CPS or CCPS is attached to the contralateral side of the patient's mouth over stable teeth. The CPS is placed on the patient using on-label dental materials (K182776) prior to the presurgical CBCT scan. A Fiducial Array (FA) with radio-opaque fiducial markers is placed on the splint prior to the CBCT scan so the virtual plan can be related to the physical space of the system using the markers. The PT is an electromechanical feedback system that is connected to the splint on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle, position, and depth to accommodate the patient movement, which will maintain the accuracy of the osteotomy.

    The subject of this submission is a design change to the sleeves in our Edentulous Patient Splint (EPS) (K200805). The EPS enables use of the NGS in fully edentulous patients. It is affixed to the anterior mandible or maxilla using standard bone screws. Like the CPS and CCPS, the EPS serves as rigid connection to the patient for robotic tracking of the patient during the procedure. The EPS is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants.

    AI/ML Overview

    The provided text discusses the Neocis Guidance System (NGS) with Edentulous Patient Splint (EPS) and a design change to its sleeves. However, it does not contain a detailed study proving the device meets acceptance criteria for performance, especially not in the context of diagnostic accuracy (e.g., sensitivity, specificity, AUC).

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (Neocis Guidance System (NGS) with Patient Splints, K200805) after a design change to the EPS sleeves. The "Performance Testing" section lists various tests conducted, primarily related to the physical and biological aspects of the device, rather than a clinical performance study.

    Therefore, many of the requested elements (like sample size for test/training sets, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, type of ground truth for training) are not available in the provided text for a clinical performance study.

    Here's what can be extracted and inferred from the text, focusing on the design change and the tests mentioned:


    Acceptance Criteria and Device Performance (Design Change Validation)

    The document describes a design change to the sleeves within the Edentulous Patient Splint (EPS) component of the Neocis Guidance System (NGS). The acceptance criteria are implicitly related to ensuring this design change does not negatively impact the safety and effectiveness of the device, and that it remains substantially equivalent to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Test/RequirementReported Device Performance/Conclusion
    Mechanical Performance (Splint)EPS Weighted Deflection Test with Optical Tracking in Sawbones®Implied to have met predefined limits for deflection, ensuring stability and accuracy. The document states "The new design is functionally the same as the predicate device."
    System AccuracyTotal System AccuracyImplied to have met accuracy requirements (likely related to navigational guidance precision), ensuring the design change does not degrade the overall system's ability to guide surgical instruments accurately.
    Risk ManagementANSI AAMI ISO 14971:2019 Medical devices Applications of risk management to medical devicesRisk analysis performed and controls implemented to mitigate risks associated with the design change. Concludes no new questions of safety or effectiveness.
    SterilizationANSI AAMI ISO 17665-1:2006/(R)2013 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical deviceSterilization process validated for the new material/design.
    BiocompatibilityANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management processBiological evaluation indicating the new materials (Neocis design titanium for sleeves, removal of stainless steel from patient-contacting materials) are safe for patient contact.
    ANSI AAMI ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityNo cytotoxicity detected.
    ANSI AAMI ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationNo irritation or skin sensitization detected.
    ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityNo systemic toxicity detected.
    ANSI AAMI ISO 10993-12: 2012 Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference MaterialsSample preparation and reference materials used were appropriate for the biocompatibility testing.
    Conclusion of Substantial Equivalence (Overall System)Overall, the design changes to the EPS sleeves have been verified using well-established methods. The new design is functionally the same as the predicate device. The subject device different questions of safety and effectiveness.The subject device is substantially equivalent to the predicate, meaning it is as safe and effective as the previously cleared device.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for performance tests like "EPS Weighted Deflection Test" or "Total System Accuracy". These are typically engineering verification tests, and the "sample size" would refer to the number of units tested.
    • Data Provenance: Not specified for these engineering tests. "Sawbones®" is mentioned, indicating laboratory testing on synthetic bone models. This is not clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for the reported tests. The tests are engineering verification tests, not diagnostic accuracy studies requiring expert-established ground truth.

    4. Adjudication method for the test set:

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned or conducted. The device is a surgical guidance system, not a diagnostic AI tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not explicitly described as a standalone algorithm performance study. The "Total System Accuracy" test would assess the device's accuracy in guiding the surgical instrument, which is its primary function (albeit with a human surgeon operating the instrument under guidance). The text focuses on the mechanical and system accuracy of the guidance mechanism itself, not a diagnostic algorithm.

    7. The type of ground truth used:

    • For mechanical tests: Engineering specifications, precision measurements, or established physical benchmarks are the "ground truth."
    • For biological tests: Standards (e.g., ISO 10993) and established laboratory protocols define the "ground truth" for material properties and effects.

    8. The sample size for the training set:

    • Not applicable. The document does not describe the development or training of an AI algorithm in the context of a "training set" for diagnostic performance.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.

    Summary of Device Performance (from the document's conclusion):
    The primary conclusion is that "The design changes to the EPS sleeves have been verified using well established methods. The new design is functionally the same as the predicate device. The subject device different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate." This implies that all the listed performance tests were successfully passed, ensuring that the modified device remains as safe and effective as its predecessor.