Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K213700

    Validate with FDA (Live)

    Date Cleared
    2021-12-13

    (20 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K201670

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MULTIX Impact is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX Impact is not intended for mammography.

    MULTIX Impact uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

    MULTIX Impact C is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact C enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX Impact C is not intended for mammography.

    MULTIX Impact C uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact C is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

    Device Description

    The MULTIX Impact (VA21) Radiography X-ray system is a floor mounted, modular system of x-ray components (x-ray tube, bucky wall stand, patient table, x-ray generator, portable wireless and fixed detectors) based on the predicate device, the MULTIX Impact (VA20, K203345).

    The MULTIX Impact C (VA21) Radiography X-ray system is a ceiling suspended, modular system of x-ray components (x-ray tube, bucky wall stand, patient table, x-ray generator, portable wireless and fixed detectors) based on the predicate device, the MULTIX Impact C (VA20, K203340).

    The following modifications have been made to the predicate devices:

      1. Upgrade software version from VA20 to VA21 to support the new features: Auto TOD Measurement, Auto Thorax Collimation, Virtual Collimation, Hybrid Image Documentation (HID).
      1. New mobile UI: Smart Remote Control (SRC).
      1. New accessory: myExam 3D Camera, to support the new software features. The myExam 3D Camera has been cleared in YSIO X.pree (K201670).
    AI/ML Overview

    The provided document is a 510(k) Summary for the Siemens MULTIX Impact and MULTIX Impact C radiographic systems. This document primarily focuses on demonstrating substantial equivalence to predicate devices based on modifications in software and accessories, rather than presenting a standalone clinical study for an AI-powered diagnostic device. Therefore, much of the requested information regarding acceptance criteria, study design for AI performance, and ground truth establishment is not present in this document.

    However, based on the information provided, here's what can be extracted and what cannot:

    General Statement:
    The document does not describe an AI-powered diagnostic device that requires specific clinical performance acceptance criteria based on accuracy, sensitivity, or specificity. Instead, it describes a radiographic system (X-ray machine) with new software features (Auto TOD Measurement, Auto Thorax Collimation, Virtual Collimation, Hybrid Image Documentation (HID)), a new mobile UI (Smart Remote Control), and a new accessory (myExam 3D Camera), and demonstrates its substantial equivalence to previously cleared predicate devices. The "acceptance criteria" discussed are related to the safety and functional performance of the radiographic system itself, and its consistency with regulatory standards and previous versions.

    Detailed Breakdown based on your Request:


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of performance acceptance criteria in the context of an AI diagnostic study (e.g., specific thresholds for sensitivity, specificity, or AUC). The "acceptance criteria" it refers to are regulatory compliance and functional testing for the X-ray system.

    Acceptance Criteria Category (Implied)Reported Device Performance (Implied)
    Regulatory ComplianceConforms to recognized standards (e.g., ANSI ES60601-1, IEC 60601-1-2, ISO 14971, IEC 62304). Software documentation for Moderate Level of Concern is included.
    Software FunctionalityTest results support that all software specifications have met the acceptance criteria. New features (Auto TOD Measurement, Auto Thorax Collimation, Virtual Collimation, HID) and new UI (Smart Remote Control) are supported.
    Risk ManagementRisk analysis completed, risk controls implemented, hazards mitigated.
    SafetyInstructions for use enable safe operation. Visual and audible warnings. System monitoring and error blocking. Adherence to industry practice for electrical and radiation hazards.
    Substantial EquivalenceDevice maintains same indications for use, similar operating environment and mechanical design as predicate devices. Performance is comparable to predicate, introduces no new safety risks.
    New Camera FunctionalitymyExam 3D Camera (Intel® RealSenseTM D400 series) and associated features (Auto Thorax Collimation, Virtual Collimation) are similar in functionality to reference device (YSIO X.pree, K201670) and bench testing concluded no impact on safety and effectiveness.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document describes "non-clinical tests" (integration and functional) and "bench testing" to support the modifications and substantial equivalence. It does not mention a "test set" in the context of diagnostic image data for performance evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. Ground truth establishment by experts is relevant for diagnostic AI performance evaluation, which is not the subject of this 510(k) summary.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. Adjudication methods are relevant for establishing ground truth in diagnostic studies, which is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC comparative effectiveness study was done. The device is an X-ray system, not an AI-assisted diagnostic software that directly affects human reader performance in interpreting images. The software updates relate to workflow and system operation (e.g., collimation, measurement, documentation) rather than diagnostic image interpretation assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    There is no indication that a standalone algorithm performance study was done. The device is an X-ray imaging system, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    This information is not provided in the document, as it is not an AI diagnostic device requiring such ground truth for performance evaluation. The "ground truth" in this context would be related to the functional correctness and safety of the X-ray system and its new features.

    8. The sample size for the training set

    This information is not provided in the document. Training sets are relevant for machine learning models, which are not explicitly described as part of the "software version: VA21" updates in a way that suggests a need for specific training set sizes for diagnostic performance claims. The software changes appear to be for operational enhancements of the X-Ray system rather than a diagnostic AI.

    9. How the ground truth for the training set was established

    This information is not provided in the document, for the reasons mentioned above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203345

    Validate with FDA (Live)

    Device Name
    MULTIX Impact
    Date Cleared
    2021-01-07

    (55 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K201670

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MULTIX Impact is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX Impact is not intended for mammography.

    MULTIX Impact uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

    Device Description

    The MULTIX Impact (VA20) Radiography X-ray system is a modular system of x-ray components (floor-mounted x-ray tube, bucky wall stand, bucky table, x-ray generator, portable wireless and fixed detectors) based on the predicate device, the MULTIX Impact (K193089). The detectors for the subject device, MULTIX Impact (VA20), are the same as the detectors of the predicate device. The following modifications have been made to the predicate device:

    • Modified tube stand (motorized) to support Ortho Function 1.
    • Modified automatic collimator 2.
      1. New bucky wall stand
      1. Upgraded software version from VA11 to VA20 to support hardware modifications
    • Modified patient table న్.
      1. Modified touch user interface (TUI)
    • Modified wireless remote control console (WRCC) with new control design 7.

    The modified system will be branded as the MULTIX Impact.

    AI/ML Overview

    The provided text is a 510(k) summary for the Siemens MULTIX Impact X-ray system. It details modifications made to a previously cleared predicate device (MULTIX Impact K193089). This document does not include information about specific acceptance criteria or the study that proves the device meets those criteria in the typical sense of a clinical performance study with defined metrics like sensitivity, specificity, or reader performance.

    Instead, the submission focuses on demonstrating substantial equivalence to a predicate device, meaning the new device is as safe and effective as a legally marketed device. The "acceptance criteria" here are primarily conformity to recognized standards and successful completion of non-clinical verification and validation testing, ensuring the modified device performs as intended and introduces no new safety risks.

    Here's a breakdown of the requested information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied by 510(k) Submission)Reported Device Performance (Summary from Submission)
    Conformance to international and national standards (e.g., IEC 60601 series, ISO 14971, NEMA PS 3.1-3.20)"MULTIX Impact (VA20) conforms to the following standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012; IEC 60601-1-3:2013; IEC 60601-1-2:2014; IEC 62366-1:2015; ISO 14971:2007; IEC 60601-1-6:2013; IEC 62304:2015; IEC 60601-2-28:2017; IEC 60601-2-54:2018; NEMA PS 3.1-3.20 (2016) and ISO 10993-1:2009." "All test results are a pass and support our claim of device safety and effectiveness."
    Functional performance of modified components (e.g., tube stand, collimator, Bucky wall stand, software)"Verification and validation testing have been successfully completed and test results show that the subject device, MULTIX Impact (VA20) with all its components, is substantially equivalent to the predicate device." "Non-clinical tests (integration and functional) were conducted on the MULTIX Impact (VA20) during product development." "The test results support that all the software specifications have met the acceptance criteria."
    Software safety and effectiveness (Moderate Level of Concern)"Software Documentation for a Moderate Level of Concern software... is also included... The performance data demonstrates continued conformance with special controls for medical devices containing software." "The risk analysis was completed, and risk controls were implemented to mitigate identified hazards."
    Equivalence to predicate device in terms of indications for use, design, material, functionality, technology, and energy source."The MULTIX Impact (VA20) is substantially equivalent to the predicate device... The subject device is within the same classification regulation, has the same indications for use, and the same mechanical design as the predicate device." "The subject device uses the same or similar components cleared in the MULTIX Impact..." "The non-clinical test data demonstrate that the MULTIX Impact (VA20) device performance is comparable to the predicate device..."
    Absence of new potential safety risks"In summary, Siemens concludes that the MULTIX Impact (VA20) does not introduce any new potential safety risks..."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes a premarket notification for a modification to an existing X-ray system. The testing mentioned is primarily non-clinical verification and validation testing of hardware and software components against engineering specifications and regulatory standards. There is no mention of a "test set" in the context of clinical image data or patient studies from specific countries or whether they were retrospective/prospective. The assessment focuses on the performance of the system as a medical device, not a diagnostic algorithm.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. As this is a 510(k) submission for an X-ray system, the "ground truth" (in the sense of expert-labeled medical images for an AI algorithm) is not relevant to the type of testing performed. The documentation focuses on engineering and regulatory compliance, not diagnostic accuracy of an AI component.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No "test set" with expert adjudication is mentioned in the context of this 510(k) submission.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This 510(k) pertains to an X-ray imaging system itself, not an AI-powered diagnostic aide. Therefore, an MRMC study and effects on human reader performance are not described.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This document does not describe an AI algorithm or its standalone performance. The device is an X-ray system, which requires human operators.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The ground truth concept, as it relates to expert-labeled medical data for AI algorithms, is not part of this submission for an X-ray system itself. The "ground truth" for the engineering tests would be the design specifications and expected functional behavior of the components.


    8. The sample size for the training set

    Not applicable. This document does not describe an AI algorithm that undergoes training.


    9. How the ground truth for the training set was established

    Not applicable. This document does not describe an AI algorithm or its training process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203340

    Validate with FDA (Live)

    Device Name
    MULTIX Impact C
    Date Cleared
    2021-01-06

    (55 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K201670, K191418

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MULTIX Impact C is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact C enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX Impact C is not intended for mammography.

    MULTIX Impact C uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact C is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.

    Device Description

    The MULTIX Impact C Radiography X-ray system is a modular system of x-ray components (ceiling suspension with x-ray tube, bucky wall stand, bucky table, x-ray generator, and portable wireless and fixed detectors) based on the predicate device, the MULTIX Impact (K193089). The detectors for the subject device, MULTIX Impact C, are the same as the detectors of the predicate device. The following modifications have been made to the predicate device:

      1. New ceiling suspension with motorized tube tilting support for ortho function and new ceiling suspension with manual tube tilting
    • Modified automatic collimator 2.
      1. New Bucky Wall Stand
    • Upgraded software version from VA11 to VA20 to support hardware modifications 4.
    • Modified patient table న్.
    • Modified touch user interface (TUI) 6.
    • Modified wireless remote-control console (WRCC) with new control design 7.

    The new system will be branded as the MULTIX Impact C.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called MULTIX Impact C, which is a radiographic X-ray system. It describes the device, its intended use, and claims substantial equivalence to a predicate device (MULTIX Impact K193089).

    However, the document focuses on demonstrating substantial equivalence through comparisons of technical characteristics and adherence to standards for the device itself (X-ray system), rather than providing details about an AI algorithm's performance or a clinical study that would involve acceptance criteria for such an algorithm. The MULTIX Impact C is designed to generate diagnostic images, but there is no mention of an AI component or software that interprets images or assists human readers in a diagnostic capacity.

    Therefore, many of the requested details about acceptance criteria, study design, expert involvement, and AI performance cannot be extracted from this document, as it describes a hardware-focused X-ray system rather than an AI-driven image analysis tool.

    Here's a breakdown based on the information available in the provided text:

    1. A table of acceptance criteria and the reported device performance

    No acceptance criteria and reported device performance directly related to an AI algorithm's diagnostic output (e.g., sensitivity, specificity, AUC) are provided. The tests described are non-clinical performance and safety tests for the X-ray system hardware and associated software.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable, as no AI-specific test set or data provenance is mentioned. The testing described is for the X-ray system's hardware and software functionality, not for AI diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no ground truth for an AI diagnostic algorithm's performance is established or discussed. The document mentions "operators are healthcare professionals familiar with and responsible for the x-ray examinations to be performed," but this refers to the users of the X-ray system, not experts establishing ground truth for AI.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no AI-specific test set or adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes an X-ray imaging system, not an AI-assisted diagnostic tool that would typically undergo an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a radiographic X-ray system; there is no standalone algorithm performance described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. Ground truth for an AI diagnostic algorithm is not discussed. The "ground truth" for the device's performance would be its conformity to engineering specifications and safety standards, as demonstrated by non-clinical testing.

    8. The sample size for the training set

    Not applicable, as there is no mention of an AI model or a training set. The software mentioned (VA20) is an upgraded version to support hardware modifications and new control features, not an AI for image analysis.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of an AI model or a training set.


    Summary of what is described regarding the device's "acceptance criteria" and testing:

    The document indicates that the device (MULTIX Impact C X-ray system) meets its "acceptance criteria" through:

    • Non-clinical performance testing: "Non-clinical tests were conducted for the MULTIX Impact C during product development. The modifications described in this Premarket Notification are supported with verification and validation testing."
    • Adherence to Standards: The device conforms to numerous international and national standards, including:
      • ES60601-1:2005/(R)2012 and A1:2012 (Medical electrical equipment – General requirements for basic safety and essential performance)
      • IEC 60601-1-3:2008+A1:2013 (Medical electrical equipment – General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment)
      • IEC 60601-1-2:2014 (Medical electrical equipment – General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests)
      • IEC 62366-1:2015 (Medical devices – Application of usability engineering to medical devices)
      • ISO 14971:2007 (Medical devices – Application of risk management to medical devices)
      • IEC 60601-1-6:2013 (Medical electrical equipment – General requirements for basic safety and essential performance – Collateral Standard: Usability)
      • IEC 62304:2015 (Medical device software – Software life cycle processes)
      • IEC 60601-2-28:2017 (Medical electrical equipment – Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis)
      • IEC 60601-2-54:2018 (Medical electrical equipment – Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy)
      • NEMA PS 3.1-3.20 (2016) (DICOM standards)
      • ISO 10993-1:2009 (Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process)
    • Software Documentation and Testing: "Software Documentation for a Moderate Level of Concern software... The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional) were conducted on the MULTIX Impact C during product development." All software specifications "met the acceptance criteria."
    • Risk Analysis: "The risk analysis was completed and risk controls were implemented to mitigate identified hazards. The test results support that all the software specifications have met the acceptance criteria."
    • Substantial Equivalence: The primary "acceptance criterion" for clearance appears to be demonstrating substantial equivalence to a predicate device (MULTIX Impact K193089), which is achieved by showing similar indications for use, mechanical design, and that "non-clinical test data demonstrate that the MULTIX Impact C device performance is comparable to the predicate device."

    In summary, the provided text is for the regulatory submission of an X-ray machine. It details the device's features, compliance with general safety and performance standards for X-ray equipment, and its similarity to a predicate device. It does not describe an AI diagnostic algorithm or a clinical study for such an algorithm's performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1