510(k) Data Aggregation

    K Number
    K233827

    Validate with FDA (Live)

    Device Name
    OxiWear
    Manufacturer
    Date Cleared
    2024-08-28

    (271 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201770

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OxiWear system is indicated for non-invasive, spot-checking and/or continuous data collection of adult (22+) and adolescent (12-21) patients who are well or poorly perfused, during motion conditions. It is intended for use in hospitals, medical facilities, home healthcare environments, and mobile environments. The recommended application site is the helix of the ear.

    Device Description

    The OxiWear System is an ear pulse oximeter that non-invasively measures and displays patient blood oxygen saturation (SpO2) and Pulse Rate for spot-checking and/or continuous data collection. The OxiWear System is intended to be used with well or poorly perfused patients of all skin types in both motion conditions. The system is intended for use in clinical and home environments for adult (22+) and adolescent (12-21) patients of all skin types who are well or poorly perfused, and can safely wear the oximeter device on the helix of the OxilVear System can be used with patients that are in both motion and stationary conditions.

    The OxilVear System contains the OxilVear Wearable, and the OxiMear Mobile Application. The OxiMear Wearable is a wireless, reusable, non-invasive ear pulse oximeter that is intended to measure SpO2 and Pulse Rate based on the amount of transmitted, reflected light through the ear. The OxiWear Wearable is powered by an internal rechargeable lithium ion battery, and attaches directly to the ear. The OxilVear Wearable is intended to be recharged by the OxiWear Charger, which is a rechargeable portable charger,

    The OxiMear Wearable is designed to be used with the OxiMeation which receives data from the OxiMear Wearable via Bluetooth connection to the user's smart phone. The OxiMear Mobile Application acts as a historical trending, live data monitoring, and text alert application is intended to be used alongside the OxiWear Wearable during daily use, as well as a data repository to refer back to previous reports containing information about the patient, pulse rate, patient contacts, and more, to present to a physician or caregiver for review.

    AI/ML Overview

    The document describes the OxiWear system, an ear pulse oximeter, and its claim of substantial equivalence to predicate devices (Nonin Medical, Inc. Models 8100Q2, 8101Q2, and 8102Q2; and Masimo Corporation's Rad-G Pulse Oximeter). The acceptance criteria are primarily defined by the performance specifications for SpO2 and Pulse Rate accuracy, and the studies conducted to prove these metrics are largely non-clinical bench and simulator tests, complemented by clinical testing for SpO2 and Pulse Rate accuracy under no motion and motion conditions.

    Here's a breakdown based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" as a separate table but compares the subject device's (OxiWear) performance specifications against those of the predicate and reference devices. These comparisons implicitly define the performance targets for acceptance.

    CharacteristicAcceptance Criteria (Predicate/Reference)Reported Device Performance (OxiWear)
    SpO2 Measurement Range70-100%70-100%
    SpO2 Accuracy, no motionPredicate: ± 3% (ARMS); Reference: Adults/Pediatrics/Infants: ± 2% (ARMS)± 3.5% (ARMS)
    SpO2 Accuracy, motionReference: ± 3% (ARMS)± 3.5% (ARMS)
    SpO2 Accuracy, low perfusionPredicate: ± 2% (ARMS); Reference: Adults/Pediatrics/Infants: ± 2% (ARMS)± 3.5% (ARMS)
    Pulse Rate Measurement RangePredicate & Reference: 25-240 BPM20-240 BPM
    Pulse Rate Accuracy, no motionPredicate & Reference: ± 3 BPM (ARMS)± 3 BPM (ARMS)
    Pulse Rate Accuracy, motionReference: ± 5 BPM (ARMS)± 5 BPM (ARMS)
    Pulse Rate Accuracy, low perfusionPredicate & Reference: ± 3 BPM (ARMS)± 3 BPM (ARMS)

    Note: For SpO2 accuracy, OxiWear's reported accuracy (± 3.5% ARMS) is slightly less precise than either predicate's specified accuracy (± 3% or ± 2% ARMS). However, the document concludes that "OxiWear has a level of accuracy comparable to an industry standard pulse oximeter," implying this difference was deemed acceptable within the context of substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document specifies "clinical testing" but does not provide specific sample sizes for the SpO2 and Pulse Rate accuracy studies.

    • Test Set Sample Size: Not explicitly stated in the provided text.
    • Data Provenance: The document states that "SpO2 validation under normal perfusion conditions was conducted under controlled laboratory conditions evaluating the OxiWear system against a cooximeter." "Motion SpO2 testing was conducted in subjects wearing the OxiWear device while on a treadmill."
      • It doesn't specify the country of origin.
      • The studies appear to be prospective clinical testing (as subjects wore the device and were evaluated).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. For pulse oximetry, the ground truth for SpO2 accuracy is typically established using a co-oximeter in a controlled desaturation study, not necessarily by human experts.

    4. Adjudication Method for the Test Set

    Not applicable/not provided. For technical performance metrics like SpO2 and Pulse Rate accuracy, ground truth is typically established by reference devices (e.g., co-oximeters) or a simulator, not through human adjudication of subjective interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC study was not conducted. This type of study is more common for diagnostic imaging AI where human readers interpret cases with and without AI assistance. The OxiWear is a sensor device measuring vital signs directly.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes. The described clinical testing for SpO2 and Pulse Rate accuracy, where the OxiWear device's measurements are compared against a reference (co-oximeter or reference pulse oximeter), represents a standalone performance evaluation of the device's algorithmic and sensor capabilities.

    7. The Type of Ground Truth Used

    • Objective Measurement from Reference Devices:
      • For SpO2 accuracy (no motion): A co-oximeter was used as the ground truth reference in controlled laboratory conditions.
      • For SpO2 and Pulse Rate accuracy (motion): A reference pulse oximeter was used as the ground truth.
      • For Low Perfusion Testing and No Motion Pulse Rate Testing: Simulators were used to establish controlled conditions, and comparisons were made against "an industry standard pulse oximeter."

    8. The Sample Size for the Training Set

    The document does not mention any "training set" in the context of machine learning or AI model development. The OxiWear device appears to be a traditional pulse oximeter, not an AI-driven diagnostic device that would require such training sets. If there are internal signal processing algorithms that might have been "trained," this information is not disclosed in the provided FDA submission summary.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as no training set for an AI/ML model is explicitly mentioned or relevant based on the description of the device as a traditional oximeter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212161

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-03-06

    (602 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K201770

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rad-97 and Accessories is a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital, hospital type facilities, mobile and home environments.

    The Rad-97 and Accessories can communicate with network systems for supplemental remote viewing (e.g., at a central station).

    The Rad-97 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.

    The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions.

    The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions.

    The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions.

    The Rad-97 and Accessories are indicated for the continuous monitoring rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions.

    In addition, the Rad-97 and Accessories are indicated to provide the non-invasive spot-checking and continuous monitoring data obtained from the Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.

    The Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.

    The optional NomoLine Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The NomoLine Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatric and infant patients.

    The optional non-invasive blood pressure (NiBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NiBP module is designed to messure for patient population described in the following table:

    Patient Population Approximate Age Range Newborn (neonate) Birth to 1 month of age Infant 1 month to 2 years of age Child 2 to 12 years of age Adolescent 12-21 years of age Adult 21 years of age and older

    Devices with Masimo technology are only to be used with Masimo sensors and cables.

    The Rad-97 and accessories are indicated for the non-invasive continuous monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions.

    PV i may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PV in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.

    The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate from Pleth (RRp) for adult and pediatric patients during no motion conditions.

    The Radical-7 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.

    The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Radical-7 and Accessories are not intended to be used as the sole basis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.

    The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions in hospitals and hospital-type facilities.

    The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions in hospital-type facilities.

    The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospital-type facilities, home environments, and transport within healthcare facilities.

    The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of Respiratory Rate from photoplethysmogram (RRp) for adult and pediatric patients during no motion conditions in hospital-type facilities, home environments, and transport within healthcare facilities.

    Device Description

    This premarket notification [510(k)] is for the authorization to market the previously cleared Rad-97 (K193626) and Radical-7 (K193242) with an additional indication for spot-checking.

    The cleared versions of the subject devices already support both continuous monitoring and spotchecking use through the activation and deactivation of alarms. This submission updates the indications to reflect both uses of the subject devices.

    Rad-97: The Rad-97 is a patient monitor capable of providing multiple parameters. The Rad-97 product family provides the integrated ability of noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Perfusion Index (Pi), Pleth Variability Index (PVi), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa), Pleth Respiration Rate (RRp), capnography parameters, and noninvasive blood pressure (NiBP) parameters.

    Radical-7: The Radical-7 is a noninvasive monitor that measures arterial oxygen saturation (SpO2), pulse rate (PR), perfusion index (Pi), Pleth Variability Index (PVi), totalhemoglobin (SpHb), carboxyhemoglobin (SpCO), total oxygen content (SpOC), methemoglobin (SpMet), acoustic respiration rate (RRa), and Pleth Respiration Rate (RRp).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Masimo Radical-7 Pulse CO-Oximeter and Accessories and the Rad-97 and Accessories. The submission's primary purpose is to update the indications for use to explicitly include "spot-checking" functionality, which the devices already supported by allowing alarm deactivation.

    The document asserts that the devices are substantially equivalent to their predicate devices (K193626 for Rad-97 and K193242 for Radical-7) because there are no changes to the device's fundamental technology, principle of operation, or performance specifications. The "spot-checking" capability was already inherent in the devices' design, as users could deactivate continuous monitoring alarms to achieve this function. Therefore, the update is primarily a labeling change to reflect existing functionality.

    Based on the provided information, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-based device is not directly applicable in the way one might expect for a new AI product. This submission is for a traditional medical device (patient monitor/oximeter), and the "study" is demonstrating substantial equivalence to a predicate device, not necessarily proving a new algorithm's performance against specific acceptance criteria for AI.

    However, we can infer the "acceptance criteria" not as novel performance thresholds, but as the maintenance of the same performance characteristics as the predicate devices, and the "proof" coming from the assertion that no changes were made to the core technology that would alter these.

    Here's an attempt to structure the information according to the requested format, interpreting "acceptance criteria" as maintaining the predicate's performance and "study" as the justification for substantial equivalence.


    Acceptance Criteria and Device Performance (Interpreted)

    The core "acceptance criteria" in this 510(k) submission are that the subject devices (Rad-97 and Radical-7) maintain substantial equivalence to their predicate devices (Rad-97 K193626 and Radical-7 K193242, respectively) and that the expanded "indications for use" to include "spot-checking" does not introduce new questions of safety or effectiveness. This is because the devices already possessed the capability for spot-checking by deactivating alarms.

    The "study" that proves the device meets these criteria is an assertion of no change in fundamental technological characteristics, principles of operation, or performance specifications compared to the predicate devices, supported by software verification and validation that confirm the integrity of the existing as-cleared software.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a substantial equivalence claim for a traditional device with an expanded indication based on existing functionality, the "acceptance criteria" are implicitly the already cleared performance specifications of the predicate devices. The reported device performance is stated to be identical to these predicate devices. The change is in the stated use (labeling), not in the core performance.

    Parameter/FeatureAcceptance Criteria (from Predicate Device Specifications)Reported Device Performance (Subject Device)
    SpO2, no motion, 70-100%Rad-97: 1.5% ARMS (adults/pediatrics/infants/neonates)**; Radical-7: 2% (adults/pediatrics/infants), 3% (neonates)Same as Predicate
    SpO2, motion, 70-100%Rad-97: 1.5% Arms (adults/infants/pediatrics/neonates)**; Radical-7: 3% ARMS (adults/pediatrics/infants/neonates)Same as Predicate
    SpO2, low perfusion, 70-100%Rad-97: 2% (adults/pediatrics/infants/neonates); Radical-7: 2% (adults/pediatrics/infants/neonates)Same as Predicate
    Pulse Rate, no motion, 25-240 bpmRad-97: 3 bpm (adults/pediatrics/infants/neonates); Radical-7: 3 bpm (adults/pediatrics/infants/neonates)Same as Predicate
    Pulse Rate, motion, 25-240 bpmRad-97: 5 bpm (adults/pediatrics/infants/neonates); Radical-7: 5 bpm (adults/pediatrics/infants/neonates)Same as Predicate
    Pulse Rate, low perfusion, 25-240 bpmRad-97: 3 bpm (adults/pediatrics/infants/neonates); Radical-7: 3 bpm (adults/pediatrics/infants/neonates)Same as Predicate
    SpCO, 1-40%Both: 3% (adults/pediatrics/infants)Same as Predicate
    SpMet, 1-15%Both: 1% (adults/pediatrics/infants/neonates)Same as Predicate
    SpHb, 8-17 g/dLBoth: 1 g/dL (adults/pediatrics)Same as Predicate
    RRa (Acoustic Respiration Rate)4-70 rpm: 1 rpm (adults/pediatrics); 4-120 rpm: 1 rpm (infants/neonates)Same as Predicate
    RRp (Pleth Respiration Rate)4-70 rpm: 3 rpm ARMS, 1 rpm Mean Error (adults/pediatrics)Same as Predicate
    CO2 (for Rad-97 with NomoLine Capnography)Single dry gasses at 22±5°C and 1013±40 hPa, 0-15 volume%: ±(0.2 volume% +2% of reading); All conditions: ±(0.3 kPa + 4% of reading)Same as Predicate
    NiBP (for Rad-97)0-300 mmHg, ISO 81060-2Same as Predicate
    Principle of OperationMasimo Rainbow SET Technology, Capnography Technology, NiBP Technology (all same)Same as Predicate
    DisplayTouchscreen LCDSame as Predicate
    Alarm TypeVisual/Audible alarmSame as Predicate
    Mode of OperationContinuous operationSame as Predicate

    Note: ARMS accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/- ARMS of the reference measurements in a controlled study.
    *Applicable with RD SET Disposable sensors.

    2. Sample Size for the Test Set and Data Provenance

    No new clinical or non-clinical testing was conducted for this submission (K212161) because the changes were limited to labeling to include existing functionality. The predicate devices' performance data implicitly serve as the "test set" and provenance.

    • Sample Size for Test Set: Not applicable for this submission as no new testing was performed. The data and sample sizes supporting the original predicate clearances (K193626 and K193242) would have established these performance metrics. The document states that the devices are "the same as the predicate cleared... with exception of the updated indications for spot-checking."
    • Data Provenance: Not explicitly stated for new data. The original predicate device data would have come from clinical studies, often including induced hypoxia studies for oximetry, which typically occur in controlled clinical environments (e.g., US, Europe). The studies for the original clearances would have been prospective to generate the performance data.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not applicable for this submission as no new ground truth was established. The performance specifications are based on the device's technical capabilities, validated against reference standards in prior clearances.
    • Qualifications of Experts: N/A for this particular submission. For the original clearances, such validation would involve highly trained clinical professionals and laboratory personnel.

    4. Adjudication Method for the Test Set

    Not applicable for this submission as no new test set requiring adjudication was generated.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI-assisted diagnostic device where human reader improvement with AI assistance would be measured. This is a traditional physiological monitor.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This is a medical device that measures physiological parameters. Its performance is inherent in its design and sensor technology, not a detached algorithm output. The "algorithm" here refers to the internal processing of physiological signals, which is proven through the device's accuracy specifications.

    7. Type of Ground Truth Used

    The ground truth for the device's performance (as established for the predicate devices) would typically be:

    • Reference Devices/Methods: In a controlled study, the device's measurements (e.g., SpO2) would be compared against a gold standard reference (e.g., co-oximeter for blood gas analysis).
    • Controlled Physiological Conditions: For SpO2 and PR, studies often involve inducing a range of oxygen saturation levels and heart rates in human volunteers under controlled conditions.

    8. Sample Size for the Training Set

    Not applicable in the context of traditional medical device validation. "Training set" is a concept primarily relevant to machine learning/AI models. The device's internal signal processing algorithms are engineered based on principles of physics and physiology, not "trained" on a dataset in the AI sense.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. See point 8.